New strategies in RF Pulse Design
Acq/Recon/Analysis Thursday, 20 May 2021
Digital Poster
3947 - 3964

Oral Session - New Strategies in RF Pulse Design
Acq/Recon/Analysis
Thursday, 20 May 2021 16:00 - 18:00
  • Calibration-free pTx of the human heart at 7T via 3D universal pulses
    Christoph Stefan Aigner1, Sebastian Dietrich1, Tobias Schaeffter1,2, and Sebastian Schmitter1,3
    1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany, 2Division of Imaging Sciences and Biomedical Engineering, King's College London, London, United Kingdom, 3Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
    This study demonstrates in vivo that 4kT-UP are highly suitable for calibration-free 3D heart FA homogenization at 7T despite large inter-subject variations due to varying age, BMI and coil placement.
    Figure 4: a) Sagittal slice of the 27 3D B1+ predictions using the default shim setting (left) and the 4kT-UP (right). Library: B1+ maps 1-22, test-cases: B1+ maps 23-27. The 4kT-UP results in a homogeneous FA in the heart ROI of all 5 test cases. Moreover, the 4kT-UP achieves a homogeneous FA also in surrounding tissues such as the aorta. b-c) Boxplot of the FA spread in the 3D heart ROI of all 27 subjects for default (b) and 4kT-UP (c) demonstrating the FA homogeneity across all B1+ predictions using the 4kT-UP pulse.
    Figure 5: B1+ predictions and reconstructed, respiration corrected 3D GRE images for two unseen test cases with 4kT-UP. These two volunteers were not part of the optimization. The 3D images are free of breathing artefacts and demonstrate, despite some differences close to the coil elements, the feasibility to achieve a homogeneous calibration-free FA of the whole heart. The remaining signal changes in the AP direction of the acquired images are a result of receive (B1-) variations.
  • Robust RF Shimming and Small-tip-angle Multi-spoke Pulse Design with Finite Difference Regularization
    Zhipeng Cao1, Adrian Paez2, Chunming Gu2, and Jun Hua2
    1Vanderbilt University, Nashville, TN, United States, 2Johns Hopkins University, Baltimore, MD, United States
    A novel magnitude-least-squares algorithm is developed for robust patient-tailored 2D multislice imaging at UHF with RF shimming and small-tip-angle multi-spoke pulse design. The algorithm is effective against excitation nulls and sub-optimal pulse solutions for brain and knee imaging.
    Fig 4. Designed flip angle distribution (top row) and acquired EPI images (bottom row) from one representative slice in the human brain at 7T with system default quadrature mode, dynamic multislice RF shimming with conventional MLS, and with proposed FD-MLS. The RF pulse design ROI is labeled in red. With fixed system gain, the FD-MLS improved the signal homogeneity and SNR over default quadrature, while not forming the null as the conventional MLS.
    Fig 5. Evaluation of FD-MLS with knee data from an 8Tx array with two different postures. The FD-MLS not only improved over null solutions (in slices marked by the green stars), but also improved over local minimum solutions (marked by green circles).
  • Off-resonance Robustness in Reduced FOV Imaging using Sheared 2DRF Excitation
    Bahadir Alp Barlas1,2, Cagla Deniz Bahadir1,2,3, Sevgi Gokce Kafali1,2,4, Ugur Yilmaz2, and Emine Ulku Saritas1,2,5
    1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey, 2National Magnetic Resonance Research Center (UMRAM), Bilkent University, Ankara, Turkey, 3Department of Biomedical Engineering, Cornell University, New York, NY, United States, 4Department of Bioengineering, University of California Los Angeles, Los Angeles, CA, United States, 5Neuroscience Graduate Program, Bilkent University, Ankara, Turkey
    The proposed sheared 2DRF pulse design technique achieves considerable reduction in pulse duration, improving the off-resonance robustness of reduced-FOV imaging via SNR enhancement while eliminating the slice coverage limitations and maintaining inherent fat suppression capability.
    FIGURE 4. In vivo imaging results in the cervical spinal cord. (a) The standard and sheared 2DRF pulses had 16.0 ms and 9.5 ms durations, respectively. (b) The localizer showing the reduced-FOV region corresponding to 200 mm X 50 mm FOV. T2-weighted EPI images acquired using (c) 1DRF pulse with additional fat suppression (cropped full FOV of 200 mm X 162.5 mm), and (d) standard and (e) sheared 2DRF pulses. Reduced-FOV imaging achieves higher resolution with reduced in-plane distortion. The sheared 2DRF pulse further improves SNR in regions with high off-resonance effects (red arrows).
    FIGURE 2. Simulated 2DRF excitation profiles for (a) the standard and (b) the sheared 2DRF designs. For the standard 2DRF pulse, the profile is periodic in the SS direction, limiting the slice coverage due to partial saturation effects. In contrast, the sheared 2DRF pulse places the replicas along the sheared axis, effectively pushing them outside the potential slice locations. (c) 1D cross-sectional excitation profiles along SS and PE directions demonstrate identical sharpness, with a slab thickness of 40 mm and a slice thickness of 4 mm.
  • Asymmetric GOIA pulses for highly selective B1 and T1 independent outer volume suppression
    Chathura Kumaragamage1, Peter B Brown1, Scott McIntyre1, Terence W Nixon1, Henk M De Feyter1, and Robin A de Graaf1
    1Department of Radiology and Biomedical Imaging, Magnetic Resonance Research Center, Yale University, New Haven, CT, United States
    An asymmetric GOIA pulse was developed with TW = 1.7% by combining hyperbolic secant, and hyperbolic tangent modulations (Tp = 6.66 ms, BW = 20 kHz). Utilizing a 4-pulse OVS method, highly selective B1 and T1-independent OVS can be achieved in vivo.
    Figure 3. (A-C) Phantom results illustrating localization performance of the RF pulses compared, selective for a 7 cm slab, and (D) illustrates traces along the X direction for the cases with a zoomed cut out illustrating the transition profile obtained for the three pulses. Excellent signal suppression throughout the slice (Mz/M0 < 0.02) and minimal perturbation outside the inversion band is seen for all pulses.
    Figure 5. An in vivo MRSI data acquired (TE/TR = 30 ms/2000 ms, 8.2 min acquisition time, 10 x 10 mm2 grid from a 5 mm axial slab) utilizing ECLIPSE localization. The anatomical image illustrates six-voxel positions on the left, and 9 voxel locations on the right, corresponding to illustrated spectra. The ellipse in red color illustrates the water ROI selected with ECLIPSE, and the blue ellipse corresponds to the lipid ROI due to chemical shift ~1.6 mm smaller in radius along the x-direction.
  • Interleaved Binomial kT-Points for Water-Selective Imaging at 7T
    Daniel Löwen1, Eberhard Daniel Pracht1, Rüdiger Stirnberg1, and Tony Stöcker1,2
    1German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany, 2Department of Physics and Astronomy, University of Bonn, Bonn, Germany
    We present a time-efficient water-selective, parallel transmit RF excitation pulse for ultra-high field applications.

    Figure 1: A comparison of different water excitation schemes. Top: RECT kT-point excitation as used in [5]. Middle: “Classical” binomial 1-1 excitation applied to kT-points pulses. The second pulse starts at an odd number of half phase cycles of fat. Bottom: Proposed interleaved version of the 1-1 binomial kT-points pulse.

    Figure 3: In vivo MP-RAGE acquisition employing a 7sh interleaved binomial kT-points pulse (Table 1) for excitation (bottom) compared to a non-selective CP-mode MP-RAGE acquisition (top). Each with (right) and without (left) fat suppression. Yellow arrows point at better homogenization with kT-points. Green arrows point at missing fat artifacts with water excitation (WE).
  • Lipid Artifact Removal by Dynamic Shimming (LARDS) with multi-coil B0 shim arrays
    Jinmin Xu1,2, Nicolas Arango2, Congyu Liao2,3, Berkin Bilgic2,3, Zijing Zhang1,2, Lawrence L Wald2,3, Setsompop Kawin2,3, Huafeng Liu1, and Jason P Stockmann2,3
    1State key Laboratory of Modern Optical Science and Engineering, Zhejiang University, Hangzhou, China, 2A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States
    A new method for improving lipid suppression using high-spatial order, rapidly-switchable B0 shim fields.  Shim currents are jointly optimized along with the lipid saturation pulse frequency offset to increase water-lipid spectral separation and improve saturation efficacy.
    Figure 3. Simulated comparison of three lipid suppression scenarios in the brain. For each case, we simulate the off-resonance histogram (c) and z-magnetization (d) for water and lipids. (e) chemical shift artifacts in brain slices due to residual lipid are also simulated (5x fat for a better contrast). Using the 32-ch AC/DC shim array, LARDS performs better than simple global homogeneity shimming because the lipid mask used for optimization is limited only to voxels within the passband of the water excitation pulse, reducing constraints on the solver.
    Figure 4. Three simulated cases of lipid suppression in the abdomen. The simulated off-resonance histogram, z-magnetization and chemical shift artifacts (5x fat for a better contrast) clearly show the benefit of LARDS compared to conventional shimming.
  • Inner Volume Excitation via Joint Design of Time-varying Nonlinear Shim-array Fields and RF Pulse
    Molin Zhang1, Nicolas Arango1, Jason Stockmann2, Jacob White1, and Elfar Adalsteinsson1,3,4
    1Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 2Athinoula A. Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 3Harvard-MIT Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, United States, 4Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, United States
    We jointly design the non-linear B0 field generated by shim array and RF pulse to do the inner volume excitation with 1/3 of the background. Compared with previous work optimizing the  static non-linear B0 field and predefined Sinc shape RF pulse, we gain 62% improvements in L2 error.
    On the left hand side, the desired pattern is shown which is a 3D volume. On the right hand, the excited results under the four different experiments settings are shown. The error is based on the pixel-wise L2 norm. We could see that proposed joint designed time-varying field and RF pulse achieves best among all experiments.
    The optimization results for different initializations for setting 3 and 4. We could see from the figure that the proposed method is more robust to the initialization with similar error ~6.2, while setting 3 is sensitive to the initialization as the error for initialization 2 is 13.85. Another thing to notice is that for the proposed method, different initializations may require different current and RF power levels.
  • iSLR: An Improved Shinnar-Le-Roux Frequency Selective Pulse Design Algorithm with Reduced Energy and More Accurate Phase Profiles
    Frank Ong1, Michael Lustig2, Shreyas Vasanawala1, and John Pauly1
    1Stanford University, Stanford, CA, United States, 2University of California, Berkeley, Berkeley, CA, United States
    We improve the SLR design process to generate pulses with lower energy (by as much as 26%) and more accurate phase profiles.
    Linear phase slice selection pulses and their transverse magnetization profiles after refocusing. The improved SLR (iSLR) pulse has a much flatter phase response than the original one. Pulse energy is reduced by 18.6% and peak is reduced by 9.1%. Note that the improved pulse is asymmetric, which shows that the proposed design compensates for the phase of α to generate a linear phase profile.
    Zero phase spin-echo refocusing pulses and their longitudinal magnetization profiles. Pulse energy is reduced by 18.6% and peak is reduced by 13.4%.
  • DeepControl: AI-powered slice flip-angle homogenization by 2DRF pulses
    Mads Sloth Vinding1, Christoph Stefan Aigner2, Jason Stockmann3,4, Bastien Guérin3,4, Sebastian Schmitter2,5, and Torben Ellegaard Lund1
    1Center of Functionally Integrative Neuroscience, Aarhus University, Aarhus, Denmark, 2Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany, 3Harvard Medical School, Boston, MA, United States, 4Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 5Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States
    In addition to localized 2D spatial-selective excitation, which it was trained for, DeepControl was able to generate optimal-control-like pulses with highly uniform FA distributions and peak voltages below the system limit, in a fraction of the time. 
    Figure 2: FA maps and corresponding B1+/B0 maps. First row: the DeepControl result. Second row: the corresponding OC result. The printed numbers in white are normalized root-mean-square errors. Row three to five, between the two dashed line, are intended to recap what missing field information results in with respect to FA maps. Using the previous version of DeepControl14, which did not account for field inhomogeneities, distortions like these should be expected. In all cases both B1+ map/B0 maps were included in the Bloch simulations (raster grid: 128x128).
    Figure 3: DeepControl vs Optimal Control (row 1 and 2 in Fig. 2) 3D and 2D histograms of voxel FA values inside the target region. The target FA was 30o. Numbers printed below the plots are mean and standard deviation FAs.
  • Uniform Magnetization Transfer contrast at 7T with Direct Saturation Control
    David Leitão1, Raphael Tomi-Tricot2, Pip Bridgen1, Tom Wilkinson1, Patrick Liebig3, Rene Gumbrecht3, Dieter Ritter3, Sharon Giles1, Ana Baburamani1, Jan Sedlacik1, Joseph V. Hajnal1,4, and Shaihan J. Malik1,4
    1Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 3Siemens Healthcare GmbH, Erlangen, Germany, 4Centre for the Developing Brain, King's College London, London, United Kingdom
    Direct Saturation Control is a novel RF pulse design method that considers the saturation of semisolids directly, rather than the flip angle. We designed pTx composite pulses to achieve spatially uniform Magnetization Transfer Ratio (MTR) maps at 7T in presence of strong B1+ inhomogeneity.
    Figure 2 - Magnetization Transfer Ratio (MTR) maps from one volunteer obtained with saturation prep-pulse applied using CP (top row) and DSatC (bottom row) modes using Target=0.6μT2 and TR=23ms. For both modes the saturation prep-pulse was applied with different number of sub-pulses: 1 (left column), 2 (middle column) and 3 (right column). The observed MTR correlates well with the predicted <B12> (Figure 3). The solution for 1 sub-pulse is not uniform, and this is also predicted from <B12>.
    Figure 5 – Results from the gradient demonstration using the non-blipped (top row) and blipped (bottom row) saturation pulses in Figure 4. (a,d) MTR maps, (b,e) <B12> of the saturation pulse and (c,f) flip angle from the RF and gradients applied.
Back to Top
Digital Poster Session - RF Pulses
Acq/Recon/Analysis
Thursday, 20 May 2021 17:00 - 18:00
  • Respiration induced B1+ changes and its compensation via respiration robust 3D kT point pulses in 7T body imaging
    Christoph Stefan Aigner1, Sebastian Dietrich1, Tobias Schaeffter1,2, and Sebastian Schmitter1,3
    1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany, 2Division of Imaging Sciences and Biomedical Engineering, King's College London, London, United Kingdom, 3Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany
    This in-vivo study demonstrates that respiration-resolved B1+ maps and respiration-robust 4kT pTx pulses are highly preferred to achieve 3D heart FA homogenization at 7T when subjects perform strong breathing.
    Figure 4: Evaluation of four different pulses optimized for the heart ROI on exhale, intermediate, inhale and optimized on all respiration states (respiration robust) for deep breathing of all 10 subjects. Depicted are boxplots containing the CVs of all respiration states. The respiration robust pulse performs best across all subjects and respiration states with the lowest median and spread.
    Figure 5: Side-by-side comparison of B1+ predictions (a) and reconstructed 3D GRE images (b) with the 4 kT-points pTx pulses optimized on inhale (left) and respiration robust (right) of subject 7. Depicted are the views of the acquired 3D volume close to the position of the B1+ prediction. The arrows point to signal drop-out regions in using the inhale pulse which are corrected by the respiration robust pulse.
  • Reducing inter-subject variability and improving accuracy of Universal Pulses using standardized (universal) pulses
    Caroline Le Ster1, Franck Mauconduit1, Aurélien Massire2, Vincent Gras1, and Nicolas Boulant1
    1Paris-Saclay University, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France, 2Siemens Healthcare SAS, Saint-Denis, France
    The pulse design process can be made transparent to the user with calibration-free universal pulses (UP). Here a new method is proposed where UPs are adjusted to scanned subjects through a fast calibration step. Adjusted SUP improve excitation performance and reduce intersubject variability.

    Figure 1: Pulse design methodology in the volunteer for the MPRAGE excitation 8° GRAPE pulse. a) A calibration matrix (L) is computed to adjust the standardized universal pulse (SUP) designed over the standardized pulse design database to the subject B1+ map. Its magnitude is close to the identity matrix while diagonal phase terms are close to 0 radian. b) RF and gradient amplitudes of the resulting adjusted SUP. c and d) Flip angle maps and histograms retrospectively simulated on the subject for the UP and adjusted SUP with the subject’s full B1+ map. NRMSEs are 9.5% and 7.5%, respectively.

    Figure 4: a) UNI contrast computed from the two inversion volumes of the MP2RAGE sequence acquired with the adjusted SUP. b to d) T1 map computed by injecting UNI signal intensities and sequence parameters in Bloch equation simulations for the MP2RAGE sequence acquired in the CP mode, CP mode with B1+ postprocessing correction and adjusted SUP, respectively. e) T1 distributions extracted from the brain for figures 4b to 4d, showing an improvement on the distributions.
  • “Universal” non-selective pulse design at 7 Tesla using a birdcage coil and a B0 shim array: Evaluation of kT-points and fully optimized pulses
    Bastien Guerin1, Eugene Milshteyn1, Yulin Chang2, Mads S Vinding3, Mathias Davids1, Wald L Lawrence1, and Jason Stockmann1
    1Massachusetts General Hospital, Charlestown, MA, United States, 2Siemens Medical Solutions, Malvern, PA, United States, 3Center for functionally integrative neuroscience, Aarhus, Denmark
    Universal kT-point pulses yield a large improvement in flip-angle uniformity compared to the standard RECT excitation, even without using pTx. Fully optimized RF + gradient pulses further improve the pulse performance however adding the shim elements in the optimization does not.
    Fig. 2: 90°, 2ms “universal” excitation pulses and their performance. For all pulses, the voltage is limited to a maximum of 400V. A: “Universal” kT-point pulse. B: “Universal” RF + gradient optimization pulse (no shim currents). C: “universal” RF + gradient + shim currents pulse (INSTANT). D: RMSE performance of the different pulse strategies and typical sagittal flip-angle map (first of the three subjects included in the universal design). The black arrow points to typical under-flip in the frontal lobe, which is largely corrected by the optimized pulses.
    Fig. 3: Flip-angle maps acquired on the last four subjects of the cohort using a standard rectangular pulse (RECT) and the “universal” kT-point and RF + gradient optimized pulses shown in Fig. 2. The target flip-angle is a uniform 90° distribution in the brain (skull has been stripped from the optimization mask). The red and black arrows point to typical under-flip in the temporal and frontal lobes, respectively.
  • Evaluating Universal and Fast Online Customized Pulses for parallel transmission using two different RF coils
    Jürgen Herrler1, Sydney Nicole Williams2, Patrick Liebig3, Shajan Gunamony4,5, Christian Meixner6, Andreas Maier7, Arnd Dörfler1, David Porter2, and Armin Michael Nagel6
    1Institue of Neuroradiology, University Hospital Erlangen, Erlangen, Germany, 2Imaging Centre of Excellence, University of Glasgow, Glasgow, Scotland, 3SIEMENS Healthineers, Erlangen, Germany, 4Institute of Neuroscience & Psychology, University of Glasgow, Glasgow, Scotland, 5MR CoilTech Limited, Glasgow, Scotland, 6Institue of Radiology, University Hospital Erlangen, Erlangen, Germany, 7Friedrich Alexander University Erlangen Nürnberg, Erlangen, Germany
    Parrallel-transmit Universal Pulses (UPs) and Fast Online Customized (FOCUS) pulses were trained and evaluated on two different radiofrequency coils. UPs may fail when used on a different coil, whereas FOCUS pulses show a more reliable pulse performance and less dependency on the coil used.
    Figure 3 MPRAGE images for a fixed adiabatic inversion pulse and various excitation pulses. Experimental images and corresponding FA simulations are shown for coil-specific CP pulses and all generated pTx excitation pulses.

    Figure 4 MPRAGE images using the UP of the respective coil for excitation and various inversion pulses. Experimental images and corresponding FA simulations are shown for coil-specific CP pulses and all generated pTx excitation pulses.

  • Motion Robust Parallel Transmission Excitation Pulse Design for Ultra-High Field MRI
    Luke Watkins1, Alix Plumley2, Kevin Murphy1, and Emre Kopanoglu2
    1Department of Physics and Astronomy, CUBRIC, Cardiff University, Cardiff, United Kingdom, 2Department of Psychology, CUBRIC, Cardiff University, Cardiff, United Kingdom
    The motion-robust pulse yielded superior performance in 97% of all magnitude and phase error metrics for 46 other off-centre positions (similar performance in remaining 3%). Similar magnitude nRMSE to the reference pulse was maintained at the centre.
    Figure 2. Magnitude and Phase profiles for a -5° roll rotation about the centre of the head. Magnitude nRMSE was reduced from 14% (reference pulse) to 5.4% (MRP). Maximum magnitude error was reduced from 64% to 20%. Phase RMSE was reduced from 17° to 3.6°, and maximum phase error from 68° to 15°. The MRP was successful at areas of high motion-induced error in magnitude and phase.
    Figure 5. (a) Magnitude nRMSE, (b) phase RMSE, (c) maximum magnitude error, (d) maximum phase error of the MRP vs the reference pulse. The green shaded area represents where the MRP outperformed the reference pulse. Cases where MRP performs worse than the reference still report very similar errors.
  • Time optimal control based design of robust inversion pulses
    Christina Graf1, Martin Soellradl1, Armin Rund2, Christoph Stefan Aigner3, and Rudolf Stollberger1
    1Graz University of Technology, Institute of Medical Engineering, Graz, Austria, 2Institute of Mathematics and Scientific Computing, University of Graz, Graz, Austria, 3Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Berlin, Germany
    We optimize short and B0- and B1-robust RF pulses by time-optimal control. The optimized pulse shows a good efficiency over a broad set of B0- and B1-variations in numerical experiments as well as extensive phantom measurements on a 3T MRI system.
    Absolute value and angle of the oil-water phantom image measured with a B1-scaling of 100% (a-b). The red line shows the position of the line plots depicted in (c) and (d). The signal intensity profile without and with inversion pulse $$$\textbf{optim}$$$ with different B1-scalings are shown in (c). In (d), the measured normalized signal intensity after application of the $$$\textbf{optim}$$$ inversion pulse by a measurement without inversion is plotted. Thus, the influence of the coil sensitivity distribution and the varying excitation flip angle is eliminated.
    Absolute value and angle of the cylindrical MR phantom image measured with a B1-scaling of 100% (a-b) and the nominal B1-map (c). The red line shows the position of the line plots depicted in (d) and (e). The signal intensity profile without and with inversion pulse $$$\textbf{optim}$$$ with different B1-scalings are shown in (d). In (e), the measured normalized signal intensity after application of the inversion pulse by a measurement without inversion is plotted. Thus, the influence of the coil sensitivity distribution and the varying excitation flip angle is eliminated.
  • 3D k-Space Domain Parallel Transmit Pulse Design
    Jun Ma1,2, Bernhard Gruber3,4, Xinqiang Yan5, and William Grissom2
    1Department of Radiology, Stanford University, Stanford, CA, United States, 2Department of Biomedical Engineering, Vanderbilt University, Nashville, TN, United States, 3A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA, United States, 4Division MR Physics, Center for Medical Physics and Biomedical Engineering, Medical University Vienna, Vienna, Austria, 5Department of Radiology and Radiological Sciences, Vanderbilt University, Nashville, TN, United States
    A k-space domain pTx pulse design algorithm is proposed, which reduces the computation time of an example 3D pTx pulse design problem by ~80%, compared to a conventional spatial domain method.
    Fig 2: Normalized excitation patterns (top row) and error maps (bottom row) in central slices for k-space domain (left) and spatial domain designs (right). The calculated RMSEs were 2.13% (spatial) and 2.31% (k-space), indicating uniform inner volume excitation while maintaining the outer volume intact. The parallelized k-space domain design required 6.7 s computation versus 30.3 s for the spatial domain method, a 78% decrease. The size of the system matrix (W) of the k-space domain design was 0.8 Gb, while the system matrix of the spatial domain design had a size of 79 Gb, a 99% decrease.
    Fig 1: (a) Central slices of the target excitation pattern used for all pulse designs, which comprised an ellipse centered on the ventricles with AP/HF/LR semi-axes of 4.8/3.2/3.2 cm. (b) The SPINS trajectory used in the designs. (c) The 24-channel loop Tx array that was simulated in a human head model to obtain B1+ maps. The array has diameter 32 cm, height 28 cm, and 16×11 cm rectangular loops in 3 rows of 8. (d) 10 ms minimum time gradient waveforms that produce the SPINS trajectory, subject to the scanner's gradient amplitude and slew rate constraints of 200 mT/m and 700 T/m/s, respectively.
  • Multidimensional RF Pulse Design with Known Spatial Encoding Imperfections
    Ziwei Zhao1, Nam G. Lee2, and Krishna S. Nayak1,2
    1Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, United States, 2Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, United States
    We describe a multidimensional small-tip RF pulse design procedure that incorporates concomitant field effects. The proposed method produces more accurate excitation patterns, especially for low field strengths, off-isocenter, and long pulse durations.
    Figure 1. Concomitant field effects approximated as a Bloch-Siegert shift. (A) Reference frame visualization of (a,b) the concomitant $$$B_c(\vec{r},t)$$$, $$$B_0$$$, and gradient field $$$\vec{G}(t)\cdot\vec{r}$$$, and Rotating frame visualization (c) showing that the $$$\tilde{B_c}(\vec{r},t)$$$ approximation can be treated as additional off-resonance in the rotating frame. (B) Detailed derivations.
    Figure 2. Concomitant field estimation accuracy. (A) (left) Original 2D RF pulse excitation profile (magnitude: top & phase: bottom). (right) Magnitude (row a, b) and phase (row c, d) of the reference profiles (a, c) and estimated profiles (b, d) using Bloch-Siegert approximation at 0.2T, 0.55T, 1.5T, 3T and 7T. NRMSE results are 4.30%, 1.60%, 0.64%, 0.33% and 0.16% from 0.2T to 7T, respectively. Reference method simulates the concomitant field in the reference frame. (B) NRMSE as a function of field strength. Note that NRMSE is < 5% for B0 ≥ 0.2 Tesla.
  • Joint optimisation of parallel transmission in 2D spin-echo based sequences
    Belinda Ding1, Iulius Dragonu2, Patrick Liebig3, and Christopher T Rodgers1
    1Wolfson Brain Imaging Centre, University of Cambridge, Cambridge, United Kingdom, 2Siemens Healthcare Limited, Firmley, United Kingdom, 3Siemens Healthineers, Erlangen, Germany
    This abstract showed that pTx spokes pulses greatly improves the spin-echo image quality at 7T when compared against traditional circularly polarised pulses. Further improvement in image quality was observed with joint optimisation algorithms.
    Figure 5: Zoomed in SE-EPI images acquired with excitation FA of 60° and refocusing FA of 120°. Significant signal dropouts can be observed in the case of CP pulses (green arrows). Excellent signal homogeneity is observed when using any of the 3 pTx optimisation methods. Signal losses at the edges (red arrows) are further recovered with matched excitation and joint optimised pulses.
    Figure 1: Pulse optimisation work flow for (2) separately optimised pTx pulses, (3) matched excitation pTx pulses and (4) jointly optimised pTx pulses.
  • Optimization of the Nominal Flip Angle in Actual Flip Angle Imaging Using Phase Difference Information
    Tsuyoshi Matsuda1, Ikuko Uwano1, Yuji Iwadate2, and Makoto Sasaki1
    1Division of Ultrahigh Field MRI, Institute for Biomedical Sciences, Iwate Medical University, Iwate, Japan, 2Global MR Applications and Workflow, GE Healthcare Japan, Hino, Japan
    To optimize the nominal flip angle (FA) value for actual FA imaging (AFI), we detected pixels representing erroneous FAs by using phase differences between two AFI images. Nominal FA values of ≥ 60° were observed to be unsuitable for the AFI at 7 T.
    Figure 3. FA and error maps of the human head calculated by the AFI method. Pixels with erroneous, aliased FAs were found in the nominal FAs of ≥ 60°.
    Figure 2. FA, phase difference, and error maps of the phantom calculated by the AFI method. FA maps determined by AFI are considerably different from those by the VFA method for nominal FAs of 90 and 130°, and the map is slightly different only at the central area in that of 50°. Pixels with erroneous FA values due to the aliasing phenomenon are readily detected by using the phase difference images by applying the threshold of 1/6 rad.
  • DeepRF-Grad: Simultaneous design of RF pulse and slice selective gradient using self-learning machine
    Jiye Kim1, Dongmyung Shin1, Juhyung Park1, Hwihun Jeong1, and Jongho Lee1
    1Department of Electrical and Computer Engineering, Seoul National University, Seoul, Korea, Republic of
    We developed a new deep reinforcement learning method that simultaneously designs an RF pulse and a slice-selective gradient waveform for slice selective inversion with reduced SAR and improved robustness to off-resonance frequency.
    Figure 1. Summary of DeepRF-Grad for a slice-selective inversion pulse. Compared to the SLR results, the DeepRF-Grad designed RF pulse and z-gradient produced a substantial reduction in SAR (62% reduction) while sustaining a similar slice profile (SLR: blue line, DeepRF-Grad: red dashed line) and satisfying hardware constraints (slew rate and maximum gradient). This idea of designing both RF and gradient to reduce SAR is similar to VERSE and comparisons including VERSE design continue in Figs 3 and 4.
    Figure 2. Comparison of a) DeepRF and b) DeepRF-Grad. Both methods use deep reinforcement learning (DRL) and gradient descent. a) DeepRF designs only RF magnitude and phase, while z-gradient is fixed as uniform. The loss function of DeepRF consists of the slice profile and SAR terms. b) In DeepRF-Grad, z-gradient is also designed simultaneously with an RF pulse. A slew rate term is added to the loss function.
  • Multi-scale Accelerated Auto-differentiable Bloch-simulation based joint design of excitation RF and gradient waveforms.
    Tianrui Luo1, Douglas C. Noll1, Jeffrey A. Fessler1, and Jon-Fredrik Nielsen1
    1University of Michigan, Ann Arbor, MI, United States
    We accelerated an auto-differentiable Bloch-simulator based joint RF and gradient design algorithm with a two stage strategy: 1) fast coarse dwell time design; 2) fine dwell time tuning. It substantially shortens designs, enabling such approaches for online applications.
    Comparison of designs on outer volume excitation objective (OV90). Our proposed method is much faster than the method in [2] (3 min vs 8.5 min). The optimized pulse waveforms and the k-space trajectory resemble that of method [2].
    Comparison of simulated excitation profiles for the pulses in Fig. 2. The goal is to excite only the spins in the outer volume. In terms of NRMSE, our proposed method performance is slightly better than that of method [2].
  • Consistency, ablation, and scalability studies of DeepRF
    Dongmyung Shin1, Jiye Kim1, Juhyung Park1, and Jongho Lee1
    1Electrical and Computer Engineering, Seoul National University, Seoul, Korea, Republic of
    A neural net powered RF design, DeepRF, is investigated. The consistency of the method is confirmed by repeating the same design. The importance of the two modules, generation and refinement, is verified through the ablation. The scalability is validated by changing a design parameter.
    An overview of DeepRF. (a) In the RF generation module, a series of RF values (i.e., actions) are generated from the RNN agent to shape an RF envelope (Nth RF), and the virtual MRI simulates a slice profile. Then, a value of the objective function (e.g., a difference between simulated and desired profiles) is calculated from which the agent changes its behavior to generate a next RF pulse ((N+1)th RF). (b) In the RF refinement module, an RF pulse is refined (Mth RF to (M+1)th RF) with respect to the objective function using RF value changes (∆RF) calculated from the Bloch graph.
    The results of the DeepRF and SLR pulse designs with different TBWs (4.3, 5.6, 6.8, and 8.1). For all the designs, the pulse shapes of the DeepRF pulses are clearly different from those of the SLR pulses (1st and 2nd columns). The slice profiles from the DeepRF and SLR pulses are almost identical (3rd and 4th columns). The SARs of the DeepRF pulses are smaller than those of the SLR pulses in all TBWs (1st column).
  • Exploring RF pulse design with deep reinforcement learning
    Xiaodong Ma1, Kamil Uğurbil1, and Xiaoping Wu1
    1Center for Magnetic Resonance Research, Radiology, Medical School, University of Minnesota, Minneapolis, MN, United States
    We expand the application of a deep reinforcement learning (DRL) pulse design framework to designing basic RF pulses and multi-band pulses. The DRL framework can be used to effectively design all types of RF pulses, improving slice profiles with reduced ripple in comparison to SLR.
    Fig.1. The deep reinforcement learning (DRL) pulse design framework. In a forward pass, the neural network takes the input target slice profile and outputs the predicted RF pulse. The Bloch simulator embedded in the system gives the produced slice profile, which in turn is used (along with the target slice profile) to evaluate the loss. In a backward pass, the gradient of the loss with respect to the weights of the neural network are calculated using backpropagation to update the neural network. The procedures described above are iterated during the training.
    Fig.5. Comparison of the SLR algorithm vs our DRL method when used to design (A) 90◦ excitation, (B) refocusing, (C) inversion and (D) 90◦ excitation multiband (MB=2, gap=50 mm) pulses. For each scenario a separate neural network was trained. Note that despite larger fluctuation observed in the predicted RF pulse, our DRL method led to visually comparable slice profiles for refocusing and inversion, while improving slice profiles for single band and multiband excitation pulses with noticeable suppression of stopband ripples.
  • RF Pulse Designs for Velocity-Selective MRA at Low Field Strengths
    Ziwei Zhao1, Nam G. Lee2, and Krishna S. Nayak1,2
    1Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, United States, 2Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, CA, United States
    Simulations indicate that velocity-selective RF pulses for intracranial MRA achieve comparable performance at 0.55T compared to 3T. We anticipate 50%-60% reduced signal loss and/or 22%-38% sharper velocity selection, potentially improving the detection of slow flow and distal vessels.
    Figure 2. Pulse Performance Tradeoff. (a) Velocity profile sharpness vs. pulse duration (left) and signal loss (right) with (b) detailed parameters of plotted designs. T1 and T2 values were used as 1121ms, 263ms at 0.55T; and 1650ms, 150ms at 3T, respectively. Normalized signal loss is defined as (M0-Mz)/M0. Note that 0.55T system (blue) shows higher fidelities of tradeoffs. The RF and gradient waveforms of two representative designs (marked “A” and “B” asterisks) are plotted in Figure 1.
    Figure 4. Impact of gradient distortions and pre-compensation. (a) GIRF predicted excitation profiles as a function of ∆f and B1+ scale. (b) GIRF pre-compensated profiles. Notice that stripe artifacts appear as spatial modulations (red arrows) and as sidelobe peaks (green arrows) in (a). These are completely resolved in (b) by using GIRF-based pre-compensation.
  • Improved B0 mapping with universal parallel transmit pulses at 7 tesla
    Jürgen Herrler1, Patrick Liebig2, Rene Gumbrecht2, Sydney Nicole Williams3, Christian Meixner4, Andreas Maier5, Arnd Dörfler1, and Armin Michael Nagel4
    1Institue of Neuroradiology, University Hospital Erlangen, Erlangen, Germany, 2SIEMENS Healthineers, Erlangen, Germany, 3Imaging Centre of Excellence, University of Glasgow, Glasgow, Scotland, 4Institue of Radiology, University Hospital Erlangen, Erlangen, Germany, 5Friedrich Alexander University Erlangen Nürnberg, Erlangen, Germany
    A universal parallel transmit (pTx) pulse was used in a B0 mapping sequence. This lead to different values in the resulting B0 maps, mainly at tissue interfaces. When used for the design of a pTx inversion pulse, these differences lead to reduced B0 artifacts in the corresponding sequence.
    Figure 4 MPRAGE images using FOCUS pTx pulses, based on B0 maps, which were acquired with either a CP pulse or a UP. The purple arrows indicate B0 related artifacts, which were weakened when using the UP-improved B0 mapping for designing the pTx inversion pulse. Different behavior of the excitation pulses could not be observed.
    Figure 3 FA simulations of FOCUS excitation (Exc) and inversion (Inv) pulses using either the B0 map acquired with a CP pulse (B0CP) or a UP (B0UP), as well as the difference maps (‘UP case – CP case’). All Bloch simulations were performed using the B0 map acquired with the UP. B0 maps acquired with CP pulses may lead to a worse performance of the FOCUS inversion pulse, especially in the sinus region (purple arrow).
  • Universal parallel transmit pulses for a 2-dimensional local excitation target pattern at 9.4T
    Ole Geldschläger1, Dario Bosch1,2, and Anke Henning1,3
    1High-field Magnetic Resonance, Max-Planck-Institute for biological Cybernetics, Tübingen, Germany, 2Biomedical Magnetic Resonance, University Hospital Tübingen, Tübingen, Germany, 3Advanced Imaging Research Center, University of Texas Southwestern Medical Center, Dallas, TX, United States
    The ‘Universal pTx Pulse’ concept for local excitation pulses was applied in vivo at 9.4T. Universal pulses for a 2D local excitation target pattern were designed. In line with  previous findings for whole-brain excitation pulses they perform just slightly worse compared to the tailored pulses.
    Figure 4: 9.4T T2*-weighted GRE acquisitions (voxel size: 0.8x0.8x0.8 mm3, matrix size: 224x280, TR = 18 ms, TE = 8ms) from the three non-database heads. Once, the respective TPs (FA7) were applied, once the UP (FA7) was applied.
    Figure 3: Bloch simulated FA profiles of the TPs for FA90 (first line of profiles), UP2D for FA90 (second line), the TPs for FA7 (third line) and UP2D for FA7 (fourth line). The eight columns on the left present the database heads. The three columns on the right present the non-database heads. The bar plot below the profiles illustrates the NRMSEs for each pulse and head.
  • Spatial localisation for mapping regional oxygen extraction fraction using parallel transmission saturation pulses at 7 T
    Yan Tong1, Peter Jezzard1, Caitlin O'Brien1,2, and William T Clarke1
    1Wellcome Centre for Integrative Neuroimaging, FMRIB Division, NDCN, University of Oxford, Oxford, United Kingdom, 2Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom
    A parallel transmission RF pulse design is presented to measure regional oxygen extraction fraction at 7 T based on the TRUST method. Initial phantom results showed the feasibility of the method, but further in vivo studies are needed for the validation of this approach.
    EPI images acquired on an oil phantom with a pulse voltage of 180V and different numbers of repetitions of the saturation module. The left, middle, and right columns show 1, 2, and 3 repetitions of the saturation module, respectively. The bottom, middle, top rows show images acquired at the bottom, in the middle, and at the top of the inversion slab, respectively.
    Schematic of the labelling strategies for the proposed method. A cylindrical ROI is chosen.