16:00 |
0489. |
Motion-insensitive sequence
for single-voxel determination of B1+ by
Bloch-Siegert shift in moving organs including the human
heart
Ayse Sila Dokumaci1, Bertrand Pouymayou1,
Roland Kreis1, and Chris Boesch1
1Depts. Radiology and Clinical Research,
University Bern, Bern, Switzerland
The determination of B1+ in moving organs, in particular
in the heart, is a largely unmet problem. Most B1+
mapping sequences are too long and/or cannot be acquired
in a (double-) triggered mode. We developed and
validated a single-voxel PRESS sequence based on the
Bloch-Siegert shift which is largely immune against
effects of motion and which can additionally be
double-triggered. Measurements in vitro and in the human
heart show an excellent performance of the sequence.
|
16:12 |
0490. |
Large Dynamic Range
Relative B1+ Mapping
Francesco Padormo1, Aaron T. Hess2,
Paul Aljabar1, Peter Jezzard3,
Matthew D. Robson2, Joseph V. Hajnal1,4,
and Peter J. Koopmans3
1Division of Imaging Sciences and Biomedical
Engineering, King's College London, London, London,
United Kingdom, 2Department
of Cardiovascular Medicine, University of Oxford,
Oxford, United Kingdom, 3FMRIB
Centre, Nuffield Department of Clinical Neurosciences,
University of Oxford, Oxford, United Kingdom,4Centre
for the Developing Brain, King's College London, London,
United Kingdom
Relative B1+ mapping relies on the validity of the SPGR
small flip angle approximation. This cannot be
guaranteed in the presence of large B1+ dynamic range.
Here we present an acquisition and reconstruction scheme
which enables relative B1+ mapping in the presence of
large dynamic range.
|
16:24 |
0491.
|
Rapid MRI System
Calibration using 3DREAM
Daniel Brenner1, Rüdiger Stirnberg1,
Eberhard Daniel Pracht1, and Tony Stöcker1,2
1German Center for Neurodegenerative Diseases
(DZNE), Bonn, Germany, 2Department
of Physics and Astronomy, University of Bonn, Bonn,
Germany
Ultra-fast 3DREAM B1 mapping provides a variety of
different calibration information, such as B1 and B0
maps as well as an image suitable for normalization of
the receive sensitivity. The DREAM FID images also allow
direct coregistration to a brain atlas, which
facilitates automatic extraction of regional calibration
data. The method is well-suited for receive and transmit
bias correction of arbitrary imaging data at high field,
as demonstrated for MP-RAGE and 3D-EPI acquired at 7T.
|
16:36 |
0492. |
Validation of Variable Flip
Angle Imaging-Based Simultaneous B1+ and T1 Mapping in the
Prostate at 3T
Novena A Rangwala1, Isabel M Dregely1,
Holden H Wu1, and Kyunghyun Sung1
1Department of Radiological Sciences,
University of California Los Angeles, Los Angeles, CA,
United States
Characterizing B1+ distributions accurately is essential
in multiparametric quantitative MRI; however, B1+
measurement sequences are additional to quantitative
protocols and increase scan duration. This study aimed
to characterize the B1+ distribution in the pelvis using
a recently proposed reference region variable flip angle
(RR-VFA) imaging based technique and compare it with a
conventionally available B1+ mapping sequence in the
prostate of healthy volunteers. The proposed technique
showed B1+ distributions consistent with the
conventional B1+ measurement technique, eliminating the
need for an additional scan for B1+ correction.
Additionally T1 measurements after B1+ correction with
the RR-VFA method had significantly lower standard
deviations, indicating improved robustness after B1+
correction.
|
16:48 |
0493. |
Direct calculation of B1+ and
B1- from
two point variable flip angle data for quantitative T1 and
PD mapping
Simon Baudrexel1,2, Ulrike Noeth2,
Sarah Reitz1,2, Johannes Christian Klein1,2,
and Ralf Deichmann2
1Department of Neurology, Goethe University
Frankfurt, Frankfurt am Main, Germany, 2Brain
Imaging Center (BIC), Goethe University Frankfurt,
Frankfurt am Main, Germany
A technique for direct calculation of the transmitted
radio frequency (RF) field (B1+) and the
receiver sensitivity profile (B1-) is
presented which is based on the acquisition of a
standard spoiled gradient echo sequence using two
different flip angles (i.e. the variable flip angle
method, VFA) without requiring any further measurements.
In a sample of 12 healthy subjects, B1+ and
B1- maps
obtained with the new method were compared with
respective maps based on existing techniques, yielding
very good agreement throughout the brain with an average
absolute deviation of 3%.
|
17:00 |
0494. |
B0 changes around the head
induced by the cardiac cycle at 7T
Lennart J. Geurts1, Vincent O. Boer2,
Tijl A. van der Velden2, Peter R. Luijten2,
Dennis W.J. Klomp2, and Jaco J.M. Zwanenburg2
1Radiology, UMC Utrecht, Utrecht, Utrecht,
Netherlands, 2Radiology,
UMC Utrecht, Utrecht, Netherlands
This study assessed B0 fluctuations caused by the
cardiac cycle using fieldprobes positioned around the
headrest in a 7T MRI bore. Six volunteers were included
and 2400 field measurements were taken at a 100ms
interval. The maximum measured cardiac field fluctuation
amplitude was 7.0 nT. On average the cardiac field
fluctuation amplitude was 27% of the respiratory field
fluctuation. Because fieldprobes were used, there was no
confounding influence of pulsatile tissue motion. We
conclude that cardiac field fluctuations can be a
measurable contributor to B0 instability.
|
17:12 |
0495.
|
Investigating the potential
of highly accelerated FatNavs for dynamic shimming
Frédéric Gretsch1, José P Marques2,
Rolf Gruetter1,3, and Daniel Gallichan1
1CIBM, EPFL, Lausanne, Vaud, Switzerland, 2Dept.
of Radiology, University of Lausanne, Vaud, Switzerland, 3Depts.
of Radiology, Universities of Lausanne and Geneva, Vaud,
Switzerland
We recently showed that fat navigators (FatNavs) can be
used for motion correction of head scans, exploiting the
natural sparsity of the fat image to allow higher
acceleration factors than water images. In this work we
investigate the potential of using a dual-echo FatNav to
also estimate dynamic changes in the B0-field due to
respiration – with the aim of applying the FatNavs to
also allow dynamic shimming. Our initial tests indicate
that the fat image is suitable for tracking
respiration-induced B0 changes – and can do so even at
acceleration factors up to 6x9 for the FatNav.
|
17:24 |
0496. |
Automatic Virtual Shimming
for Robust Fat Suppression in Subtractionless First-Pass
Peripheral Angiography
Holger Eggers1 and
Tim Leiner2
1Philips Research, Hamburg, Germany, 2Department
of Radiology, University Medical Center Utrecht,
Utrecht, Netherlands
To apply Dixon methods for background suppression in
first-pass peripheral angiography, the water-fat
separation has to be highly reliable over large field of
views. This is difficult to achieve in the lower legs,
because large main magnetic field offsets and gradients
occassionally arise from poor resonance frequency
determination or shimming. An approach based on virtual
shimming is proposed in the present work to
automatically compensate for such offsets and gradients.
It is shown to improve the robustness of the fat
suppression in subtractionless first-pass peripheral
angiography in demanding cases.
|
17:36 |
0497. |
Fast B1 inhomogeneity
correction in bSSFP imaging using transient-state signal
Min-Oh Kim1 and
Dong-Hyun Kim1
1Electrical and electronic engineering,
Yonsei University, Seoul, Seoul, Korea
B1+ inhomogeneity induces different signal modulations
spatially, therefore, it needs to be corrected to make
the acquired images more reliable. Here, an extremely
fast B1 mapping method is proposed to correct B1
inhomogeneity induced image modulation in bSSFP imaging
using early transient-state signal.
|
17:48 |
0498. |
Respiration Induced B0
Variation in Double Echo Steady State Imaging (DESS) in the
Breast
Catherine J Moran1, Kristin L Granlund1,
Bragi Sveinsson1,2, Marcus T Alley1,
Bruce L. Daniel1, and Brian A Hargreaves1
1Radiology, Stanford University, Stanford,
CA, United States, 2Electrical
Engineering, Stanford University, Stanford, CA, United
States
The Double Echo Steady State (DESS) sequence has been
shown to provide diffusion weighted breast images
without the severe distortion of EPI DWI. However, DESS
images in the breast demonstrate prominent ghosting
artifacts, which, in the low SNR second echo can
compromise image quality. We investigated respiration
induced B0 in the breasts as the source of this artifact
and present initial results of correction of the
artifact using a DC navigator.
|
|