Fat-insensitive T2water measurement using multiple Dixon turbo spin-echo acquisitions with effective echo time increments
Ruaridh M Gollifer1,2, Tim JP Bray1,3, Margaret Hall-Craggs1,3, and Alan Bainbridge2
1Centre for Medical Imaging (CMI), University College London, London, United Kingdom, 2Department of Medical Physics and Bioengineering, University College London Hospital, London, United Kingdom, 3Department of Imaging, University College London Hospital, London, United Kingdom
The proposed fat-insensitive T2water measurement, based on Dixon TSE with effective TE increments, is accurate over a range of T2 and fat fraction values and could provide a quantitative alternative to the widely-used STIR sequence for imaging inflamed and/or oedematous bone and soft tissue.
Figure 4: Healthy volunteer scan of TSE Dixon sequence for water only images with corresponding T2 map.
Figure 3: T2water values for TSE Dixon compared to MRS for FFs of 0, 30 and 40% and agar concentrations of 2% (blue), 3% (green) and 4% (red).