UVC Based Wireless Patient Bore Disinfection Utilizing Scanner RF Transmission
Devavrat Likhite1, Rob Amerling1, Leon Lee1, and Saban Kurucay1
1GE Healthcare, Waukesha, WI, United States
The presented UVC based wireless patient bore disinfection fixture, utilizing scanner RF transmission, is able to disinfect the patient bore and table of a 3T GE MR system in 2-minutes and a 1.5T GE MR system in 6-minutes (UVC dose >40mJ/cm2).
Figure
4 UVC fixture operating and disinfecting the inner surfaces of the
bore and patient table on a 3T Signa Premier system
Table
1 shows the used UVC dots and corresponding UVC dose measured on various GE MR
systems(Color representation
in the table may not be accurate due to image capture challenges. Refer to the corresponding
numerical dose value presented)