2311
Input Impedance Comparison of MR-Conditional Cardiac Implantable Pulse Generators at the 1.5T MR Frequency of 63.87 MHz
Jason Meyers1, David Prutchi1, and Ramez Shehada2
1Impulse Dynamics (USA) Inc., Marlton, NJ, United States, 2Medical Technology Laboratories, La Mirada, CA, United States
The input impedances at 63.87 MHz of six MR-conditional cardiac IPGs were measured and found to be low and comparable, suggesting that impedance is not intentionally added by device manufacturers to limit RF-induced electrode heating.
Figure 2: This test fixture was constructed to allow direct measurement of the impedance between an IPG’s header connector block and Can. Also shown are the IPGs modified to provide direct access to terminal blocks to allow this measurement.
Figure 1: Simplified diagram showing the RF current circulation paths that cause tissue heating.