Optimized Structural Imaging at 7T using ME-MP2RAGE
Eberhard Daniel Pracht1, Philipp Ehses1, Hankyeol Lee2, Kamil Uludag2,3, and Tony Stoecker1,4
1German Center for Neurodegenerative Diseases, Bonn, Germany, 2Center for Neuroscience Imaging Research, Institute for Basic Science & Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of, 3Techna Institute & Koerner Scientist in MR Imaging, University Health Network, Toronto, ON, Canada, 4Department of Physics and Astronomy, University of Bonn, Bonn, Germany
We present an optimized, scan-time efficient multi-echo MP2RAGE sequence (ME-MP2RAGE) for improved brain segmentation at ultra-high field. The proposed method removes susceptibility induced image distortions along the readout direction.
Figure 2: ME-MP2RAGE acquisition
Left: Composite image 1 (odd_odd-even_even echo). Image distortions are significantly reduced compared to the MP2RAGE acquisition.
Right: A further decrease in image distortions is achieved using composite image 2 (odd_even - odd_even echoes).
Animated GIF images are depicted as well for better illustration of locally induced image distortions.
Figure1: MP2RAGE acquisition
Left: Difference image (odd -even echo). Image distortions induce signal enhancements/attenuations at tissue boundaries.
Right: Corresponding animated GIF illustrating the image distortions in greater detail.