Artifacts & Corrections
Acq/Recon/Analysis Thursday, 20 May 2021
Oral

Oral Session - Artifacts & Corrections
Acq/Recon/Analysis
Thursday, 20 May 2021 12:00 - 14:00
  • Left-Right Intensity Asymmetries Systematically Vary Across MR Scanners and Introduce Diagnostic Uncertainty
    Arvin Arani1, Christopher G. Schwarz1, Matthew C. Murphy1, Joshua D. Trzasko1, Jeffrey L. Gunter1, Matthew L. Senjem1, Heather J. Wiste1, Kiaran P. McGee1, Matthew A. Bernstein1, John Huston III1, and Clifford R. Jack Jr.1
    1Mayo Clinic, Rochester, MN, United States
    This study shows that left-right intensity asymmetries are system specific, systematic, can mimic disease and create diagnostic uncertainty, and that they impact multiple sequences (T1-weighted and FLAIR).
    Localized regions of inhomogeneity observed on the same patient in both 3D FLAIR and T1-w diagnostic images. In this case the localized region (red arrows) was thought to be suspicious of potential herpes encephalitis and follow-up imaging was conducted.
    Two experienced radiologist’s asymmetry scores (assessment localized to the hippocampus), are plotted against the percent difference in intensity measured in the hippocampus (left side > right side) with automated atlas-based segmentation. The same 30 image volumes from 30 different patients were scored. The color of the data points corresponds to the response of each radiologist to the question: “Is clinical follow-up required?” where: definitely not (black +), uncertain (blue +), definitely yes (red +).
  • Off-Resonance Self-Correction by Implicit B0-Encoding
    Franz Patzig1, Bertram Wilm1, and Klaas Paul Pruessmann1
    1Institut for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland
    We demonstrate the capability of coil-arrays to encode temporal information to be utilized to obtain an estimate of the B0 background field and self-correct single-shot images using standard EPI and spirals (up to R=4).
    Figure 3. Self-correction of images acquired with single-shot spiral and EPI (R=2) trajectories: blurring and geometrical distortions could be effectively mitigated. The algorithm was initialized with no information on B0 given. Overlays of the self-corrected and the uncorrected images show vast differences. The corrected images show only minor differences to the reference images and in some areas even more details in the spiral images. The estimated field maps are similar for spiral and EPI trajectories in both slices.
    Figure 5. Single-shot self-correction shown for spiral-out and EPI read-outs for varying undersampling factors. In the fully sampled cases field maps with 15x15 coefficients were fitted; in the undersampled scenarios (R=2-4) maps modelled with 13x13 coefficients were fitted to increase the regularization. B0-introduced blurring and geometrical distortions in spirals and EPIs (lines added for assistance), respectively, could effectively be reduced without requiring any prior knowledge on B0.
  • Improved Fat and Water Depiction in Musculoskeletal MRI by Control of Through-Slice Chemical-Shift Artifacts in 2D Turbo-Spin-Echo Imaging at 7 T
    Constantin von Deuster1,2, Stefan Sommer1,2, Christoph Germann3,4, Natalie Hinterholzer2, Robin M. Heidemann5, Reto Sutter3,4, and Daniel Nanz2,4
    1Siemens Healthcare, Zurich, Switzerland, 2Swiss Center for Musculoskeletal Imaging (SCMI), Balgrist Campus, Zurich, Switzerland, 3Radiology, Balgrist University Hospital, Zurich, Switzerland, 4University of Zurich, Zurich, Switzerland, 5Siemens Healthcare, Erlangen, Germany
    MSK Imaging at 7 T using Turbo-Spin-Echoes (TSE) suffers from severe through-slice chemical-shift artifacts due to unmatched, low to moderate RF-Pulse bandwidths. Increasing and matching RF-bandwidths to 1500 Hz for both pulse types allows reliable non-fat suppressed MSK imaging.
    a) TSE Imaging parameters: TR/TE: Repetition/Echo Time, FoV: Field of View, Resol: Spatial Resolution, Aver./Conc.: Averages/Concatenations, iPat: Grappa Acceleration Factor, ROBW: Readout Bandwidth, SAR: Specific Absorption Rate, b) evaluation criteria for qualitative analysis of through-slice chemical-shift artifacts (CSA).
    The chemical-shift artifact (*) width (green double arrow) was measured at two locations (black arrows): (i) at 50 % condyle height and (ii) separated by 10 mm along the posterior direction (a). b) Stronger artifacts due to unmatched and moderate RF-bandwidths (UMB), compared to smaller artifacts obtained with matched and increased RF-bandwidths (MIB) (c). The phase encoding direction is along the head-feet direction (dashed white arrow). d) Artifact size for both locations and UMB/MIB. e) Artifact rating for bone and muscle tissues and UMB/MIB (*: significant difference).
  • Concomitant field compensation using additional oscillating gradients in a double diffusion encoding imaging sequence
    Julian Rauch1,2, Frederik B. Laun3, Theresa Palm3, Jan Martin3,4, Maxim Zaitsev5,6, Mark E. Ladd1,2,7, Peter Bachert1,2, and Tristan A. Kuder1
    1Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Division of Physical Chemistry, Lund University, Lund, Sweden, 5Medical Physics, Department of Radiology, Faculty of Medicine, Medical Center University of Freiburg, Freiburg, Germany, 6High Field Magnetic Resonance Center, Center for Medical Physics and Biomedical Engineering Medical University of Vienna, Vienna, Austria, 7Faculty of Medicine, Heidelberg University, Heidelberg, Germany
    Additional oscillating gradients in a double diffusion encoding imaging sequence remove the concomitant phase induced by the used bipolar gradients and correct related artifacts.
    Figure 1: Schematic representation of the proposed method in a spin-echo double diffusion encoding sequence. a) Bipolar diffusion-weighting gradients (blue) are used for double diffusion encoding, here for instance along the physical x and y axes. These gradients induce additional concomitant fields which cause severe artifacts. b) Oscillating gradients (green) are implemented to null the concomitant fields induced by the bipolar gradients.
    Figure 4: Comparison of acquired magnitude images for the long phantom cylinder orientated along the z-axis. The images were corrected for diffusion attenuation. a) Without concomitant field compensation, the image exhibits strong artifacts, especially signal voids (q = 46 mm-1). b) No concomitant field related artifacts can be seen when the concomitant phase is removed by the oscillating gradients (q = 46 mm-1). c) Image acquired with q = 0 mm-1 as reference.
  • Spectrally-encoded multi-spectral imaging (SEMSI) for off-resonance correction near metallic implants.
    Daehyun Yoon1, Philip Lee2, Krishna Nayak3, and Brian Hargreaves1
    1Radiology, Stanford University, Stanford, CA, United States, 2Electrical Engineering, Stanford University, Stanford, United Kingdom, 3Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, United States
    We present a novel spectrally-encoded multi-spectral imaging approach, denoted SEMSI, that can correct for signal translation and pile-up artifacts near metallic implants without the limitations of conventional view-angle tilting.
    Figure 4. Comparison of 2D SE, MAVRIC-SL, and SEMSI. Images on the same slice location by 2D spin-echo (A), Product 3D MSI (MAVRIC-SL) (B), and the proposed SEMSI sequence (C). The proposed SEMSI sequence achieves comparable off-resonance correction with improved sharpness compared to the MAVRIC-SL sequence.
    Figure 2. SEMSI Image Reconstruction. The reconstructed slices images are first FFTed along the echo time dimension after windowing to generate spectral images. Then based on the center frequency of each spectral bin, the spectral images are re-registered to its original location and combined to form the final image.
  • Cancellation of streak artifacts using the interference null space (CACTUS) for radial abdominal imaging
    Zhiyang Fu1,2, Maria I Altbach1,3, and Ali Bilgin1,2,3
    1Department of Medical Imaging, University of Arizona, Tucson, AZ, United States, 2Department of Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States, 3Department of Biomedical Engineering, University of Arizona, Tucson, AZ, United States
    In radial imaging, streaks due to gradient inhomogeneities can obscure pathology. We introduce a streak cancellation method which can be combined with iterative reconstructions and compares favorably with earlier methods as shown in in vivo abdominal experiments.
    Figure 1: Illustration of CACTUS algorithm.
    Figure 4: (a) T2 maps of two representative subjects using locally low rank (LLR) only, LLR with once-pruning ACS, LLR with twice-pruning ACS, and LLR with CACTUS; (b) ROIs on the LLR T2 maps and (c) mean and standard deviation (in parenthesis) of T2 values in millisecond within ROI. Rectangular bounding boxes are placed on both arms for the estimation of interference array correlation statistics.
  • Improved dynamic distortion correction for fMRI using single-echo EPI, a fast sensitivity scan and readout-reversed first image (REFILL)
    Simon Daniel Robinson1,2,3, Beata Bachrata3,4, Korbinian Eckstein3, Saskia Bollmann1, Steffen Bollmann5, Siegfried Trattnig3, Christian Enzinger2, and Markus Barth5
    1Centre for Advanced Imaging, University of Queensland, Brisbane, Australia, 2Department of Neurology, Medical University of Graz, Graz, Austria, 3Department of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria, 4Christian Doppler Laboratory for Clinical Molecular MR Imaging, Medical University of Vienna, Vienna, Austria, 5School of Electrical Engineering and Information Technology, University of Queensland, Brisbane, Australia
    An improved dynamic distortion correction - REFILL - which requires only a 4s pre-scan and a single reversed-readout-encoded EPI volume, provides fieldmaps which accurately reflect field changes due to motion and physiology encountered in fMRI. 
    Fig 5: The effect of shim change on the accuracy of a conventional (static) GE-based fieldmap and the dynamic EPI-based REFILL fieldmap. An imposed shim change, mimicking the effect of motion or physiology in fMRI, leads to additional distortion in frontal regions at 7T (top row, left). The change in field is not captured in the static fieldmap, leading to an incomplete correction (second column). The dynamic REFILL fieldmap for this volume is accurate, and gives a complete correction (third column) - see outline transferred from the distortion-free reference (top row, right).

    Fig.2: Dependence of the readout gradient, φG, on field strength, sequence, receiver bandwidth and echo time. φG was close to linear and quite consistent over phase-encode lines (left) and slices (not shown). It was larger at 3T (red symbols) than 7T (black symbols) and followed opposing trends with TE and bandwidth at the two field strengths.

  • Laser Heating Induced Susceptibility Artifacts Cause Significant Temperature Erros in PRF Shift-based MR Thermometry
    Ziyi Pan1, Meng Han2, Yawei Kuang2, Hao Sun2, Kai Zhang3, Yuan Lian1, Yishi Wang4, Wenbo Liu2, Guangzhi Wang5, and Hua Guo1
    1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2Sinovation Medical, Beijing, China, 3Beijing Tiantan Hospital, Capital Medical University, Beijing, China, 4Philips Healthcare, Beijing, China, 5Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China
    We demonstrate that laser heating induced magnetic susceptibility changes can lead to significant temperature errors in PRF shift-based thermometry. We proposed a new temperature imaging algorithm based on the multi-echo GRE sequence to correct the susceptibility errors.
    FIG. 5 Representative zoomed-in temperature images of the epilepsy patients obtained during MRgLITT treatment. The temperature maps were superimposed onto the post-ablation T1w+Gd images. (a) 38 years old, female, hippocampal sclerosis. (b) 34 years old, male, focal cortical dysplasia (FCD). Results without susceptibility correction (top row, traditional algorithm, TE=19ms) and with correction (bottom row, multi-echo based algorithm, multiple TEs) are both demonstrated.
    FIG. 4 Selected thermal images of a Doberman. From left to right: (1) single-echo TE=4ms and (2) TE=19ms temperature images calculated via the traditional PRF algorithm; (3) the multi-echo (TE=4/9/14/19ms) combined temperature image using the proposed multi-echo algorithm. Note that the proposed method outperforms its single-echo counterparts, providing robust correction of the susceptibility induced artifacts at the heating center, and effective suppression of the CSF flow-induced artifacts near the ventricles (artifacts pointed by the yellow arrows).
  • Identifying the source of spurious echoes in single voxel 1H MR Spectroscopy
    Zahra Shams1, Dennis W.J. Klomp1, Vincent O. Boer2, Jannie P. Wijnen1, and Evita C. Wiegers1
    1Department of radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark
    We analyzed the possible locations of which stimulated echoes could arise, where the dephasing of the local magnetic field inhomogeneity opposed the crusher gradients. This leads to signal rephasing.
    Figure 3. Maximum intensity projection (MIP) maps of all possible pathways for (A) the optimized crusher scheme and (B) the same scheme with reverse polarities of the gradients along FH direction. The MIP maps depict the number of occurrence among all pathways per B0 map (PB-2nd) voxel where φcrush matches -φinhom.gradient. The red arrows point to the locations with more than 3 occurrences through pathways.
    Figure 4. Spectra with PB-2nd VOI shim setting. (A) Time and frequency domain signal from a voxel of 2 × 2 × 2 cm3. (B) Comparison of spectra (NSA=1) acquired from CSI (spatial resolution = 27 cm3, matrix size = 7 × 7 × 6) and SV MRS (NSA=16) with the same localization (semi-LASER, TE = 31 ms). The frequency domain signals from the central slice, contain the VOI (red box), of a matrix size of 7 × 7 × 6 (left). All the spectra in the grid are scaled by a factor of 10 compared to the VOI. Resulting spectrum from SV MRS which was placed on the VOI can be seen in the right-bottom of the figure.
  • Markerless optical head tracking system using facial features
    Toru Sasaki1, Ryuichi Nanaumi1, Mitsuo Nishimura1, Kazuhiko Fukutani1, Shuichi Kobayashi1, Kazuya Okamoto2, Hiroshi Kusahara2, and Kazuto Nakabayashi2
    1Medical Products Technology Development Center, R&D Headquarters, Canon Inc., Tokyo, Japan, 2Advanced MRI Development PJ Team, Canon Medical Systems Corp., Kanagawa, Japan
    A markerless stereo optical system tracking facial features such as skin textures and eyebrows, was developed. Despite the limited apertures of the head coil, the optical system can track the volunteer’s facial features during scanning.
    Fig. 1. Optical tracking system designed for the 64ch head coil. The four MR-compatible cameras record the patient’s face in the head coil through the mirror.
    Fig. 2. Face images obtained from the four MR-compatible cameras. Green squares indicate the facial features for tracking.
Back to Top
Digital Poster Session - Artifacts & Corrections Related to the Acquisition
Acq/Recon/Analysis
Thursday, 20 May 2021 13:00 - 14:00
  • Automated Radial Streaking Artifact Suppression with RGB-STAR
    Rohit Chacko Philip1, Ali Bilgin1,2,3, and Maria I Altbach1,2
    1Medical Imaging, University of Arizona, Tucson, AZ, United States, 2Biomedical Engineering, University of Arizona, Tucson, AZ, United States, 3Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States
    Streaking artifacts due to gradient nonlinearities in radial imaging are suppressed using an automated region growing algorithm, which eliminates the need for manual selection of ROIs for interference covariance matrix selection.
    Qualitative results on a representative axial slice of an abdominal MRI scan reconstructed (a) without any streak suppression algorithm (reference), and with (b) SW, (c) CR+OP, (d) B-STAR (i.e., manual ROI), (e) RGB-STAR (Auto Center), and (f) RGB-STAR (Auto Arms).
    Bland-Altman plots showing agreement between B-STAR and the two automated RGB-STAR interference covariance matrix generation methods.
  • Aliasing Artifact Reduction in Spiral Real-Time MRI
    Ye Tian1, Yongwan Lim1, Ziwei Zhao1, Dani Byrd2, Shrikanth Narayanan1,2, and Krishna S. Nayak1
    1Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, United States, 2Department of Linguistics, University of Southern California, Los Angeles, CA, United States
    We present and evaluate two methods that mitigate a common artifact in mid-sagittal spiral MRI of speech production, arising from gradient nonlinearity and an ineffective anti-aliasing filter.
    Figure 2. Illustration of spiral aliasing artifacts in a numeric phantom and in-vivo. The numeric phantom is based on XCAT with SPGR contrast (a), using simulated coil sensitivities (b), and gradient non-linearity maps of the GE Zoom coil. Simulated image shown in (c) illustrates the distortion and the hotspot outside the FOV (green arrow). Simulated spiral reconstructions without (d) and with (e) readout LPF both contain spiral aliasing artifacts within the FOV (blue arrows). This behavior closely matches in-vivo data, with a representative example shown in (f).
    Figure 1. Flowcharts of artifact reduction methods. The data was temporally combined and reconstructed to 2.5×FOV to obtain coil images. Root sum-of-squares coil combination was performed, and the resulted image was smoothed. A mask M1 was automatically detected around the pixel with the highest intensity outside the FOV by thresholding. The LF method can then be performed by applying masks M1 and M2 during reconstruction. For the ES method, M1 is applied on the coil images, and k-space of the aliasing source was estimated and subtracted from the raw k-space for further processing.
  • Streaking artifact reduction of free-breathing undersampled stack-of-radial MRI using a 3D generative adversarial network
    Chang Gao1,2, Vahid Ghodrati1,2, Dylan Nguyen3, Marcel Dominik Nickel4, Thomas Vahle4, Brian Dale5, Xiaodong Zhong6, and Peng Hu1,2
    1Department of Radiological Sciences, University of California Los Angeles, Los Angeles, CA, United States, 2Department of Physics and Biology in Medicine, University of California Los Angeles, Los Angeles, CA, United States, 3Department of Bioengineering, University of California Los Angeles, Los Angeles, CA, United States, 4MR Application Predevelopment, Siemens Healthcare GmbH, Erlangen, Germany, 5MR R&D Collaborations, Siemens Medical Solutions USA, Inc., Cary, NC, United States, 6MR R&D Collaborations, Siemens Medical Solutions USA, Inc., Los Angeles, CA, United States
    We developed a 3D residual generative adversarial network to remove streaking artifacts of undersampled stack-of-radial MRI. We have shown the feasibility of the network with 3.1x to 6.3x acceleration factors and 6 different echo times.
    Figure 4. Example image quality of the single-echo test results with 6.3x to 3.1x accelerator factors. The input shows the undersampled images, the output shows the images after our network destreaking, and the target shows the fully-sampled ground truth images.
    Figure 5. Example image quality of the multi-echo test results of a 4.2x acceleration factor. The input shows the undersampled images, the output shows the images after our network destreaking, and the target shows the fully-sampled ground truth images.
  • A Method for Correcting Non-linear Errors in Radial Trajectories
    Hideaki Kutsuna1, Sho Kawajiri2, and Hidenori Takeshima3
    1MRI Systems Development Department, Canon Medical Systems Corporation, Kanagawa, Japan, 2MRI Systems Development Department, Canon Medical Systems Corporation, Tochigi, Japan, 3Advanced Technology Research Department, Research and Development Center, Canon Medical Systems Corporation, Kanagawa, Japan
    The authors propose a new method to use a non-linear model for correcting errors in radial trajectories. The proposed method suppresses streak artifacts and image shadings which appear with conventional linear correction.

    Illustration of the estimation procedure for the trajectory shifts.

    Blue arrows represent the first blade.

    Red arrows represent the second blade.

    Reconstructed images obtained from the stack-of-stars volunteer scan, under free-breathing

    (a) Image reconstructed without correction

    (b) Image reconstructed with the linear correction

    (c) Image reconstructed with the non-linear correction.

    Comparing (b) to (c), streak artifacts (arrow 1) and image shading (arrow 2) are suppressed in (c).

    The scan parameters: TE=1.3ms, TR=3.6ms, FOV=35cm*35cm, matrix-size=256*256, sampling-pitch=6$$$\mu$$$s, number-of-slices=50, slice-thickness=4mm, number-of-trajectories=800.

  • Fast Image Reconstruction for Non-Cartesian Acquisitions in the Presence of B0-inhomogeneities
    Mirco Grosser1,2 and Tobias Knopp1,2
    1Section for Biomedical Imaging, University Medical Center Hamburg-Eppendorf, Hamburg, Germany, 2Institute for Biomedical Imaging, Hamburg University of Technology, Hamburg, Germany
    We propose a fast algorithm for the reconstruction of non-cartesian acquisitions with B0-inhomogeneity. The proposed method uses a new SVD-based approximation of the B0-aware imaging operator in combination with diagonal k-space preconditioning.
    Figure 4: Two reconstructed slices of the brain dataset. Reconstructions were performed using TS with $$$L=10$$$ (second column), TS with $$$L=20$$$ (third column) and the proposed SVD-based method (fourth column).
    Figure 5: The upper part of the figure shows the normalized root mean squared deviation (NRMSD) of the reconstructed slices. The middle part displays the number of basis functions $$$L$$$ employed by the SVD based approximation. The bottom plot shows the NRMSD in dependence of the iterations of the PDHG solver.
  • FID-navigated phase correction for multi-shot 3D EPI acquisitions
    Tess E Wallace1,2, Tobias Kober3,4,5, Simon K Warfield1,2, and Onur Afacan1,2
    1Computational Radiology Laboratory, Boston Children's Hospital, Boston, MA, United States, 2Radiology, Harvard Medical School, Boston, MA, United States, 3Advanced Clinical Imaging Technology, Siemens Healthcare, Lausanne, Switzerland, 4Radiology, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland, 5LTS5, Ecole Polytechnique Federale de Lausanne, Lausanne, Switzerland
    FIDnavs embedded in a 3D EPI sequence can rapidly and accurately estimate shot-to-shot B0 field changes up to second order, resulting in substantially improved image quality and tSNR in the presence of spatiotemporal B0 variations.
    Figure 4. Axial slice through 3D EPI image in a volunteer acquired in the reference position (a), combined image from reference position and nose touching (b), and augmented reconstruction using ground-truth MGRE field maps (c) and FIDnav field estimates (d). Difference images relative to the reference image (e–g). FID-navigated augmented reconstruction successfully compensates for artifacts in the combined image with B0 variations.
    Figure 2. Comparison of field difference maps (in Hz) measured from multi-gradient-echo (MGRE) data and FIDnavs for changing shim settings in the ADNI phantom. FID-navigated field estimates are in excellent agreement with ground-truth pixel-wise field maps.
  • Temporal Oscillation in the Phase Error as an Unresolved Source of Ghosting in EPI at 7T
    Pål Erik Goa1 and Neil Peter Jerome2
    1Dept. of Physics, NTNU, Trondheim, Norway, 2Siemens Healthineers, Trondheim, Norway
    Significant temporal variations in the linear offset in odd-even k-space lines is observed in Echo-Planar Imaging at 7T with classic Nyquist ghost as a result. Preliminary analysis indicates the issue can be corrected using extended navigators the captures the oscillatory behavior. 
    Figure 3: Experimental results for three different MR-systems. Top row: single channel measured values of line-by-line δkRO- δkRO,nav. For the 3T system there are no oscillations or offset observed. For the two 7T systems, the temporal development is quite similar, with both initial oscillations and a significant offset between the navigator-measured δkRO and the equilibrium line-by-line value. Middle row: reconstruction based on standard navigator-method. Bottom row: reconstruction using line-by-line δkRO correction. W/L = [0, 0.1*median(phantom)].
    Figure 2: Appearance of Nyquist ghost for different temporal development of δkRO. Top row: plot of δkRO in units of ΔkRO as function of kPE. Middle row: resulting images with window/level (W/L) = [0, max], bottom row: same images but with W/L = [0,0.1*max]. Increasing the linear phase offset (columns 1-2) produces the expected Nyquist ghosting; an oscillating offset introduces more complex ghosting dependent on the frequency (columns 3-4), where a decaying oscillation will have a reduced effect according to the decay time (column 5).
  • Automatic detection of “fat suppression-like” artifacts in brain diffusion MRI
    Stefano Tambalo1,2, Riccardo Pederzolli2, Andrea Spagnolo2, Lisa Novello1, and Jorge Jovicich1
    1CIMeC, University of Trento, Trento, Italy, 2Department of Radiology, G.B. Rossi Hospital, University of Verona, Verona, Italy
    We propose a tool that automatically detects “poor fat-suppression like” artifacts in diffusion MRI data. The tool indicates the specific image slices and volumes where the artifact is present, with high specificity (0.977) and sensitivity (0.889).
    Figure 3. Representative images with different level of artifact intensity and corresponding Dice Index values as a function of posterior-anterior template shift. From top to bottom: high, medium, and low level. The prominent peak in each graph corresponds to the best-matching offset of template and artifact.
    Figure 2. A brief overview of the tool processing steps from input image to artifact detection. a) input image; b) skull mask and template; straightened image; d,e) image gradient; f, g, h) image gradient merged and shifted for edge enhancement; i) base image for template matching; j) overlay of base image and template. A detailed description of each step is given in the text.
  • Effects of Phase-Encoding Directions on Diffusion MRI Reproducibility
    Grayson Clark1, M. Okan Irfanoglu1, and Carlo Pierpaoli1
    1QMI, NIBIB/NIH, Bethesda, MD, United States
    The reproducibility of DTI derived metrics varies among the data acquired with different phase-encoding directions. Data acquired using LR/RL phase-encoding directions were superior in reproducibility to AP/PA data in the whole brain although regional variability was also observed.
    Figure 1. Population-level standard deviation maps for TR for all four phase-encoding directions. As evident in the temporal lobes of AP and PA data, ghost artifacts originating from the eyes significantly increased the variability in these regions. Additionally, in the Pons region, the variability was again significantly higher in AP and PA data compared to RL and LR. The reproducibility at mid-brain level was mostly similar for all phase-encoding directions except a small region of high variability near the caudate for the PA data.
    Number of regions each PED produced the lowest and highest variabilities. The columns labeled “# lowest" and "# highest" variability indicate the number of ROIs where a given PED yielded the lowest or highest raw standard deviation values. The columns “# clearly best” and "worst” show the number of times the difference between the automatically computed "more reproducible" and "less reproducible" cluster centroids was more than 20%. These values indicate an overall trend where LR and RL PED generally produced more reproducible results and PA PED was the least reproducible.
  • Simultaneous super resolution and distortion correction for Single-shot EPI DWI using deep learning method
    Xinyu Ye1, Yuan Lian1, Pylypenko Dmytro1, Yajing Zhang2, and Hua Guo1
    1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2MR Clinical Science, Philips Healthcare, Suzhou, China
    We demonstrate that the deep learning-based method can jointly increase resolution and correct distortions for SSh-EPI DWI. Based on the results, our method provides more consistent DWI images with the undistorted high-resolution reference images.
    Fig. 3. Comparison between state-of-the-art methods with the proposed method using the test subject. 2 representative slices of b0 and mean DWI images are selected. The red arrowheads point at regions where residual distortion and blurring artifacts still exist after distortion correction.
    Fig. 2. Validation results of the proposed method. Input LR SSh-EPI, T2w-TSE, reference PSF-EPI and the reconstructed results are shown in the figure. b0 and mean DWI results show consistency with the reference image.
  • Convolutional Neural Network for Slab Profile Encoding (CPEN) in Simultaneous Multi-slab (SMSlab) Diffusion Weighted Imaging
    Jieying Zhang1, Simin Liu1, Yuhsuan Wu1, Yajing Zhang2, and Hua Guo1
    1Center for Biomedical Imaging Research, Department of Biomedical Engineering, Tsinghua University, Beijing, China, 2MR Clinical Science, Philips Healthcare, Suzhou, China
    CPEN outperforms NPEN in boundary artifacts correction for SMSlab. The results of CPEN shows fewer residual artifacts and higher SNR, and its computation is much faster. It’s meaningful when a large number of DWI volumes are acquired.
    Figure 3. The correction results of the images with 1.3mm (a) and 1mm (b) isotropic resolutions. The uncorrected images (1st column), SMSlab NPEN (2nd column), CPEN (3rd column) and the reference images (4th column) are shown. In each sub-figure, the upper row shows the b0 images, and the bottom row shows the b=800 s/mm2 images. The acquisition time is marked on top of each column. The nominal SNRs of the reference and the SMSlab images are comparable.
    Figure 4. MD (a) and FA (b) maps of the 32-direction dataset. 1 mm isotropic resolution and b=1000 s/mm2 are used. The acquisition time is 33 mins. The uncorrected images (1st column), and the results of CPEN (2nd column) are shown.
  • Validation of a Retrospective Eddy Current Correction Algorithm for Advanced Diffusion MRI
    Paul I Dubovan1,2, Jake J Valsamis1,2, and Corey A Baron1,2
    1Medical Biophysics, Western University, London, ON, Canada, 2Center for Functional and Metabolic Mapping, Robarts Research Institute, London, ON, Canada
    Considering the time-varying behavior of eddy currents in a computational eddy current correction model results in improved image quality for an advanced diffusion MRI technique known as Oscillating Gradient Spin Echo.
    FIG. 3. Quantitative assessment of eddy current correction quality between FSL eddy only (FSL), our time-constant model (TCEDDY), our time-varying model (TVEDDY), and field monitored eddy current correction (FM). For both PGSE and OGSE acquisitions, the MSE was measured between all volumes with inverse distortions of both subjects, normalized by the mean MSE of each subject without correction, and averaged over all directions. Statistically significant changes in MSE between adjacent correction techniques is indicated by *(p < 0.01).
    FIG. 2. Qualitative comparison between eddy current corrupted OGSE slice without correction, and after correction with FSL eddy, TCEDDY, TVEDDY, and field monitored eddy current modelling. Yellow arrow: blurring artefact corrected by modelling eddy current decay.
  • Retrospective Eddy Current Artifact Reduction for Balanced SSFP Cine Imaging via Deep Learning
    Cynthia Chen1, Christopher Sandino2, Adam Bush3, Frank Ong2, and Shreyas Vasanawala3
    1California Institute of Technology, Pasadena, CA, United States, 2Electrical Engineering, Stanford University, Stanford, CA, United States, 3Radiology, Stanford University, Stanford, CA, United States
    Eddy currents due to changing magnetic fields reduce diagnostic image quality, especially in SSFP acquisitions. We propose a retrospective deep learning method that successfully reduces eddy current artifacts in 2D cardiac cine imaging.
    Figure 4. Prospectively acquired data with optimal view ordering but apparent eddy current artifacts and prediction following application of 3D U-Net. (Left) A 2D cardiac cine dataset is prospectively acquired with 12X acceleration, then reconstructed using a model-based deep learning approach. Eddy current artifacts manifest as residual aliasing across the image. (Right) The images on the right are corrected using our 3D U-Net approach, resulting in reduction of eddy current artifacts.
    Figure 2. Data processing pipeline. A) Eddy current effects of random phase encoding orderings were simulated by scaling each k-space line by a random amplitude in a randomly chosen range. B) 3D-Unet Architecture with an additive skip connection to enable residual learning. The input and output sizes are Nx224x224x32.
  • Dispersing FID artifact uniformly by modulating phase of 180 degrees pulse of Spin Echo sequence with quadratic function.
    Kosuke Ito1 and Atsushi Kuratani1
    1Healthcare Business Unit, Hitachi, Ltd., Kashiwa, Japan
    Dispersing FID artifact in spin echo image uniformly by modulating phase of 180 degrees pulse quadratic function was proposed. Signal intensity of artifact is smaller, uniform along phase encode direction, and artifact was not confirmed on images. Higher acceleration factor was applied.

    Figure 3 Volunteer T1 weighted image

    (a): conventional phase control

    (b): proposed phase modulation

    (c): difference between (a) and (b), Window width is 1/5 of (a) and (b)

    (d): line profile on FID artifact

    Figure 1. RF phase of proposed method (a), and signal intensity of FID artifact
  • Removal of Gibbs ringing artefacts for 3D acquisitions using subvoxel shifts
    Thea Bautista1, Jonathan O'Muircheartaigh2,3,4,5, Joseph V Hajnal1,3, and J-Donald Tournier1,3
    1Department of Biomedical Engineering, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2Forensic & Neurodevelopmental Sciences, King's College London, London, United Kingdom, 3Centre for the Developing Brain, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 4Department of Perinatal Imaging & Health, School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 5MRC Centre for Neurodevelopmental Dirorders, King's College London, London, United Kingdom
    The subvoxel shifts method for the removal of Gibbs ringing artefacts can trivially be extended to 3D with simple modifications, and performs well on numerical phantom and in vivo data. 
    overview of the 3D subvoxel shifts approach. Each panel shows three orthogonal cuts through the 3D image. The approach is identical to the original 2D version, but applies the filtering and subvoxels shifts along all three spatial axes, with modified filters appropriate for the 3D case. For each axis, the image is Fourier filtered to attenuate high frequencies in the directions orthogonal to the axis of interest. Gibbs ringing removal is then applied using 1D subvoxel shifts along that axis. The resulting three images are then summed together to give the final output.
    demonstration of the proposed method in practice, using a T2-weighted 3D acquisition (details). The top two rows show 3 orthogonal cuts through the original image with evident Gibbs ringing artefacts and the output of the proposed method. The bottom two rows show magnified sections of the same images for closer inspection.
  • Ringing Artifacts Reduction with Low-Pass Filtered Deblurring Kernels
    Dinghui Wang1 and James G. Pipe1
    1Radiology, Mayo Clinic, Rochester, MN, United States
    Ringing artifacts can occur in spiral imaging at regions with spatially rapidly changing field map, especially with long readout. The proposed deblurring method with a modified model can substantially reduce the ringing artifacts. 
    Figure 4 Comparison of results. A mask computed based on the Laplacian of Δf0 was used to combine the results of the standard deblurring (a) and the proposed method (c). The combined results are shown in (b). Water and fat in (a) were also low-pass filtered (d). Note that the residual ringing inside the yellow boxes is significantly less in (c) (d) compared to (d).
    Figure 5 Results of 2D T2-weighted images with spiral readout length of 45 ms. (a) is the blurred TE image. the field map and and its gradient and laplacian are shown in (b)-(d), respectively. A mask computed based on the Laplacian of Δf0 (d) was used to combine the results of the standard deblurring (e) and the deblurring with low-pass filtered kernels (g). The combined results are shown in (f). The ringing artifacts (pointed by the arrows) in (e) are mitigated in (f) and (g).
  • Removal of Partial Fourier-Induced Gibbs (RPG) Ringing artifacts in MRI
    Hong-Hsi Lee1, Dmitry S Novikov1, and Els Fieremans1
    1New York University School of Medicine, New York, NY, United States
    We investigate and remove the Gibbs-ringing artifact caused by partial Fourier (PF) acquisition and zero filling interpolation in MRI data. With the understanding of oscillating convolution kernels due to the PF acquisition, the ringings in magnitude images can be robustly removed.
    Fig 3. Numerical simulations of Gibbs-ringing removal in 2D magnitude images of the Shepp-Logan phantom with phase. The first two rows show images of the signal magnitude before and after ringing removal, and the third row shows residual maps. For pf = 7/8 and 6/8, the RPG pipeline robustly removes PF-induced ringings, whereas for pf = 5/8 the corrected image still has non-trivial amount of ringings after applying RPG, potentially caused by the interaction of the images phase in high resolution ground truth and the convolution kernel in Eq. (1).
    Fig 4. Mean diffusivity (MD) map baed on diffusion-weighted-images (DWIs) before and after Gibbs-ringing removal using local subvoxel-shifts (SuShi) and RPG, with the residual of MD before and after corrections. The unit of MD is $$$\mu$$$m$$$^2$$$/ms.
Back to Top
Digital Poster Session - Artifacts & Corrections Related to the System & Sample
Acq/Recon/Analysis
Thursday, 20 May 2021 13:00 - 14:00
  • Development and evaluation of a numerical simulation approach to predict metal artifacts from passive implants in MRI
    Tobias Spronk1,2,3, Oliver Kraff1, Gregor Schaefers3,4, and Harald H Quick1,2
    1Erwin L. Hahn Institute for MR Imaging, University of Duisburg-Essen, Essen, Germany, 2High-Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany, 3MRI-STaR Magnetic Resonance Institute for Safety, Technology and Research GmbH, Gelsenkirchen, Germany, 4MR:comp GmbH, Testing Services for MR Safety & Compatibility, Gelsenkirchen, Germany
    This study presents a numerical simulation approach that predicts the MR image artifact size of passive medical implants with a high accuracy.
    (A) Simulated spin echo and gradient echo images of a titanium rod acquired on a 3 T MRI system. (B) Measured spin echo and gradient echo image of the titanium rod. (C) Overlay of the artifacts from image (A) and (B). Red color: both artifacts are congruent; yellow: measured artifact only; blue: simulated artifact only.
    This figure compares the artifact size of the titanium (A-C) and stainless steel (D-F) rod of the simulations (light grey) and the measurements (black) in mm2 separated by the different field strengths, (A,D) 1.5 T, (B,E) 3 T, (C,F) 7 T. Artifact sizes are shown for different implant orientations in regard to B0 (para: parallel and perp: perpendicular), sequences (GRE and SE), slice orientations in regard to implant orientation (cor, sag, tra), and phase-encoding direction (FH, AP).
  • Realistic Simulation of MRI Metal Artifact and Field Strength Dependence
    Kübra Keskin1, Brian Hargreaves2,3,4, and Krishna S. Nayak1,5
    1Electrical and Computer Engineering, University of Southern California, Los Angeles, CA, United States, 2Radiology, Stanford University, Stanford, CA, United States, 3Electrical Engineering, Stanford University, Stanford, CA, United States, 4Bioengineering, Stanford University, Stanford, CA, United States, 5Biomedical Engineering, University of Southern California, Los Angeles, CA, United States
    We demonstrate a pipeline for realistic simulation of in-vivo MRI around metal.  We use this simulator to predict the performance of multi-spectral imaging on emerging low-field systems (e.g. 0.55 Tesla) and predict optimal imaging parameters.
    Figure 1: Simulation Pipeline. 3D XCAT body and implant masks are utilized to create tissue parameter maps. The susceptibility map is used to calculate the field shift. K-space is simulated using the field shift, proton density (PD) and MSI sequence parameters with phase encoding in both left-right and anterior-posterior directions. Complex Gaussian noise is added in k-space. Spectral bin images are reconstructed and combined with quadrature summation.
    Figure 2: MAVRIC-SL simulation of total hip arthroplasty. Central coronal slices from 3D volumes are shown for B0 = 0.2T, 0.55T, 1.5T, 3T, and Nbins = 12, 18, 24. Other parameters: BW/pixel = 625Hz; RF bandwidth = 2.25kHz and bin spacing 1kHz. Implant contours are shown with yellow outlines. Artifacts around the implant are indicated with red arrows. A small artifact adjacent to implant surface at 0.55T is shown with green arrow. Total scan times are 3:53, 5:49, and 7:45 for Nbins of 12, 18, and 24, respectively.
  • Joint estimation of image content and field inhomogeneity for artifact correction near metallic implants
    Alexander R Toews1,2, Daehyun Yoon1, and Brian A Hargreaves1,2,3
    1Radiology, Stanford University, Stanford, CA, United States, 2Electrical Engineering, Stanford University, Stanford, CA, United States, 3Bioengineering, Stanford University, Stanford, CA, United States
    Conventional multispectral imaging mitigates geometric distortion near metal at the expense of readout blur and residual intensity artifacts. Phantom and simulation results demonstrate the proposed method reduces readout blur and in some cases eliminates residual intensity artifacts.
    Figure 3: Simulated phantom study of a ball with magnetic susceptibility 900 ppm at 3T. Input images exhibit readout blur and ripple artifact due to VAT. Ripple artifacts are reduced in the initial image (red arrow) and eliminated in the final estimation of $$$m$$$. Grid lines are sharper in the final image (yellow arrow), indicating mitigation of the VAT blur effect. Final $$$f$$$ agrees well with the true field within the region of excited spins.
    Figure 5: Results in shoulder phantom shown for an off-center coronal slice. Ripple-like intensity artifacts are apparent in the input and initial images (red arrow), but significantly reduced in the final image. Differences in readout resolution are less apparent than in Figure 4 due to the absence of grid lines.
  • Accurate mUlti-echo phase image wiTh uneven echO spacing and Ultra-High Dynamic Range (AUTO-HDR)
    Yuheng Huang1,2, Xinheng Zhang1,2, Serry Fradad1, Lu Meng1, Ghazal Yoosefian1, Linda Azab1, Xinqi Li3, Alan Kwan4, Rohan Dharmakumar1, Hui Han1, and Hsin-Jung Yang1
    1Biomedical Imaging Research Institute, Cedars Sinai Medical Center, Los Angeles, CA, United States, 2Bioengineering, UCLA, Los Angeles, CA, United States, 3Columbia University, New York, NY, United States, 4Cardiology, Cedars Sinai Medical Center, Los Angeles, CA, United States
    An accurate and reliable phase mapping algorithm with Ultra-High dynamic range was developed using a multi-echo GRE sequence with uneven echo spacing and iterative reconstruction.
    Figure 1. Image processing flow chart and the uneven echo spacing. The image processing pipeline is presented in panel A. Phase images were first combined with a graph-cut algorithm to conduct high SNR phase maps. The central frequency and unwrapping artifact were further calibrated from the derivative of the phase difference maps between the uneven DTEs. The phase accumulation from the uneven echo spacing is presented in panel B.
    Figure 5. In-vivo images from a healthy human volunteer with an ICD on the chest. In the presents of ICD, the compromised image SNR and strong phase accumulation from the off-resonance source lead to obvious phase unwrapping artifacts and central frequency shifts in the peer algorithms (arrows). The proposed AUTO-HDR shows a much smoother field map around the ICD and higher SNR at a region further away from the ICD. The improved field map demonstrates the capability of AUTO-HDR in resolving fast varying phase change while preserving high SNR in human subjects with metallic implants.
  • Directional effect of frequency-encoding gradient on T2WI-STIR imaging: a phantom study to evaluate the metal artifact
    Xu Lulu1, Shen Yong2, Dou Weiqiang3, and Qi Liang1
    1Radiology, Jiangsu Province Hospital, Nanjing, China, 2GE Healthcare, MR Enhanced Application China, Beijing, China, 3GE Healthcare, MR Research China, Beijing, China
    The directional effect of the frequency-encoding gradient on T2WI-STIR was different from the performance on the T1WI.
    Fig.1 (a-b) Under the default window width and window level, the images show the artifacts, including the low (long arrow) and high (arrowhead) signal intensities. The screw without protons displays a low signal intensity (short arrow).
    Fig.1 (a-b) Under the default window width and window level, the images show the artifacts, including the low (long arrow) and high (arrowhead) signal intensities. The screw without protons displays a low signal intensity (short arrow).
  • Image blending method to produce consistent SNR in images denoised using a convolutional neural network
    Anuj Sharma1 and Andrew J Wheaton1
    1Magnetic Resonance, Canon Medical Research USA, Inc., Mayfield Village, OH, United States
    We propose an image blending method that uses target SNR as a quantitative metric to produce natural-looking, denoised images. The proposed method is shown to produce denoised images with consistent SNR values in head, spine and knee applications.
    Figure 2: Results from the head axial FLAIR experiments. The denoising CNN reduced image noise at the cost of significant smoothing. Blended images demonstrate the use of proposed method to generate images at different target SNR values. The target and measured SNR values closely match each other in all the blended images.
    Figure 3: Results from the lumbar spine sagittal STIR and knee sagittal IW experiments. Images processed by the denoising CNN appear artificially smooth. Images blended using the proposed method have a natural appearance. The target and measured SNR values closely match each other.
  • Denoising MR 2D COSY spectra using a joint sparse parametric Matching Pursuit (jSPaMP)
    Boris Mailhe1, Zahra Hosseini2, Bing Ji3, Hui Mao3, and Mariappan S. Nadar1
    1Digital Technology and Innovation, Siemens Healthineers, Princeton, NJ, United States, 2MR R&D Collaboration, Siemens Medical Solutions USA Inc., Atlanta, GA, United States, 3Department of Radiology and Imaging Sciences, Emory University, Atlanta, GA, United States
    2D COSY data can be denoised using a combination of parametric fitting in time and nonparametric joint sparsity in the j-coupling direction.
    Figure 5: Comparison of the high SNR spectrum obtained with NSA of 96 (left) and a jSPaMP denoised spectrum with NSA of 16 (right).
    Figure 2: Description of the proposed jSPaMP algorithm.
  • Characterising the variance and reproducibility of low rank denoising methods for spectroscopic data
    William T Clarke1 and Mark Chiew1
    1Wellcome Centre for Integrative Neuroimaging, NDCN, University of Oxford, Oxford, United Kingdom
    Despite visually denoising MRS data it is not clear whether low rank denoising decreases uncertainty in metabolite concentrations. We use simulation and in vivo reproducibility to show that denoising is beneficial, but much less than is apparent.
    Figure 1. Denoising of a single identical Lorentzian peak with independent noise. The black line shows the noiseless signal, the red line shows a single voxel’s data after denoising (no denoising applied in “Noisy”, all spectra mean shown in “Avg”). The Shaded grey region shows ± two standard deviations of all 128 spectra. All denoising approaches show high apparent noise level reduction, but LP denoising does not show a reduction in peak amplitude variance.
    Figure 5. Voxel-wise maps of mean metabolite concentration (left, red/yellow) and standard deviation of metabolite concentration (blue) across the ten repeated acquisitions for each denoising case. Data is shown for three metabolites and is overlaid on a T1w structural image. Colour scales are fixed for each metabolite.
  • A Survey of Faraday Cage Attenuation Measurements of Clinical MRI Systems
    Francesco Padormo1, Joe Martin1, Jane Ansell1, Elizabeth Gabriel1, Laurence H. Jackson1, Caitlin O'Brien1, Simon Shah1, David Price1, and Geoff Charles-Edwards1
    1Guy's & St Thomas' NHS Foundation Trust, London, United Kingdom
    We present faraday cage attenuation measurements on eleven MRI systems, to assess the need to replace the cage when replacing/upgrading an MRI system. 
    Figure 1 - Faraday cage attenuation measurements of eleven clinical MRI systems: measurements were acquired at 65MHz, 128MHz and 297MHz at the magnet room door, console room window and penetration panel. Box and whisker plots adjacent to the data show the median in red, interquartile range within the box, whilst the whiskers show the range of data excluding outliers, which are indicated with a red cross.
    Figure 2 - Summary of RF attenuation measurements in eleven clinical MRI systems
  • Usefulness of Deep Learning Based Denoising Method for Compressed Sensing in Pituitary MRI
    Takeshi Nakaura1, Hiroyuki Uetani1, Kousuke Morita1, Kentaro Haraoka2, Akira Sasao1, Masahiro Hatemura1, and Toshinori Hirai1
    1Diagnostic Radiology, Kumamoto University, Kumamoto, Japan, 2Cannon Medical Systems Japan, Tochigi, Japan
    The DLR based denoising method can improve the image quality of T2WI of pituitary with CS as compared with the conventional wavelet based denoising method, and the difference became more noticeable at higher denoising levels.
    Representative case. (upper row) Conventional wavelet based denoising method, (lower row) DLR based denoising method. The DLR based denoising method offered decrease in image noise, artifact, and clear depiction of pituitary grand and small objects as compared with the conventional wavelet based denoising method.
    A box plot shows the SNR of CSF. There was a progressive increase in SNR of CSF with DLR based denoising method with increase of the denoising level in any location. On the other hand, the SNR of conventional wavelet-based method was not increased at high denoising levels (4-5).
  • Evaluation of noise reduction performance using deep learning reconstruction: A phantom study
    Hitoshi Kubo1, Yuya Abe2, Tomoya Yokokawa2, Seira Yokoyama2, and Koji Hoshi2
    1Fukushima Medical University, Fukushima, Japan, 2Hoshi General Hospital, Koriyama, Japan
    SNRs were increased higher significantly by DLR in all SNR ranges. Increasing ratio of SNR was varied by means of parameter settings. Combination of the DLR parameters affected varies to SNR, SSIM, and spatial resolution of the images.
    Signal-to-Noise ratios on the images of 2 mm, 4 mm, and 8 mm of slice thickness with various parameters of Deep Learning Reconstruction (DLR) were shown. DLR had a powerful performance to increase SNR in all SNR ranges.
    Structural similarity (SSIM) of the images were shown. Decreasing of SSIM higher in thinner slice images is compared to that in thicker slice images. Edge enhancement recovered SSIM higher, especially in thinner slice images.
  • Suppressing the ballistocardiography artifacts on EEG collected inside MRI using the dynamic modeling on heartbeats
    Hsin-Ju Lee1,2, Hsiang-Yu Yu3,4,5, Cheng-Chia Lee4,5,6, Chien-Chen Chou3,4, Chien Chen3,4, Wen-Jui Kuo5,7, and Fa-Hsuan Lin1,2,8
    1Physical Sciences Platform, Sunnybrook Research Institute, Toronto, ON, Canada, 2Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada, 3Department of Epilepsy, Neurological Institute, Taipei Veterans General Hospital, Taipei, Taiwan, 4School of Medicine, National Yang-Ming University, Taipei, Taiwan, 5Brain Research Center, National Yang-Ming University, Taipei, Taiwan, 6Department of Neurosurgery, Neurological Institute, Taipei Veterans General Hospital, Taipei, Taiwan, 7Institute of Neuroscience, National Yang-Ming University, Taipei, Taiwan, 8Department of Neuroscience and Biomedical Engineering, Aalto University, Espoo, Finland

    Using both simulations and empirical data at 3T, we demonstrated that the dynamic modeling of heartbeats (DMH) method can suppress the BCG artifacts on the EEG collected inside MRI more efficiently than Optimal Basis Set method in both epilepsy and steady-state visual evoked potential data.

  • Robust spatial and temporal unwrapping for accurate quantitative susceptibility mapping
    Alex Ensworth1, Véronique Fortier1,2, Jorge Campos Pazmino1, and Ives R Levesque1,2,3,4
    1Medical Physics Unit, McGill University, Montreal, QC, Canada, 2Biomedical Engineering, McGill University, Montreal, QC, Canada, 3Research Institute of the McGill University Health Centre, Montreal, QC, Canada, 4Gerald Bronfman Department of Oncology, McGill University, Montreal, QC, Canada
    Quality-guided region-growing phase unwrapping with temporal median phase restoration is more accurate than Laplacian-based methods for quantitative susceptibility mapping. 
    Five echoes after QGRG spatial unwrapping are shown across the columns. From left to right, the first echo is at 5.0 ms, followed by 10.4 ms, 15.8 ms, 21.2 ms and then 26.6 ms. The top row is without temporal unwrapping (TU), the middle row is with temporal unwrapping and the bottom row is a plot of the line profiles for each column. Dataset C2 was used.
    Six different techniques (three Laplacian, top row, and three path-following, bottom row) were used to spatially unwrap phase data from dataset C2. The output from each unwrapping technique is shown with an axial cross-section of the brain, and a line profile beneath it. Panel A represents the Laplacian from the STI suite, B is the Laplacian from the MEDI toolbox, C is the Laplacian from a collaborator, D represents QGRG, E represents SEGUE and F represents region growing from the MEDI toolbox.
  • Off-resonance correction of non-cartesian SWI using internal field map estimation
    Guillaume Daval-Frérot1,2,3, Aurélien Massire1, Mathilde Ripart2, Boris Mailhe4, Mariappan Nadar4, Alexandre Vignaud2, and Philippe Ciuciu2,3
    1Siemens Healthcare SAS, Saint-Denis, France, 2CEA, NeuroSpin, CNRS, Université Paris-Saclay, Gif-sur-Yvette, France, 3Inria, Parietal, Palaiseau, France, 4Digital Technology and Innovation, Siemens Healthineers, Princeton, NJ, United States
    We propose to approximate the $$$\Delta{B_0}$$$ field map of a single echo acquisition using long echo time. The estimated 3D field maps, produced at 0.6 mm isotropic under 3 minutes, are then used for $$$\Delta{B_0}$$$ correction with performances similar to acquired ones.
    Figure 4 : Comparison of the SWI with mIP without correction (A), with correction using the acquired field map (B) and with correction using the estimated field map (D) with the same slices as Fig3. (C) and (E) are respectively the difference between (A) and (B), and between (A) and (D).
    Figure 1 : Pipeline of the field map estimation followed by an off-resonance correction. The magnitude and associated mask estimated with the adjoint are used to weight and stabilize the phase unwrapping algorithm.
  • Characterizing the acquisition protocol dependencies of B0 field mapping and the effects of eddy currents and spoiling
    Divya Varadarajan1,2, Mukund Balasubramanian2,3, Daniel J. Park1, Thomas Witzel4, Jason P. Stockmann1,2, and Jonathan R. Polimeni1,2,5
    1Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 2Department of Radiology, Harvard Medical School, Boston, MA, United States, 3Boston Children’s Hospital, Boston, MA, United States, 4Qbio Inc., San Carlos, CA, United States, 5Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, United States
    Here we investigate the accuracy of the standard B0 field map acquisition, demonstrate that the estimated fields vary with several acquisition parameters, and investigate sources of these errors.
    Figure 1. The figure shows the variation of fieldmaps with TR and plots the associated statistics of the bias.
    Figure 5. Slice select gradient reversal along with PE and FE gradient reversal (a) cancels most of the bias in the fieldmap (b) and reduces to a mean of 0.5 Hz and a 97th percentile of ~2 Hz (c).
  • Application of Voxel Spread Function Method for Correction of Magnetic Field Inhomogeneity at 7T
    seyedeh nasim adnani1, Thomas Denney Jr.1, Alexander Sukstansky2, Dmitriy Yablonskiy2, and adil bashir1
    1electrical and computer engineering, auburn university, auburn, AL, United States, 2radiology, Washington university school of medicine in St. Louis, St. Louis, MO, United States
    Ultra high-field imaging scenarios suffer from increased field inhomogeneities compared with 3T scanners or below. Voxel Spread Function addresses this issue and is feasible for ultra high-field MRI imaging and compensates for the decreased T2* that is the outcome of the inhomogeneities.
    Representative T1 weighted image and T2* maps from the in vivo experiment, a) before correction and b) after macroscopic field inhomogeneity correction. Average increase in the T2* values in both GM and WM is evident. GM/WM T2* contrast is also increased.
    Proton density (a) and examples of T2* maps calculated without (b) and with F-term correction (c). Regions substantially affected by the B0 field inhomogeneity are indicated by red arrows. F-function correction significantly reduces magnetic field inhomogeneity artifacts on T2* maps and sharpens T2* contrast in gray and white matter.
  • Correction of transmit-field induced signal inhomogeneity in 3D MP-FLAIR at 7T
    Jan Ole Pedersen1, Vincent O. Boer2, Oula Puonti2, Jaco M. Zwanenburg3, and Esben Thade Petersen2,4
    1Philips Healthcare, Copenhagen, Denmark, 2Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital, Hvidovre, Denmark, 3Department of Radiology, University Medical Center Utrecht, Utrcecht, Netherlands, 4Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby, Denmark
    A bias-field correction based on simulations and measurements of B1+ was developed to ease radiological assessment of MP-FLAIR at 7T.  In addition to showcasing the algorithm, the abstract exemplifies the high dependency of MP-FLAIR towards B1+ through simulations and volunteer brain images.

    Figure 4:
    Simulated versus measured dependency of MP-FLAIR image intensity to B1+. Dots: Measured MP-FLAIR intensities as a function of measured DREAM B1+. Colors depict different RF gains. Other sources of signal loss (e.g. B1-/B0, fluid-content) cause the measured intensities to “fill out” the area under the curve. Black line: Simulated FLAIR intensities. Increasing discrepancy between simulation and measurements is seen below 60% and above 140% of B1+, which is ascribed to inherent bias in the measured B1+. Red line: The MP-FLAIR dependency to B1+ used for bias-field correction.

    Figure 3:
    5 repetitions of an MP-FLAIR scan with different RF gains (0.6 – 1.4). For low RF gains, only the mid brain (where B1+ is relatively large) experience close to nominal flip-angles. For RF gains higher than 0.8, the mid brain shows as hypo-intense due to flip-angles being larger than the nominal flip-angle. The lower occipital lobe (where B1+ is relatively small), show as hypo-intense for all but the highest RF gain, where it experiences close to nominal flip angles and shows as hyper-intense, due to the surrounding tissue experiencing higher than nominal flip-angles.

  • Integrated Spin-Echo EPI scans for Fast Simultaneous B1 and B0 mapping in the Human Brain
    Sofia Chavez1,2
    1Centre for Addiction and Mental Health (CAMH), Toronto, ON, Canada, 2Psychiatry, University of Toronto, Toronto, ON, Canada
    By combining the requirements for B1 mapping using the DAM and reversed phase encode blips for B0 mapping like that used by topup, we were able to map B1 and B0 from integrated scans.
    Fig.1 Cartoon depicting the proposed acquisitions and processing pipeline. The images contained in the purple square (top left) represent the four SE-EPI scans that need to be acquired (each in 15 s).
    Fig.4 Sample of the B1 map produced using the proposed method compared to the B1 map produced using a typical SE-DAM protocol in our Centre (as per ref.2). The horizontal and vertical profiles are plotted on the right to show the very good agreement between these two B1 maps.
  • Optimisation of MRI Bias Field Correction Algorithms on Whole Brain and Atrophy Measurements
    Kain Kyle1,2 and Chenyu Wang1,2
    1Sydney Neuroimaging Analysis Centre, Sydney, Australia, 2University of Sydney, Sydney, Australia
    In this investigation we compared bias correction techniques by estimating brain volume change in healthy controls.  We found are proposed iterative WM N4 bias correction technique resulted in the lowest percentage brain volume change.
    Figure 1. Example of bias correction applied to native T1-WI. Maps were generated by dividing the corrected image by the native image. Bias correction methods N3 (top left), N4 (top right), N4W(bottom left) and N4ITER(bottom right).
    Figure 2. Summary of iterative WM N4 bias correction technique. 1) Initial WM mask is generated with FSL-FAST on native T1-WI and used for N4Wcorrection. 2) WM mask is segmented from N4 corrected image with FSL-FAST. 3) Volume of WM mask from step 2, is compared to the volume of the WM mask from the previous iteration, or initial input WM mask if it is the first iteration. 4) If the change in volume is greater than 0.05%, the updated mask is used to perform N4Won the native T1-WI. 5) If the change in volume is less than 0.05 percent, the process is terminated.