|
Computer # |
|
3212. |
49 |
Q matrix approach to
control implant heating by transmit array coils
Frank Seifert1, Gerd Weidemann1,
and Bernd Ittermann1
1Physikalisch-Technische Bundesanstalt (PTB),
Braunschweig und Berlin, Germany
The hazard of excessive local tissue heating due to
metallic implants can be mitigated when utilizing the
additional degrees of freedom of a transmit coil array.
In this proof of principle study we used the single
Q-matrix associated with the proximate tissue volume of
a metallic implant to determine the 'worst case'
steering conditions for implant heating. From a simple
projection of the voltage vector to be applied to a sub
space which is orthogonal to this 'worst case' voltage
vector new steering conditions are determined which
result in much lower implant heating. To demonstrate the
efficacy of this method we performed thermal simulations
using a 7T 8-channel head coil model. This approach can
simply be embedded into B1-shimming procedures.
|
3213.
|
50 |
Local SAR elevations in the
human head induced by high-permittivity pads at 7 Tesla
Thomas M Fiedler1, Mark E Ladd1,2,
and Andreas K Bitz1
1Medical Physics in Radiology, German Cancer
Research Center (DKFZ), Heidelberg, Germany, 2Erwin
L. Hahn Institute for Magnetic Resonance Imaging,
University Duisburg-Essen, Essen, Germany
It has been shown that specific configurations of
high-permittivity pads can lead to an increase of local
SAR. The electric field distribution depends on several
parameters, e.g. polarization and distribution of the
incident field, tissue distribution, and the actual
geometry and position of the pad as well as its
permittivity. In this work, we investigate several
combinations of these parameters by RF field simulations
in two body models and with two excitation scenarios to
determine under which conditions SAR elevations are
likely to occur or can be avoided.
|
3214. |
51 |
A Comparison Between
Three-Point Dixon Sequences and Label Fusion Techniques for
Water-Fat Separation in High-Field MRI Local SAR Estimation
Angel Torrado-Carvajal1,2, Esra A. Turk2,3,
Joaquin L. Herraiz2,3, Yigitcan Eryaman2,4,
Juan A. Hernandez-Tamames1,2, Elfar
Adalsteinsson5,6, Larry L. Wald4,6,
and Norberto Malpica1,2
1Medical Image Analysis and Biometry Lab,
Universidad Rey Juan Carlos, Mostoles, Madrid, Spain, 2Madrid-MIT
M+Vision Consortium, Madrid, Spain, 3Research
Laboratory of Electronics, Massachusetts Institute of
Technology, Cambridge, MA, United States, 4Martinos
Center for Biomedical Imaging, Dept. of Radiology, MGH,
Charlestown, MA, United States, 5Dept.
of Electrical Engineering and Computer Science,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 6Harvard-MIT
Health Sciences and Technology, Massachusetts Institute
of Technology, Cambridge, MA, United States
In this work we compare the results of B1+ field and SAR
distribution obtained by using patient-specific 3PD
images and two label fusion estimation approaches over a
T1-weighted volume for fat and water segmentation. B1+
field distributions were found to be almost the same for
the three models. An IDEAL label-fusion approach
provides very similar SAR distribution results to the
patient-specific approach. A CT label-fusion approach
provides an increased SAR distribution map. The use of
label fusion techniques to estimate the fat and water
separation in MRI images allows an accurate segmentation
with a similar accuracy as patient-specific 3PD
sequences.
|
3215. |
52 |
Local SAR Estimation for
Parallel RF Transmit at 7T Using Directional Couplers
Matthew Restivo1, C.A.T van den Berg1,
Alexander Raaijmakers1, Peter Luijten1,
and Hans Hoogduin1
1University Medical Center Utrecht, Utrecht,
Netherlands
Parallel RF Transmit (pTx) in high field MRI is limited
by the difficulty in correctly estimating the local
specific absorption rate (SAR) given the complicated
interferences of the electric field distributions
produced by the various elements. Directional couplers
allow for current to be measured easily on the
transmission line, but parallel matching circuitry
transforms the phase and amplitude of the current before
entering the element. We propose a method for converting
directional coupler values to tx element current values,
thus allowing for the correct scaling and linear
combination of simulated fields. This is necessary in
order to accurately predict local SAR without overly
complex simulations.
|
3216. |
53 |
Anatomical Models of
Pregnant Women in 3T pTx Body Coils: Evaluation of SAR and
B1+ Optimization in Various Imaging Positions
Manuel Murbach1, Esra Neufeld1,
Eugenia Cabot1, Earl Zastrow1,
Juan Corcoles2, Wolfgang Kainz3,
and Niels Kuster1,4
1ITIS Foundation, Zurich, Switzerland, 2Department
of Electronic and Communication Technology, Universidad
Autónoma de Madrid (UAM), Madrid, Spain,3Center
for Devices and Radiological Health (CDRH), US Food and
Drug Administration (FDA), Silver Spring, MD, United
States, 4Swiss
Federal Institute of Technology (ETH), Zurich,
Switzerland
In this study, we investigate the effects of pTx
multitransmit body-coils at 3 T on anatomical models of
a pregnant woman with her fetus at three different
stages of gestation. Various imaging positions, regions
of interest (ROI), and pTx coil excitation vectors are
evaluated for effect on SAR and B1+ uniformity. The B1+
uniformity can increase by a factor of >4. The more
equalized field distribution for large ROI typically
leads to lower local SAR. However, the worstcase
excitations identified lead to extremely high local SAR
enhancement, emphasizing the need for robust management
of exposure safety.
|
3217. |
54 |
Multi-body-model Method for
Design of Mismatch-insensitive SAR-aware Parallel Transmit
RF Pulses
Mihir Pendse1 and
Brian Rutt1
1Radiology, Stanford University, Stanford,
CA, United States
Patient local SAR is usually controlled by way of
estimations from numerical simulations on a library of
tissue models. However, due to the variability between
subjects in both anatomy shape and positioning within
coil, mismatch between patient and model must be taken
into account. We demonstrate that by using multiple body
models with moderate degrees of mismatch it is possible
to achieve significantly lower SAR than with a single
model. The implication of this result is that, despite
significant intersubject variability, relatively sparse
body model libraries can be used without sacrificing
robustness of SAR estimation.
|
3218. |
55 |
Temperature Sensor Implant
for Analysis of RF Safety of Active Implantable Medical
Devices under MRI
Berk Silemek1,2, Volkan Acikel1,2,
and Ergin Atalar1,2
1Bilkent University, Ankara, Turkey, 2Umram,
Ankara, Turkey
The aim of this study is to present a prototype of the
implantable temperature sensor which will be used in
animal studies to emulate the situation of a patient
with an Active Implantable Medical Device inside the MRI
and investigate possible tissue damage due to the RF
induced tissue heating. The prototype records
temperature information during MRI scan and transmits
the desired data wirelessly, while implant is still
inside of the body, to the remote without any surgery
operation.
|
3219. |
56 |
A Phantom Designed
Specifically for Local SAR Validation - permission withheld
Matthew Restivo1, Ronald Mooiweer1,
C.A.T van den Berg1, Alexander Raaijmakers1,
Frank Simonis1, Peter Luijten1,
and Hans Hoogduin1
1University Medical Center Utrecht, Utrecht,
Netherlands
We built a phantom which exhibits significant localized
heating and provides good conditions for measuring
temperature rise with MR thermometry and/or using fiber
optic temperature probes. Measurements with this phantom
allow for local SAR validation for RF coil models, a
necessary step prior to relying on SAR simulations for
in-vivo safety decisions.
|
3220. |
57 |
An approach to
Temperature-based Virtual Observation Points for Safety
Assurance and Pulse Design
Giuseppe Carluccio1,2, Cem Murat Deniz1,2,
and Christopher Michael Collins1,2
1Radiology, Center for Advanced Imaging
Innovation and Research (CAI2R), New York University
School of Medicine, New York, New York, United States,2Radiology,
Bernard and Irene Schwartz Center for Biomedical
Imaging, New York University, New York, New York, United
States
Safety guidelines are provided by IEC to set maximum
temperature and maximum SAR achievable during MRI scans.
SAR-based safety evaluation of multichannel transmit
systems has been accelerated using Virtual Observation
Points to enable rapid RF safety monitoring and RF pulse
design based on SAR constraints. In this work, using a
matrix formalism to compute temperature, we propose a
method to identify temperature based VOPs. Temperature
based VOPs enable fast parallel transmit RF pulse design
with iterative or non-iterative methods based on strict
temperature constraints.
|
3221. |
58 |
Torque and Translational
Force Estimation for Ferromagnetic Objects: the Saturation
Effect
Vahid Ghodrati1 and
Abbas Nasiraei Moghaddam1,2
1BME, Amirkabir University of Technology
(Tehran Polytechnic), Tehran, Tehran, Iran, 2School
of Cognitive Sciences, Institute for Research in
Fundamental Sciences (IPM), Tehran, Iran
Ferromagnetic objects in the MR-environment experienced
linear magnetic region or saturation region. almost
ferromagnetic objects placed in the saturation region in
the commercial MR-scanners. Therefore, it is necessary
to considered saturation region for these objects to
have a better estimation of the torque and force that
acted on ferromagnetic objects. Here, the approximated
strength for the magnetic field that can be saturated
the almost ferromagnetic objects will be presented.
Also, it will show that just by using the linear region,
the approximation of force will be 1.5 times
overestimated in comparison by considering of real
states for these objects.
|
3222. |
59 |
Experiments and Analysis of
Virtual Observation Points at 7T
Yujuan Zhao1, Tiejun Zhao2, and
Tamer Ibrahim1
1University of Pittsburgh, Pittsburgh,
Pennsylvania, United States, 2Siemens
Medical Solutions USA, Pittsburgh, Pennsylvania, United
States
PTX excitation may create distinct hot spots inside the
human body by constructive interference of electric
fields generated by the respective coil elements driven
with individual amplitudes and phases. Monitoring global
and local peak SAR is a challenging task. In this study,
the system global power and the local SAR from Virtual
Observation Points (VOP) model in conjunction with
rigorous RF modeling that incorporates coupling are
demonstrated and verified in-vivo by experiments
acquired using the Parallel RF transmission system. The
input simulated SAR model is also verified
quantitatively using B1+ maps as well as local VOP SAR
monitoring.
|
3223. |
60 |
Breast Tissue Expanders
with Magnetic Ports: Clinical Experience at 1.5-Tesla
Nanda Deepa Thimmappa1, Christina Y Ahn2,
Silvina P Dutruel1, Joshua L Levine3,
Srikanth Reddy Boddu1, and Martin R Prince1
1Radiology, Weill Cornell Medical College,
New York, NY, United States, 2NY
Langone Medical Center, Department of Plastic Surgery,
New York, New York, United States, 3Department
of Plastic Surgery, New York Eye and Ear Infirmary of
Mount Sinai, New York, New York, United States
There are anecdotal reports of tissue expander damage or
dislodgement during MRI and are currently labeled “MR
Unsafe”. At our institution, abdomen-pelvis MRI/MRA is
frequently requested in these patients to map perforator
vessels supplying abdominal fat/skin to plan autologous
breast reconstruction. Since these tissue expanders will
be removed at the time of breast reconstruction,
surgeons performing these operations feel that the
benefit of pre-operative MRA outweighs the risk of any
implant migration or local tissue damage potentially
caused by MRI. In this study we retrospectively review
breast tissue expander safety for 73 patients who
underwent abdominal and pelvic MR for autologous breast
reconstruction.
|
3224. |
61 |
An Algorithm for
Maximum-SAR Targeted RF Hyperthermia
Mihir Pendse1 and
Brian Rutt1
1Radiology, Stanford University, Stanford,
CA, United States
While shorter electromagnetic wavelengths and resulting
SAR hotspots are a source of concern in ultra-high field
MRI, the same phenomenon can be desirable in the field
of targeted RF hyperthermia. In this work, we describe
the “maxSAR” pulse design problem and implement a
nonconvex interior-point algorithm to solve it. We
demonstrate in simulation that an 8-loop pTx array
operating at 298 MHz (7T), can be used to intentionally
produce localized hotspots at arbitrarily located target
regions while still constraining the SAR at healthy
voxels to be within the local SAR regulatory limit.
|
3225. |
62 |
Effect of 3T MRI noise on
adults hearing observed by the dynamic auditory brainstem
response test
Huan Li1, Yan An1, Qinli Sun1,
Yanyan Li1, Pan Cao1, Miaomiao
Wang1, Jianxin Guo1, and Jian Yang1
1Department of Radiology, The First
Affiliated Hospital of Medical College,Xi'an Jiaotong
University, Xi'an, Shaanxi, China
Purpose Magnetic resonance imaging (MRI) as a diagnostic
method have been widelyPurpose Magnetic resonance
imaging (MRI) as a diagnostic method have been widely
clinical application for over twenty years. A prominent
disadvantage of MRI is the high level of noise produced
during scanning process. The noise is hazardous for
patients and operators. It may cause the temporary or
permanent hearing threshold shift. This study is aimed
to explore whether the noise of 3T MRI has an influence
on normal adults hearing. Methods We performed head MRI
to 29 healthy volunteers who had passed the Auditory
Brainstem Response (ABR) hearing test . We performed the
ABR audiometry within 24 hours before the MRI
examination for the first time, and carry on the second
audiometry within 20 minutes after the MRI scanning,
Among all the 29 cases, 26 showed a hearing threshold,
and we review their hearing in 25 days later. The
results were applied Wilcoxon matched-pairs test, SPSS
18. Results There was significant difference of left and
right ear hearing between before and 20 minutes after
MRI (p<0.01) and no significant difference of both ear
hearing between before MRI and the review in 25 days
later. Conclusion The noise lasted close to 1 hour from
3T MRI scanning has a transient effect on adults
hearing.clinical application for over twenty years. A
prominent disadvantage of MRI is the high level of noise
produced during scanning process. The noise is hazardous
for patients and operators. It may cause temporary or
permanent hearing threshold shift. This study is aimed
to explore whether the noise of 3T MRI has an influence
on adults hearing. Methods We performed head 3T MRI to
those 29 cases of healthy volunteers who had passed the
Auditory Brainstem Response (ABR) hearing test . We
performed the ABR audiometry within 24 hours before the
MRI examination for the first time, and carry on the
second audiometry within 20 minutes after the MRI
scanning, and we review their hearing 25 days later. The
results were applied Wilcoxon matched-pairs test, SPSS
18. Results The left and right ear¡¯s hearing before and
after MRI , the tests were taken to be significant at
p<0.01. There have not significantly difference in both
ear¡¯s hearing before MRI and the review. Conclusion The
noise that 3T MRI produced has only a transient effect
on adults hearing.
|
3226. |
63 |
RF-induced heating in MRI
of tissue around an aneurysm clip near the middle cerebral
artery at 7 T under consideration of the Pennes bioheat
equation
Yacine Noureddine1,2, Oliver Kraff1,
Mark E. Ladd1,3, Karsten Wrede4,
Gregor Schaefers2, and Andreas K. Bitz3
1Erwin L. Hahn Institute for MRI, University
Duisburg-Essen, Essen, NRW, Germany, 2MR:comp
GmbH, MR Safety Testing Laboratory, Gelsenkirchen, NRW,
Germany, 3Medical
Physics in Radiology, German Cancer Research Center
(DKFZ), Heidelberg, BW, Germany, 4Clinic
for Neurosurgery, University Hospital Essen, Essen, NRW,
Germany
Synopsis: For safety assessment of metallic aneurysm
clips, RF-induced heating resulting from electric field
elevations in the tissue is the major concern with
respect to patient safety. In this study, RF and thermal
simula-tions were conducted to determine RF heating
around an aneurysm clip at 7T with a standard RF
transmit coil using both a homogeneous and heterogeneous
model of the head.
|
3227. |
64 |
MRI planning for SAR
management in pTx systems
Joaquin L. Herraiz1, Yigitcan Eryaman1,2,
Esra Abaci Turk1, Angel Torrado-Carvajal1,3,
Adrian Martin1,4, Emanuele Schiavi1,4,
Bastien Guerin5, Elfar Adalsteinsson1,6,
Lawrence L. Wald5,7, Juan A.
Hernandez-Tamames1,3, and Norberto Malpica1,3
1Madrid-MIT M+Vision Consortium in RLE,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 2Center
for Magnetic Resonance Research,Department of Radiology,
University of Minnesota, Minneapolis, MN, United States, 3Medical
Image Analysis and Biometry Laboratory, Universidad Rey
Juan Carlos, Madrid, Spain, 4Dept.
of Applied Mathematics, Universidad Rey Juan Carlos,
Mostoles, Madrid, Spain, 5Martinos
Center for Biomedical Imaging, Dept. of Radiology, MGH,
Charlestown, MA, United States, 6Dept.
of Electrical Engineering and Computer Science,
Harvard-MIT Health Sciences and Technology,
Massachusetts Institute of Technology, Cambridge, MA,
United States, 7Harvard-MIT
Health Sciences and Technology, Massachusetts Institute
of Technology, Cambridge, MA, United States
Current approaches to acquire clinical 7-T MR data
within acceptable SAR limits require imaging with
reduced flip angle and/or number of slices, which
reduces SNR/Contrast/Detectability, thereby undermining
the potential benefits of high-resolution imaging.
Parallel transmission RF pulse design has recently shown
to be able to manage global and local SAR in 7-T systems
if a correct patient head and torso model is known.
Nevertheless, it is unclear if patient-specific SAR
reduction methods could be implemented in a feasible way
in the clinical practice in the near future. In this
work, we propose an effective protocol for enabling
patient-specific SAR modeling and reduction to utilize
the full capability and clinical promise of 7-T MR
imaging.
|
3228. |
65 |
Fast, Thermal Dose-based
Exposure Safety Supervision
Esra Neufeld1, Manuel Murbach1,
and Niels Kuster1,2
1IT'IS Foundation for Research on Information
Technologies in Society, Zurich, Switzerland, 2Swiss
Federal Institute of Technology (ETHZ), Zurich,
Switzerland
A fast method to estimate exposure safety during MRI
based on thermal dose (CEM43) has been developed that is
capable of considering the transient nature of heating,
exposure duration, local SAR, and the impact of
perfusion as well as thermoregulation. Thus it overcomes
problems associated with traditional SAR-based limits.
The assumptions and relationships underlying the model
have been verified and necessary constants derived using
simulations. Comprehensive uncertainty analysis has been
performed and the predictions have been validated
against measurement data and detailed simulations of
realistic scan sequences and exposure scenarios.
|
3229. |
66 |
Signal changes in dentate
nuclei with 10 or more gadolinium-based contrast
administrations: comparison of linear versus macrocytic
contrast agents
Daisy Q Huang1, Martin Prince1,
George Shih1, and Yan Cao1
1Radiology, New York Presbyterian
Hospital/Weill Cornell, NY, New York, United States
This retrospective study explores the correlation
between high T1 signal in dentate nuclei and multiple Gd
contrast administrations. Signal changes vary with type
of contrast agent used, suggesting a multifactorial
basis beyond just number of Gd administrations.
|
3230. |
67 |
Radiofrequency-induced
heating of intracranial stereo-EEG electrodes during MRI: a
phantom study
Annie Papadaki1,2, David Carmichael3,
Mark James White1,2, Hoskote Chandrashekar1,
Tarek Yousry1,2, Beate Diehl4,5,
Louis Lemieux4, and John Stephen Thornton1,2
1Lysholm Department of Neuroradiology,
National Hospital for Neurology and Neurosurgery, UCLH,
London, London, United Kingdom, 2Department
of Brain Repair and Rehabilitation, UCL Institute of
Neurology, London, London, United Kingdom, 3Imaging
and Biophysics Unit, UCL Institute of Child Health,
London, United Kingdom, 4Department
of Clinical and Experimental Epilepsy, UCL Institute of
Neurology, London, United Kingdom, 5Department
of Neurophysiology, National Hospital for Neurology and
Neurosurgery, UCLH, London, United Kingdom
In this study we assessed temperature changes (∆T)
during MRI at the tip of Stereo-electroencephalography
(SEEG) electrodes in a phantom. Fibre optic thermometry
measurements at the tip of two electrodes were performed
at 1.5T and 3T, during a high SAR TSE sequence and an
fMRI EPI sequence. Different electrode tail terminations
were tested to match possible clinical scenarios.
Scanning with electrode tails only, the maximum ∆T did
not exceed 1°C at 1.5T, while it exceeded the guideline
limits at 3T. Connecting extension cables and varying
the cable termination has a significant effect in
RF-induced heating
|
3231. |
68 |
Trial of safe working
procedure against occupational SMF exposure - evaluation of
its effectiveness in occupational SMF exposure levels and
work performances among 3 T MRI system users -
Sachiko Yamaguchi-Sekino1, Masaki Sekino2,
and Toshiharu Nakai3
1National Institute of Occuaptional Safety
and Health, Japan, Kawasaki, Kanagawa, Japan, 2Graduate
School of Engineering, The University of Tokyo, Tokyo,
Japan, 3Neuroimaging
& Informatics, National Center for Geriatrics and
Gerontology, Aichi, Japan
Occupational exposure to the static magnetic field (SMF)
is an arising problem in operators of magnetic resonance
imaging (MRI) system. Although the individual motion
control is recommended for MR workers to prevent these
temporal effects, more feasible safe working procedure
is required to achieve the compliance monitoring. The
present study set 30 cm of the approach restricted area
from the end of 3 T MRI system instead of the individual
motion control. The average of the maximum SMF exposure
among subjects was decreased in 25 % by this action (t-test, p<0.01).
And motion analysis by motion capture indicated that
work performances were not affected under this action.
These results suggest that this simple safe working
procedure is applicative for MR workers to reduce
occupational SMF exposure without remarkable changes in
work performances. And the quantification of MR workers’
motions by motion capture provides the useful
information for electromagnetic field dosimetry.
|
3232. |
69 |
Effect of Cranial Fixation
Plates on Brain MR Imaging at 7T in Neurosurgical Patients
Bixia Chen1,2, Tobias Schoemberg1,2,
Oliver Kraff1, Andreas K. Bitz1,3,
Harald H. Quick1,4, Mark Edward Ladd1,3,
Ulrich Sure2, and Karsten Henning Wrede1,2
1Erwin L. Hahn Institute for Magnetic
Resonance Imaging, University Duisburg-Essen, Essen,
NRW, Germany, 2Department
of Neurosurgery, University Hospital Essen, University
Duisburg-Essen, Essen, NRW, Germany, 3Medical
Physics in Radiology, German Cancer Research Center
(DKFZ), Heidelberg, BW, Germany, 4High
Field and Hybrid MR Imaging, University Hospital Essen,
University Duisburg-Essen, Essen, NRW, Germany
The importance of ultra-high-field MRI in neuroimaging
is increasing rapidly. Titanium cranial fixation plates
(CFP) are commonly used during neurosurgical operations,
and have been considered MR-conditional in simulations
and head model studies. We evaluated imaging artifacts
in multiple sequences correlated to CFP in vivo. Five
patients were examined before, within 72 hours after,
and 3 months after surgery at 3T, and in a 7T whole-body
MR system using an 8-channel RF head coil. CFP caused
minor artifacts in TOF and MPRAGE at 7T, depiction of
adjacent brain tissue was impaired in SWI due to
susceptibility artifacts, comparable with 3T artifacts.
|
3233. |
70 |
RF safety assessment of a
bilateral 4-channel Tx/Rx 7T breast coil
Thomas M Fiedler1, Aaron S Kujawa1,
Frank Resmer2, Patrick Stein2,
Titus Lanz2, Mark E Ladd1,3, and
Andreas K Bitz1
1Medical Physics in Radiology, German Cancer
Research Center (DKFZ), Heidelberg, Germany, 2RAPID
Biomedical GmbH, Rimpar, Bavaria, Germany, 3Erwin
L. Hahn Institute for Magnetic Resonance Imaging,
University Duisburg-Essen, Essen, Germany
Numerical simulations for a bilateral 4-channel
transmit/receive breast coil were performed to
investigate the optimum excitation mode of the coil and
to determine the maximum permissible input power. An
exact model of the breast coil was generated and
validated by a comparison of measured and simu-lated B1+
distributions in a phantom. An anatomical heterogeneous
body model was used to obtain realistic specific
absorption rates (SAR). B1+ shimming was performed to
find an excitation mode with optimal B1+ homogeneity.
|
3234. |
71 |
Direct SAR Mapping by
Thermoacoustic Imaging: Experimental Proof-of-Concept
Simone Angela Winkler1, Paul Picot2,
Michael Thornton2, and Brian K Rutt1
1Dept. of Radiology, Stanford University,
Stanford, CA, United States, 2Endra
Inc., Ann Arbor, Michigan, United States
We propose a new concept for direct measurement of
specific absorption ratio (SAR), to be used as a safety
assessment / monitoring tool for MRI. The concept uses
short bursts of RF energy and measures the resulting
thermoacoustic excitation pattern by an ultrasound
transducer array, followed by image reconstruction to
yield the 3D SAR distribution. We conducted an
experimental study to test the feasibility of this novel
concept for direct SAR mapping. The promising results
encourage us to further develop the method for
application to in-vivo local SAR mapping.
|
3235. |
72 |
An Investigation on IEC
Head SAR Limit on Orbit Heating
Xin Chen1, Charles Poole2, Michael
Steckner1, and Robert Brown2
1MR, Toshiba Medical Research Institute USA,
Inc., Mayfield Village, OH, United States, 2Department
of Physics, Case Western Reserve University, Cleveland,
OH, United States
Numerical simulations are used to investigate thermal
impacts of 3.2W/kg head SAR limit for brain imaging. No
significant SAR or temperature rise are observed in
brain tissues or eyes.
|
|