Field Interactions & Safety |
Tuesday 21 April 2009 |
Room 323ABC |
16:00-18:00 |
Moderators: |
Penny Anne Gowland
and Greig C. Scott |
|
|
|
16:00 |
298. |
Influence of Strong Static
Magnetic Fields on Myocardial Mechanics: Evaluation
Applying Ultra-Wideband Radar |
|
|
Florian Thiel1,
Mathias Hein2, Jürgen Sachs2,
Ulrich Schwarz2, Tomasz Lindel1,
Frank Seifert1
1Physikalisch-Technische Bundesanstalt (PTB),
Berlin, Germany; 2Ilmenau University of
Technology |
|
|
ECG is excessively used
for triggering MR data acquisition to image the
heart. It is well established that the ECG is
corrupted by the magneto-hydrodynamic effect (MHD).
Hence, there is increasing difficulty to use the ECG
for MR-triggering especially at B0-fields beyond 1.5
T. To investigate whether the myocardial contraction
is also influenced by the static magnetic field, we
propose a novel method, based on an ultra-wideband
radar technique (UWB radar) to monitor the global
myocardial dynamics inside a MR scanner. |
|
|
|
16:12 |
299. |
Effects of 7 Tesla MRI
Examination on Postural Stability |
|
|
Jens Matthias
Theysohn1,2, Andreas K. Bitz1,
Oliver Kraff1,2, Stefan Maderwald1,2,
Marcus Gerwig3, Oliver Kastrup3,
Dagmar Timmann3, Franz Schmitt4,
Harald H. Quick1,2, Elke Ruth Gizewski1,2,
Michael Forsting1,2, Mark E. Ladd1,2,
Susanne C. Ladd1,2
1Erwin L. Hahn Institute for Magnetic
Resonance Imaging, University Duisburg-Essen, Essen,
Germany; 2Department of Diagnostic and
Interventional Radiology and Neuroradiology,
University Hospital Essen, Essen, Germany; 3Department
of Neurology, University Hospital Essen, Essen,
Germany; 4Siemens Healthcare, Erlangen,
Germany |
|
|
Ultra high-field MRI (7
Tesla and above) generates more temporary
side-effects compared to 1.5T and 3T, e.g.
dizziness. Postural stability was quantitatively
measured before and after a 7 Tesla examination of
the head. Twenty-five volunteers underwent a
Romberg’s test. Stability with eyes closed was
significantly reduced directly after the MR
examination, but was fully recovered 15 minutes
after examination. Physiological effects by
orthostatic regulation were excluded. Stability
results for tests with eyes open show no significant
difference. The results show that exposition to a 7
Tesla examination causes only a temporary
dysfunction or over-compensation of the vestibular
system. |
|
|
|
16:24 |
300. |
Numerical Evaluation of SAR
Within Whole-Body Pregnant Woman Models in MRI
Birdcage Coil |
|
|
Zhangwei Wang1,
George X. Xu2, Victor Taracila1,
Jason Jin1, Fraser J. Robb1
1GE Healthcare Coils, Aurora, OH, USA; 2Department
of Biomedical Engineering, Rensselaer Polytechnic
Institute, Troy, NY, USA |
|
|
Whole body pregnant
woman models at different gestational stages were
adopted to investigate the SAR variation at 1.5T.
The highest SAR is occurred in the mother's
peripheral tissues in all pregnancy phase. And the
maximum local SAR of the fetus is below IEC
limitation. The results show that the local maximum
SAR1g and SAR10g can be better indications as
limitation factor other than the whole body average
SAR. Future work will be done by using more
pregnancy stages and higher resolution models. |
|
|
|
16:36 |
301. |
Optimized Signal Intensity and
T1r Based NMR Thermometry for Ultra-High Field RF
Safety Applications |
|
|
Devashish Shrivastava1,
Timo Liimatainen1, Ute Goerke1,
Jeramy Kulesa1, Timothy Hanson1,
Shalom Michaeli1, J. Thomas Vaughan1
1CMRR, Radiology, University of Minnesota,
Minneapolis, MN, USA |
|
|
Good linear correlations
were obtained between the core temperatures, and the
'optimized' signal intensities and the longitudinal
magnetization relaxation rate constants in the
rotating frame (i.e., T1r) in an in vivo
porcine brain at 3 tesla (T). The correlations were
obtained to measure RF heating induced temperatures
with sub-degree Celsius accuracy in porcine brains
at ultra-high fields. |
|
|
|
16:48 |
302. |
Real-Time Global and Local SAR
Monitoring for Parallel Transmission Systems |
|
|
Ingmar Graesslin1, Sven Biederer2,
Bjoern Annighoefer3, Hanno Homann1,
Henry Stahl1, Peter Vernickel1,
Ulrich Katscher1, Dennis Glaesel4,
Paul Harvey5
1Philips Research Europe, Hamburg, Germany;
2University of Lübeck, Lübeck, Germany;
3TU Hamburg-Harburg, Hamburg, Germany;
4Research Establishment for Applied
Science, Wachtberg, Germany; 5Philips
Healthcare, Best, Netherlands |
|
|
A comprehensive RF
safety concept for parallel transmission MRI systems
can be realized by verifying the SAR limits before
scanning and monitoring RF signals during scanning.
However, using a pre-calculated safety margin
over-estimates the actual SAR. This abstract extends
the concept of pre-calculated safety margins with a
real-time global and local SAR monitoring system.
This new concept allows for a significantly
increased permissible RF duty cycle and improves the
detection of violations of SAR limits or any unsafe
condition. The impact of patient motion on SAR is
analyzed, and a scan termination in case of unsafe
system operation is demonstrated. |
|
|
|
17:00 |
303. |
Safety and Initial Results
with a Dedicated Human 7T Eye Coil |
|
|
Peter A. Wassenaar1,
Kathryn Richdale2, Katharine Teal
Bluestein1, John Christoforidis3,
Titus Lanz4, Michael Vincent Knopp1,
Petra Schmalbrock1
1Radiology, The Ohio State University,
Columbus, OH, USA; 2Optometry, The Ohio
State University, Columbus, OH, USA; 3Ophthalmology,
The Ohio State University, Columbus, OH, USA; 4Rapid
MR International, LLC, Columbus, OH, USA |
|
|
Ultra-high field MRI
allows higher resolution for examination of the fine
structures of the eye. This work focuses on the
safety testing and implementation of a custom
surface coil at 7T. A transmit head coil was used
with a 4cm receive coil. Safety was insured through
both active and passive decoupling. Testing
indicated good homogeneity of the imaging region and
signal drop off as expected for the coil diameter.
Animal testing demonstrated that tissue heating did
not exceed 0.3°C, even with high RF power. The use
of a custom coil yielded high quality images of the
human eye in vivo. |
|
|
|
17:12 |
304. |
RF Safety Assessment of a 7T
Head Coil Using Thermal Modeling with Discrete
Vessels |
|
|
Astrid L.H.M.W. van Lier1, Cornelis
A.T. van den Berg1, Dennis W.J. Klomp1,
Bas W. Raaymakers1, Jan J.W. Lagendijk1
1Radiology, Radiotherapy, Nuclear Medicine,
University Medical Center Utrecht, Utrecht,
Netherlands |
|
|
The RF safety assessment
of a commercially available 7T head coil using
thermal simulations with a discrete vessels model is
shown. This model incorporates the heating of blood
in the arteries and veins as it traverses through a
heated volume, which is an addition to Pennes’ bio
heat transfer equation. Comparing the maximum SAR
with the maximum temperature rise within the head it
was concluded that the restrictions based on volume
integrated SAR calculations are over-conservative,
which results in the sub-optimal imaging. |
|
|
|
17:24 |
305. |
Reduction of Implant RF
Heating by Modification of Electric Field
Distribution |
|
|
Yigitcan Eryaman1,
Haydar Celik1, Burak Akin1,
Ergin Atalar1
1Electrical Engineering, Bilkent University,
Ankara, Turkey |
|
|
In this work, the
optimum EM field that reduces the RF heating of
implant leads is calculated. The resulting field
also satisfies a given transmit target sensitivity
constraint in an axial plane. Among infinite number
of solutions satisfying these constraints, the one
that minimizes the average SAR is presented. The
field distribution which satisfies the electric
field and the target transmit sensitivity
constraints is realized by using a linearly
polarized birdcage coil. Finally, heating
experiments of resonant length wires is conducted by
using linearly polarized and quadrature coils. |
|
|
|
17:36 |
306. |
Detecting Unsafe Device
Coupling Using Reversed RF Polarization |
|
|
William Overall1,
Pascal Stang1, John Pauly1,
Greig Scott1
1Electrical Engineering, Stanford University,
Stanford, CA, USA |
|
|
Patients with long-wire
implants are ineligible for MR scans due to the risk
of RF burns arising from RF fields coupling to their
device. We propose the use of a low-power prescan to
quantify the level of coupling and determine the
burn risk for a particular patient in a particular
scanner. This is accomplished through use of
reversed RF polarization imaging, which produces
images with signal proportional to RF coupling.
Simulations and experimental data suggest that this
technique can be reliably used to determine burn
risk in a prescan setting. |
|
|
|
17:48 |
307. |
Characterization of the
Relationship Between MR-Induced Distal Tip Heating
in Cardiac Pacing Leads and the Electrical
Performance of Novel Filtered Tip Assemblies |
|
|
Robert S. Johnson1,
Holly Moschiano1, Robert Stevenson1,
Scott Brainard2, Sam Ye2,
Joseph E. Spaulding1, Warren Dabney1
1Cardiac & Neurology, Greatbatch, Inc.,
Clarence, NY, USA; 2Cardiac & Neurology,
Greatbatch, Inc., Plymouth, MN, USA |
|
|
The impedance of
inductor-capacitor band stop filters were
characterized with impedance spectroscopy. If was
found the impedance of the device was controllable
with proper selection of the inductor and capacitor
values. The filters were incorporated in implantable
leads and the temperature rise of the distal end was
measured during a MRI scan with an RF-intensive
imaging protocol. The minimum heating was 9°C with
the highest impedance filter and the maximum heating
was 51°C with a control lead. Systematic changes in
the lead configuration and location reduced the
heating to near instrumentation noise level. |
|
|
|
|