Electronic Posters
: Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
Pulse Sequences - RF
Tuesday May 10th
Exhibition Hall |
13:30 - 15:30 |
Computer 118 |
13:30 |
4408. |
Simultaneous B1 and
B0 mapping
using dual echo actual flip angle imaging (DE-AFI)
Claudia Lenz1, Oliver Bieri1,
Klaus Scheffler1, and Francesco
Santini1
1Radiological Physics, University
of Basel Hospital, Basel, Switzerland
Only recently, actual flip angle imaging (AFI)
has been introduced as a fast and robust 3D
method for mapping the B1 transmit field by
measuring the spatial variations of the
effective flip angle. The AFI pulse sequence
consists of a dual TR conventional spoiled
gradient echo pulse sequence, where TR2 >
TR1. In this work, a second echo has been
added to the standard AFI timing diagram,
which enables additional B0 mapping by
reconstructing phase difference maps based
on the phase images of the two acquired
echoes. In vivo results of fast simultaneous
B1 and B0 mapping using dual echo AFI are
presented.
|
14:00 |
4409. |
T1-nonlinearity corrections for fast
Transmit-Array B1+-mapping
of the human brain in the small-tip-angle regime
Martijn Anton Cloos1,2, Nicolas
Boulant1, Guillaume Ferrand2,
Michel Luong2, Christopher J
Wiggins1, Denis Le Bihan1,
and Alexis Amadon1
1LRMN, CEA, DSV, I2BM, NeuroSpin,
Gif-Sur-Yvette, ile-de-France, France, 2CEA,
DSM, IRFU, Gif-Sur-Yvette, ile-de-France,
France
Currently, B1+-mapping
is particularly difficult due to the
combination of time and conservative
specific absorption rate (SAR) constraints
applicable to parallel transmission studies
involving human subjects. Measuring relative
B1+-maps in the
low-tip-angle regime provides a low SAR
solution. Inherently, this method requires a
relatively long TR (0.2-1.0s) to remain in
the domain where the signal intensity is
linearly dependent for a large range of
flip-angles. Considering 3D tailored
excitation pulses, such TR values require
too much time for transmit-array B1+-mapping
and pulse validation. To tackle the
aforementioned problems, an optimized
version of this method for the
quantification of non-selective excitation
pulses is presented.
|
14:30 |
4410. |
An
experimental comparison of B1-mapping Techniques
at two field strengths
Rolf Pohmann1
1Magnetic Resonance Center, Max
Planck Institute for Biological Cybernetics,
Tübingen, Germany
B1 mapping was shown
|
15:00 |
4411. |
Fast B1 Mapping
using a STEAM-based Bloch-Siegert Preparation
Pulse
Kay Nehrke1, and Peter Börnert1
1Philips Research Laboratories,
Hamburg, Germany
Fast B1-mapping is an essential prerequisite
for multi-element transmit applications like
e.g. RF-shimming or multi-dimensional RF
pulse design. However, the acquisition speed
of B1-mapping sequences is typically limited
by SAR constraints, relaxation times, or
characteristic sequence properties. In this
work, a novel STEAM-based preparation
sequence is presented, which employs the
recently introduced Bloch-Siegert B1-mapping
approach. The preparation sequence stores
the Bloch-Siegert phase shift along the
longitudinal magnetization, which allows the
fast readout of the stimulated echo by a
subsequent train of small angle pulses. The
flexibility and versatility of this concept
is demonstrated in experiments on phantoms
and in vivo.
|
Wednesday May 11th
|
13:30 - 15:30 |
Computer 118 |
13:30 |
4412. |
Gradient and
frequency modulated excitation for a tailored
spatial trajectory with two-dimensional time
encoding for Fourier-free imaging
Angela Lynn Styczynski Snyder1,
Curt Corum2, Steen Moeller2,
Nathan Powell3, and Michael
Garwood2
1Department of Biomedical
Engineering, University of Minnesota,
Minneapolis, MN, United States, 2Department
of Radiology, University of Minnesota, 3Department
of Neuroscience, University of Minnesota
A frequency-swept RF pulse and modulated
gradients can be used to move a resonance
region through space to generate sequential
excitation and subsequent echo formation
with time-dependence, which allows image
formation without use of the Fourier
transform. This two-dimensional time
encoding (2DTE) has the unique and important
property that each region in space can be
treated independently, which is potentially
of great benefit for solving problems that
are inherently spatial in nature, such as
compensating for B0 and B1 inhomogeneities,
which is of increasing interest for higher
field strengths.
|
14:00 |
4413. |
Simultaneous
Bloch Siegert B1+ and
T2 mapping
in one experiment using a multi spin echo
sequence
Volker Sturm1, Thomas Christian
Basse-Lüsebrink1,2, Thomas Kampf1,
Guido Stoll2, and Peter Michael
Jakob1
1Experimental Physics 5,
University of Würzburg, Würzburg, Germany, 2Neurology,
University of Würzburg, Würzburg, Germany
A novel method for B1+ mapping
based on the Bloch-Siegert (BLS) shift was
recently introduced. BLS-based B1+ mapping
employs off-resonant pulses before signal
acquisition to encode B1 information
into the signal phase. In the present study,
BLS B1+ mapping
was extended to CPMG-based Multi-Spin-Echo (MSE).
Through only one experiment, this method
simultaneously provides the data needed for
B1+ mapping
with the data necessary for T2-quantification.
Ex vivo phantom and in vivo experiments were
performed to investigate the proposed
method.
|
14:30 |
4414. |
A novel
B1-insensitive outer volume suppression pulse
Travis Benjamin Smith1, and
Krishna S Nayak1
1Electrical Engineering,
University of Southern California, Los
Angeles, CA, United States
Reduced field-of-view imaging enables
accelerated acquisitions or finer
resolutions than standard prescriptions.
Outer volume suppression reduces signals
outside a region of interest and attenuates
aliasing energy when a reduced field-of-view
is prescribed. We present a new design for
B1-insensitive outer volume suppression and
demonstrate its performance at 3 Tesla in a
phantom and in vivo with cardiac-gated
scans.
|
15:00 |
4415. |
Time
Interleaved Acquisition of Modes (TIAMO): an
Analysis of SAR and Image Contrast Implications
Stephan Orzada1,2, Stefan
Maderwald1,3, Benedikt A. Poser1,4,
Sören Johst1,2, Mark E. Ladd1,2,
Stephan Kannengiesser5, and
Andreas K. Bitz1,2
1Erwin L. Hahn Institute for
Magnetic Resonance Imaging, Essen, NRW,
Germany, 2Department
of Diagnostic and Interventional Radiology
and Neuroradiology, University Hospital
Essen, Essen, NRW, Germany, 3University
of Duisburg-Essen, Essen, NRW, Germany, 4Donders
Institute for Brain, Cognition and Behaviour,
Centre for Cognitive Neuroimaging, Radboud
University Nijmegen, Nijmegen, Netherlands, 5Siemens
Healthcare Sector, Erlangen, Germany
High static field strengths of 7 Tesla and
above are challenging due to the severe B1
inhomogeneities caused by the short
wavelength. Several methods have been
proposed to tackle this problem including RF
Shimming and Transmit SENSE. Recently, a
technique called Time-Interleaved
Acquisition of Modes (TIAMO) has been
proposed to reduce the impact of B1
inhomogeneity. In this work the influence of
TIAMO on the image contrast as well as on
the time-averaged SAR at 7 Tesla is
addressed in detail, showing that TIAMO is
superior to RF shimming in terms of image
homogeneity and SAR.
|
|
|
Electronic
Posters : Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
B1 & Mapping
Monday May 9th
Exhibition Hall |
14:00 - 16:00 |
Computer 119 |
14:00 |
4416. |
Saturated Double Angle
Method with radial sampling
Liyong Chen1,2, and Edward V.R. DiBella1,2
1Utah Center for Advanced Imaging Research,
Department of Radiology, University of Utah, Salt Lake
City, Utah, United States, 2Department
of Bioengineering, University of Utah, Salt Lake City,
Utah, United States
Saturated double angle method with radial sampling was
proposed to get B1+ map.
|
14:30 |
4417. |
A new phase-based method
for rapid 3D B1 mapping
using double RF pulses
Yulin V Chang1
1Mechanical Engineering, Washington
University, St. Louis, MO, United States
A fast, easy-to-implement, phase-based 3D B 1 mapping
method is presented. This method uses a pair of
orthogonal RF pulses to generate a flip-angle-dependent
phase deviation of the magnetization vector from /4.
It is robust to T 1 and
T 2 relaxations
and the B 0 field
inhomogeneity.
|
15:00 |
4418. |
Comparison of four phase
based methods for the B1+ mapping
at 7T
Flavio Carinci1,2, Federico von
Samson-Himmelstjerna1,3, Davide Santoro1,
Tomasz Lindel1,4, Matthias Dieringer1,5,
Frank Seifert1,4, Jan Sobesky3,6,
and Thoralf Niendorf1,5
1Berlin Ultra-High Field Facility (BUFF), Max
Delbrück Center for Molecular Medicine (MDC), Berlin,
Germany, 2Department
of Physics, Insubria University, Como, Italy,3Center
for Stroke Research Berlin (CSB), Charitè
Universitaetsmedizin Berlin, Berlin, Germany, 4Physikalisch-Technische
Bundesanstalt (PTB), 5Experimental
and Clinical Research Center (ECRC)), Charitè Campus
Berlin, Berlin, Germany, 6Department
of Neurology, Charitè Universitaetsmedizin Berlin,
Berlin, Germany
In ultra-high field MRI the increase of signal to noise
ratio comes together with longer T1 and shorter T2*
relaxation times, higher specific absorption rates and
higher B1 field inhomogeneities. Phase-based methods for
B1 mapping have been shown to be more accurate than
magnitude-based methods as the phase of the signal is
insensitive to T1 relaxation effects and coil
sensitivity profiles. In this work we compare four
phase-based methods taken from literature and adapted to
7T MRI: the Mugler method, the Morrell method, the
Santoro method and the Sacolick method. The methods are
compared in terms of the sensitivity to the B1 and B0
inhomogeneities, SAR levels and repetition times, using
simulations together with phantom and in vivo
experiments.
|
15:30 |
4419. |
Reduction of Required
Gradient Spoiler Size For AFI B1 Mapping
Kim Shultz1, Greig Scott1, and
John Pauly1
1Electrical Engineering, Stanford University,
Stanford, CA, United States
AFI is a powerful, fast technique for B1 mapping, useful
for quantitative imaging and parallel transmit pulse
design. It is made more difficult by the requirement of
very large gradients to spoil the transverse
magnetization at the end of each TR. We present a
technique to correct for the error that occurs with
small spoiler gradients based on the expected
reconstruction error seen in simulations for various
tissues. This allows accurate AFI results with more
reasonable gradient sizes. Additionally, the error seen
in lipids is smaller than that seen with very large
spoiler gradients.
|
Tuesday May 10th
|
13:30 - 15:30 |
Computer 119 |
13:30 |
4420. |
On the Effectiveness of RF
Spoiling at 7T
Douglas A C Kelley1,2, and Janine M Lupo2
1Global Applied Science Laboratory, GE
Healthcare, San Francisco, CA, United States, 2Radiology
and Biomedical Imaging, University of California, San
Francisco, San Francisco, CA, United States
RF spoiling is a common technique for suppression of
transverse coherence, but T2 reduction, T1 elongation
and RF inhomogeneity in human tissue at 7T alter the
effectiveness of the technique in quantitative imaging
methods like Actual Flip Imaging.
|
14:00 |
4421. |
Asymmetric field
distribution in B1+ and B1- maps
are caused by phase differences in field components in the
laboratory frame
Hidehiro Watanabe1, Nobuhiro Takaya1,
and Fumiyuki Mitsumori1
1Environmental Chemistry Division, National
Institute for Environmental Studies, Tsukuba, Ibaraki,
Japan
Magnitude of B1+ has
an asymmetric profile in human brain at high field. We
investigated the source of this asymmetry through simple
change of the formula on the relationship of RF fields
in the rotating and laboratory frames. |B1+|
can be expressed as the magnitude and the phase
difference of the RF field components in the laboratory
frame. While the former component has symmetric profile,
the latter has asymmetry in human brains at 4.7 T. From
these results, we concluded that the asymmetries in |B1+|
and |B1-*| maps are
derived from the phase difference in the laboratory
frame.
|
14:30 |
4422. |
3D Slab Selective AFI
utilizing a thin slab approach
Christopher Thomas Sica1, and Christopher M
Collins1
1Radiology, The Pennsylvania State
University, Hershey, Pennsylvania, United States
AFI is an established B1+ mapping technique that can
suffer from slice profile effects, and is thus typically
implemented with 3D non-selective RF pulses,
necessitating 3D encoding of a large volume and long
scan duration. This work presents a 3D slab-selective
AFI technique with a significant reduction in scan time.
A thin slab is excited and a small number of slices are
encoded along the slab select axis. Slices within the
central portion of the slab are kept and outlying slices
are discarded. This approach was investigated using
simulation and experiment in a phantom. Agreement
between non-selective and slab-selective AFI is very
good.
|
15:00 |
4423. |
Sa2RAGE sequence
improvements and in-vivo brain RF-shimming at 7 Tesla
Florent Eggenschwiler1, Arthur William Magill1,2,
Tobias Kober1, Rolf Gruetter1,3,
and José Pedro Marques1,2
1EPFL, Laboratory for Functional and
Metabolic Imaging, Lausanne, Vaud, Switzerland, 2University
of Lausanne, Department of Radiology, Lausanne, Vaud,
Switzerland,3Universities of Geneva and
Lausanne, Department of Radiology, Switzerland
The accuracy of the recently proposed B1+-mapping
sequence: Saturation prepared with 2 RApid Gradient
Echoes (Sa2RAGE) was improved in terms of T1-insensitivity
by considering partial Fourier encoding as well as a
parallel imaging technique (GRAPPA). This sequence was
then used to map the B1+ field
distribution obtained when performing in-vivo brain
RF-shimming at 7 Tesla. Shimming was performed by
solving the Magnitude Least Squares (MLS) problem and
the B1+ homogeneity
improvement throughout the brain was demonstrated by
showing that the distribution of the B1+ values
after shimming is more centered on the mean B1+,
illustrating the validity of the Sa2RAGE sequence as
well.
|
Wednesday May 11th
|
13:30 - 15:30 |
Computer 119 |
13:30 |
4424. |
Statistical Analysis of B1
Mapping Techniques
Daniel Joseph Park1, Ahsan Javed1,
Neal Kepler Bangerter1,2, Mohammad Mehdi
Khalighi3, and Glen R Morrell2
1Electrical and Computer Engineering, Brigham
Young University, Provo, UT, United States, 2Department
of Radiology, University of Utah, Salt Lake City, UT,
United States, 3Global
Applied Science Laboratory, GE Healthcare, Menlo Park,
CA, United States
A statistical analysis of the Bloch-Siegert (BS),
phase-sensitive (PS), dual-angle (DA), and actual flip
angle (AFI) B1 mapping techniques is presented. The
analysis employed a Monte Carlo simulation of the
statistical performance of each technique based on the
Bloch equations. The phase-sensitive B1 method is shown
to yield consistently lower mean-bias estimates of the
flip angle with a smaller standard deviation than each
of the other techniques, even in low-SNR situations.
|
14:00 |
4425. |
Theoretical and
Experimental Efficiency and Optimization of Flip Angle
Mapping Techniques
Trevor Wade1,2, Charles McKenzie2,3,
and Brian Rutt4
1Robarts Research Institute, London, Ontario,
Canada, 2Biomedical
Engineering, The University of Western Ontario, London,
Ontario, Canada, 3Medical
Biophysics, The University of Western Ontario, London,
Ontario, Canada, 4Department
of Radiology, Stanford University, Stanford, California,
United States
A theoretical propagation of noise analysis was
conducted on 3 flip angle ( )
imaging techniques (the saturated double angle method,
actual flip angle imaging, and the non-inverted double
angle Look-Locker method). This analysis was used to
predict the optimal imaging parameters, and the imaging
efficiency ( to noise
ratio (ANR) normalized by scan time). The predicted ANR
was then compared to the experimentally measured value
for a range of flip angles and T1 values.
|
14:30 |
4426. |
A short TR, MFA approach
to simultaneous B1+ and T1 mapping
Christopher Thomas Sica1, and Christopher M
Collins1
1Radiology, The Pennsylvania State
University, Hershey, Pennsylvania, United States
A novel, simultaneous B1+ and T1 mapping method is
presented here. The proposed method relies upon a series
of spoiled GRE scans acquired with a fixed TR and
increment of the flip angle between scans. A non-linear
fit is applied to the signal expression M0(1-E1)sin[λθ]
/ (1 – E1cos[λθ]) to obtain both B1+ and T1. The
proposed method is compared to the Actual Flip Angle
Imaging (AFI) B1+ mapping method. The B1+ mapping
results show excellent agreement with AFI, with error
between the two methods typically on the order of
several percent.
|
15:00 |
4427. |
B1-Mapping with the
Transient Phase of SSFP
Carl Ganter1, Marcus Settles1,
Klaus Scheffler2, and Oliver Bieri2
1Department of Radiology, Technische
Universität München, Munich, Germany, 2Division
of Radiological Physics, University of Basel Hospital,
Basel, Switzerland
In rapid SSFP sequences, the approach toward dynamical
equilibrium (transient phase) typically consists of a
smooth decay and a superimposed damped oscillation.
Recent theoretical studies for the FID signal of
unbalanced SSFP showed that the frequency of the latter
is closely connected to the actual flip angle and
depends only weakly on tissue parameters. In this work,
we show that fast and accurate B1-mapping, based on a
frequency analysis of the transient phase of SSFP, can
be implemented in a conceptually simple and flexible
manner.
|
Thursday May 12th
|
13:30 - 15:30 |
Computer 119 |
13:30 |
4428. |
Fast 3D B1 mapping
with single-shot EPI
Jay Moore1,2, Marcin Jankiewicz1,3,
Adam W Anderson1,4, and John C Gore1,4
1Institute of Imaging Science, Vanderbilt
University, Nashville, TN, United States, 2Department
of Physics and Astronomy, Vanderbilt University,
Nashville, TN, United States, 3Department
of Radiology and Radiological Sciences, Vanderbilt
University, 4Department
of Biomedical Engineering, Vanderbilt University
A workflow is demonstrated for B1 mapping
using a multi flip-angle fitting technique in
conjunction with a single-shot EPI read-out. Resulting
field maps are shown to differ minimally when compared
to the same measurements made with a low EPI factor.
With the protocol presented here, all necessary
measurements for whole-brain mapping are acquired in
less than 2 minutes.
|
14:00 |
4429. |
In-vivo RF Receiver
Sensitivity Measurement Using Phase-Based B1+ Mapping on a
Reverse-Oriented Subject
Seung-Kyun Lee1, and William Thomas Dixon1
1GE Global Research, Niskayuna, NY, United
States
A method is described to measure the receiver
sensitivity of a circularly polarized transmit and
receive RF coil by mapping the transmit RF field
strength on a reverse-oriented subject. A phase-based RF
field mapping method allows robust RF mapping unaffected
by tissue T1 or image contrast. The method is verified
on a multi-slice abdominal scan of a volunteer at 3 T.
|
14:30 |
4430. |
Multi-Slice B1+ Shimming
for 7T MRI
Andrew T Curtis1, Kyle M Gilbert1,
Martyn L Klassen1, Joseph S Gati1,
and Ravi S Menon1
1Centre for Functional and Metabolic Mapping,
University of Western Ontario, London, Ontario, Canada
Leveraging multi-channel transmit arrays to produce
efficient and uniform RF fields becomes a difficult task
as field strength increases, thanks to the well known
wave effects, and greater coupling between the coils and
the sample, especially in tight-fitting arrays. In this
situation, the traditional geometric driving phase
configuration fails to produce the expected maximum
excitation efficiency. It is demonstrated that utilizing
the extra degrees of freedom present in multi-slice
imaging protocols allows calculation of B1+ shim
solutions that can markedly reduce transmitted power
(and consequently global SAR) by shimming to recover
efficient excitation modes.
|
15:00 |
4431. |
RF Pulse Optimization for
Bloch-Siegert B1+ Mapping
Mohammad Mehdi Khalighi1, Brian K. Rutt2,
Manojkumar Saranathan2, and Adam B. Kerr3
1Global Applied Science Laboratory, GE
Healthcare, Menlo Park, CA, United States, 2Department
of Radiology, Stanford University, Stanford, CA, United
States,3Department of Electrical Engineering,
Stanford University, Stanford, CA, United States
B1+ mapping
by the Bloch-Siegert (BS) method has been shown to be
fast and accurate; however, at high field it suffers
from high SAR and long TE, which results in lengthened
scan times and signal loss due to B0 inhomogeneity.
We have designed a new BS RF pulse that can be applied
closer to the water resonance yet still limits
excitation within the on-resonant band of interest.
Comparison of BS B1+ mapping
using the new pulse versus the conventional Fermi pulse
in both phantoms and human brain shows improved B1+ map
quality with substantial reductions in both TE and SAR.
|
|
|
Electronic
Posters
: Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
Tailoring Excitation with Parallel Transmission & Advanced
Pulse Design
Monday May 9th
Exhibition Hall |
14:00 - 16:00 |
Computer 120 |
14:00 |
4432. |
Relaxation-Enhanced
Multiple Inner-Volume Imaging Using Parallel 3D
Spatially Selective Excitation
Johannes Thomas Schneider1,2, Martin Haas2,
Wolfgang Ruhm1, Juergen Hennig2,
and Peter Ullmann1
1Bruker BioSpin MRI GmbH, Ettlingen,
Germany, 2Dept.
of Radiology, Medical Physics, University Medical
Center Freiburg, Freiburg, Germany
Recently, first examples of 3D Parallel Spatially
Selective Excitation (PEX) have been presented. This
technique allows the restriction of the generated
transverse magnetization to specified 3D target
volumes, while outside these volumes the
magnetization is fully restored into the
longitudinal direction by the end of the pulse.
Exploiting this feature, Multiple Inner-Volume
Imaging (MIVI) allows interleaved excitation and
data acquisition of multiple different inner-volumes
within one repetition time. Compared to successive
imaging of multiple inner-volumes in individual
experiments, MIVI results in reduced scan times
and/or enhanced SNR due to prolonged T1-relaxation
periods.
|
14:30 |
4433. |
Selective Excitation
of Arbitrary Three-Dimensional Targets on a Human MR
System using Parallel Transmit
Martin Haas1, Jeff Snyder1,
Johannes T Schneider1,2, Peter Ullmann2,
Denis Kokorin1,3, Hans-Peter Fautz4,
Jürgen Hennig1, and Maxim Zaitsev1
1Department of Radiology Medical Physics,
University Medical Center Freiburg, Freiburg,
Germany, 2Bruker
BioSpin MRI GmbH, Ettlingen, Germany, 3International
Tomography Center, Novosibirsk, Russian Federation, 4Siemens
Healthcare, Erlangen, Germany
Spatially selective excitation of an arbitrarily
shaped three-dimensional target volume is
demonstrated on a human MR scanner at 3T,
accelerated by means of 8-channel parallel
transmission. The good excitation fidelity and
effective restoration of magnetization outside the
volume of interest to the longitudinal direction
allows for a reduction of the readout field of view
in all three dimensions, resulting in an increase of
resolution by a factor of two in all three
directions while maintaining the measurement time.
|
15:00 |
4434. |
Sparse Parallel
Transmit Excitation Trajectory Design for Rapid
Inner-Volume Excitation
Cem Murat Deniz1,2, Dong Chen3,
Leeor Alon1,2, Ryan Brown1,
Hans-Peter Fautz4, Daniel K Sodickson1,
and Yudong Zhu1
1Center for Biomedical Imaging,
Department of Radiology, NYU School of Medicine, New
York, NY, United States, 2Sackler
Institute of Graduate Biomedical Sciences, NYU
School of Medicine, New York, NY, United States, 3Center
for Mathematical Science, Technical University of
Munich, Munich, Germany, 4Siemens
Medical Solutions, Erlangen, Germany
Tailored inner-volume excitation on whole-body
scanners is currently limited by long 3D RF pulses.
Effective pulse length reduction with parallel
transmission requires careful selection of the
excitation k-space trajectory. In this work, two
methods of determining sparse excitation
trajectories were compared for parallel transmit
pulse design in the small-tip angle and
large-tip-angle regimes: a) an Orthogonal Matching
Pursuit (OMP) algorithm, and b) a one-step basic
thresholding approach. Good inner-volume excitation
with a pulse length of less than 9ms was achieved
using an eight-channel transmitter on a whole-body
human 7T scanner.
|
15:30 |
4435. |
Volume Localization
using Adiabatic Inversion Pulses in FAIR Imaging
Ziqi Sun1, Sergey Petryakov1,
George Caia1, Alex Samouilov1,
and Jay L Zweier1
1Davis Heart & Lung Research Institute,
The Ohio State University, Columbus, Ohio, United
States
Adiabatic inversion pulses have been widely used in
slice- and whole-volume spin inversion in
perfusion-weighted MRI. To date, no reports are
available for using selective adiabatic inversion
pulses for localized volume inversion in perfusion
measurements. In this study, we proposed a 3D volume
localization using adiabatic inversion pulses for
FAIR imaging on a flow phantom. In comparison to the
conventional slice-inversion FAIR technique, the
perfusion rate measured using the localized
volume-inversion FAIR method is in the acceptable
accuracy and generates enhanced T1-weighting
contrast.
|
Tuesday May 10th
|
13:30 - 15:30 |
Computer 120 |
13:30 |
4436. |
Large Tip Angle
Segmented RF Design for Multi-Dimensionally Selective
Imaging and Spectroscopy with Parallel Transmit
Martin Haas1, Jeff Snyder1,
Peter Ullmann2, Jürgen Hennig1,
and Maxim Zaitsev1
1Department of Radiology Medical Physics,
University Medical Center Freiburg, Freiburg,
Germany, 2Bruker
BioSpin MRI GmbH, Ettlingen, Germany
Multi-dimensionally selective RF excitation suffers
from artifacts related to the typically long pulse
duration. In particular, applications in MR
spectroscopy would significantly benefit from
shorter pulses due to increased bandwidth. Previous
approaches to mitigate this problem using segmented
selective excitation across several repetitions have
been limited to small tip angles. In this work, a
segmented RF design algorithm for large tip angle
pulses, with additional parallel transmit
acceleration, is described for multi-dimensionally
selective excitation and demonstrated in simulations
and a phantom. The method is based on a segment-wise
optimal control optimization and not limited to 2D
or a particular segmentation.
|
14:00 |
4437. |
Flexibly shaped
saturation band excitation using 7T parallel transmit
system
Borjan Gagoski1, Khaldoun Makhoul2,3,
Dieter Ritter4, Kawin Setsompop2,3,
Josef Pfeuffer4, Himanshu Bhat5,
Philipp Hoecht5, Michael Hamm5,
Ulrich Fontius4, Lohith Kini1,
Joonsung Lee1, Lawrence L Wald2,6,
and Elfar Adalsteinsson1,6
1Electrical Engineering and Computer
Science, Massachusetts Institute of Technology,
Cambridge, MA, United States, 2A.A.
Martinos Center for Biomedical Imaging, Department
of Radiology, Massachusetts General Hospital,
Charlestown, MA, United States, 3Harvard
Medical School, Boston, MA, United States, 4Siemens
Healthcare, Erlangen, Germany, 5Siemens
Healthcare, Charlestown, MA, United States, 6Harvard-MIT
Division of Health Sciences and Technology, MIT,
Cambridge, MA, United States
Parallel RF transmission (pTx) offers significant
flexibility over single-channel excitations for
creating arbitrarily shaped spatial excitation
patterns, which can be designed for undersampled
spiral-based excitation k-space trajectories played
during the pTx RF waveform. In this work we validate
the feasibility of using 7T 8-channel pTx for
flexibly shaped saturation bands that suppresses a
user-defined spatial shape, in combination with a
B1+ mitigated pTx slice-selective "spokes"
excitation.
|
14:30 |
4438. |
In vivo Zoom Imaging
using Transmit SENSE
Ingmar Graesslin1, Sebastian Boetzl1,
Ulrich Katscher1, Kay Nehrke1,
Bjoern Annighoefer2, Giel Mens3,
and Peter Börnert1
1Philips Research Laboratories, Hamburg,
Germany, 2TU
Hamburg-Harburg, Hamburg, Germany, 3Philips
Healthcare, Best, Netherlands
Zoom imaging allows an increase of the spatial
resolution in the targeted region of interest (ROI)
without increasing the scan time. Alternatively,
maintaining spatial resolution, zoom imaging allows
a reduction of the scanning time by reducing the
field of view (FOV) to the ROI. For both techniques,
potential backfolding artefacts from body regions
outside the reduced FOV can be suppressed via local
excitation. This local excitation can be performed
by multi-dimensional RF pulses, which can be
accelerated using spatial transmit sensitivity
encoding instead of full gradient encoding. This
paper presents (first) zoom imaging volunteer
experiments carried out on an 8-transmit channel 3T
MRI system, with a fully integrated real-time SAR
validation prior to the scan allowing the safe use
of spatially selective Transmit SENSE RF pulses and
their fast calculation for optimal workflow.
|
15:00 |
4439. |
Practical
Considerations for the Design of Parallel Transmission
Pulses at ultra high field
Tiejun Zhao1, Hai Zheng2,
Yik-Kiong Hue3, Tamer Ibrahim2,3,
Yongxian Qian3, and Fernando Boada2,3
1Siemens Medical Solutions, Pittsburgh,
Pennsylvania, United States, 2Bioengineering,
University of Pittsburgh, Pittsburgh, Pennsylvania,
United States, 3Radiology,
University of Pittsburgh, Pittsburgh, Pennsylvania,
United States
Parallel transmission was recently proposed as one
of the methods to accelerate multidimensional
selective excitation using multiple coils driven
with independent waveforms. The parallel excitation
could be a potential solution for mitigating or
solving the RF and SAR problems at the ultra-high
field (e.g, 7T). One of the popular methods was
proposed by Grissom et al. In this work, we propose
two novel methods that the artifacts due to finite
gradient raster step can be successfully compensated
to avoid tilted and excitation errors. Significant
improvement can be obviously seen from both
simulations and experiments on 7T scanner.
|
Wednesday May 11th
|
13:30 - 15:30 |
Computer 120 |
13:30 |
4440. |
Characterization and
correction of eddy currents for ultra high field
parallel transmission with RF pulse design
Hai Zheng1, Tiejun Zhao2,
Yongxian Qian3, Tamer Ibrahim1,3,
and Fernando Boada1,3
1Bioengineering, University of
Pittsburgh, Pittsburgh, Pennsylvania, United States, 2Siemens
Medical Solutions, Pittsburgh, Pennsylvania, United
States, 3Radiology,
University of Pittsburgh, Pittsburgh, Pennsylvania,
United States
Hardware and experimental imperfections can severely
impact the excitation patterns in ultra high field
parallel transmission. In this work, we propose a
new eddy current correction method combining
existing eddy current characterization method for
parallel transmission RF pulse design to compensate
the overall distortion induced by gradient and RF
coils as well as the system delay to reduce the
distortions in parallel transmission. Our proposed
compensation method is a straightforward and easy to
implement, but yields an evident improvement in the
performance.
|
14:00 |
4441. |
Parallel Transmission
in Human Brain at 9.4T Counteracting Eddy Current
Induced Excitation Errors in RF Pulse Design
Xiaoping Wu1, Gregor Adriany1,
Kamil Ugurbil1, and P-F. Van de Moortele1
1CMRR, Radiology, University of
Minnesota, Minneapolis, MN, United States
|
14:30 |
4442. |
An Interleaved
Spatial-Spectral Pulse for Imaging Large Chemical-shift
Components
Jing Chen1, Jing An2, and Yan
Zhuo1
1State Key Laboratory of Brain and
Cognitive Science, Inst. of Biophysics, Chinese
Academy of Sciences, Beijing, China, People's
Republic of, 2Siemens
Healthcare, MR Collaboration NE Asia, Siemens Mindit
Magnetic Resonance, China, People's Republic of
This work describes an interleaved spatial-spectral
RF pulse for use with large excitation/suppression
difference frequency, for example, exciting water
while suppressing fat on a 7T system. The excitation
is divided into three repetitions to fill the
excitation k-space in an interleaving pattern.
Interleaving allows driving the gradient at a lower
slew rate. Therefore, this method has the advantages
of less eddy current distortion, relative ease of
implementation, and thinner slice thickness.
|
15:00 |
4443. |
RF energy reduction by
parallel transmission using large-tip-angle composite
pulses
Rene Gumbrecht1,2, Elfar Adalsteinsson3,4,
Paul Müller2, and Hans-Peter Fautz1
1Siemens Healthcare, Erlangen, Germany, 2Department
of Physics, Friedrich-Alexander University,
Erlangen, Germany, 3Electrical
Engineering and Computer Science, Massachusetts
Institute of Technology, Cambridge, MA, United
States, 4Harvard-MIT
Division of Health Sciences and Technology,
Massachusetts Institute of Technology, Cambridge,
MA, United States
For ultra high field strengths of 7T and above, peak
RF power and SAR are limiting factors for high
performance imaging. One way to keep these
quantities low is to prolong the RF pulses. However,
long RF pulses are prone to B0 homogeneities because
the bandwidth of such pulses is reduced. In this
work, we investigate the potential of prolonging
large-tip-angle composite RF pulses for power and
SAR reduction by using pTx pulse design methods
which account for local B0 field variations.
|
Thursday May 12th
|
13:30 - 15:30 |
Computer 120 |
13:30 |
4444. |
B1 inhomogeneity
mitigation in the human brain at 7T with selective
pulses by using Average Hamiltonian Theory
Nicolas Boulant1, Martijn Cloos1,
and Alexis Amadon1
1NeuroSpin, CEA Saclay, Saclay, France
The design of the strongly modulating pulses,
developed by the same authors and initially
non-selective, has been adjusted to create selective
pulses by using average Hamiltonian theory. Such a
theory formulates an approximate analytical
expression of the dynamics which allows navigating
quickly in some pulse parameter space. Once a pulse
is found to perform sufficiently well using that
approximation, it can be modified with a few
iterations by calculating more accurately the spin
evolution. Using that technique, in-vivo results are
reported which show good mitigation of the B1
inhomogeneity problem over an axial slice in the
human brain at 7T.
|
14:00 |
4445. |
Non-Slice Selective
Uniform Tipping RF Pulse Design for 3D MRI at High Field
Hui Liu1,2, and Gerald B. Matson1,3
1Center for Imaging of Neurodegenerative
Diseases (CIND), Veterans Affairs Medical Center,
San Francisco, CA, United States, 2Northern
California Institute for Research and Education, San
Francisco, CA, United States, 3University
of California, San Francisco,CA, United States
Although high-field MRI offers increased
signal-to-noise (SNR), the non-uniform tipping
produced by conventional RF pulses leads to
spatially varying contrast such as a bright center,
and sub-optimal S/N, thus complicating the
interpretation of the MR images. The aim of this
research was to develop non-slice-selective (NSS) RF
pulses with immunity to B1 inhomogeneity and
resonance offset for a full range of tip angles. To
accomplish this, we developed an optimization
routine to design RF pulses with a desired range of
immunity to B1 inhomogeneity and to resonance
offset. Simulations were validated by phantom tests
and an in vivo human study. The resulting pulses
were more efficient (in terms of length) than
previous pulses in the literature. These pulses have
promise for 3D MRI experiments at high field.
|
14:30 |
4446. |
T2-Weighting
Enhancement using Pseudo-Echoes Generated by Selective
Adiabatic Refocusing Pulses in a CPMG Pulse Sequence
Ziqi Sun1
1Davis Heart & Lung Research Institute,
The Ohio State University, Columbus, Ohio, United
States
Selective adiabatic excitation pulses have been
utilized to generate pseudo-echoes to improve image
contrast and quality. In this study, pseudo-echo
images were obtained using a customized CPMG pulse
sequence incorporated with selective adiabatic
refocusing pulses that alternate frequency sweep
directions between each of the echoes. Apparent T2
time constants were measured from the pseudo-echo
images using the customized CPMG sequence.
T2-weighting was significantly increased using the
sequence in comparison to the CPMG sequence using
amplitude modulated selective refocusing pulses.
|
15:00 |
4447. |
Fast Spin Echo Imaging
with Quadratic Phase-Modulated non-CPMG Echo Train in
Parallel Transmit – a Simulation Study
Seung-Kyun Lee1, Mika W Vogel2,
William A Grissom2, Graeme C McKinnon3,
and Patrick H Le Roux4
1GE Global Research, Niskayuna, NY,
United States, 2Advanced
Medical Applications Laboratory, GE Global Research,
Munich, Bavaria, Germany, 3Applied
Science Lab, GE Healthcare, Waukesha, WI, United
States, 4Applied
Science Lab, GE Healthcare, Palaiseau, France
In parallel transmit, phase-relaxed RF pulse design
significantly improves image quality in gradient
echo and spin echo images. Here we present a
simulation study showing that phase-relaxed 90-180
RF pulses can also be used in fast spin echo imaging
with Le Roux’s quadratic phase modulation
implemented on all transmit channels.
|
|
|
Electronic
Posters
: Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
Quantitative MRI
Monday May 9th
Exhibition Hall |
14:00 - 16:00 |
Computer 121 |
14:00 |
4448. |
Experimental
Evaluation of RF Non-uniformity Correction in the
Mapping of the Proton Density
Vincent Gras1, Zaheer Abbas1,
and Nadim Jon Shah1,2
1Institute of Neuroscience and Medicine
4, Medical Imaging Physics, Forschungszentrum Jülich
GmbH, Jülich, Germany, 2Faculty
of Medicine, Department of Neurology, RWTH Aachen
University, Aachen, Germany
Mapping of the proton density (PD) is a common MRI
application, with interesting perspective in medical
research. At 3 Tesla, PD imaging supposes an
accurate correction of the sample-dependent RF
non-uniformity, which is measured using a dedicated
protocol termed B1 mapping. Such measurement is
possible if one assumes 1) the linear relationship
between flip angle and B1 and 2) the NMR reciprocity
principle. From these 2 assumptions, it follows a
unique correction scheme. In this work, the accuracy
of the correction is evaluated to 2%, setting an
upper limit to the precision of the PD measurement
achievable with this general approach.
|
14:30 |
4449. |
Quantitative Water
Content Mapping at 1.5 and 3 Tesla Field Strength
Vincent Gras1, Zaheer Abbas1,
Anna-Maria Oros-Peusquens1, Klaus Hans
Manfred Möllenhoff1, Fabian Keil1,
Miriam Rabea Kubach1, and Nadim Jon Shah1,2
1Institute of Neuroscience and Medicine
4, Medical Imaging Physics, Forschungszentrum Jülich
GmbH, Jülich, Germany, 2Faculty
of Medicine, Department of Neurology, RWTH Aachen
University, Aachen, Germany
A 1.5T and 3T quantitative water content imaging
protocol is proposed, extending a previously
developed method, suited to 1.5T only. The benefit
of working at 3T is a noticeable gain in SNR
compared to 1.5T, although refinements are needed
for the accuracy to be preserved. The gradient echo
approach, envisaged in this work, allows for the
detection of myelin water, exhibiting short
relaxation times. Furthermore, the required short
scanning time, 15 minutes, is particularly suited to
the clinical practice. The accuracy of the method is
evaluated to 2%, making this method eligible for the
detection of cerebral oedema of various origins.
This method is presented and in vivo results at 1.5T
and 3T are shown.
|
15:00 |
4450. |
Quantitative Magnetic
Resonance Imaging in Light-Chain (AL) Amyloidosis:
Preliminary Experience
Stephan William Anderson1, Jennifer
Ellis-Ward2, Erskine Hawkins3,
James A Hamilton4, Carl J O'Hara5,
Lawreen H Connors6, Jorge A Soto1,
David C Seldin2, and Hernan Jara1
1Radiology, Boston University Medical
Center, Boston, MA, United States, 2Hematology
and Medical Oncology, Boston University Medical
Center, 3Boston
University School of Medicine, 4Physiology
and Biophysics, Boston University Medical Center, 5Pathology
and Laboratory Medicine, Boston University Medical
Center, 6Biochemistry,
Boston University School of Medicine
Purpose: To evaluate the effects of amyloid
deposition on T2 and apparent diffusion coefficient
(ADC) measurements in human autopsy specimens of
myocardium, liver, spleen, and kidney of patients
with AL amyloidosis using 11.7T MRI. Methods: T2 and
ADC values of tissues with amyloid deposition were
compared to control specimens. Results Significant
differences in T2 and ADC values were seen when
comparing control and involved specimens of
myocardium, spleen and kidney (all p<0.002).
Conclusion: Deposition of amyloid protein in human
autopsy specimens substantially affects ADC values
in myocardium, liver, spleen, and kidney and T2
values in myocardium, spleen and kidney.
|
15:30 |
4451. |
Characterization of
Modified Look Locker (MOLLI) using Bloch simulations and
corroboration with scan measurements
Neville D Gai1, Christian Stehning2,
Marcelo Nacif1, and David A Bluemke1,3
1Radiology & Imaging Sciences, National
Institutes of Health, Bethesda, MD, United States, 2Philips
Research Europe, Hamburg, Germany, 3NIBIB,
Bethesda, MD, United States
The MOLLI technique for T1 mapping has found favor
in cardiac imaging due to its relative robustness to
motion. A subsequent work explored optimization of
parameters based purely on performing phantom scans
while varying scan parameters. Such an approach
could fail to be comprehensive in the covered
parameter space or parameters affecting accuracy.
Here, we develop Bloch simulation based
characterization of MOLLI and evaluate T1 accuracy
for several scan parameters. We corroborate T1
values obtained from simulation with values obtained
from scanning phantoms. Based on simulations and
scanning, it is shown that the simulation is a
valuable tool in understanding the accuracy of T1
values given different scan conditions.
|
Tuesday May 10th
|
13:30 - 15:30 |
Computer 121 |
13:30 |
4452. |
Comparison of
Different EPI-based Approaches to Measure T2’ in Human
Brain for the Purpose of Oxygenation Measurements
Thomas Christen1, Heiko Schmiedeskamp1,
Matus Straka1, Roland Bammer1,
and Greg Zaharchuk1
1Department of radiology, Stanford
University, Stanford, California, United States
We compared in this study three different EPI-based
approaches for measuring T2’ : (1) a subtraction of
T2 from T2* relaxation rates that are derived from
separate multiecho sequences; (2) an asymmetric spin
echo (ASE) sequence; (3) a combined multiple
gradient echo and spin echo (SAGE) sequence. The
results obtained in the brain of 4 subjects suggest
that the different approaches give similar T2’
values. They however showed spatial differences. The
highest-quality maps were obtained with the ASE
method. The SAGE approach, while containing spatial
artifacts, could be used in a dynamic approach with
high time resolution.
|
14:00 |
4453. |
On the T1 of fat
calculated from a segmented Look Locker scout scan and
its implications in cardiac imaging
Neville D Gai1, Christian Stehning2,
Saman Nazarian3, Evrim Turkbey1,
and David A Bluemke1,4
1Radiology & Imaging Sciences, National
Institutes of Health, Bethesda, MD, United States, 2Philips
Research Europe, Hamburg, Germany, 3Division
of Cardiology, Johns Hopkins University, Baltimore,
United States, 4NIBIB,
Bethesda, MD, United States
T1 mapping can differentiate between diffuse
fibrosis and normal myocardium. Several longitudinal
cardiac studies do not have a dedicated T1 mapping
scan as part of the protocol. However, segmented
Look-Locker (LL) with b-SSFP is typically used as a
scout scan. T1 values may then be derived from such
a scan. Certain diseases exhibit fatty infiltration
as part of their etiology. Based on simulations and
scanning, we show that fat T1 quantification so
obtained can be highly erroneous based on the scan
parameters used. Therefore, conclusions based on
observed cardiac T1 values in cases where fatty
infiltration is substantial might be fraught with
complications.
|
14:30 |
4454. |
Accurate T1 Measurement
with IR-prepared Segmented Gradient Echo and A New
Regression Algorithm
Haosen Zhang1, Kevin Hitchens1,
Qing Ye1, EriK B. Schelbert2,
and Chien Ho1
1Pittsburgh NMR Center for Biomedical
Research, Department of Biological Science, Carnegie
Mellon University, Pittsburgh, PA, United States, 2Department
of Medicine, University of Pittsburgh, Pittsburgh,
PA, United States
In this study, an inversion recovery (IR)-prepared
segmented gradient echo sequence was developed to
sample the T1 recovery curve in Look-Locker scheme
at 7 T. Via a multi-variable regression algorithm
corrected for the saturation effect induced by the
á-train, we have obtained more accurate T1 values
than with the conventional three-parameter fit
(average error of -0.53% vs. -5.4% separately,
compared to T1s measured with IR-SE) in agarose
phantoms simulating human myocardium and blood.
|
15:00 |
4455. |
Non-Exponential T2* Decay
in White Matter
Peter van Gelderen1, Jacco A de Zwart1,
Jongho Lee1, Pascal Sati2,
Daniel S Reich2, and Jeff H Duyn1
1Advanced MRI section, LFMI, NINDS,
National Institutes of Health, Bethesda, MD, United
States, 2Translational
Neuroradiology Unit, Neuroimmunology Branch, NINDS,
National Institutes of Health, Bethesda, MD, United
States
A number of methods have been proposed to determine
the myelin distribution in human brain, which may
report on pathology. One potentially promising
method is measurement of the T2* relaxation
curve ,which may be affected by restricted mobility
of myelin associated water. Here we investigated
this possibility using multi-gradient echo MRI at 3T
and 7T in normal brain. In myelin-rich regions, we
observed relaxation characteristics more consistent
with effects from susceptibility variations rather
than from mobility restrictions. This may complicate
the interpretation, as susceptibility effects are
affected by additional factors independent of myelin
content.
|
Wednesday May 11th
|
13:30 - 15:30 |
Computer 121 |
13:30 |
4456. |
Fast Radio-frequency
Enforced Steady State (FRESS) Spin Echo MRI for
Quantitative T2 Mapping
Jerry S. Cheung1, Enfeng Wang1,2,
XiaoAn Zhang2, Emiri Mandeville3,
Eng H. Lo3, A. Gregory Sorensen1,
and Phillip Zhe Sun1
1Athinoula A. Martinos Center for
Biomedical Imaging, Department of Radiology, MGH and
Harvard Medical School, Charlestown, MA 02129,
United States, 2Department
of Radiology, 3rd Affiliated Hospital, Zhengzhou
University, China, People's Republic of, 3Neuroprotection
Research Laboratory, Department of Radiology and
Neurology, MGH and Harvard Medical School,
Charlestown, MA 02129, United States
Transverse relaxation time (T2) is a
basic but very informative MRI parameter, widely
used in imaging to examine a host of diseases
including multiple sclerosis, stroke, and tumor.
When the repetition time (TR) is very long, T2 can
be derived by fitting T2-weighted images
as a function of echo time (TE). However, short TR
is often used to minimize scan time, which may
introduce non-negligible errors in T2 measurement.
Our study proposed a fast RF-enforced steady state (FRESS)
spin echo MRI sequence, which saturates the
magnetization after the spin echo and ensures a
TE-independent steady state for accurate T2 mapping.
|
14:00 |
4457. |
Quantitative T1 estimation
using Tissue Specific Imaging
Arezou Koohi1, and Vasiliki N Ikonomidou1
1Electrical and Computer Engineering,
George Mason University, Fairfax, VA, United States
This study presents the application of Tissue
Specific Imaging (TSI), a double inversion recovery
variant that produces three single tissue type
images, to the quantitative mapping of tissue T1 values.
The methodology presented, which can add value to
the application of TSI, is stable over a wide range
of T1 values,
up to 4500 ms for usually encountered
signal-to-noise ratio values.
|
14:30 |
4458. |
Single-slice mapping
of submillisecond T2 using
spin echo prepared ultra-short echo time imaging
Stefan Kirsch1, and Lothar R Schad1
1Computer Assisted Clinical Medicine,
Heidelberg University, Mannheim, Germany
Measurement of the transversal relaxation time T2 is
one of the most established techniques to
characterize material samples or biological tissue
by means of NMR. If molecular motion is restricted,
direct spin-spin interactions become relevant and
extremely short T2 values
are expected. Here we present a slice-selective MRI
method for mapping of submillisecond T2.
The method utilizes a spin echo preparation followed
by slice-selective ultra-short echo time imaging.
Mapping of submillisecond T2 could
be useful in studies on material samples, short T2 biological
tissue like bone, tendon or cartilage and on
quadrupolar nuclei like 23Na, 35Cl,
and 17O.
|
15:00 |
4459. |
Effect of the Slice
Profile on the T1 Measurement
with Steady-State Magnetization
Jung-Jiin Hsu1
1Radiology, University of Miami School of
Medicine, Miami, Florida, United States
The accuracy of the T1 measurement
depends on the accuracy of the flip angle (FA). The
FA is non-uniform in a slice across the
slice-thickness direction, which can result in
serious T1 measurement
error. Although it has been recognized that
measuring the FA regional inhomogeneity in a volume
is important, the effect of the slice profile is
still less known. In this work, the slice-profile
effect is studied systematically for various common
experimental conditions. The results provide
essential guidelines for experiment design and scan
parameter selection for the T1 measurement
with steady-state magnetization (e.g., DESPOT).
|
Thursday May 12th
|
13:30 - 15:30 |
Computer 121 |
13:30 |
4460. |
T2* myelin water
imaging with bmGESEPI for macroscopic field
inhomogeneity compensation
Yoonho Nam1, Eung-Yeop Kim2,
Dosik Hwang1, and Dong-Hyun Kim1
1Electrical & Electronic Engineering,
Yonsei University, Seoul, Korea, Republic of, 2Radiology,
Yonsei University, Seoul, Korea, Republic of
Myelin water imaging is a useful tool for studying
white matter diseases. So far, a multi-exponential
T2 analysis using multi-echo spin echo sequence has
been mostly used for myelin water imaging. But,
using a multi-echo spin echo sequence has some
limitations such as high SAR, small coverages.
Therefore, several studies have been conducted using
multi-echo gradient echo sequences recently.
However, T2* decay curve measured by gradient echo
sequence can be easily distorted by macroscopic
field inhomogeneity. In this study, the bmGESEPI
method was applied for myelin water imaging to
remove the effects of macroscopic field
inhomogeneity.
|
14:00 |
4461. |
Simulation of the
Filtering Effect of the FLASH Readout on Saturation
Recovery T1 Evaluation
Moritz Cornelius Berger1, Wolfhard
Semmler1, and Michael Bock1
1Medical Physics in Radiology, German
Cancer Research Center (DKFZ), Heidelberg, Germany
Fast saturation recovery turboFLASH imaging
introduces a saturation-time dependent filter on the
magnetization-prepared k-space. To mitigate these
filtering effects for reliable T1 measurements,
multiple k-space segmentation schemes were simulated
to characterize the tradeoff between the accuracy of
T1 determination and total measurement time.
|
14:30 |
4462. |
Rapid T2 Mapping
of Mouse Heart Using CPMG Sequence and Model-based
Compressed Sensing Reconstruction
Yong Chen1,2, Wen Li1,2, and
Xin Yu1,2
1Department of Biomedical Engineering,
Case Western Reserve Univ, Cleveland, OH, United
States, 2Case
Center for Imaging Research, Case Western Reserve
Univ, Cleveland, OH, United States
In this study, we developed a novel method for rapid
T2 mapping of mouse heart in vivo. Our results
demonstrate that a high temporal resolution of ~15
seconds can be achieved by combining a fast
multi-echo spin-echo sequence with a model-based
compressed sensing reconstruction.
|
15:00 |
4463. |
Multi-slice
Look-Locker T1 mapping
for the mouse heart
Adrienne E Campbell1,2, Anthony N Price3,
Bernard M Siow1, Jack A Wells1,
Mark F Lythgoe1, and Roger J Ordidge2
1Centre for Advanced Biomedical Imaging,
Division of Medicine and Institute of Child Health,
University College London, London, United Kingdom, 2Department
of Medical Physics and Bioengineering, University
College London, London, United Kingdom, 3Robert
Steiner MRI Unit, Imaging Science Department,
Hammersmith Hostpital, Imperial College London,
London, United Kingdom
A time-efficient multi-slice ECG gated Look-Locker
sequence is developed for multi-slice T1 mapping in
the mouse heart. With this sequence, multi-slice T1
maps can be generated for the mouse heart in less
than 10 minutes. Multi-slice T1 mapping was tested
in phantoms and in a CD-1 mouse, and it was found
that T1 measurements agreed with single-slice T1
measurements within 4% for the phantom and within
10% in vivo. This sequence will have great
applicability for multi-slice arterial spin
labelling perfusion measurements to study cardiac
disease models.
|
|
|
Electronic
Posters
: Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
Electromagnetic Tissue Property Mapping
Monday May 9th
Exhibition Hall |
14:00 - 16:00 |
Computer 122 |
14:00 |
4464. |
Electrical conductivity
imaging of brain tumours.
Astrid L.H.M.W. van Lier1, Johannes M.
Hoogduin2, Daniel L. Polders2,
Vincent O. Boer2, Jeroen Hendrikse2,
Pierre A. Robe3, Peter A. Woerdeman3,
Jan J.W. Lagendijk1, Peter R. Luijten2,
and Cornelis A.T. van den Berg1
1Radiotherapy, UMC Utrecht, Utrecht,
Netherlands, 2Radiology,
UMC Utrecht, Utrecht, Netherlands, 3Neurosurgery,
UMC Utrecht, Utrecht, Netherlands
Electrical conductivity mapping is a new technique to
generate MRI contrast. Over the last years, the theory
and practical implementation of the method has been
shown, and measurements of phantoms and healthy
volunteers were presented. Here we present for the first
time conductivity mapping in brain tumour patients.
Distinct differences between the conductivity of some
tumours compared to surrounding tissue were observed.
The potential of this conductivity mapping for tumour
detection, characterization and staging will be further
investigated.
|
14:30 |
4465. |
Electrical impedance
tomography using magnetic resonance as the voltage source
Michiro Negishi1, Tangji Tong1,
Peter Brown1, Terrence Nixon1, and
R Todd Constable1,2
1Diagnostic Radiology, Yale University, New
Haven, CT, United States, 2Neurosurgery,
Yale University, New Haven, CT, United States
We present a novel impedance imaging method called
Magnetic Resonance Driven Electrical Impedance
Tomography (MRDEIT), where magnetic resonance is used to
apply voltages and the signals from surface electrodes
or surface RF detectors are analyzed to obtain the
complex permittivity distribution in the target volume.
In the current study, we test whether the RF voltages
measured by surface electrodes changes when the
conductance of the phantom is altered as the theory
predicts. The experiment results showed that a higher
conductivity results in a higher average electrode
voltage and that the spatial voltage profile from the
Finite Element Method based simulations.
|
15:00 |
4466. |
In vivo conductivity
mapping using double spin echo for flow effect removal
Narae Choi1, Minoh Ghim1,
Seungwook Yang1, Sang-Young Zho1,
and Dong-Hyun Kim1,2
1Electrical and Electronic Engineering,
Yonsei University, Sinchon dong, Seoul, Korea, Republic
of, 2Radiology,
Yonsei University, Sinchon dong, Seoul, Korea, Republic
of
The electric properties tomography (EPT) is used to
evaluate tissue conductivity in vivo by analyzing B1
map. The B1 map, however, is hampered by attenuation of
signal amplitude from the flow within a voxel. This
study focuses on the reconstruction of in vivo
conductivity map using double spin echo signal for flow
compensation.
|
15:30 |
4467. |
Rapid estimation of
conductivity and permittivity using Bloch-Siegert B1 mapping
at 3.0T
Selaka Bandara Bulumulla1, Seung-Kyun Lee1,
Teck Beng Desmond Yeo1, W Thomas Dixon1,
and Thomas K Foo1
1GE Global Research, Niskayuna, New York,
United States
Tissue conductivity and permittivity are critical to
estimating local RF power deposition in the human body
during MR imaging. These electrical properties may also
have diagnostic value as malignant tissue types have
shown higher permittivity and conductivity than
surrounding healthy tissue. In this work, we explore
rapid estimation of conductivity and permittivity using
Bloch-Siegert B1 mapping at 3.0T. In a 24cm axial plane
FOV, 128x128 resolution, permittivity was obtained in
approximately 1 ½ min. Obtaining conductivity required
an additional 1min.
|
Tuesday May 10th
|
13:30 - 15:30 |
Computer 122 |
13:30 |
4468. |
MREIT and EPT: a
comparison of two conductivity imaging modalities
Dong-Hyun Kim1, Min-oh Ghim1, Ohin
Kwon2, Hyung Joong Kim3, Jin Keun
Seo4, and Eung Je Woo3
1Electrical and Electronic Engineering,
Yonsei University, Seoul, Korea, Republic of, 2Mathematics,
Konkuk University, Korea, Republic of, 3Biomedical
Engineering, Kyung Hee University, Korea, Republic of, 4Mathematics,
Yonsei University, Seoul, Korea, Republic of
Magnetic resonance electrical impedance tomography
(MREIT) and MR electrical properties tomography (MREPT)
are medical imaging modalities capable of visualizing
the electrical permittivity and conductivity
distribution of conducting objects at different
operating frequencies. MREIT provides conductivity
information at low frequency while MREPT is related to
the conductivity at the Larmor frequency. Here both
experiments are carried out to show the contrast
differences in the two imaging methods.
|
14:00 |
4469. |
Mechanism of Conductivity
Image Contrast in MREIT: Numerical Simulation and Phantom
Experiment
Young Tae Kim1, Tong In Oh1, Atul
Singh Minhas1, Hyung Joong Kim1,
Jin Keun Seo2, Oh In Kwon3, and
Eung Je Woo1
1Biomedical Engineering, Kyung Hee
University, Yongin, Gyeonggi, Korea, Republic of, 2Computational
Science and Engineering, Yonsei University, Seoul,
Korea, Republic of,3Mathematics, Konkuk
University, Seoul, Korea, Republic of
MREIT utilizes MR phase images affected by externally
injected currents to reconstruct conductivity images.
Understanding its contrast mechanism could be difficult
due to complexities in associated mathematical
expressions of bioelectromagnetism. We constructed
stable conductivity phantoms not affected by the ion
diffusion process placing thin insulating hollow
cylinders with holes of different diameters inside a
saline tank. When we extracted induced magnetic flux
density images from MR phase images, their slopes
changed with conductivity contrasts. Analyzing measured
and computed magnetic flux density images of the
phantoms, we could quantitatively validate the contrast
mechanism in MREIT.
|
14:30 |
4470. |
Quantitative
Susceptibility Imaging using L1 regularized reConstruction
with Sparsity Promoting Transformation: SILC
Deqiang Qiu1, Greg Zaharchuk1,
Shangping Feng1, Thomas Christen1,
Kyunghyun Sung1, and Michael E. Moseley1
1Lucas Imaging Center, Stanford University,
Stanford, CA, United States
We describe a novel method (SILC) for reconstructing
susceptibility distribution from phase maps using L1
regularized iterative method with a sparsity promoting
transformation. Both simulations and application to in
vivo human brain imaging are presented. The SILC method
was also compared to a kernel modification method.
|
15:00 |
4471. |
In vivo whole brain
susceptibility mapping using compressed sensing
bing Wu1, Wei Li1, and Chunlei Liu1
1Brain imaging and analysis center, Duke
University, Durham, NC, United States
A novel suscepibility mapping method based on compressed
sensing is proposed, in which the ill-conditioned
k-space regions are estimated using compressed sensing.
No prior knowledge of the underlying susceptibility map
is required. Much lower lelvel of streaking artifacts
are received comparing to the direct threshold method
using in vivo brain data set.
|
Wednesday May 11th
|
13:30 - 15:30 |
Computer 122 |
13:30 |
4472. |
Regularized Quantitative
Susceptibility Mapping for Phase-based Regional Oxygen
Metabolism (PROM) at 7T
Audrey Peiwen Fan1, Berkin Bilgic1,
Thomas Benner2, Bruce R Rosen2,3,
and Elfar Adalsteinsson1,3
1Electrical Engineering and Computer Science,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 2Radiology,
Athinoula A. Martinos Center for Biomedical Imaging,
Charlestown, MA, United States, 3Health
Sciences and Technology, Harvard-MIT, Cambridge, MA,
United States
Venous oxygen saturation (Yv) is an important indicator
for brain function and absolute quantification of Yv is
critical to estimation of the cerebral metabolic rate of
oxygen (CMRO2). Here we implement an l1-regularized
quantitative susceptibility mapping technique to measure
Yv in cerebral veins at 7T without assumptions about
vessel orientation and geometry. QSM measurements in
parallel vessel segments resulted in mean Yv = 61.2% and
were comparable to mean Yv = 62.1% estimated with MR
susceptometry. QSM measurements in curved, in-plane
vessel segments resulted in mean Yv = 66.7%, which lies
in expected physiological range.
|
14:00 |
4473. |
A theoretical analysis of
the Morphology Enabled Dipole Inversion (MEDI) method: using
anatomical information to improve the calculation of
susceptibility
Tian Liu1,2, Weiyu Xu3, Amir
Salman Avestimehr3, and Yi Wang1,2
1Biomedical Engineering, Cornell University,
Ithaca, NY, United States, 2Radiology,
Weill Cornell Medical College, New York, NY, United
States, 3School
of Electrical and Computer Engineering, Cornell
University, Ithaca, NY, United States
The inverse problem from magnetic field to
susceptibility source is ill-posed. Even in a noise-free
senario, there are infinite number of solutions
fulfilling the observation, and a little error in the
input data will result in severe noise propagation in
the final outcome. In this abstract, we provide a
detailed analytical and numerical analysis to
demonstrate that the Morphology Enabled Dipole Inversion
(MEDI) approah provides a unique and accurate solution
under ideal conditions, and the error of the
reconstruction is tightly bounded by the error in the
gradient echo image in the presence of Gaussian noise.
|
14:30 |
4474. |
Fast in vivo
susceptibility imaging using compressed sensing and parallel
imaging
bing Wu1, Wei Li1, and Chunlei Liu1
1Brain imaging and analysis center, Duke
University, Durham, NC, United States
Susceptibility map may be derived from the image phase
information, however obtaining the image phase using a
3D SPGR sequence is usually a time consuming process. A
novel image reconstruction method is proposed to
accelerate the susceptibility imaging by exploiting the
complementary properties of compressed sensing and
parallel imaging. We show that qualitatively and
quantitatively accurate whole brain susceptibility map
at 1mm isotropic resolution with a TE of 40ms may be
obtained within 8 minutes.
|
15:00 |
4475. |
Susceptibility mapping:
computation of the field map using water-fat separation at
7T
Ildar Khalidov1, Tian Liu1, Martin
R. Prince1, and Yi Wang1
1Radiology, Weill Cornell Medical College,
NYC, NY, United States
Susceptibility mapping in animals is a promising
technique for contrast agent quantification. To get a
susceptibility map, the inverse problem solver requires
high-quality field map as input. To estimate the field,
the water-fat separation problem in the conditions of
strong field inhomogeneity at 7T is solved in this
abstract. We use the combination of the VARPRO method
with a Markov chain to enforce spatial consistency
between field map values. We expect the method to be
useful for contrast quantification in high-field animal
MRI.
|
Thursday May 12th
|
13:30 - 15:30 |
Computer 122 |
13:30 |
4476. |
Improving Susceptibility
Mapping of Veins Using a K-space Iterative Approach
Jin Tang1, Saifeng Liu1, Jaladhar
Neelavalli2, and E Mark Haacke2,3
1School of Biomedical Engineering, McMaster
University, Hamilton, Ontario, Canada, 2The
MRI Institute for Biomedical Research, Detroit,
Michigan, United States,3Academic Radiology,
Wayne State University, Detroit, Michigan, United States
Mapping susceptibility from field perturbation data is a
difficult inverse problem. Here we present a unique
k-space iteration/image processing approach which
dramatically reduced reconstruction streak artefacts
caused by an ill-posed problem of inverse filter and
simultaneously improved the accuracy of susceptibility
quantification. This method could potentially be used
for quantitative in vivo venous oxygen saturation
measurement using SWI data.
|
14:00 |
4477. |
Susceptibility Mapping in
Rat Deep Brain Structures using UHF MRI
David A Rudko1, L M Klassen1,
Sonali N de Chickera2, Greg A Dekaban2,
and Ravi S Menon1
1Centre for Functional and Metabolic Mapping,
Robarts Research Institute, London, Ontario, Canada, 2Biotherapeutics
Research Group, Robarts Research Institute, London,
Ontario, Canada
In this study, microscopic field (µB 0) maps
were used to asses the Generalized Lorentzian model
(GLmodel) of field shifts in healthy rat brain at 9.4 T.
Experimentally measured µB 0 at
multiple sampling orientations was within measured error
tolerance compared to the GLmodel shifts in subcortical
GM and cerebral veins. The enhanced complexity of white
matter field shifts was addressed by examining: (i) the
elliptical cylinder and (ii) the circular cylinder
approximations for axonal bundle fields. The circular
cylinder approximation produced negative field shifts
comparable to µB[ 0 estimates
and the fields around phospholipid-like susceptibility
inclusions. and
R2 * maps
were reconstructed independently to estimate iron
concentration around susceptibility inclusions.
|
14:30 |
4478. |
Susceptibility Mapping of
Human Brain Reflects Spatial Variation in Tissue Composition
Wei Li1, Bing Wu1, and Chunlei Liu1,2
1Brain Imaging & Analysis Center, Duke
University, Durham, NC, United States, 2Radiology,
Duke University, Durham, NC, United States
A novel susceptibility mapping method is developed,
which is based on two complementary equations, i.e., the
Fourier relationship between phase and susceptibility,
and its k-space first-order derivatives. This method
allows high quality susceptibility mapping of human
brain in vivo. The resultant susceptibility maps allow
excellent visualization of deep nuclei and white matter
fiber bundles, which reflects spatial variation in
tissue composition, especially iron and myelin
distribution. This method is robust and efficient, thus
providing a convenient tool for routine susceptibility
mapping for the study of brain physiology and
neurological diseases.
|
15:00 |
4479. |
Susceptibility
quantification in MRI using phase gradient mapping
Luning Wang1, and Qun Zhao1
1Department of Physics and Astronomy,
University of Georgia, Athens, GA, United States
Susceptibility quantification in MRI is a novel
technique that can be involved in many applications,
such as tracking stem cells labeled with
super-paramagnetic iron oxide (SPIO) nanoparticles.
Because of the tight relationship between susceptibility
and magnetic field inhomogeneity, most of the published
techniques employ the MR phase images to estimate
susceptibility distributions. However, the wrapping
effect in MR phase maps is a headachy problem and has to
be solved before any further data processing. In this
work, we present a new technique that implements the
phase gradient mapping (PGM) method to directly
calculate the susceptibility distribution.
|
|
|
Electronic
Posters
: Pulse Sequences, Reconstruction & Analysis
|
Click on
to view the abstract pdf and click on
to view the video presentation. |
Pulse Sequences - Contrast Mechanisms
Monday May 9th
Exhibition Hall |
14:00 - 16:00 |
Computer 123 |
14:00 |
4480. |
Feasibility of myelin
water fraction quantification using multi-component gradient
echo sampling of spin echoes
Yann Gagnon1,2, Neil Gelman1,2,
and Jean Théberge1,2
1Medical Biophysics, University of Western
Ontario, London, Ontario, Canada, 2Lawson
Health Research Institute, London, Ontario, Canada
A new acquisition scheme is proposed consisting of
symmetrically sampling, using gradient echoes, the
rephrasing and subsequent dephasing parts of multiple
spin echoes, termed multi-component gradient echo
sampling of spin echoes (mcGESSE). The theoretical
ability of this method to calculate the MWF is
demonstrated in a two-component model using simulated
data and realistic temporal signal to noise profiles.
Implementation of this sequence would allow for
whole-brain MWF data in a clinically relevant scan time
of approximately 10 to 20 minutes.
|
14:30 |
4481. |
2D Multi-slice
Quantitative Myelin Water Imaging at 3T
Junyu Guo1, Qing Ji1, and Wilburn
E. Reddick1
1Radiological Sciences, St Jude Children's
Research Hospital, Memphis, TN, United States
Myelin water fraction (MWF) provides a direct indicator
of myelin structure and component change due to white
matter diseases. Conventionally, the data acquisition
time for generating a MWF map of a single slice takes
about 25 minutes. In this study, we proposed a 2D
multi-slice acquisition scheme with much higher
efficiency and capturing the small contribution of the
myelin-water signal. This acquisition scheme could cover
most of the brain volume in less than 12 minutes. We
also provided a weighted regularized nonnegative least
squares (wrNNLS) algorithm to generate reliable
parametric maps.
|
15:00 |
4482. |
Simulation of Double
Pulsed Field Gradient Experiments
Gregory T. Baxter1, Evren Ozarslan2,3,
Peter J. Basser2, and Lawrence R. Frank1,4
1Radiology, UCSD, La Jolla, CA, United
States, 2STBB
/ PPITS / NICHD, National Institutes of Health,
Bethesda, MD, United States, 3Center
for Neuroscience and Regenerative Medicine, USUHS,
Bethesda, MD, United States, 4VASDHS,
La Jolla, CA, United States
Double pulsed field gradient (DPFG) MR experiments are
growing in popularity owing to their ability to reveal
new structural features of tissue. Here, we simulate
DPFG experiments in restricted environments and compare
these results against theory. We then simulate diffusion
in arrays of microcapillaries both filled and surrounded
by water. We find significant changes in the MR signal
as a function of cylinder spacing.
|
15:30 |
4483. |
Intermolecular
double-quantum coherence imaging without coherence selection
gradients
Yanqin Lin1, Guiping Sheng1,
Congbo Cai1, Shuhui Cai1, Jianhui
Zhong2, and Zhong Chen1
1Department of Physics, Xiamen University,
Xiamen, Fujian, China, People's Republic of, 2Department
of Imaging Sciences, University of Rochester, Rochester,
NY, United States
Intermolecular double-quantum coherences (iDQCs) have
some unique properties and have been applied for novel
contrast of MRI. In the conventional iDQC imaging
experiments, the distant dipolar field (DDF) is
generated by coherence selection gradients (CSGs).
However, it is found that DDF can arise from certain
sample geometry in the absence of CSGs. Here, we
demonstrate that sample geometry plays an important role
in generating DDF in the absence of CSGs, and this DDF
can yield stronger iDQC signal than that from the CSGs
if an appropriate phase cycling scheme is adopted. This
will facilitate iDQC practical applications.
|
|