Ultrafast motion-minimized shoulder MRI with a deep learning constrained Compressed SENSE reconstruction
Jihun Kwon1, Masami Yoneyama1, Takashige Yoshida2, Kohei Yuda2, Yuki Furukawa2, Johannes M. Peeters3, and Marc Van Cauteren3
1Philips Japan, Tokyo, Japan, 2Tokyo Metropolitan Police Hospital, Nakano, Japan, 3Philips Healthcare, Best, Netherlands
The accelerated CS-AI showed comparable
image quality to that of the reference method. The acquisition time for CS-AI was
about 40 seconds, which is less than one-third of the reference method.
Figure
2. Fat-suppressed
PDw axial images for (a) C-SENSE reference with acceleration factor 2, (b)
SENSE (c) C-SENSE, and (d) CS-AI with acceleration factor 4. The yellow arrows
denote suprascapular nerve.
Figure
4. Fat-suppressed
T2w coronal images for (a) C-SENSE reference with acceleration factor 2, (b)
SENSE, (c) C-SENSE, and (d) CS-AI with acceleration factor 4. The yellow arrows
denote synovial fluid.