ISMRM 21st
Annual Meeting & Exhibition
○
20-26 April 2013
○
Salt Lake City, Utah, USA |
ELECTRONIC POSTER
SESSION • ENGINEERING |
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (10:30-11:30) Exhibition Hall |
|
Computer # |
|
4338.
|
25 |
NIST/ISMRM MRI System
Phantom T1 Measurements on Multiple MRI Systems
Kathryn E. Keenan1, Michael A. Boss1,
Edward F. Jackson2, Seon-joo Kown3,
Dominique L. Jennings3, and Stephen
Russek1
1National Institute of Standards and
Technology, Boulder, CO, United States, 2Imaging
Physics, The University of Texas M.D. Anderson
Cancer Center, Houston, TX, United States, 3Massachusetts
General Hospital, Boston, MA, United States
An MRI system phantom was developed through
collaboration between the ISMRM and NIST. We used
this phantom to answer: 1) how does measured T1
compare to the known T1; and 2) is there variation
in the measured T1 across multiple MRI systems? For
more information on the NIST/ISMRM phantom, please
visit: http://collaborate.nist.gov/mriphantoms/bin/view/MriPhantoms/MRISystemPhantom.
|
4339. |
26 |
A Novel fMRI-Compatible
Device to Quantify the Cortical Response to
Walking-Related Foot Sole Pressures
Ying Hao1,2, Kai Zhang1, Ye
Wang1, Xiaoying Wang1,3, Jing
Fang1,4, Jue Zhang1,4, and
Brad Manor1,2
1Academy for Advanced Interdisciplinary
Studies, Peking University, Beijing, Beijing, China, 2Gerontology
Department, Beth Israel Deaconess Medical Center,
Boston, MA, United States, 3Radiology
Department, Peking University First Hospital,
Beijing, Beijing, China, 4College
of Engineering, Peking University, Beijing, Beijing,
China
This novel foot sole stimulator elicited more
simulated walking pressure stimulus to the foot sole
compared to the former version, meanwhile, it
remains no interfered with image quality and did not
cause severe motion artifacts. This system is thus
feasible for fMRI studies to explore functional
brain networks involved in the perception and
modulation of the real foot sole somatosensation.
Combination of this new tool with functional MRI
(fMRI) will afford insight into the functional brain
networks underlying the afferent feedback under
walking.
|
4340. |
27 |
Motion Phantom for
Real-Time MRI
Sebastian Schaetz1, Markus Untenberger1,
Aaron Niebergall1, and Jens Frahm2
1Biomedizinische NMR Forschungs GmbH, Max
Planck Institute for Biophysical Chemistry,
Göttingen, Lower Saxony, Germany, 2Biomedizinische
NMR Forschungs GmbH, Max Planck Institute, Göttingen,
Lower Saxony, Germany
A novel metal-free motion phantom based on a simple
pneumatic rotary motor with optical rotational speed
measurement is presented. The phantom facilitates
the simulation of various motion speeds from 0.05 to
0.5m/s by air-pressure adjustment. It is a suitable
tool for for both qualitative and quantitative
evaluation of real-time MRI methods.
|
4341. |
28 |
A Single-Magnet Fast
Field-Cycling Whole-Body MRI System with Detection at
0.2 T
Gareth R. Davies1, Kerrin J. Pine1,
and David J. Lurie1
1Aberdeen Biomedical Imaging Centre,
University of Aberdeen, Aberdeen, Scotland, United
Kingdom
Fast Field-Cycling (FFC) MRI systems differ from
conventional scanners by their ability to adjust
rapidly the magnetic field strength B0 during the
pulse sequence. This offers unique image contrast,
based on the field-dependence of the spin-lattice
relaxation time (T1-dispersion) giving, for example,
increased sensitivity to changes in protein. We have
built a whole-body FFC-MRI system with a detection
field of 0.2 T. This has required novel designs of
magnet, power supplies, and control hardware and
software. Design issues and solutions and initial
results are presented. The scanner will allow
clinical applications of FFC to be explored.
|
4342.
|
29 |
Construction of a 1.5 T
Cryogen Free MRI System
Sergei Obruchkov1 and
Robin Dykstra1
1Victoria University, Wellington, New
Zealand
A 260 mm bore 1.5T MRI system was designed around a
cryogen-free superconducting magnet. The system was
designed for small animal imaging applications in
cancer research, to be located were cryogen
availability and costs are prohibitive for
conventional magnet systems.
|
4343. |
30 |
PNS Safety of the
Composite Gradient System
-permission withheld
K Craig Goodrich1, Seong-Eun Kim1,
Joshua D. Kaggie1, J. Rock Hadley1,
William B. Handler2, Blaine A. Chronik3,
and Dennis L. Parker1
1UCAIR, University of Utah, Salt Lake
City, Utah, United States, 2Physics
and Astronomy, Univ of Western Ontario, London,
Ontario, Canada, 3Physics
and Astronomy, Western University, London, Ontario,
Canada
This work was done to establish PNS limits for
simultaneous operation of a 3 axis head/neck
gradient insert and standard gradient system
(composite mode operation). PNS thresholds were
measured for each gradient system as well as
combined operation for each axis separately and
combined. Experimental results suggest that body and
insert gradient thresholds are independent of each
other, allowing for increased combined gradient
field strength and slew rates in composite mode
until either constituent threshold is reached. These
results show a definite safety advantage for
composite gradient mode operation allowing larger
gradients and slew rates.
|
4344. |
31 |
Design Study for Making
Split Gradient Coils More Robust
Michael S. Poole1, Nadim Jon Shah1,2,
and Rob Hawkes3
1INM-4, Forschungszentrum Juelich,
Juelich, Nordrhein-Westfalen, Germany, 2Neurologische
Klinik, Universitaetsklinikum Aachen, Aachen,
Nordrhein-Westfalen, Germany, 3Wolfson
Brain Imaging Centre, Addenbrookes Hospital,
Cambridge, Cambridgeshire, United Kingdom
The redesign of split gradient coils for a hybrid
MR-PET system is considered. Numerous changes in the
coil design are assessed for their performance with
the aim of making the coil more robust. New design
strategies and a slackening of design constraints
are also assessed for their ability to mitigate any
performance loss from making the coils more robust.
A linking annulus was found to be essential, but we
enforce radial connections for simplifying the
construction. A complex relationship exists between
the design constraints and the coil performance,
which needs to be considered carefully before a
final design is constructed.
|
4345. |
32 |
Simple, Accurate and
Efficient Multilayer Integral Method for Eddy Current
Simulation in Thin Volumes of Arbitrary Geometry
Produced by MRI Gradient Coils
Hector Sanchez Lopez1, Fabio Freschi2,
Adnan Trakic1, Elliot Smith1,
Jeremy Herbert1, Miguel Fuentes1,
Stephen Wilson1, Limei Liu1,
Maurizio Repetto2, and Stuart Crozier1
1ITEE, The University of Queensland,
Brisbane, QLD, Australia, 2Department
of Energy, Politecnico di Torino, Torino, Torino,
Italy
We present a new accurate and efficient eddy current
simulation method capable of calculating induced
currents in finite thickness conducting volumes of
arbitrary geometry induced by arbitrary arrangements
of gradient coils. The method has been
experimentally validated using a z-gradient coil and
its performance tested against COMSOL and the
Fourier Network method. We present an example to
demonstrate the capabilities of the method in terms
of predicting the induced currents, power losses and
pre-emphasis simulations using the excited
eigenvalue corresponding to the surrounding
structure. The method is accurate and fast enough to
be performed in a laptop Intel corei7 CPU.
|
4346. |
33 |
Theoretical Design of
Shim Arrays with Irregular Coil Geometry
Peter T. While1 and
Jan G. Korvink1,2
1Department of Microsystems Engineering
(IMTEK), Laboratory for Simulation, University of
Freiburg, Freiburg im Breisgau, Baden-Württemberg,
Germany, 2Freiburg
Institute for Advanced Studies (FRIAS), University
of Freiburg, Freiburg im Breisgau,
Baden-Württemberg, Germany
Conventional shim systems are comprised of a set of
dedicated coils for inducing low-order spherical
harmonic fields. In contrast, shim arrays consist of
multiple simple coil structures, such as circular
loops, arranged on a uniform grid. These elements
generate an alternative set of non-orthogonal basis
fields that offer the potential for high-order
shimming. In this study, a design concept is
presented for generating shim arrays with irregular
element geometry that afford field accuracy and
efficiency advantages over the use of regular
circular loops for inducing low-order fields.
|
4347.
|
34 |
Flanged-Edge Transverse
Gradient Coil Design for a Hybrid LINAC-MRI System
Limei Liu1, Hector Sanchez-Lopez1,
Feng Liu1, and Stuart Crozier1
1The University of Queensland, Brisbane,
Queensland, Australia
Combined Linear accelerator (LINAC) - MRI systems
can provide image-guided radiotherapy treatment.
These systems require splitting of the MRI scanner
to provide a central gap large enough to ensure dual
access for the accelerator and the patient. This
raises technical difficulties for maintaining high
gradient coil performance. In this research, a
dedicated split transverse gradient coil was
designed with a flange connected to the central coil
end and compared to existing coil designs. It was
found that a flanged-edge coil design produced
manageable coil performance and eddy currents.
|
4348. |
35 |
Very Low Field MRI with
Giant Magnetoresistance Based Sensors
Quentin Herreros1, Hadrien A. Dyvorne2,
Paolo Campiglio1, Amala Demonti1,
Guenaelle Jasmin-Lebras1, Myriam
Pannetier-Lecoeur1, and Claude Fermon1
1CEA, Gif-sur-Yvette, France, 2Mount
Sinai School of Medicine, New York, NY, United
States
Avoiding numbers of high field drawbacks, very low
field MRI requires sensitive detection devices .We
have developed a novel magnetic sensor based on a
combination of spin electronics and superconductors,
labelled mixed sensor, and reaching sensitivities in
the fT range at low frequencies down to 1kHz. This
mixed sensor was inserted in a small scale prototype
system and in vivo images were successfully acquired
at 7mT, showing high performance and versatility of
the sensor for very low field MRI.
|
4349. |
36 |
Low Acoustic RF Coil
Design
Saikat Saha1 and
Washington De Lima2
1GE, Waukesha, WI, United States, 2GE,
Florence, SC, United States
During high field MR scans, patients become
uncomfortable due to the high acoustic noise and
vibration produced inside the patient bore. This
noise can be as high as ~125dBA depending on the
field strength of the scanner and PSD being used.
The main source of noise inside the scanner is due
to the gradient coil (when pulsed) vibrating in a
static magnetic field generated by the magnet. This
noise is transmitted through the RF body coil to the
patient ears. Another major source of noise is
vibration of the RF body coil itself. RF body coil
vibrates due to the eddy current generated by the
gradient coil pulsing on the copper sections of the
RF coil. Some of the previous work, which addresses
the issue of acoustic noise and vibration have been
documented in [1-2]. The design described here
addresses the issue of acoustic noise by
significantly reducing the vibration which in turn
reduces the noise by upto 9 dBA in a conventional
Birdcage RF coil without compromising the RF
performance.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (10:30-11:30) Exhibition Hall |
RF Array Coil Coupling & RF Coils for Multinuclear & Hybrid
Systems
|
Computer # |
|
4350. |
37 |
Comparison of Arrays with
Various Mutual Impedances: Noise Correlation, SNR and
Parallel Imaging Performance
Adam Maunder1, Mojgan Daneshmand1,
Pedram Mousavi1, B. Gino Fallone2,
and Nicola De Zanche2
1Electrical and Computer Engineering,
University of Alberta, Edmonton, Alberta, Canada, 2Oncology,
University of Alberta, Edmonton, Alberta, Canada
In arrays, mutual resistance is the source of intrinsic
noise correlation, which has been assumed to be
unalterable using preamplifier decoupling and lossless
decoupling networks. By using inter-element capacitors
cancellation of both mutual resistance and reactance is
achieved between adjacent coils. Array performance
comparisons are made using two-and four-coil arrays in
the presence of preamplifier decoupled a) with the
mutual impedance removed, b) only the mutual reactance
removed and c) with no additional decoupling. On
average, the differences in SNR and parallel imaging
g-factor are minimal.
|
4351. |
38 |
Evaluation of Capacitive
and Transformer Decoupling Methods Using in Non-Overlapped
Array at 7.0T
Hongbae Jeong1, Kyoung-Nam Kim1,
Suk-Min Hong1, Joshua Haekyun Park1,
Myung-Kyun Woo1, Young-Bo Kim1,
and Zang-Hee Cho1
1Neuroscience Research Institute, Gachon
University, Incheon, Korea
We have conducted research on comparison between
capacitive decoupling network and transformer decoupling
methods with 8-channel transceive array. Three coils
were desing including hybrid version of two decoupling
methods, in terms of SNR and noise correlation matrix on
7T.
|
4352. |
39 |
Analysis of Imaged Object's
Permittivity, Conductivity, Size and Position Effects on
Optimal Capacitive, Inductive and Transformer Decoupling
Schemes in RF Coil Arrays
Volkan Emre Arpinar1 and
L. Tugan Muftuler1,2
1Department of Neurosurgery, Medical College
of Wisconsin, Milwaukee, WI, United States, 2Center
for Imaging Research, Medical College of Wisconsin,
Milwaukee, WI, United States
Although transformer, capacitive and inductive
decoupling techniques have been widely used in transmit
or receive coil arrays, a comparative analysis of their
tolerances to load variances were not reported. In this
study full wave electromagnetic simulations with
realistic coil models are used to understand the
robustness of each decoupling method to the variations
in size and electrical properties of the loading object.
|
4353. |
40 |
Mutual Coupling Study Using
a Pair of Fractal Loop RF Coils
Seunghoon Ha1 and
Orhan Nalcioglu1,2
1Tu & Yuen Center for Functional Onco-Imaging,
University of California Irvine, Irvine, California,
United States, 2Department
of Cogno-Mechatronics Engineering, Pusan National
University, Pusan, Korea
In this work, we introduced a fractal loop RF coil
keeping the lower radiation loss than the corresponding
loop RF coil, which may provide more adjustment of the
coil positioning. With designed a pair of fractal loop
RF coils, we investigated optimizing overlap locations
and the magnetic vector field propagation on the
position. In parallel with the simulation study, we
built the pair of the fractal loop coil, measureed the
isolation, and scanned phantom images. The results
proved that the fractal loop RF coil had an alternative
in designing RF coil by comparing with other circular
loop coils.
|
4354.
|
41 |
A Five Channel Receive
Array for Cardiac Imaging Using Hyperpolarized 13C
at 3T.
William Dominguez-Viqueira1, Angus Z. Lau1,
Albert P. Chen2, and Charles H. Cunningham1,3
1Imaging, Sunnybrook Research Institute,
Toronto, Ontario, Canada, 2GE
Healthcare, Toronto, Ontario, Canada, 3Medical
Biophysics, University of Toronto, Toronto, Ontario,
Canada
Hyperpolarized-13C substrates have become a promising
tool to study real-time cardiac-metabolism in-vivo.
Multi-channel receiver systems can provide similar SNR
performance to a single surface coil, but with increased
coverage. A single channel and a 5-channel array 13C
receive coil were simulated in Matlab and compared with in-vivo measurements.
SNR improvements at the base of the heart, were achieved
by using the 5-channel coil array. This coil array may
be suitable for human cardiac 13C
studies and parallel imaging reconstruction in the near
future.
|
4355.
|
42 |
23Na/1H In-Vivo Renal
MRI of Rodent Kidney at 3T by Using a Double-Tuned
Transceiver Resonator System
Raffi Kalayciyan1, Matthias Malzacher1,
Sabine Neudecker2, Norbert Gretz2,
and Lothar R. Schad1
1Computer Assisted Clinical Medicine,
Heidelberg University, Mannheim, Germany, 2Medical
Research Center, Heidelberg University, Mannheim,
Germany
23Na-MRI at high-field strengths (>3T) has proven to be
a unique modality for monitoring renal function
non-invasively after pharmacological interventions in
animal models. Nevertheless, performing in vivo studies
on clinical MR-systems simplifies transferring
newly-developed 23Na-MRI methods to clinical diagnostic
imaging, although the lower field strengths represent a
major drawback. In this work we developed a transceiver
double-resonant (23Na/1H) RF resonator system, and
integrated it using an RF interface at 3T human scanner
in order to monitor furosemide-induced 23Na signal
changes in renal tissue.
|
4356.
|
43 |
An Improved Surface Coil
Design for Proton Decoupled Carbon-13 Magnetic Resonance
Spectroscopy
Eulalia Serés Roig1, Guillaume Donati1,
Martin Meyerspeer2, Lijing Xin1,3,
Rolf Gruetter4,5, and Arthur W. Magill1,3
1Laboratory of Functional and Metabolic
Imaging, Ecole Polytechnique Fédérale de Lausanne,
Lausanne, Vaud, Switzerland, 2Center
for Medical Physics and Biomedical Engineering, Medical
University of Vienna, Vienna, Vienna, Austria, 3Department
of Radiology, University of Lausanne, Lausanne, Vaud,
Switzerland, 4École
Polytechnique Fédérale de Lausanne, Lausanne, Vaud,
Switzerland, 5Department
of Radiology, Universities of Lausanne and Geneva,
Lausanne, Geneva, Switzerland
Carbon-13 MRS is an inherently low-sensitivity
technique, making detection signal-to-noise ratio (SNR)
critical. An important step in the optimization of SNR
is the design of the RF coil. The probe must be capable
of simultaneous transmission at the 1H frequency while
receiving at the 13C frequency for J-decoupling.
Especially when operating at high-field, this makes SAR
limitations challenging. In this abstract we demonstrate
a quadrature-13C/quadrature-1H surface coil, using traps
to decouple the 13C loops from the 1H loops. This design
increases the 13C detection sensitivity without
increasing power deposition at the 1H frequency. We
compare the performance to a standard
linear-13C/quadrature-1H coil using glycogen detection
in the human calf at 7T. An improvement in SNR by a
factor of 1.8 was shown.
|
4357. |
44 |
Numerical Optimization of a
3-Channel Array Coil for 31P
Functional Spectroscopy at 7T
Andre Kuehne1,2, Ewald Moser1,2,
and Elmar Laistler1,2
1Center for Medical Physics and Biomedical
Engineering, Medical University of Vienna, Vienna,
Austria, 2MR
Centre of Excellence, Medical University of Vienna,
Vienna, Austria
A 3-channel 31P
coil array for 7 T brain spectroscopy of the visual
cortex is optimized numerically. The coil size &
distance to the head are varied to find the optimum coil
geometry. For each configuration, the whole 2D
excitation phase parameter space is evaluated w.r.t. B 1 and
local SAR using a fast SAR calculation routine based on
local power correlation matrices. Three different
excitation modes are compared w.r.t. B 1+/ (P),
B 1+/ (SAR)
and field homogeneity.
|
4358. |
45 |
Analysis of Tissue
Properties and MRI Signals in the Head for PET/MRI
Attenuation Correction
H. Michael Gach1, Julie C. Price2,
Denise K. Davis2, Sungkyu Jung3,
Jonathan P. Carney2, James A. Ruszkiewicz2,
Brian J. Lopresti2, Charles M. Laymon2,
and Chester A. Mathis2
1Radiology & Bioengineering, University of
Pittsburgh, Pittsburgh, PA, United States, 2Radiology,
University of Pittsburgh, Pittsburgh, PA, United States, 3Statistics,
University of Pittsburgh, Pittsburgh, PA, United States
Positron emission tomography (PET) requires photon
attenuation correction to accurately reconstruct the PET
images. PET-MR systems use MRI to image and
differentiate tissue and generate attenuation or
mu-maps. Tissues are typically delineated into four
categories (air, bone, fat, and water) using Dixon
fat/water or ultra-short TE (UTE) pulse sequences and
then assigned the applicable mu-value. Unfortunately,
these sequences have limitations in differentiating the
tissue types and cannot image air (due to its negligible
mass and proton densities) and compact bone (due to its
T2 < 20 µs). A regression analysis based on brain tissue
physical parameters and MRI signals indicates that
tissue mu values can be accurately calculated for all
brain tissues based on a combination of the data,
particularly the tissue’s proton density and magnetic
susceptibility.
|
4359. |
46 |
Novel and Compact PET
Insert for Simultaneous PET/MR Imaging of Small Animals
Konrad Lankes1, Hans Post1, Hans
Wehrl2, Chih-Chieh Liu2, Mosaddek
Hossain2, Bernd Pichler2, and Sven
Junge1
1Bruker BioSpin MRI, Ettlingen, Germany, 2Preclinical
Imaging and Radiopharmacy, University of Tuebingen,
Tuebingen, Germany
Over the last view years combined, simultaneous PET and
MR imaging has become more importance, especially in the
domain of monitoring metabolic processes. and has been
developed as an excellent tool in the area of clinical
and preclinical research. This work discusses the
development of a highly efficient and compact PET/MR
imaging system for small animals, allowing studies of
rats and mice. The performances with respect of imaging
quality, temporal and special resolution of PET- and
MR-imaging have been carefully measured on the bench and
phantoms. In-vivo studies are presented to show the
possibilities of PET/MR imaging in small animal
research.
|
4360. |
47 |
Development of a Novel RF
Body Coil Integrated with MR-Compatible PET Detector
-permission withheld
Gary H. Glover1, Craig S. Levin1,
Anton Linz2, William T. Peterson2,
and Sri Harsha Maramraju2
1Stanford University School of Medicine,
Stanford, CA, United States, 2GE
Healthcare, Waukesha, WI, United States
A hybrid PET/MR prototype for simultaneous PET/MR
imaging has been developed. The scanner integrates a
high performance Time-of-Flight MR-compatible PET system
with a modified widebore MR scanner. A novel RF body
coil design enables installation of the 25 cm axial
length PET detector inside a conventional gradient coil
with negligible impact on B1 homogeneity. Other
measurements of signal-to-noise ratio, image temporal
stability and geometric distortion parameters with the
PET system powered off and on demonstrate negligible
image degradation. Furthermore, PET energy and timing
resolution measurements were not affected by MRI and
confirm bilateral compatibility of the two modalities.
|
4361. |
28 |
MRI Compatible Orthotopic
Breast Cancer Window Chamber Model for Multi-Modality
Imaging
Arthur F. Gmitro1, Rachel L. Schafer2,
and Huimin Leung3
1Medical Imaging, University of Arizona,
Tucson, Arizona, United States, 2Biomedical
Engineering, University of Arizona, Tucson, Arizona,
United States, 3College
of Optical Sciences, University of Arizona, Tucson,
Arizona, United States
A new orthotopic breast cancer window chamber animal
model system for multi-modality imaging is described and
presented. This platform system can be used for MRI,
nuclear, photoacoustic, and optical imaging of breast
cancer xenografts. High-resolution optical imaging
allows cancer processes to be studied at the
cellular/molecular level and correlated with
observations obtainable with MRI at the tissue level.
The new model system is especially applicable to the
study of cancer processes such as growth, invasion,
metastasis, and therapeutic response. The system is also
useful for the validation and application of imaging
biomarkers for therapeutic development.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (11:30-12:30) Exhibition Hall |
|
Computer # |
|
4362. |
25 |
Distribution of RF Traps to
Reduce RF Heating with Endoluminal Coils: An Experimental
Study
Jean-Marie Verret1, Frank Pilleul2,
Olivier Beuf3, and Cecile Rabrait4
1CREATIS, Villeurbanne, France, Metropolitan, 2Hospices
Civils de Lyon, Lyon, France, Metropolitan, 3CREATIS,
Lyon, France, Metropolitan, 4General
Electric Healthcare, Velizy, France, Metropolitan
There are many techniques considered in the literature
to reduce RF heating. A small number of passive RF traps
incorporated in the reception cable present excellent
results in the case of an endoluminal coil at 3T. The
historical design of RF trap such as LC filters is
efficient. Due to the distribution of induced currents
in the cable, a close attention should be paid to the
location of the RF traps along the cable. Optimal
placememt can also narrow down the number of RF traps to
a minimum.
|
4363. |
26 |
Investigation of the
Transition from Travelling Wave to Near Field Excitation
Using Crossed Dipole Antennas
Karthik Lakshmanan1, Martijn A. Cloos2,
and Graham C. Wiggins1
1NYU, Newyork, Newyork, United States, 2New
York University School of Medicine, Newyork, Newyork,
United States
Crossed Dipoles have been used to create travelling
waves in the bore of the magnet which can be used for MR
imaging. Travelling wave excitation is relatively
inefficient compared to conventional volume resonators.
If the crossed dipole is placed close to the object,
however, efficiency increases while excitation pattern
changes. In this study we compared the excitation
profile and efficiency of a crossed dipole antenna pair
as a function of distance between the antenna and the
object imaged. The efficiency of the antenna
configuration showed an inverse proportionality to the
distance between the antenna and the Imaging object.
When the distance between the antenna pair and the
imaging object was less than quarter wavelength the
excitation pattern changed from a travelling pattern
with peak low in the head and with neck excitation to
more of a near field excitation with the excitation
primarily confined to the top of the head.
|
4364. |
27 |
A Numerical Approach to the
Development of MRI Radiofrequency Coils
Gemma R. Cook1, Martin John Graves1,
and David J. Lomas1
1Department of Radiology, University of
Cambridge, Cambridge, United Kingdom
This educational presentation is aimed at scientists
interested in radiofrequency coil design. It details the
steps involved in the simulation of coils and the ways
the results aid in the development of optimised designs.
General methods to arrange a model geometry with
appropriate boundary conditions and approximations are
included and the ways in which performance metrics such
as SNR, g-factor and SAR are calculated from
electromagnetic field results explained. The Finite
Element Method for EM field solutions will be described
to assess the benefits of a simulation-based assessment
of coil designs; examples use COMSOL Multiphysics (COMSOL
Inc., Sweden).
|
4365. |
28 |
3T Transceiver Quadrature
Surface Coil Design for Cervical Spine Spectroscopy
Wonje Lee1, Oded Gonen2, Daniel
Sodickson1, and Graham Wiggins1
1Radiology, NYUMC, New York, NY, United
States, 2Radiology,
New York University, New York, NY, United States
MRS in the cervical spinal cord is particularly
challenging because of the small size and deep
anatomical location, limiting the signal available and
placing stringent demands on the accuracy of spatial
localization. Due to the use of high bandwidth
refocusing pulses and the inclusion of additional water
suppression the body coil transmission with a receive
only coil may impose limitations. Motivated by the need
of a local transceiver coil aiming for the C-spine
spectroscopy a new design of 3T TR quad surface coil is
presented, and its efficiency performance is evaluated,
compared to the commercial head-neck receive coil.
|
4366. |
29 |
A Transceive RF Coil for
Imaging Tissue Specimen at 3T Based on PCB Design
-permission withheld
Roland Müller1, André Pampel1,
Toralf Mildner1, Henrik Marschner1,
and Harald E. Möller1
1Max Planck Institute for Human Cognitive and
Brain Sciences, Leipzig, Germany
Goal of the project was to design a transceive RF coil
for imaging small tissue specimen on a human-scale 3T MR
scanner. Besides high SNR, high homogeneity and
stability of the B1+ field, a major objective was to
minimize RF heating. Two perpendicular Helmholtz coil
pairs were combined to obtain a quadrature coil. To
achieve a high degree of accuracy during coil
fabrication, the whole 3D design was made from printed
circuit board material. Differential-mode sheath waves
were easily and aperiodically suppressed by connecting
the coax cable shields with a few resistors.
|
4367. |
30 |
A New UHF Transceiver
Antenna Design: Modified Folded Dipole
Wonje Lee1, Daniel Sodickson1, and
Graham Wiggins1
1Radiology, NYUMC, New York, NY, United
States
The linear electric dipole antenna has been suggested in
favor of its current distribution by electrodynamics,
reporting potential benefits of better penetration depth
and efficiency at 7T. However, it exhibits strong
loading sensitivity when close proximity to the
conductive object due to its open structure, which
hampers applications in subject dependent body imaging.
To aim for practical transceiver body array
implementation at 7T, a new antenna design, calling
“modified folded dipole (mf-dipole)” is presented. The
evaluation of the proposed antenna exhibits less loading
sensitivity, and favorable B1 patterns, proposing a
solution for challenging 7T body imaging.
|
4368. |
31 |
7T Tx Body Coil with
Rx-Only Insert: Primarily Results
Shailesh B. Raval1, Yujuan Zhao1,
Tiejun Zhao2, Narayanan Krishnamurthy1,
Sossena Wood1, and Tamer S. Ibrahim1
1University of Pittsburgh, Pittsburgh, PA,
United States, 2Siemens
Medical Solutions, Pittsburgh, PA, United States
. The goal of this study is to use multi-channel
transmit coil to target and excites the nuclei in
spatially uniform manner in conjunction with a
receive-only insert. In this preliminary study, we
utilized 7 Tesla human MRI scanner with a 12-channel
transmit array combined with 16-channel receive-only
body array that target kidney and liver
|
4369. |
32 |
Transmit-Only Receive-Only
Operation of a Switch-Tuned 13C-1H
Radiofrequency Coil for Improved in Vivo 13C
Spectroscopy
Heeseung Lim1, Kundan Thind1,
Francisco M. Martinez-Santiesteban1, and
Timothy J. Scholl1,2
1Medical Biophysics, Western University,
London, Ontario, Canada, 2Imaging
Research Laboratories, Robarts Research Institute,
Western University, London, Ontario, Canada
This work reports the integration of a 13C
surface receive coil with a switched-tuned 13C-1H
birdcage RF coil to enable combined transmit-only
receive-only (TORO) operation to improve sensitivity for
in vivo 13C
spectroscopy. The 13C-1H
birdcage coil is operated in 1H
mode as a transmit-receive (TR) volume coil for
anatomical imaging. Those images are subsequently
co-registered with chemical shift imaging data obtained
from 13C
TORO operation. The TORO combination produces four times
higher 13C
signal-to-noise ratio than the switched-tuned RF coil
alone operating in TR mode.
|
4370. |
33 |
Development of Cross-Pole
RF Tx Array for Breast Imaging at 7T
Junghwan Kim1, Narayan Krishnamurthy1,
Yujuan Zhao1, Tiejun Zhao2,
Kyongtae Ty Bae1, and Tamer S. Ibrahim1
1University of Pittsurgh, pittsburgh, PA,
United States, 2Siemens
Medical Solutions, pittsburgh, PA, United States
We have developed a two-sided RF resonator based on
8-channel array design for breast MR imaging at 7T. The
RF resonator was evaluated by means of modeling and
simulations as well as successful empirical testing
involved with a breast phantom and two normal
volunteers. Future studies include optimization of
phases and amplitudes for B1 shimming and combination
with Rx array insert.
|
4371. |
34 |
Feasibility Numerical Study
of the Travelling Wave MRI at 3T
Fabian Vazquez1 and
Alfredo O. Rodriguez2
1Fac Ciencias, UNAM, Mexico, DF, Mexico, 2Dep
Ing. Elec, UAM Iztapalapa, Mexico, DF, Mexico
The travelling wave magnetic resonance imaging (twMRI)
approach has been proved successful at 3T and 7T for
large magnet bores. However, there is still some
controversy regarding the feasibility of the twMRI being
able to generate images a lower frequency than 300 MHz.
To study this, we chose the parallel-plate waveguide
because it is able to propagate any frequency for the
principal mode, and its cut-off frequency is zero. Then,
the PPWG Poynting vector fields with a human phantom and
the magnetic fields for the principal mode were
numerically computed at 3T.
|
4372. |
35 |
Characterizing Bazooka
Baluns
-permission withheld
M. Arcan Erturk1,2 and
AbdEl-Monem M. El-Sharkawy2
1Department of Electrical & Computer
Engineering, Johns Hopkins University, Baltimore,
Maryland, United States, 2Russell
H. Morgan Department of Radiology and Radiological
Sciences, Johns Hopkins University, Baltimore, MD,
United States
Baluns or cable traps are used to suppress common mode
currents on the outer shield of coaxial cables
connecting MRI coils. Their main function is to enhance
the SNR performance of coils, reduce coupling and ensure
RF safety. Here, we demonstrate a method to calculate
the expected impedance and resonance frequency of
bazooka baluns. The method is verified experimentally
where bench measurements agree with our calculations
from 50MHz up to at least 150MHz.
|
4373. |
36 |
Demonstration of High
Efficiency on Coil RF Amplifier for 7 Tesla MRI
Hugo Kroeze1, Jeroen Stroeve1,
Michel Italiaander1, Ingmar Voogt1,
Giel Mens1, Peter R. Luijten1, and
Dennis W. J. Klomp1
1Radiology, UMC Utrecht, Utrecht, Utrecht,
Netherlands
At high field MRI it is usefull to place the RF
amplifiers directly on the coil. To avoid excessive heat
in the bore, these amplifiers must be very efficient. A
class-E amplifier was constructed with an efficiency of
80%. Class-E amplifiers can not amplify amplitude
modulated signals, a modulator was added. This modulator
was driven from a gradient waveform board. The amplifier
was directly connected to a 50 mm loop coil. Images of a
phantom and a wrist were made using this setup. This
demonstrates that efficient on coil amplifiers can be
connected to MRI scanners without major hardware
modification.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (11:30-12:30) Exhibition Hall |
Dielectrics & Novel RF Concepts
|
Computer # |
|
4374. |
37 |
Near-Field Wave Impedance
Matching with High-Permittivity Dielectric Materials for
Optimum Transmittance in MRI Systems
Giuseppe Carluccio1, Sukhoon Oh1,
Qing Yang2, Danilo Erricolo3, Ray
Weiluo2, and Christopher Michael Collins1
1Radiology, New York University, New York,
New York, United States, 2Radiology,
Pennsylvania State University, Hershey, Pennsylvania,
United States, 3Electrical
and Computer Engineering, University of Illinois at
Chicago, Chicago, Illinois, United States
A method to find the optimum permittivity of the
dielectric pads that maximizes the matching by analyzing
with a simple 1D analytical model the near field wave
impedance of the fields generated by a magnetic loop
which propagate through different materials. The
permittivity values that maximize the matching in this
simplified model are compared with the permittivity
values that maximize the B1 field in 3D simulations to
investigate the correlation between matching and high
permittivity dielectric pads.
|
4375. |
38 |
Electromagnetic Analysis of
Dielectric Shimming Using High Permittivity Materials
Wyger M. Brink1 and
Andrew Webb1
1Radiology, Leiden University Medical Center,
Leiden, Netherlands, Zuid-Holland, Netherlands
In this work the field patterns resulting from a high
permittivity pad have been analyzed and a theoretical
framework for describing these effects has been
evaluated.
|
4376. |
39 |
A Novel 7T Transmit Array
Using TE 01 Mode
Dielectric Resonators
Jonathan Y. Lu1, Xiaoliang Zhang2,3,
and Brian K. Rutt4
1Electrical Engineering Department, Stanford
University, Stanford, CA, United States, 2Department
of Radiology and Biomedical Imaging, University of
California-San Francisco, San Francisco, CA, United
States, 3UC
Berkeley/UCSF joint graduate group in Bioengineering,
San Francisco, CA, United States, 4Radiology
Department, Stanford University, Stanford, CA, United
States
We propose a novel transmit array coil design for use at
7T employing an array of dielectric resonators. Our
design is composed of cylinders filled with deionized
water, which after excitation by copper loops can
function as transmit array elements. Using FDTD
simulations, we have found that this design is easy to
tune to high frequency, uses low cost material, and
shows improved decoupling and B1+ efficiency
compared to a loop array of matched dimension.
|
4377. |
40 |
In-Vivo Evaluation of a New
High Dielectric Constant Material for Local Enhancement of B1+ and
SNR at 3T
Christopher T. Sica1, Sebastian Rupprecht1,
Ray Luo2, Zhipeng Cao2, Raffi
Sahul3, Seongtae Kwon3, Michael T.
Lanagan4, Christopher Collins5,
and Qing Yang1,6
1Radiology, Penn State College of Medicine,
Hershey, Pennsylvania, United States, 2Bioengineering,
Pennsylvania State University, University Park,
Pennsylvania, United States,3TRS Technologies
Inc, State College, Pennsylvania, United States, 4Engineering
Science and Mechanics, Pennsylvania State University,
University Park, Pennsylvania, United States, 5Radiology,
New York University, New York, New York, United States, 6Neurosurgery,
Penn State College of Medicine, Hershey, Pennslyvania,
United States
A new high dielectric constant material, in the form of
monolithic blocks of relative permittivity ~800, is
evaluated for local enhancement of B1+ and
SNR at 3T. Phantom studies show up to a two fold
increase in SNR close to the blocks along with gains in
transmit efficiency. An in-vivo study of the brain
showed SNR gains up to 40% in a localized region close
to the blocks, a reduction of transmission power by 50%,
and local gains in transmit efficiency. Parallel imaging
performance in the brain was evaluated and was minimally
affected by the presence of the HDC blocks.
|
4378. |
41 |
Analytical Study on the
Effects of High Dielectric Material Surrounding Spherical
Sample with Coil Array for Parallel Imaging
Wei Luo1,2, Giuseppe Carluccio3,
Zhipeng Cao4,5, Yang X. Qing1, and
Christopher Michael Collins1,3
1Radiology, The Pennsylvania State
University, Hershey, PA, United States, 2Engineering
Science & Mechanics, The Pennsylvania State University,
University Park, PA, United States, 3Radiology,
New York University, New York, NY, United States, 4Radiology,
Pennsylvania State University, Hershey, PA, United
States, 5Bioengineering,
The Pennsylvania State University, University Park, PA,
United States
Effects of high dielectric material (HDM), that has a
size larger than the coil element of a parallel receive
array, are theoretically studied at 3T. The results
suggest that the gain from receive sensitivity of the
array brought by the HDM overcomes the enhanced
g-factor, which can promise to improve signal-to-noise
ratio even when the HDM has a size which is larger than
the coil element of a parallel receive array.
|
4379. |
42 |
Manipulating B1 Spatial
Distribution at 7 Tesla with Dielectric Pads at a Distance
from a Transmit Surface Coil
Manushka V. Vaidya1,2, Sukhoon Oh1,
Christopher Michael Collins1,2, Daniel K.
Sodickson1,2, and Riccardo Lattanzi1,2
1Bernard and Irene Schwartz Center for
Biomedical Imaging, New York University School of
Medicine, New York, NY, United States, 2Sackler
Institute of Graduate Biomedical Sciences, New York
University School of Medicine, New York, NY, United
States
The use of dielectric pads between the radiofrequency
(RF) coil and sample has been proposed to “focus” the B1 field
into the sample to improve transmit efficiency. In this
study, we investigated how dielectric pads placed at a
distance from the RF coil affect the B1+
spatial distribution inside the sample. We performed
numerical simulations of the B1+ distribution
inside a uniform cylinder at 7T for various positions of
the dielectric pad with and without a surrounding
shield. Manipulating B1 spatial
distribution with dielectric pads can be advantageous
for various MR applications, including improving RF
homogeneity at ultra-high fields.
|
4380. |
43 |
Analysis of the Effect of
an External High Dielectric Sleeve on the Performance of a
Head Coil at 128 MHz
Bu S. Park1, Joshua W. Guag1,
Leonardo M. Angelone1, and Sunder S. Rajan1
1CDRH/DP, FDA, Silver Spring, MD, United
States
We present experimental and numerical simulation results
showing that high dielectric materials (HDMs) located
outside of a RF head coil decrease the electric field
(E-field, |E|) with minimum change of the RF magnetic
field (B1+) at 3T MRI imaging. Compared to previous
research using HDMs located between the RF coil and
sample, our method locating the HDM outside the coil
allowed an increased sample size and more flexibility of
HDM thickness optimization. Numerical simulation results
showed more than 30% decrease in the local SAR at the
boundary region of the head model. Validating
experimental results showed a 21% decrease in the
maximum |Etotal| using a HDM made of distilled water.
|
4381. |
44 |
Maximized Local B1+ Using
Optimized Dielectric Pad at 7 T: Numerical Optimization and
Experimental Validation
Sukhoon Oh1, Wei Luo2, Bei Zhang1,
Cem Murat Deniz3, Michael T. Lanagan4,
Graham C. Wiggins1, and Christopher Michael
Collins1
1Center for Biomedical Imaging, School of
Medicine, New York University, New York, New York,
United States, 2Center
for NMR Research, Radiology, College of Medicine, The
Pennsylvania State University, Hershey, Pennsylvania,
United States, 3New
York University, New York, New York, United States, 4Engineering
Science and Mechanics, The Pennsylvania State
University, University Park, Pennsylvania, United States
Dielectric materials with high permittivity have been
used to improve signal homogeneity in high field MR
images and enhance SNR near the pad. Permittivity,
geometry, dimension, and location of dielectric pad are
variables to be optimized to maximize the effect of a
dielectric pad. In this study, we report the
permittivity optimization of dielectric pad to maximize
the local B1+ in a phantom with electromagnetic field
simulations at 7T and verification with flip angle
mapping experiments at 7T when the dielectric pad is
present with different permittivity values.
|
4382. |
45 |
Dipole Antenna Without
Ceramic Substrate and Still Low SAR: The Fractionated Dipole
Antenna.
Alexander Raaijmakers1, Ingmar Voogt1,
Dennis W. J. Klomp1, Peter R. Luijten1,
and Nico van den Berg1
1Imaging Division, UMC Utrecht, Utrecht,
Netherlands
RF coil array elements for body imaging at ultra-high
field strengths need to be designed as antennas. An
example of such a design is the single-side adapted
dipole antenna (SiSiAD). Its drawback: it is heavy. In
this study, we present another type of array element
that has been designed according to radiative
principles: the fractionated dipole antenna. It consists
of a 30 cm dipole where the legs are split into 3
segments, interconnected by 20-50 nH inductors.
Simulations and measurements show that the fractionated
dipole antenna has similar or better B1+/SAR
performance than the SiSiAD, while being lighter and
cheaper.
|
4383. |
46 |
A 16-Element Highly
Flexible RF Array Coil for 3T MRI
Yunsuo Duan1, Bradley S. Peterson1,
Feng Liu1, and Alayar Kangarlu1
1MRI Research, Department of Psychiatry,
NYSPI/Columbia University, New York, NY, United States
The SNR of array coils depend on the filling factor of
the coils because the sensitivity of each coil element
dramatically attenuates with the increase of the
separation between the coil element and the object to be
imaged. A desirable array coil therefore is expected to
be positioned closest to the object. This requires that
the array coil must be highly flexible in order to fit
for objects of variable dimensions and the coil built in
a way that its flexibility does not affect its
performance. Although some adjustable array coils have
been proposed, their flexibility is limited because of
the extreme difficulty in decoupling between coil
elements with the change of coil geometries. We present
a 16-element array coil with high flexibility and
provide a comparison study between the highly flexible
array coil and a rigid cylindrical coil based on our
research on high decoupling. The flexible array coils
are able to significantly improve the SNR and
homogeneity of images.
|
4384. |
47 |
A Novel Alternating
Impedance Transceiver Coil for 7T MRI
James F. Stack, Jr.1
1Remcom, Inc., State College, PA, United
States
Particle Swarm Optimization (PSO) and finite difference
time domain (FDTD) simulations are employed to design a
novel alternating impedance microstrip resonator (MSR)
transceiver array for 7T MRI. The resulting design uses
non-uniform spacing of impedance elements to control the
B1+ field distribution and improve the peak B1+ value.
Comparisons are shown between the optimized geometry, a
traditional straight MSR and an alternating impedance
MSR with uniformly spaced circular impedance elements.
|
4385. |
48 |
Current-Line Solution for
Understanding and Predicting B1 /
B1+ Behavior
and Investigating Central Focusing
Gianluigi Tiberi1, Mauro Costagli1,
Riccardo Stara1,2, James Tropp3,
and Michela Tosetti1,4
1Fondazione Imago7, Pisa, PI, Italy, 2Univerisity
of Pisa, Pisa, PI, Italy, 3GE
Healthcare Technologies, Fremont, CA, United States, 4Stella
Maris Scientific Institute, Pisa, Pi, Italy
We introduce a novel analytical approach based on the
theory of cylindrical waves radiated by a filament of
a-c current. The approach permits to calculate B1/ B1+
inside volume resonators loaded with cylinders.
Moreover, it permits to separate the solution to single
line source problem (primordial solution) and the
composite one (the summations of primordial solutions
accordingly to the resonator driving). This feature is
of fundamental importance, because only by looking at
single line solution it is possible to have insight into
central focusing, investigating and discriminating among
phenomena of dielectric resonance, standing wave and
multi-source interference.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (10:30-11:30) Exhibition Hall |
|
Computer # |
|
4386. |
49 |
Brain Arrays for Neonatal
and Premature Neonatal Imaging at 3T
Azma Mareyam1, Duan Xu2, Jonathan
R. Polimeni1,3, Eli Siskind1,
Vijay Alagappan4, James N. Blau1,
Boris Keil1,3, Nan Tian2, Wei Zhao1,3,
and Lawrence L. Wald1,3
1A.A. Martinos Center for Biomedical Imaging,
Department of Radiology, Charlestown, Massachusetts,
United States, 2UCSF
School of Medicine, San Francisco, California, United
States, 3Harvard
Medical School, Boston, Massachusetts, United States, 4GE
Healthcare, Cleveland, Ohio, United States
Imaging premature and neonatal babies in incubators is
challenging due to time constraints, motion artifacts
and the need for high resolution. These issues call for
increased detector sensitivity to achieve high spatial
resolution and highly-parallelized array image
acquisition. In this work we have constructed and tested
two 16-channel brain arrays sized for roughly 32 and 36
weeks gestational age as well as a 32-channel array for
term neonates. The SNR and g-factors of these arrays are
compared to each other and to an adult sized coil.
|
4387. |
50 |
The Dual Purpose CAPTAIN
(Cardiac Adult / Pediatric Torso Assembly for in-vivo Imaging)
-permission withheld
Modhurin Banerjee1, Clyve Konrad Follante1,
Aleksey Zemskov1, Rodney Bills1,
Kolman Juhasz1, Steven G. Lee1,
Victor Taracila1, and Fraser Robb1
1MRI, GE Healthcare, Aurora, Ohio, United
States
It is current practice to use adult coil systems for
pediatric imaging, which is problematic due to a variety
of difficulties encountered in workflow/preparation,
emergency egress, coil sensitivity-anatomy mismatch, and
the necessity to anesthetize the patient. A 32 channel
thoracic coil system with application to the 0-7 year
old patient population was developed from a prototype
originally created by Shreyas Vasanwalla at
Stanford/Lucinda Bell Children’s Hospital. A uniform
array was found (both through simulation and experiment)
to give the best SNR and g factor performance, offering
a significantly improved scan time and patient
experience relative to existing coil systems.
|
4388. |
51 |
A Scalable Constellation
Coil Design for 3T Body Imaging
Xing Yang1, CemMurat Deniz1, Ye Li1,
Ryan Brown1, and Yudong Zhu1
1Center for Biomedical Imaging, Department of
Radiology, NYU School of Medicine, New York, NY, United
States
Based on the field equivalence principles of
electrodynamics, a previous study1 introduced a unifying
solution to RF coil, suggesting that a single surface
structure and a surface current driving mechanism may
enable not only emulations of but higher performance
than any externally applied RF coil. A full-wave EM
simulation included therein further demonstrated a
particular 128-port constellation coil structure’s
potential as a unifying solution. In an effort to
leverage these results and possibly improve today’s
clinical MR systems that have varying numbers of
parallel RF channels, the present study explored a
constellation coil design that was adapted to suit a
small as well as a large number of channels. MR imaging
experiments on a clinical 3T scanner were conducted to
quantify SNR behavior of the same coil as the number of
implemented ports increases. Commercial phased array
coils were used as comparison references.
|
4389. |
52 |
8 Channel Transmit and 16
Channel Receive Constellation Coil for 7T MRI
Ye Li1, Xing Yang1, Cem Murat
Deniz2, and Yudong Zhu1
1NYU School of Medicine, New York, New York,
United States, 2New
York University, New York, New York, United States
In this work, we propose a novel 8-channel transmit and
16-channel receive constellation coil for 7T MRI. The
gradient-echo images were acquired with phase shimming.
The SNR was calculated and then normalized by flip
angle. Transmit power to achieve 90 degree flip angle on
central coronal plane was measured. Transmit and
reception performance of the constellation coil were
compared with that of two commercial coils that is each
composed of a birdcage transmit coil and a high-density
receive-only phased array. Our results show that the
constellation coil can achieve excellent SNR and
transmit efficiency with a single structure.
|
4390. |
53 |
High Filling Factor Vs.
High Channel Count: Which One Wins in 3T Breast Imaging?
Luca Marinelli1, Eric Fiveland1,
Keith Park1, Kenneth Rohling1, and
Ileana Hancu2
1GE Global Research Center, Niskayuna, NY,
United States, 2GE
Global Research, Niskayuna, NY, United States
The tradeoffs between high filling factor and high
number of coils in 3T breast imaging were considered.
Experiments performed in vitro and in vivo indicate
better performance (higher SNR and lower g-factor) for
the cases in which the receive array is allowed to
conform to a woman's anatomy. Results also suggest that
flexible, high channel count breast arrays may be one of
the most promising avenues to increase the SNR of breast
images.
|
4391. |
54 |
31 Channel, Flexible Breast
Coil for 3T Imaging
Ileana Hancu1, Eric Fiveland2,
Keith Park2, Kenneth Rohling2,
Seung-Kyun Lee2, and Luca Marinelli2
1GE Global Research, Niskayuna, NY, United
States, 2GE
Global Research Center, Niskayuna, NY, United States
A novel, 31 channel, flexible breast coil for 3T imaging
is presented. The performance of this coil, in vitro and
in vivo is demonstrated. A comparison with the GE, 8
channel coil is also performed.
|
4392. |
55 |
A Dedicated 8-Channel
Flexible Array Coil for Ocular fMRI
Marcos Alonso Lopez Terrones1,2, Markus
Düring1,3, Daniel Gareis1, and
Peter M. Jakob3,4
1NORAS MRI products, Höchberg, Bavaria,
Germany, 2Health
Services of Durango, Durango, Durango, Mexico, 3Experimental
Physics 5, University of Wuerzburg, Wuerzburg, Bavaria,
Germany, 4Research
Center Magnetic Resonance Bavaria, Wuerzburg, Bavaria,
Germany
In ocular imaging, the development of dedicated MRI
coils is useful to monitoring the visual cortex and to
determinate the damage in the visual pathway before a
surgery. We developed an 8-channel flexible array for
ocular fMRI. The coil is comprised of two pieces with
four channels each. SNR-maps and in-vivo images were
compared with a conventional head-coil at 1.5T. The
8-channel flexible array shows 200% SNR improvement at
the surface and 25% at the center of the phantom.
In-vivo images acquired with the 8-channel flexible
array have also shown better image quality throughout
the whole visual pathway.
|
4393. |
56 |
A 12-Channel Dorsal
Receive-Only Body Array for 7 Tesla.
Ria Pradhan1, Andreas K. Bitz1,
Stefan Maderwald1, Oliver Kraff1,
Mark E. Ladd1, and Stephan Orzada1
1The Erwin L. Hahn Institute for Magnetic
Resonance Imaging, Universität Duisburg - Essen, Essen,
Nord Rhein Westfalen, Germany
A 12-channel dorsal receive-only array was built for use
with an 8-channel transmit/ receive coil. The 12
elements were geometrically decoupled. The array was
evaluated through measurements with a vegetable oil
phantom and a volunteer. The g-factor was calculated
from the SNR of the images using the pseudo multiple
replica method. Maps of the g-factor using only the
8-channel transmit/ receive body coil were compared to
those obtained using both the 8-channel coil and the
12-channel array. As expected, the parallel imaging
capability was enhanced by the 12-channel array.
|
4394. |
57 |
MR Spectroscopic Imaging
(MRSI) Study Using the Diaper Shaped RF Coil Array.
Seunghoon Ha1 and
Orhan Nalcioglu1,2
1Tu & Yuen Center for Functional
Onco-Imaging, University of California Irvine, Irvine,
California, United States, 2Department
of Cogno-Mechatronics Engineering, Pusan National
University, Pusan, Korea
MRI and MR spectroscopic imaging (MRSI) is used for
detailed anatomic and metabolic evaluation of the
prostate. Especially, for successful MRSI acquisition,
the research about RF coil, non-invasive, enable to
provide both competitive SNR and large FOVs on prostate
MRI may be a prerequisite. In this study, we proposed a
new non-invasive coil with a competitive SNR for
prostate MRSI and compare its performance with other
commercial coils. The proposed coil showed MR spectra
having a competitive SNR when comparing with the
endorectal coil as well as generated more stable spectra
than endorectal coil on the overall prostate phantom
area.
|
4395. |
58 |
Unilateral 8ch Receiver
Array for Ultra High Resolution Time-Resolved 3D CE-MRA of
the Hand
Paul T. Weavers1, Thomas C. Hulshizer1,
Phillip J. Rossman1, Phillip M. Young2,
and Stephen J. Riederer1
1MR Research Laboratory, Mayo Clinic,
Rochester, Minnesota, United States, 2Radiology,
Mayo Clinic, Rochester, Minnesota, United States
Hand and arm MR angiography presents a particular
challenge due to the spatial and temporal resolution
demands. When unilateral vs. bilateral imaging is
clinically indicated, unilateral acquisition allows not
only reduced FOV but also reduced patient discomfort in
positioning. A modular, eight-channel receiver coil is
described which provides the high SNR necessary for
highly accelerated (R=8) 2D SENSE time-resolved
contrast-enhanced MRA, allowing very high (0.6 mm
isotropic) spatial resolution, arterial phase imaging.
|
4396. |
59 |
High Temporal and Spatial
Resolution Breast MR Imaging at 7T; Feasibility Study Using
8-To-1 Channel Tx-Only Array Combined with 8 Channel Rx-Only
Insert
Junghwan Kim1, Yujuan Zhao1,
Narayan Krishnamurthy1, Tiejun Zhao2,
Kyongtae Ty Bae1, and Tamer S. Ibrahim1
1University of Pittsurgh, pittsburgh, PA,
United States, 2Siemens
Medical Solutions, pittsburgh, PA, United States
We have developed the eight Tx only array combined with
eight Rx only array insert based on TTT design for
breast MR imaging at 7T. RF coil was tested on three
normal volunteers with and without the Rx only insert.
T1W with fat-saturation breast MRI was successfully
acquired and demonstrated.
|
4397. |
60 |
7-Channel Half-Cylinder
Shaped Transmit Coil with 32-Channel Receiver Array for
Multipurpose Head Imaging at 7T
Johanna J. Bluemink1, Martijn Lunenburg2,
Wouter Koning2, Ingmar J. Voogt2,
Michel Italiaander2, Peter R. Luijten2,
Natalia Petridou2, and Dennis W.J. Klomp2
1Radiotherapy, UMC Utrecht, Utrecht, Utrecht,
Netherlands, 2Radiology,
UMC Utrecht, Utrecht, Netherlands
A half-cylinder coil array is presented that combines 7
detunable transmit coil elements with a high density
receiver array of 32 detunable elements. The setup
allows optimized focused B1+ and B1- in for instance
cerebellum or visual cortex, while providing maximized
space for visual stimulation. Being independently
flexible with B1 steering, the setup also allows
scanning with standard immobilization devices as used in
radiotherapy.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (11:30-12:30) Exhibition Hall |
|
Computer # |
|
4398. |
61 |
Z-Shim RF Coil Design
Enhances Parallel Transmit Performance in Body Imaging at 3T
Xiaoping Wu1, Jinfeng Tian1,
Sebastian Schmitter1, Brian Hanna1,
Joseph Pfeuffer2, Michael Hamm2,
Juergen Nistler2, University of Minnesota
University of Minnesota Vaughan1, Kamil
Ugurbil1, and Pierre-Francois Van de Moortele1
1CMRR, Radiology, University of Minnesota,
Minneapolis, MN, United States, 2Siemens
Healthcare, Erlangen, Germany
There has been an increasing interest in designing RF
arrays with coil elements distributed in all three
dimensions so as to also allow B1+ manipulation along
the Z direction. Such Z-shim coils have been shown to
provide improved RF transmit efficiency for brain
imaging, as compared to a conventional coil which has
only in-plane encoding capability. In this study, we
evaluate the performance of a Z-shim RF array designed
for body imaging at 3T and compare it to a conventional
array, by designing parallel transmit (pTX) RF pulses
with power and SAR regularizations based on
electromagnetic simulations of the two coils. Our
results show that the Z-shim RF array can give rise to
large improvement of flip angle homogenization and/or
reduction of SAR for pTX when used for body imaging at
3T, as compared to the conventional coil without
Z-encoding capability.
|
4399. |
62 |
Simultaneous Excitation of
Distinct Electromagnetic Modes Using a Tx Array
Yujuan Zhao1, Sossena Wood1,
Tiejun Zhao2, Narayanan Krishnamurthy1,
and Tamer S. Ibrahim1
1University of Pittsburgh, Pittsburgh,
Pennsylvania, United States, 2Siemens
Medical Solutions, Pittsburgh, Pennsylvania, United
States
The inhomogeneous distribution of the magnetic field B1+
at ultrahigh field imaging (≥7T MRI) is a significant
challenge for the RF coil design. In this work, a new
excitation paradigm is presented utilizing 20-channel
Tic-Tac-Toe (TTT) based RF transmit array design where
the 20 elements are combined into 5 different groups in
order to excite 20 distinct modes with longitudinal
spatial selectivity. Any 5 of these modes can be
simultaneously excited. The optimization shimming method
has been used to find a uniform excitation pattern under
SAR constraints. The results are successfully tested on
a 7T MRI scanner using phantoms and in-vivo (7 human
subjects.)
|
4400. |
63 |
Densely-Populated
Transceiver Surface Coil Array for the Human Brain Studies
at 7 T.
Nikolai I. Avdievich1, Jullie W. Pan1,
and Hoby P. Hetherington1
1Neurosurgery, Yale University, New Haven,
CT, United States
Transceiver surface coil arrays improve transmit
performance and B1 homogeneity for head imaging up to
9.4T. Arrays with larger number of smaller elements
further improve SNR and parallel imaging. Overlapping
adjacent elements optimizes both the loading and the
penetration depth, but complicates the decoupling.
First, the adjacent coils generate substantial mutual
resistance and cannot be decoupled using common methods,
which compensate only for the mutual reactance. Second,
in a densely-populated overlapped array, coupling exists
between non-adjacent elements. We have developed a
single-row (1x16) overlapped surface coil transceiver
head array with decoupling of both the adjacent and the
next-to-one neighboring elements.
|
4401. |
64 |
Construction of a 4-Channel
Transmit Neck Array for PCASL Tagging at 7 Tesla and
Comparison with a Head Coil.
Konstantinos Papoutsis1, James A. Meakin1,
Aaron T. Hess2, Jamie Near3,
Stephen J. Payne4, David Edwards4,
and Peter Jezzard1
1FMRIB Centre, University of Oxford, Oxford,
Oxfordshire, United Kingdom, 2Department
of Cardiovascular Medicine, University of Oxford,
Oxford, Oxfordshire, United Kingdom,3McGill
Univeristy, Montreal, Quebec, Canada, 4Department
of Engineering Science, University of Oxford, Oxford,
Oxfordshire, United Kingdom
In this study, a 4-channel neck array has been
constructed and used at 7 Tesla as a tagging coil for
the pCASL sequence. A comparison of the tagging
efficiency and SAR was made between tagging with a
1Tx/32Rx head coil vs tagging with the neck array.
Electromagnetic characterization of the neck coil helped
to address local SAR risk by estimating the worst case
transmission pattern. The neck array provided 70%
tagging efficiency with 80% SAR whereas tagging with the
head coil generated a tagging efficiency of 30% and a
global SAR of approximately 60%. Tagging with the neck
coil is the preferable method as long as the SAR is
supervised at all times.
|
4402. |
65 |
An Algorithm for Automatic
Optimisation of Transmit Array Coil Tune and Match Applied
in a Cardiac TEM Coil at 7T
Christopher T. Rodgers1, Graeme A. Keith1,
Aaron T. Hess2, Carl Snyder3,
University of Minnesota University of Minnesota Vaughan3,
and Matthew D. Robson1
1Univ Oxford, Oxford, United Kingdom, 2University
of Oxford, Oxford, United Kingdom, 3CMRR,
Univ Minnesota, Minneapolis, MN, United States
7T body MRI requires transmit array coils. The popular
TEM resonator design requires time consuming ~30min
manual adjustments for each subject. This can be
mechanised using piezo-actuators to drive the adjustment
capacitors. We implement a system for monitoring coil RF
characteristics in situ using RF pulses from the
scanner. We characterise 4 potential optimisation
metrics before introducing a simple, robust, efficient
algorithm that automates coil tuning in less than 5
minutes. We close by explaining how auto-tuning may give
not just faster, but actually superior coil tuning by
optimising at isocentre and by monitoring power
reflected through all RF channels.
|
4403. |
66 |
Imaging the Kidneys at 7T
Using 8Tx/32Rx Abdominal Coil and RF Shimming of Individual
Slices
Niravkumar Darji1, Martin Haas2,
Gopesh Patel1, Oliver Speck1, Inge
Brinkmann2, and Michael Bock2
1Department for Biomedical Magnetic
Resonance, magdeburg, Sachsen anhalt, Germany, 2Department
of Radiology, Freiburg, Baden-Württemberg., Germany
B1 field inhomogeneity become increasingly problematic
in ultrahigh-field (UHF) strengths (7T). Parallel
transmition is introduced to reduce B1 inhomogeneity. In
abdominal imaging region like Kidney or spine shows more
structural visibility after performing RF shimming on
individual channels of the 8Tx/32Rx abdominal RF coil.
The T2* maps also reflect the more homogeneous signal
distribution after RF shim perfomance.
|
4404. |
67 |
Parallel Transmission
Experiments Using an Extensible RF Pulse Generator
Andre Kuehne1,2, Patrick Waxmann2,
Werner Hoffmann2, Harald Pfeiffer2,
Reiner Seemann2, Frank Seifert2,
and Bernd Ittermann2
1Medical University of Vienna, Vienna,
Vienna, Austria, 2Department
for Medical Metrology, Physikalisch-Technische
Bundesanstalt, Berlin, Germany
An extensible multichannel small signal chain and its
application to transmit SENSE experiments at 3T is
presented. Up to 20 channels can be driven by a 2x4HE
19" units, with the possibility of interconnecting and
synchronizing multiple of these nodes.It is capable to
work with any scanner, given a clock and trigger signal
is provided, and can be used for a wide range of
frequencies up to 300 MHz. The obtained images are
indistinguishible from images generated with the
manufacturer's small signal chain, proving its
feasibility.
|
4405. |
68 |
New Results for Digital
Beamforming in MRI
Emad Ebbini1, Lance DelaBarre2,
University of Minnesota University of Minnesota Vaughan3,
and Anand Gopinath4
1University of Minnesota, Minneapolis,
Minnesota, United States, 2Radiology,
University of Minnesota, Minneapolis, Minnesota, United
States, 3CMRR,
Radiology, University of Minnesota, Minneapolis,
Minnesota, United States, 4ECE,
University of Minnesota, Minneapolis, Minnesota, United
States
Recent novel results on digital beam forming of MRI
signals are presented. These show that digital beam
forming has gain, enhances contrast and reduces signal
to noise ratio.
|
4406. |
69 |
Effect of Shielding on
Surface Coil Loops at 7T
Gillian Haemer1 and
Graham Wiggins1
1The Bernard and Irene Schwartz Center for
Biomedical Imaging, NYU Medical Center, New York, NY,
United States
Radiation power loss, or the radiated power of a coil
that is not absorbed into the sample, is assumed to be a
problem at high field in MRI, as it is known to increase
with the frequency (ω) and coil-bounding area (A): RLOSS
α A2 * ω4 [1,2]. Various authors suggest that the
effects of radiation loss, such as decreased SNR and
coil Q, can be mitigated by adding RF shielding to coils
used at high field.
|
4407. |
70 |
Integration of 2-Channel
Parallel Transmission with Forced Current Excitation for
Improved B1 Homogeneity
in Breast Imaging at 7T
Sergey Cheshkov1,2, Ivan E. Dimitrov1,3,
Wouter Koning4, Joseph Rispoli5,
Mary P. McDougall5,6, Steve Wright5,6,
and Craig R. Malloy1,2
1Advanced Imaging Research Center, University
of Texas Southwestern Medical Center, Dallas, TX, United
States, 2Radiology,
University of Texas Southwestern Medical Center, Dallas,
TX, United States, 3Philips
Medical Systems, Cleaveland, OH, United States, 4UMC
Utrecht, Utrecht, CX, Netherlands, 5Biomedical
Engineering, Texas A&M University, College Station, TX,
United States, 6Electrical
Engineering, Texas A&M University, College Station, TX,
United States
Ultra-high field MR offers the promise of higher
spatial/temporal resolutions in imaging, and improved
sensitivity and quantification in MRS. Capitalizing on
these advantages requires resolving the issues of
decreased B1 homogeneity.
We report on improved B1 homogeneity
at 7T achieved by combining the flexibility of 2-channel
parallel transmission with the insensitivity to loading
of Forced Current Excitation breast coil. The overall
transmit design is favorable for high quality breast
imaging at high static magnetic field. Additionally,
because of its insensitivity to coil loading this
approach may allow for better patient throughput since
breast size and composition vary widely across
population.
|
4408. |
71 |
20-To-8 Channel Tx Array
with 32-Channel Adjustable Receive-Only Insert for 7T Head
Imaging
Tamer S. Ibrahim1, Yujuan Zhao1,
Narayanan Krishnamurthy1, Shailesh B. Raval1,
Tiejun Zhao2, Sossena Wood1, and
Junghwan Kim1
1University of Pittsburgh, Pittsburgh, PA,
United States, 2Siemens
Medical Solutions, Pittsburgh, PA, United States
The use of multi-channel arrays at 7T is currently
common for alleviating RF inhomogeneities and addressing
RF safety concerns. In this work we integrate 32-channel
Rx-only insert with 20-to-8 channel Tx array in order to
achieve safe and robust 3D homogeneous excitation across
different subjects.
|
4409. |
72 |
A Shielded 8 Channel TxRx
Head Array with Triangular Elements and Second Order
Decoupling
Gang Chen1, Martijn A. Cloos2,
Karthik Lakshmanan3, Daniel Sodickson3,
and Graham Wiggins3
1The Sackler Institute of Graduate Biomedical
Sciences, New York University School of Medicine, New
York, NY, United States, 2The
Bernard and Irene Schwartz Center for Biomedical
Imaging, New York University School of Medicine, New
York, NY, United States, 3The
Bernard and Irene Schwartz Center for Biomedical
Imaging, New York University Langone Medical Center, New
York, NY, United States
As more advanced Parallel Transmit (PTx) technology at
high field continues to emerge, the development of
compatible Transmit-Receive(TxRx) arrays with good
transmit and receive sensitivity as well as decoupling
has become a greater priority. We describe here a
shielded version of the 8 channel PTx suitable
triangular array for head imaging at 7T MRI designed to
increase B1+ coverage along z direction as well as
keeping diverse B1 profiles for the individual elements.
In-vivo experiments demonstrated the benefits and
potential of using this Triangular array with PTx
system.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (10:30-11:30) Exhibition Hall |
|
Computer # |
|
4410. |
73 |
Improving B1-Based SAR
Determination Via Iterative Determination of Missing Field
Components
Ulrich Katscher1 and
Karim Djamshidi1
1Philips Research Europe, Hamburg, Germany
The additional degrees of freedom available in parallel
transmission hamper straight-forward SAR estimations as
applied for single channel transmission. As an
alternative to the usually applied model-based SAR
estimation, SAR can be estimated from individually
measured B1 maps. Overall, the accuracy of this B1-based
SAR determination is satisfying, however, could be
further increased if the hitherto unknown longitudinal
component of B1 can be determined. This study
investigates an iterative estimation of this component
and its impact on the resulting local SAR determination,
based on spherical and realistic patient models for an
eight-element RF transmit array at 3 T.
|
4411. |
74 |
A Low-Power, Offline
Prescreen to Detect and Suppress Dangerous Currents
Christopher W. Ellenor1, Pascal P. Stang1,
John M. Pauly1, and Greig C. Scott1
1Electrical Engineering, Stanford University,
Stanford, CA, United States
We demonstrate the feasibility of “offline” screening of
patients for dangerous interactions between imaging
fields and implanted wires. We show that using a
standard imaging coil outside the MRI environment, and
only a few milliwatts of power, we can measure the
impedance spectra of the coil, which shows
characteristic distortions in the case of a resonant
interaction. Furthermore, we show that this same
spectral information can be used to determine the
phasing of the drive ports which can null the induced
current during a subsequent scan.
|
4412. |
75 |
Parallel Transmit Pulse
Design with Implant-Friendly Modes
Yigitcan Eryaman1,2, Bastien Guérin3,
Elfar Adalsteinsson4,5, and Lawrence L. Wald2,5
1Research Laboratory of Electronics,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 2Martinos
Center for Biomedical Imaging, Dept. of Radiology, MGH,
Charlestown, MA, United States, 3Martinos
Center for Biomedical Imaging, Dept. of Radiology,
Massachusetts General Hospital, Charlestown, MA, United
States, 4Dept.
of Electrical Engineering and Computer Science,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 5Harvard-MIT
Health Sciences and Technology, Cambridge, MA, United
States
8 channel transmit array is used to reduce the SAR
around a generic implant model in a uniform phantom. For
this purpose, implant friendly modes of the transmit
array are generated. Using these modes, the optimum
least square 3-spokes pulse design solution is
calculated with an optimization approach which
explicitly constrains both global and local SAR.
|
4413. |
76 |
Reference-Free Detection of
RF Unsafe Conditions and Countermeasures for Implantable MR-Conditional
Devices
Ingmar Graesslin1, Johannes Achtzehn1,
Sascha Krueger1, Peter Vernickel1,
Kay Nehrke1, and Steffen Weiss1
1Philips Research, Hamburg, Germany
The presence of implants mostly represents a
contraindication for MR examinations. Here, a concept
based on parallel transmit technology is used to detect
potentially RF-unsafe conditions due to implants by
monitoring the currents in transmit coils. The original
method had required a reference scan without device.
Here, the method was improved by exploiting the
left-right symmetry of patients to avoid the reference
scan. Hence, the method may be applied to patients with
implants. Basic feasibility of this concept is shown in
pseudo-in vivo and phantom studies by detecting coupling
to pacemaker leads. Tip heating could be reduced by a
factor of 6 by changing the transmit phases.
|
4414. |
77 |
Evaluation of Virtual
Observation Points for Local SAR Monitoring of Multi-Channel
Transmit RF Coils at 7 Tesla
Andreas K. Bitz1, Rene Gumbrecht2,3,
Stephan Orzada1, Hans-Peter Fautz2,
and Mark E. Ladd1,4
1Erwin L. Hahn Institute for MRI, University
of Duisburg-Essen, Essen, Germany, 2Siemens
Healthcare, Erlangen, Germany, 3Department
of Physics, Friedrich-Alexander University, Erlangen,
Germany, 4Department
of Diagnostic and Interventional Radiology and
Neuroradiology, University Hospital Essen, Essen,
Germany
SAR monitoring for multi-Tx approaches in high-field MRI
is a challenging task, since several channels have to be
supervised simultaneously and the local SAR has to be
considered. For computationally efficient online local
SAR supervision, the concept of virtual observation
points (VOPs) has been proposed. Here, VOP models are
evaluated for realistic exposure configurations with two
7T RF coils, an 8-channel head and an 8-channel body
coil, and for three body models. For all configurations
and for a reasonable maximum overestimation of about
10%, the number of VOPs is smaller than 500, for which
online monitoring has been successfully tested.
|
4415. |
78 |
Imaging Healthy Volunteers
and Patients with Tattoos or Permanent Make-Up at 7T: A
Retrospective Study
Yacine Noureddine1,2, Andreas K. Bitz1,3,
Markus Thürling1,4, Gregor Schaefers2,
Mark E. Ladd1,3, and Oliver Kraff1,3
1Erwin L.Hahn Institute for MRI, University
Duisburg-Essen, Essen, NRW, Germany, 2MR:comp
GmbH, MR Safety Testing Laboratory, Gelsenkirchen, NRW,
Germany, 3Department
of Diagnostic and Interventional Radiology and
Neuroradiology, University Hospital Essen, Essen, NRW,
Germany, 4Department
of Neurology, University Duisburg-Essen, Essen, NRW,
Germany
In this study we describe our experience with scanning
77 subjects with tattoos or permanent make-up at 7T.
Only local transmit RF coils were used in the individual
measurements. The results indicate safe imaging and no
adverse events.
|
4416. |
79 |
In Silico and in Vitro
Investigation of Temperature Elevation Close to an Aneurysm
Clip at 7T
Yacine Noureddine1,2, Oliver Kraff1,3,
Mark E. Ladd1,3, Karsten Wrede4,
Gregor Schaefers2, and Andreas K. Bitz1,3
1Erwin L.Hahn Institute for MRI, University
Duisburg-Essen, Essen, NRW, Germany, 2MR:comp
GmbH, MR Safety Testing Laboratory, Gelsenkirchen, NRW,
Germany, 3Department
of Diagnostic and Interventional Radiology and
Neuroradiology, University Hospital Essen, Essen, NRW,
Germany, 4Clinic
for Neurosurgery, University Hospital Essen, Essen, NRW,
Germany
For safety assessment of metallic aneurysm clips,
RF-induced heating resulting from electric field
elevations in the tissue is the major concern with
respect to patient safety. In this study, RF and thermal
simulations were conducted to determine RF heating
around an aneurysm clip at 7T using a simplified model
of the head. Validation of the simulations was supported
by RF field measurement.
|
4417. |
80 |
Circuit Model for Implant
Electrode and Lead-Electrode Impedance Matching
Volkan Acikel1,2 and
Ergin Atalar1,2
1Electrical and Electronics Engineering,
Bilkent University, Ankara, Turkey, 2National
Magnetic Resonance Research Center (UMRAM), Ankara,
Turkey
In our previous studies we modeled the lead of the
implant using Modified Transmission Line Method (MoTLiM)
and solve the Radio Frequency (RF) induced currents on
the leads. In this study we propose an electrode model
that completes our electromagnetic model of active
implants under MRI. Also with this model electrode and
lead impedances are defined considering their RF
scattering and tissue interaction properties. It is
shown that although putting series inductor between lead
and electrode is a working technique to prevent heating,
failure in choosing inductance value can cause increase
in the heating.
|
4418. |
81 |
Application of Rotating RF
Coil Array in B1 Shimming
with Strict Local SAR Constraints
Jin Jin1, Feng Liu1, Adnan Trakic1,
Ewald Weber1, and Stuart Crozier2
1University of Queensland, St Lucia,
Queensland, Australia, 2The
University of Queensland, St Lucia, Queensland,
Australia
This work investigated the possibility of using a
four-channel mechanically rotating RF coil array (RRFCA)
to shim the transmit magnetic fields for high-field MRI,
while strictly controlling local specific absorption
rate (SAR). Since the numerous available degrees of
freedom in both the temporal and the spatial domain are
exploited, a notably higher excitation accuracy was
achieved compared with the stationary eight-channel
phased-array coil. The results of this primarily
theoretical study also signify that, since only a low
channel count is required, the geometry of RRFCA can be
optimised more flexibly to improve field penetration and
therefore excitation efficiency and performance.
|
4419. |
82 |
Safety Evaluation of
Algorithms to Optimize Transmit Efficiency for Local
Excitation with a Transmit Array
Giuseppe Carluccio1 and
Christopher Michael Collins1
1Radiology, New York University, New York,
New York, United States
Recently methods have been presented to provide optimum
driving currents for transmit array which maximize the
circularly polarized component of the B1 field in a
small region of interest and minimize the overall power
transmitted. In this work we evaluate safety parameters,
such as the 10g local average SAR and temperature
increase, when some of these methods are applied for two
different regions of interest in a body-sized transmit
array.
|
4420. |
83 |
Online Local SAR
Supervision for Transmit Arrays at 7T
Rene Gumbrecht1,2, Ulrich Fontius1,
Holger Adolf1, Thomas Benner1,
Franz Schmitt1, Elfar Adalsteinsson3,4,
Lawrence L. Wald4,5, and Hans-Peter Fautz1
1Siemens Healthcare, Erlangen, Germany, 2Department
of Physics, University of Erlangen, Erlangen, Germany, 3Electrical
Engineering and Computer Science, Massachusetts
Institute of Technology, Cambridge, MA, United States, 4Harvard-MIT
Division of Health Sciences and Technology,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 5Department
of Radiology, A. A. Martinos Center for Biomedical
Imaging, Charlestown, MA, United States
Local SAR and its safe supervision are dominant limiting
factors for high performance pTX applications. Safety
concepts typically provide a SAR prediction strategy
based on numerical simulations. Only sequences that will
not exceed SAR limits are allowed to be executed. To
capture system fault during runtime, an additional
online supervision was proposed to ensure the correct
execution of the RF pulses. In this study, an online
local SAR calculation and supervision system using
measured RF pulse shapes was built. Local SAR is
calculated in near real-time based on a compressed
electric field model of the current coil and patient.
|
4421.
|
84 |
On Patient-Specific Models
for Accurate SAR Estimations at 7 T
Jin Jin1, Feng Liu1, Ewald Weber1,
and Stuart Crozier2
1University of Queensland, St Lucia,
Queensland, Australia, 2The
University of Queensland, St Lucia, Queensland,
Australia
To more effectively predict the patient SAR values for
high-field MRI applications, a novel approach of
creating patient-specific tissue model is demonstrated
in this study. First, a matching model from a
high-resolution image- and tissue-library is selected
for the target patient. The library tissue distribution
is then warped to match the patient’s geometry as the
corresponding library image is registered to the
low-resolution scans of the patient. Results from
studying the models’ 1-gram SAR distribution using
finite-different time-domain method suggest that the
developed patient model can predict regions of elevated
SAR within the patient with remarkable accuracy.
|
|
|
ELECTRONIC POSTER
SESSION • ENGINEERING
Thursday, 25 April 2013 (11:30-12:30) Exhibition Hall |
|
Computer # |
|
4422. |
85 |
Functional MRI Using
Implanted Stereotactic EEG Electrodes --- Heating
Investigation for Safety
Pallab K. Bhattacharyya1, Jorge A.
Gonzalez-Martinez2, Mark J. Lowe1,
Dileep Nair1, Richard Prayson1,
Myron Zhang1, and Stephen E. Jones1
1Cleveland Clinic, Cleveland, OH, United
States, 2Cleveland
Clinic Foundation, Cleveland, OH, United States
As a safety measure of performing MRI with implanted
stereotactic EEG (SEEG) electrodes, we investigated the
effect on electrode heating by the geometry and
configuration of the electrodes and associated
connections by measuring radio frequency induced heating
of a SEEG electrode embedded within a phantom inside an
intraoperative MRI suite, with the electrode directly
connected to stimulus hardware located in the control
room. Once the safety of the procedure was ensured, fMRI
scans were performed on human subjects with implanted
SEEG electrodes as well. No damage to brain slice was
confirmed by pathology.
|
4423. |
86 |
Safe Online Local SAR
Calculation for Transmit Arrays Using Asynchron Data
Processing
Rene Gumbrecht1,2, Thomas Benner1,
Ulrich Fontius1, Holger Adolf1,
Andreas K. Bitz3, and Hans-Peter Fautz1
1Siemens Healthcare, Erlangen, Germany, 2Department
of Physics, Friedrich-Alexander University, Erlangen,
Germany, 3Erwin
L. Hahn Institute for Magnetic Resonance Imaging,
University Duisburg-Essen, Essen, Germany
A RF supervision concept that considers local SAR is an
essential pre-requisite for high performance pTX
applications. It needs to provide at least one online RF
supervision component that detects violation of local
SAR limits in real-time. Using the MR signal processing
unit for that fails to match real-time requirements. In
this study, an online local SAR calculation was
implemented on the image reconstruction PC of a MRI
system. To overcome latency issues, a real-time logic in
conjunction with a real-time global SAR supervision unit
is used to guarantee patient safety at all times
|
4424. |
87 |
A Novel Method for
Experimental Assessment of Antenna Safety Using MR
Thermometry
Cem Murat Deniz1,2, Leeor Alon2,3,
Xing Yang3, Gene Young Cho1,2,
Christopher Michael Collins2,3, Daniel K.
Sodickson2,3, and Yudong Zhu2,3
1Department of Radiology, Bernard and Irene
Schwartz Center for Biomedical Imaging, New York
University, New York, NY, United States, 2Sackler
Institute of Graduate Biomedical Sciences, New York
University School of Medicine, New York, NY, United
States, 3Department
of Radiology, Bernard and Irene Schwartz Center for
Biomedical Imaging, New York University School of
Medicine, New York, NY, United States
Even though simulations have been commonly used for
safety evaluation of antenna structures, often complex
antenna-body structures are difficult to simulate and
the simulations do not align with the experimental
results. In this work, we propose an alternative method
of monitoring RF safety by experimentally measuring the
temperature change using MR thermometry. We show good
agreement between optical temperature measurements and
MR thermometry and simulation studies. Heating pattern,
|E| and SAR fields from a simulated simple dipole
antenna matches well with experimental results both in
pattern and in magnitude. Strength of this method is in
its simplicity, accuracy and speed.
|
4425. |
88 |
Predicting Long-Term
Temperature Increase from Time-Dependent SAR Levels with a
Single Short-Term Temperature Response
-permission withheld
Giuseppe Carluccio1, Zhipeng Cao2,
and Christopher Michael Collins1
1Radiology, New York University, New York,
New York, United States, 2Bioengineering,
Pennsylvania State University, Hershey, Pennsylvania,
United States
In a patient exam typically a series of pulse sequences
with very different power levels and SAR levels are
applied. Temperature increase caused by SAR absorption
may damage tissues, and development of fast temperature
computation methods could be useful. We propose a rapid
method to predict temperature increase over an entire
patient exam with time-varying SAR levels applied, after
characterizing the temperature tissue response to a
short SAR pulse.
|
4426. |
89 |
Safely Assessing RF Heating
Potential of Conductive Devices Using Image-Based Current
Measurements
Greg Griffin1, Kevan Anderson2,
and Graham A. Wright1,2
1Medical Biophysics, University of Toronto,
Toronto, Ontario, Canada, 2Imaging
Research, Sunnybrook Research Institute, Toronto,
Ontario, Canada
This study presents a technique to safely assess the RF
heating potential of a conductive device in the MR bore.
RF artifacts acquired with a low power sequence are
analyzed and used to predict heating induced by a higher
power sequence. Experimental validation in a phantom is
presented for several configurations.
|
4427. |
90 |
Scanner Specific
Relationship Between Displayed SAR and Heating Adjacent to
Deep Brain Stimulation Leads: An in Vitro Study
Robert Dawe1 and
Leo Verhagen Metman1
1Rush University Medical Center, Chicago, IL,
United States
The guideline for head specific absorption rate (SAR)
during MRI of patients with implanted deep brain
stimulation (DBS) leads is difficult to interpret
because SAR estimation techniques vary among scanner
manufacturers. The purpose of this study was to
establish a relationship between displayed head SAR and
local tissue heating in our particular scanner. Heating
of approximately 4.9˚ C per 1.0 W/lb. of displayed head
SAR was observed, confirming that potentially dangerous
heating of DBS leads is possible. At the same time, the
results suggest that the SAR guideline for implanted DBS
leads may be overly conservative.
|
4428. |
91 |
Mobile Phone RF Safety
Testing Using Magnetic Resonance Imaging
Leeor Alon1,2, Cem Murat Deniz1,2,
Gene Young Cho1,2, Xing Yang1,
Christopher Collins1, Yudong Zhu1,2,
and Daniel Sodickson1,2
1Department of Radiology, Bernard and Irene
Schwartz Center for Biomedical Imaging, New York
University, New York, NY, United States, 2Sackler
Institute of Graduate Biomedical Sciences, New York
University, New York, NY, United States
Local specific absorption rate (SAR) quantification is
important for ensuring safe usage of RF transmitting
devices. Regulatory bodies around the world mandate RF
safety testing and more than $1B is spent annually on
testing. In contrast to the gold-standard measurement
schemes that mechanically move a probe and measure the
electric field generated by the wireless devices, this
work utilizes MR thermometry to quantify power
deposition from mobile phones, allowing rapid
acquisitions with higher resolution than before. The
results indicate that safety testing of RF emitting
devices such as mobile phones can be robustly conducted
using MRI.
|
4429. |
92 |
B1-Based SAR Estimation for
Human Brain Imaging with Average Brain Property Values
Substitution
Xiaotong Zhang1, Jiaen Liu1,
Sebastian Schmitter2, Pierre-Francois Van de
Moortele2, and Bin He1
1Department of Biomedical Engineering,
University of Minnesota, Minneapolis, MN, United States, 2Center
for Magnetic Resonance Research, University of
Minnesota, Minneapolis, MN, United States
Recently proposed Electrical Properties Tomography
technique, which aims to reconstruct the electrical
properties of biological tissues via measured B1,
provides a feasible way for quantitative
subject-specific SAR estimation. However, Laplacian over
B1 is involved in the inverse algorithm, which may
severely suffer from noise-contamination in measured B1,
and deteriorate the reliability of SAR estimation. In
this study, by using multi-channel coil at 7T for brain
imaging, the performance of local SAR calculation has
been evaluated when literature-reported average
electrical property and mass density values are employed
as a substitution, which intends for a more robust and
reliable SAR estimation approach.
|
4430. |
93 |
A Novel MRI Based
Electrical Properties Measurement Technique
Volkan Acikel1,2, Oytun Ulutan1,
Ali Caglar Ozen1,2, Burak Akin2,
Yigitcan Eryaman3,4, and Ergin Atalar1,2
1Electrical and Electronics Engineering,
Bilkent University, Ankara, Turkey, 2National
Magnetic Resonance Research Center (UMRAM), Ankara,
Turkey, 3Research
Laboratory of Electronics, Massachusetts Institute of
Technology, Cambridge, MA, United States, 4Martinos
Center for Biomedical Imaging, MGH, Charlestown, MA,
United States
Most of the Magnetic Resonance Imaging studies require
accurate knowledge of electrical properties of the
phantoms. There are several techniques for measuring
electrical properties of phantoms but there is lack of
accuracy analysis. A new method to measure electrical
properties of gel phantom was proposed based on the lead
tip heating phenomena and used for verification of home
made coaxial transmission line measurement fixture.
Electrical properties of 3 different gel phantoms were
measured using both methods and results are compared.
Although the proposed method requires long MRI
experiment, it can be used for verification of other
electrical properties measurement techniques.
|
4431. |
94 |
Efficient Method to
Evaluate the Heating Risk of Coils
Lin Liu1, Liang Xuan2, and Qiong
Zhang3
1MR, GE Company, Beijing, Beijing, China, 2MR,
GE company, Waukesha, WI, United States, 3MR,
GE company, Beijing, Beijing, China
Safety requirements, has been as MRI system first and
foremost of indicators. In direct contact with patients
or volunteers parts, the coil is undoubtedly the most in
need of extra attention. Thermal test is one of the
important coil integrated tests and coil security
evaluation. In order to effectively assess the
performance of the coil, engineers usually use worse
clinical protocol scanning coils up to 2.5 hours; and
during the entire testing process, the highest coil
surface temperature should not be higher than 41
degrees. To save test time , evaluate and avoid damage
caused by overheating of the coil in the testing
process, we have designed a new method that is based on
using the 1st one hour measured data for fitting and
extrapolation, thus, the whole test can be shorten to
only one hour, and also the risk of damaging coil is
reduced. Validation against 44coils with 52 modes using
two scanners has proved that this method can effectively
assess coil characteristics and reduce overheating risk.
|
4432. |
95 |
Development of a
Patient-Specific, Three-Dimensional, Heterogeneous, Radio
Frequency Thorax Phantom
Clifton Haider1, David Holmes1,
Robert Sainati1, Daniel Schwab1,
Daniel V. Litwiller2, Desmond Yeo3,
and Barry Gilbert1
1Biomedical Engineering, Mayo Clinic,
Rochester, MN, United States, 2Global
Applied Science Laboratory, GE Healthcare, Rochester,
MN, United States, 3GE
Global Research, GE, Niskayuna, NY, United States
Methods were developed to create phantoms with specific
and highly repeatable conductivity and permittivity
matching those of desired tissue. Numerical models
results were highly similar with experimental results.
Complex 3D organ models accurately represented the
target anatomy and RF material properties.
|
4433. |
96 |
A Simple and Conservative
Method to Include Variation of Core Blood Temperature with
Pennes’ Bioheat Equation
Giuseppe Carluccio1 and
Christopher Michael Collins1
1Radiology, New York University, New York,
New York, United States
Pennes’ bioheat equation is one of most used models to
predict temperature increase in body tissues. We present
a simple and conservative method to include the
variation of blood temperature increase in the Pennes’
bioheat equation due to the whole-body SAR absorption.
Plots are provided showing the core blood temperature
over time, and the spatial temperature increase
distribution for a cross-section of the body obtained
with both the Pennes’ bioheat equation having constant
value of blood temperature, and the method with the time
dependent blood temperature here presented.
|
|
|
|