0620
Rapid calibration scan for estimating temporally-varying eddy currents in diffusion imaging using a time-resolved PEPTIDE imaging approach
Merlin J Fair1 and Kawin Setsompop1,2
1Radiological Sciences Laboratory, Stanford University, Stanford, CA, United States, 2Electrical Engineering, Stanford University, Stanford, CA, United States
The time-resolved PEPTIDE approach is investigated for use as a fast (<30s) calibration scan for the estimation of eddy current induced phase evolutions. Simulation, phantom and in vivo work demonstrates the potential accuracy of such a technique, including up to a b-value of 5000s/mm2.
Figure 5 – (a) Single-shot data: phantom & in-vivo. The low SNR of b=5000 in-vivo data is apparent in both EPI and PEPTIDE (PEP), but image phase across time is still measurable. (b) Estimated phantom eddy fields, measured by FSL and PEPTIDE. First 3 directions for PEPTIDE are direct estimates (red square), while others are calculated using a linear model. (c) Estimated in-vivo eddy fields (with same diffusion-directions and acq. params as in phantom). “Phantom PEP-ref”: PEPTIDE-estimated field in the phantom, masked by the in-vivo brain FOV, for reference with the in-vivo results.
Figure 1 - a) EPTI/PEPTIDE uses a zig-zag sampling trajectory to cover a large ky-t section per acquisition shot, with B0-informed-GRAPPA used to fill-in the missing data. PEPTIDE acquires rotated EPTI shots (blades) to cover the full k-t space. With full k-t data, each time-point has consistent phase and signal decay, so the images in the resolved time-series are free from typical-EPI distortion and blurring. b) The phase of the time-series images for a DWI acquisition can be compared with that of a diffusion-free (b=0) acquisition, to estimate eddy current induced phase changes.