3356
MRI System Phantom: assessing the MR visible thermometer, scanner geometric distortion corrections, and effect of fill conductivity
Stephen E Russek1, Kathryn E Keenan1, Karl F Stupic1, Teryn S Wilkes2, Ramesh Karki3, and Todor Karaulanov4
1NIST, Boulder, CO, United States, 2Intermountain Neuroimaging Consortium, University of Colorado, Boulder, CO, United States, 3University of Colorado Anschutz, Radiological Sciences, Aurora, CO, United States, 4CaliberMRI, Inc., Boulder, CO, United States
MRI System Phantom: Demonstrated MR-readable thermometer for temperature corrections, geometric distortion analysis down to 1 part in 1000, and fill-conductivity effects on MR measurements
Figure 1. (A) Photo of the system phantom in the 32-channel head coil, (B) Axial slice showing the phantom plates; fiducial, relaxation time, and proton density/ signal-to-noise ratio arrays; resolution and slice profile insets; and the LC thermometer. (C) Top plate of the phantom showing the serial number and phantom orientation. Here, since the phantom is rotated from its default orientation right/left (R/L) corresponds to anterior/posterior (A/P) directions for a human in the head coil. (D) Image of LC thermometer with 4mm regions of interest and transition temperatures marked.
Figure 3. Geometric distortion data showing the difference of the apparent position of each fiducial sphere from the real position, as a function of distance from phantom center. (A), (B) distortions with software corrections turned off and turned on, respectively. The points are color coded to indicate where they are in the phantom.