1288
Reproducibility and Multi-vendor Accuracy Comparison of T1- and T2- Mapping Using 2D and 3D Synthetic MRI with GRAPPA, SENSE and Compressed Sense
Maarten Naeyaert1, Tim Vanderhasselt1, Marcel Warntjes2, and Hubert Raeymaekers1
1Radiology, Universitair Ziekenhuis Brussel, Brussels, Belgium, 2SyntheticMR AB, Linköping, Sweden
Comparison of 2D and 3D simultaneously acquired T1, T2 and PD maps on 3T scanners of different vendors shows almost-general agreement between scanners, sequence and with and without GRAPPA, SENSE, or compressed sensing. T1 and T2 values approach the reference values and are reproducible.
Figure 3: Measured versus reference T1 (left) and T2 (right). Colour of the symbols indicates vendor, the shape indicates acceleration used, and filled vs hollow indicates 2D vs 3D respectively.
Figure 4: Difference between measured and reference T1 (left) and T2 (right) values. Colour of the symbols indicates vendor, the shape indicates acceleration used, and filled vs hollow indicates 2D vs 3D respectively.