Systems Engineering I
Engineering/Interventional/Safety Wednesday, 19 May 2021
Oral

Oral Session - Systems Engineering I
Engineering/Interventional/Safety
Wednesday, 19 May 2021 16:00 - 18:00
  • A portable brain MRI scanner based on a 72 mT, 35 kg "Halbach-bulb" magnet and external gradient coil
    Clarissa Zimmerman Cooley1,2, Jason Stockmann1,2, and Lawrence L Wald1,2,3
    1Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Harvard–MIT Division of Health Sciences and Technology, Cambridge, MA, United States
    We present a portable MRI scanner for brain imaging using a 35 kg Halbach bulb magnet and external gradient coils. In vivo human T2-weighted images are demonstrated.
    Prototype halbach-bulb scanner is shown. The order of the components from innermost to outermost are: RF Tx/Rx coil, B0 shim magnets, RF shield, main B0 magnet, and gradient coils. During the scan, the patient’s head rest on the RF coil. The rest of the system (B0 magnet, RF shield, and gradient coils) is on rails and slides over the coil and patient.
    3D T2-weighted images of the brain in a healthy adult volunteer. 3D RARE sequence, resolution = 1.5mm x 2.8mm x 7mm,TR/TEeff =3000ms/134ms, acquisition time = 25 min (6 averages). A) Images reconstructed assuming linear encoding fields resulting in distortion, b) Images reconstructed using the generalized encoding method employing the measured fieldmaps in (c). C) combined measured B0 variation map and readout gradient maps for image reconstruction.
  • Simultaneous imaging of hard and soft biological tissues in a low-field MRI scanner
    José Miguel Algarín1, Elena Díaz2, Pepe Borreguero1, Fernando Galve1, Daniel Grau2, Juan Pablo Rigla2, Rubén Bosch1, José Manuel González2, Eduardo Pallás1, Miguel Corberán1, Carlos Gramage1, Alfonso Ríos2, José María Benlloch1, and Joseba Alonso1
    1I3M, CSIC, Valencia, Spain, 2Tesoro Imaging, Valencia, Spain

    In the present work we have demonstrated the capability of our new low-cost “DentMRI – Gen I” scanner to simultaneously image hard and soft biological tissues; we show that human teeth can be imaged with high resolution at low magnetic fields; and we have devised a new pulse sequence (DRaNSSE).

    (a) Picture of the scanned rabbit head. (b) Picture of a rabbit skull, taken from Gabrielle Ochnik, Pinterest. (c) Top: single slices for 90 µs dead time; middle: the same slices for 1 ms dead time; bottom: difference between the above images.
    (a) 2 dimensional slices of four human teeth embedded in a piece of pork ham (PETRA). (b) Photograph of the sample.
  • Design and Construction of a Low-Cryogen, Lightweight, High-Performance, Head-only  Compact 7T MRI
    Thomas K.F. Foo1, Mark Vermilyea2, Minfeng Xu2, Anbo Wu2, Yihe Hua2, Wolfgang Stautner2, Ye Bai2, Justin Ricci2, Doug Kelley3, John III Huston4, Yunhong Shu4, Matt A Bernstein4, Christopher Hess5, and Duan Xu5
    1GE Global Research, Niskayuna, NY, United States, 2GE Research, Niskayuna, NY, United States, 3GE Healthcare, Fairfax, CA, United States, 4Mayo Clinic, Rochester, MN, United States, 5University of California - San Francisco, San Francisco, CA, United States
    A feasible and practical design for a low-cryogen, high performance, head-only 7T MRI scanner was completed. Construction of that system is underway and is expected to be completed by the end of 2021 to generate first images.
    Figure 1: (a) Proposed C7T magnet showing the 8-coil configuration, cold heads, and helium tanks. The size of the C7T relative to an average person is also shown for scale. (b) One of the 2 large main coils being wound with superconducting NbTi wire.
    Figure 4: Diffusion imaging performance at different b-values showing (a) diffusion TE time, and (b) diffusion pulse width (d) for gradient configurations of whole-body 7T, C3T, C7T, and MAGNUS. The performance improvement between the C7T gradient and the whole-body 7T is clear.
  • Design and Development of a Next-Generation 7T human brain scanner with high-performance gradient coil and dense RF arrays.
    David A Feinberg1,2,3, Peter Dietz4, Chunlei Liu1,5, Kawin Setsompop6, Pratik Mukherjee7,8, Lawrence L Wald9,10,11, An T Vu7,8, Alexander JS Beckett1,2, Ignacio Gonzalez Insua4, Martin Schröder4, Stefan Stocker4, Paul H Bell12, Elmar Rummert4, and Mathias Davids9,10,13
    1Helen Wills Neuroscience Institute, University of California, Berkeley, CA, United States, 2Advanced MRI Technologies, Sebastopol, CA, United States, 3Department of Cognitive Neuroscience, Maastricht University, Maastricht, Netherlands, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Department of Electrical Engineering and Computer Sciences, University of California, Berkeley, CA, United States, 6Radiological Sciences Laboratory, Stanford University, Stanford, CA, United States, 7Radiology, University of California, San Francisco, CA, United States, 8San Francisco Veteran Affairs Health Care System, San Francisco, CA, United States, 9A.A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 10Harvard Medical School, Boston, MA, United States, 11Harvard-MIT Division of Health Sciences Technology, Cambridge, MA, United States, 12Siemens Medical Solutions USA, Inc, Cary, NC, United States, 13Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Heidelberg, Germany
    The scanner is integrated with an asymmetric head gradient coil designed with 3-layers of windings to achieve in-vivo Gmax 200 mT/m and Smax 900 T/m/s. The 128-channel receiver system enables use of high density arrays for greater sensitivity in cortex and higher accelerations.
    Figure 2 - Left) Photo of gradient coil with 3 stepped layers. The overall mechanical length of the gradient coil is 160 cm to mount easily inside magnet and for better dynamics. The total extent of the wires in Z direction is 119 cm. Right) Plot of wiring pattern showing 3 layers as primary (Pri), intermediate (Mid) and secondary (Sec) layers in stepped design.
    Figure 1 - Photos of the scanner. A) Front side of scanner showing RF cables connecting the 96-ch Rx, 16-ch Tx coil to 128-ch interface box. B) gradient coil cover with shoulder cut outs with the coil interface box in front. The front bore is 60 cm diameter for subject access. C) Photo of energy chain between two 64-ch RF receiver interfaces. D) The rear bore is 39 cm diameter. The coil interface box is positioned in the rear of the magnet bore when the RF coil is positioned at isocenter. The flat flexible energy chain shields over 270 cables and is seen connected to the coil interface box.
  • Towards 8ch multi transmit with high power ultrasonic spirals and 72ch receive setup for ultimate spatial encoding at 7T
    Dimitri Welting1, Edwin Versteeg1, Ingmar Voogt2, Joost van Straalen3, Martijn Heintges3, Marco Rietveld3, Jeroen C.W. Siero1,4, and Dennis W.J. Klomp1
    1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Wavetronica, Utrecht, Netherlands, 3Prodive Technologies, Son, Netherlands, 4Spinoza Centre for Neuroimaging Amsterdam, Amsterdam, Netherlands
    A light weight setup is constructed to boost the spatiotemporal resolution for fMRI experiments. Using a modified gradient amplifier and 2-axis insert gradient coil in combination with a 72 channel setup successful MR signals were recorded.
    Figure 1: The components of the setup used. The NG500 1.1 gradient amplifier (A). (B) consists of the 2-axis insert gradient, an 8 channel transceive dipole array, a 64 channel receive array and an agar-based head phantom. The middle images show a more detailed view without insulation tape of the used insert gradient and receiver coils.
    Figure 4: Effect of the oscillating gradient on the MR-signal. Left: the MR-signal with the amplifier off. Middle: the amplifier on. This data was acquired with a slice-selective 1D acquisition showing only one of the simultaneously acquired 72 channels. Right: raw data controlling all 5 gradients.
  • In vivo Two-photon Magnetic Resonance Imaging of Human Brain at 3T
    Jianshu Chi1, Victor Han1, and Chunlei Liu1,2
    1Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, CA, United States, 2Helen Wills Neuroscience Institute, University of California, Berkeley, Berkeley, CA, United States
    We acquired gradient echo images of an in vivo human brain at 3T with all of the standard single-photon excitations replaced by two-photon equivalents. The resulting images were similar with some slight differences in contrast.
    Figure 3: Single- and two-photon GRE brain images of a healthy volunteer. Left panel: Single-photon image with a flip angle of 44.5 degrees vs. a TR of 350 ms (Top) and a TR of 150 ms (Bottom). Right panel: Two-photon image with a $$$B_{1xy}$$$ amplitude corresponding to an equivalent two-photon 44.5-degree flip vs. a TR of 350 ms (Top) and a TR of 150 ms (Bottom). Some brain structure contrast is seen to be enhanced. Other parameters: FOV 27.6 cm, slice thickness 3 mm, matrix 256x160, TE 10ms.
    Figure 2: Left: equipment room set up. Right: $$$B_{1z}$$$ coil in a 3T scanner.
  • PNS optimization of a high-performance asymmetric gradient coil for head imaging
    Mathias Davids1,2,3, Peter Dietz4, Gudrun Ruyters4, Manuela Roesler4, Valerie Klein1,3, Bastien Guerin1,2, David A Feinberg5,6, and Lawrence L Wald1,2,7
    1Martinos Center for Biomedical Imaging, Boston, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 4Siemens Healthineers, Erlangen, Germany, 5Advanced MRI Technologies, Sebastopol, CA, United States, 6Brain Imaging Center and Helen Wills Neuroscience Institute, University of California, Berkeley, CA, United States, 7Harvard-MIT, Division of Health Sciences and Technology, Cambridge, MA, United States
    Informing the design phase of a high-strength head gradient using PNS modeling allowed alteration of the coil windings to balance head and body PNS, which greatly improved PNS thresholds and usability of the coil performance. Results were validated in a constructed coil using PNS experiments.
    Figure 4: Top: Number of reported stimulations in different body parts during the experimental study. Bottom: Maps of predicted PNS hot-spots in the male and female model given in terms of PNS oracle and equivalent PNS thresholds (inverse of the PNS oracle) for a 400 us rise time trapezoidal waveform. Every blob corresponds to an activation hot-spot, with both color and size corresponding to the strength of the activation.
    Figure 1: Photo of the final constructed Impulse gradient coil and 3D rendering of the three-layer winding pattern (primary, intermediate, and secondary layers of all axes combined). The gray sphere corresponds to the 20 cm region-of-linearity (ROL). Note that the constructed coil is longer (160 cm) than the wire extent (119 cm) to improve mechanical properties and simplify the MR system assembly.
  • Wireless Body Sensor Data Acquisition Platform for Motion Tracking
    Leanna Pancoast1,2, Douglas Brantner1,2, Roy Wiggins1,2, Jerzy Walczyk1,2, and Ryan Brown1,2
    1Center for Biomedical Imaging, NYU Grossman School of Medicine, New York City, NY, United States, 2Center for Advanced Imaging Innovation and Research, NYU Grossman School of Medicine, New York City, NY, United States
    Auxiliary sensors can help overcome the low temporal resolution in MRI, but often require cumbersome cabling. We describe an MRI-compatible sensor data acquisition platform for wireless motion tracking and correlate accelerometer respiratory motion data with optical measurements.
    Figure 4: In vivo respiratory data from the wireless accelerometer (blue) and video (orange) show excellent agreement during normal breathing (~0.25Hz, R>0.9) and lack of interference from MRI through closely matching waveforms and derived respiratory frequencies. The accelerometer and video had slightly lower correlation (R>0.86) during heavy breathing (~0.8Hz), possibly because the respiratory frequency approached the sensor’s current sampling frequency (2 Hz). Note the video (60 fps) was resampled to match the time basis of the accelerometer to display their correlation.
    Figure 2: Left panel: Photos alongside frequency maps show devices that are compatible or incompatible with local use because field disturbance is limited (a,b,c,e) or not limited (d,f) to voxels within 2cm of the surface, respectively. Right panel: Noise spectra using the ATWINC1500 show similar characteristics at baseline and with the shielded wireless device for 23.5±0.25MHz, but show interference at 24MHz (likely caused by a harmonic from the 2.4GHz WiFi signal).
  • Head Motion Tracking using an EEG-system and Retrospective Correction of High-Resolution T1-weighted MRI
    Malte Laustsen1,2,3, Mads Andersen4, Rong Xue3,5,6, Kristoffer H. Madsen2,7, and Lars G. Hanson1,2
    1Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs. Lyngby, Denmark, 2Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark, 3Sino-Danish Center, University of Chinese Academy of Sciences, Beijing, China, 4Philips Healthcare, Copenhagen, Denmark, 5State Key Laboratory of Brain and Cognitive Science, Beijing MR Center for Brain Research, Institute of Biophysics, Chinese Academy of Sciences, Beijing, China, 6Beijing Institute for Brain Disorders, Beijing, China, 7DTU Compute, Technical University of Denmark, Kgs. Lyngby, Denmark
    Motion tracking based on carbon wire loops show close similarity to interleaved navigators (MAD: [0.13,0.33,0.12]mm, [0.28,0.15,0.22]deg) yielding similar improvement to image sharpness (ΔAES: 12%) after retrospective correction of T1w 3D images, without sequence modification.
    Figure 3: Retrospective motion correction of a structural scan with instructed movement (Table 1: subject 2, trial 4) with moderate motion of 1-2mm (x,y,z) and 1-2deg (θ,ϕ,ψ). Tracking curves for CWLs and NAVs (red, blue respectively) show high degree of similarity (low mean absolute difference, MAD), with largest discrepancy (gray) in y-translations. A substantial increase to visual sharpness (e-l) is evident after retrospective correction using either tracking method. Both methods lead to an increase to average edge strength (AES) of 12.8/9.9% for CWLs and NAVs, respectively.
    Figure 1: Measured wire loop signal $$$\mathit{v}_i(t)$$$ for part of a navigator consisting of a weighted sum of columns of $$$\dot{\tilde{\pmb{\mathit{G}}}}(t)$$$, which resemble filtered time derivatives of gradient activity, and are found by sampling the sequence in a static phantom pre-scan with only one active gradient component $$$\mathit{g}_x(t)$$$, $$$\mathit{g}_y(t)$$$, $$$\mathit{g}_z(t)$$$, or active RF. Each wire loop ($$$i=[1,2,\ldots ,I]$$$) measures a unique weighted sum dependent on loop position, orientation, and geometry.
  • Beat Pilot Tone: Exploiting Preamplifier Intermodulation of UHF/SHF RF for Improved Motion Sensitivity over Pilot Tone Navigators
    Suma Anand1 and Michael Lustig1
    1Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, CA, United States
    Our motion sensing Beat Pilot Tone (BPT) requires no on-subject hardware and induces a beat frequency in the receiver via two tones (2.4GHz, 2.528GHz). BPT obtains improved sensitivity (20x) to motion compared to PT and does not impact image SNR.
    Figure 1: A schematic illustration of (a) Beat Pilot Tone (BPT) vs. (b) Pilot Tone (PT) acquisitions. PT has a 2.347m wavelength ($$$f_{PT} = 127.8MHz$$$), while BPT has a 12.5cm wavelength ($$$ f_T = 2.4GHz $$$); thus, there is much more interaction between BPT and the cm-scale motion of the body. c) BPT uses two frequencies, $$$f_T$$$ and $$$f_T+f_{BPT}$$$, where $$$f_{BPT}=f_{larmor} + f_{offset}$$$ (e.g., $$$f_{offset} = 100kHz$$$). A signal with a beat frequency $$$f_{BPT}$$$ is mixed in the coil preamplifier due to intermodulation; this signal is the BPT.
    Figure 4: a) A volunteer was asked to move their head up and down during the scan, a motion in the range of 2cm. b) BPT from the 3 most modulated coils are compared to the PT, with the coil arrangement in c). The arrows show up-down motion, which is clearly visible in BPT, but barely so in PT. d) The two most modulated body coils (17 and 20) for the abdominal scan. e) The BPT and PT magnitudes and phases were filtered with a moving average and scaled to match the bellows signal, which they match closely.
Back to Top
Digital Poster Session - Hot-Wired Systems: Gradients & Magnets
Engineering/Interventional/Safety
Wednesday, 19 May 2021 17:00 - 18:00
  • Gradient Characterization for High Performance Gradients
    Nastaren Abad1, Afis Ajala1, Yihe Hua1, and Tom K.F Foo1
    1General Electric Global Research, Niskayuna, NY, United States
    In this study a field monitoring system was used to characterize a head-only, high-performance gradient coil (MAGNUS),by determining the gradient transfer function. The modulation function indicated good correspondence for arbitrary waveforms and non-cartesian k-space trajectories.
    Figure 2. Magnitude (A), time-domain (B) and phase (C) of measured gradient IRF for the three axes, without pre-emphasis compensation. The arrows highlight channel specific known mechanical resonances at lower frequencies. No filtering has been used on this data. Notably, measurement noise in the IRF increases towards the higher frequencies, where the power of the input waveform is null.
    Figure 4. Representative images for two different readouts for a 2D spiral acquisition. For the shorter readout (A) both monitored and IRF demonstrate distinct sharpening of the grid lines and decreased blurring at the edges, compared to nominal reconstruction, also evident in the difference images (B). For the longer readout (C), error in correction for the field monitored measurement is a result of the probes dephasing during the long readout (D). IRF predicted trajectory correction can be used to compensate for this.
  • Measuring the Gradient Impulse Response Function (GIRF) for UTE Imaging on an MR-Linac
    Rosie Goodburn1, Tom Bruijnen2, Wajiha Bano1, Uwe Oelfke1, and Andreas Wetscherek1
    1Radiotherapy and Imaging, The Institute of Cancer Research, London, United Kingdom, 2Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands
    UTE images acquired on an MR Linac using a multi-echo stack-of-stars sequence showed higher image quality when offline reconstruction was based on k-space trajectories that were corrected with gradient delays and measured GIRFs.
    Figure 4: Phantom UTE images reconstructed with GIRF/delay correction a) off/off; b) off/on; c) on/off; d) on/on. Note artifacts: halo (solid arrows), dark boundaries (dotted arrows), and bright boundaries (dashed arrows).
    Figure 3: First-order GIRFs in frequency space measured for the MR-Linac at our institution. Represented in terms of magnitude (arbitrary units) and complex phase components for physical X, Y, and Z gradients of the system.
  • Phantom-based high-resolution measurement of the gradient system transfer function
    Hannah Scholten1, Manuel Stich2, and Herbert Köstler1
    1Department of Diagnostic and Interventional Radiology, University of Würzburg, Würzburg, Germany, 2Siemens Healthcare, Erlangen, Germany
    We present a new phantom-based measurement approach to determine the gradient system transfer function with good SNR and a high frequency resolution below 10 Hz.
    Figure 2: Magnitude of the linear self-term of the GSTF in x-direction, evaluated with three different datasets: All 18 readouts from all six measurements (top), only the first readout of each of the six measurements (middle) and only the first readout of the first measurement (bottom). For better visibility, the curves are plotted with an offset of 0.05 between each other.
    Figure 1: Schematic sequence timing of the first and second measurement. The durations and amplitudes of the RF pulses, gradients and readouts are not displayed to scale. For clarity, only three of the 16 triangles are depicted. Each readout had a duration of approximately 25 ms.
  • Calibration of Concomitant Field Compensation using Phase Contrast MRI
    Thomas K.F. Foo1, Louis Frigo2, Myung-Ho In3, Nastaren Abad1, Vincent B Ho4, and Matt A Bernstein3
    1GE Research, Niskayuna, NY, United States, 2GE Healthcare, Waukesha, WI, United States, 3Mayo Clinic, Rochester, MN, United States, 4Uniformed Services University of the Health Sciences, Bethesda, MD, United States
    A single-sided bipolar encoding, phase difference method was used to calculate the transverse gradient offsets for asymmetric gradient coils by fitting the measured phase values. This was used for pre-emphasis compensation for the zeroth and first order concomitant gradient fields.
    Figure 2: Phase error from a single-sided encoding experiment for (a) MAGNUS unit#2, and (b) the C3T system. The difference between the measured and the EM-derived nominal offset of 12 cm is shown.
    Figure 4: Phase difference images for the single-sided encoding experiment with g3 = 71 mT/m (a) before applying the zeroth order frequency correction (ROI phase = 4.2$$$\pm$$$0.1 radians), and (b) after correction (ROI phase = 0.1$$$\pm$$$0.1 radians). A zero phase is expected for the phase difference in the stationary phantom.
  • Direct Comparison of Gradient Modulation Transfer Functions and Acoustic Noise Spectra of the same MRI at High- (3T) and Lower-Field  (0.75T)
    Hannes Dillinger1, Sebastian Kozerke1, and Christian Guenthner1
    1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland
    On lower-field (0.75T), mechanical resonances are reduced leading to increased gradient fidelity and reduced audible noise. Maxima in the acoustic spectrum can directly be related to mechanical resonances in the GMTF.
    Figure 2: Comparison of Gradient Modulation Transfer Functions’ (GMTF) self terms for 0.75T and 3T configurations. Mechanical resonances are identified by local dips (anti-resonances) and peaks (resonances) in the GMTF that deviate from the ideal low-pass GMTF. Resonances are generally less pronounced for 0.75T compared to 3T. A reduction by 60% for 0.75T is found for the antiresonance in the z axis GMTF.
    Figure 4: Comparison of relative microphone amplitude spectra (RMAS) for the chirp sequence at 0.75T and 3T. For each gradient axis, RMAS were normalized to the maximal value. RMAS are reduced for 0.75T by up to 70% depending on the gradient axis compared to 3T. In addition, peaks in the RMAS correspond to the mechanical resonances identified in the GMTF.
  • PSF-based reconstruction for removal of artifacts caused by misalignment between a silent gradient insert and the body gradients
    Edwin Versteeg1, Dennis W.J. Klomp1, and Jeroen C.W. Siero1,2
    1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Spinoza centre for neuroimaging Amsterdam, Amsterdam, Netherlands
    Positioning of a silent gradient axis driven in synergy with the body gradients is prone to misalignment due to operator errors. Thismisalignment with the body-gradients can be estimated and corrected using a PSF based reconstruction, which significantly reduces ghosting artifacts.
    Figure 3: Reconstruction results without and with inclusion of the PSF-mapping (top row). Here, the zoomed FOV is shown (112 x 112 mm2 to show more detail in the phantom). The differences between the different reconstructions (bottom row). Note that we can clearly see the large effect of the 0th order PSF component on the ghosting. The horizontal lines in all images are caused by a stimulated echo.
    Figure 4: Reconstruction results for the 0th order correction on a large head phantom. Due to the large phantom size, large B1+-inhomogeneities are visible in the bottom of the phantom. In addition, some air bubble induced susceptibility artifacts and signal pile-up due to field non-linearities are visible in the bottom of the image
  • Eddy current correction for field probes mounted in a head coil
    Jennifer Nussbaum1, Maria Engel1, and Klaas Paul Pruessmann1
    1Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland
    When performing concurrent field monitoring with NMR field probes and a head coil, the field measurement can be corrupted by eddy currents on the head coil that are induced by the MR gradients. In this work, we provide a one-time calibration solution to correct for this issue.
    EPI reconstructed with concurrent field monitoring (without SENSE and without B0 correction). Left: reconstructed using corrected probe signal for the trajectory calculation, middle: reconstructed without probe signal correction, right: difference image. The ghosts disappear when the concurrently monitored probe phases, used to calculate the trajectory, are corrected.
    Phase correction of an x-gradient frequency swept pulse (20 mT/m). In the zoom it can be seen that the corrected signal (red) coincides with the signal that is measured in absence of the head coil (blue).
  • Design and Implementation of High Switching Frequency Gradient Power Amplifier Using eGaN Devices
    Soheil Taraghinia1, Volkan Acikel2, Reza Babaloo1,3, and Ergin Atalar1,3
    1UMRAM, Bilkent University, Ankara, Turkey, 2Aselsan A.S., Ankara, Turkey, 3Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey
    A single stage H-bridge gradient amplifier utilizing eGaN devices with 150 V/ 50 A voltage/ current ratings and 1 MHz switching frequency is implemented for an insert gradient array system. A single stage LC low pass filter with 50 kHz cut-off frequency is designed to attenuate the ripple current. 
    Figure 4. (a) Measured trapezoidal current waveform with 50 A flat top amplitude and 0.25 A/us slew-rate with filter. (b) Temperature of eGaN transistors to 45° C from room temperature for the current on (a) with 10% duty cycle and five minutes of operation. Zoomed version of the current is shown in Fig. 5.
    Figure 5. (a) Zoomed current (top) and voltage (bottom) applied to the coil in Fig. 4 without (a) and with (b) filter. Peak-to-peak ripple current is reduced more than twelve times by using the LC filter. Small resistance of the coil (about 200 mΩ) requires small voltage levels and PWM duty cycle.
  • Droop compensation of gradient current waveforms in gradient array systems
    Reza Babaloo1,2, Soheil Taraghinia2, and Ergin Atalar1,2
    1Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey, 2National Magnetic Resonance Research Center (UMRAM), Ankara, Turkey
    This work represents the gradient array system with a nonlinear model and utilizing the inverse of the acquired model to regulate the gradient currents and compensate for the droop in a feed-forward open-loop configuration.
    Fig.4. Experiment results. Comparison of the coils’ currents for the linear (uncompensated) and nonlinear (compensated) models. The reference inputs are trapezoid waveforms with 50A and 10A at the flat-top for ch-1 and ch-2, respectively. The rise time is the same for both channels (200μs).
    Fig.1. (a) The circuit diagram of the two-channels gradient array. (b) The hardware setup including control card (VC707 evaluation board), custom-built GPAs and two channels array coils.
  • Fast and quiet MPRAGE using a silent gradient axis at 7T – subject experience and qualitative image assessment
    Sarah M Jacobs1, Edwin Versteeg1, Leonie NC Visser2, Anja G van der Kolk1,3, Dennis WJ Klomp1, and Jeroen CW Siero1,4
    1Department of Radiology and Nuclear Medicine, University Medical Center Utrecht, Utrecht, Netherlands, 2Alzheimer Center Amsterdam, Department of Neurology, Amsterdam Neuroscience, Vrije Universiteit Amsterdam, Amsterdam UMC, Amsterdam, Netherlands, 3Department of Radiology, the Netherlands Cancer Institute, Amsterdam, Netherlands, 4Spinoza Centre for Neuroimaging Amsterdam, Amsterdam, Netherlands
    We have shown preliminary evidence that our silent gradient axis and readout module incorporated into a T1-weighted MPRAGE sequence is perceived more quiet and positive and delivers images of largely acceptable quality.
    Figure 2: (A) Means (SD) of reported sound level ratings directly after the scan (immediate) and after all scans (delayed) for the quiet compared to the standard scan; (B) Means (SD) of comfort level, overall experience and willingness to undergo scan again ratings for the quiet compared to the standard scan. Asterisk indicates statistical significance.
    Figure 3: T1-weighted axial images of the standard (top row) and the quiet (bottom row) scans of all five subjects.
  • Gradient induced electric field within a shoulder cut-out gradient coil built for head and neck imaging.
    Arjama A Halder1,2, Eric J Lessard1,2, William B Handler1, and Blaine A Chronik1,2
    1xMR Labs, Physics and Astronomy, London, ON, Canada, 2Medical Biophysics, Western University, London, ON, Canada
    Simulations to identify gradient induced electric field within a shoulder cut-out gradient coil from wire patterns are performed. The maximum electric field seen within the coil is less than 0.03 V/m at 1A.
    Figure 3. The simulated electric field on the yz plane at the isocenter within the x and y axes coils with 1A of current flowing through the coil, a. showing the contribution from the x axis gradient coil and b. showing the contribution from the y axis coil.
    Figure 2. The simulated electric field on the xy plane at the isocenter of both the x and y coil along a specific line with a. evaluated at the point shown in Figure 1a. and b. evaluated at the point shown in Figure 1b. with 1A of current flowing through the coil.
  • Approaching order of magnitude increase of gradient strength: Non-linear breast gradient coil for diffusion encoding
    Sebastian Littin1, Tristan A. Kuder2, Feng Jia1, Arthur Magill2, Philipp Amrein1, Frederik B. Laun3, Sebastian Bicklehaupt4, Mathias Davids5,6,7, Valerie Klein5,7, Mark E. Ladd2, and Maxim Zaitsev1,8
    1Department of Radiology, Medical Physics, University Freiburg, Faculty of Medicin, Freiburg, Germany, 2Medical Physics in Radiology, German Cancer Research Center, Heidelberg, Germany, 3Department of Radiology, MR Physics, University Medical Center Erlangen, Erlangen, Germany, 4Department of Radiology, University Medical Center Erlangen, Erlangen, Germany, 5Department of Radiology, A. A. Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 6Harvard Medical School, Boston, MA, United States, 7Computer Assisted Clinical Medicine, Heidelberg University, Medical Faculty Mannheim, Heidelberg, Germany, 8High Field MR Center Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria
    The aim of this project is to design and implement a non-linear single channel breast gradient coil for diffusion encoding. In vivo diffusion weighting with gradient strengths above 1 T/m is feasible in the female human breast.
    Fig 4: Diffusion-Weighted images (top row) measured in a dedicated phantom within the same slice for different amplitudes demonstrate the diffusion encoding capability. Measurements were performed with spin-echo diffusion deploying 10ms duration for each gradient pulse. ADC maps (bottom row) calculated using measured field maps to correct for inhomogeneities. Depicted are ADC maps acquired with same currents which results in different diffusion weighting closer to the bottom of the cup. Positions are given in the frame of reference of the scanner.
    Fig 5: Breast gradient coil during construction. Coil windings from square profile litz wire were manually wound onto a 3D-printed former (upper left). All components were combined into a housing (right) including water cooling, temperature sensors, and a second, non-functional cup. The coil was cast in epoxy (lower left) under vacuum.
  • Towards a More Power Efficient Two-Channel Biplanar Z-Gradient Coil Design Using Target Field Method
    Haile Baye Kassahun1, Sadeq S Alsharafi1, Ahmed M Badawi1, and AbdEl-Monem M El-Sharkawy1
    1Systems and Biomedical Engineering, Cairo University, Cairo, Egypt
    A   power-efficient two-channel biplanar Z-gradient coil is designed and compared to a conventional coil design. It was shown that the dissipated power can be lowered by more than 25% for the specific coil dimensions.
    Figure 2: Coil track location of the upper disc of the two-channel coil.
    Figure 5: Dissipated power of two-channel coil designs in comparison to the conventional coil design.
  • Design of Biplanar Matrix Z2 Nonlinear Gradient Coil with An Open Structure for O-Space imaging
    Congcong Liu1,2, Shi Su1, Ye Li1,2, Xin Liu1,2, Hairong Zheng1,2, Dong Liang1,2,3, and Haifeng Wang1,2
    1Paul C. Lauterbur Research Center for Biomedical Imaging, Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences, Shenzhen, China, 2Shenzhen College of Advanced Technology, University of Chinese Academy of Sciences, Shenzhen, China, 3Research Center for Medical AI, Shenzhen Institute of Advanced Technology,Chinese Academy of Sciences, Shenzhen, China
    A new type of openstructure nonlinear matrix  gradient coil is designed. The linear gradient coil in traditional MRI can be eliminated by adopting time-segment control of the matrix coil element, thereby making the whole system more concise and low-cost.
    Fig. 1 The 3D model diagram of the proposed biplanar matrix coil and the open-structure low-field MRI system. The green and red boxes indicate the specific installation location of the coil. The red and green arrows indicate the biplanar matrix coil and the low-field MRI system, respectively.
    Fig. 4 Contour and field of the proposed biplanar Matrix Z2 Nonlinear Gradient Coil (a) Matrix coil composed of coil elements with the same structure. The green and red boxes indicate the position of the coil, as seen as Fig. 1. (b) The model diagram of the different CP’s positions of the Z2 spherical harmonic field being moved at different times in (a). The energization sequence of the biplanar matrix coil of 4×4 element is from t8 to t9, and t7 to t10 follow the arrow direction.
  • MRI Hybrid Gradient Coil Equipped with a Programmable Z-Array and Conventional X- and Y- Elements
    Manouchehr Takrimi1 and Ergin Atalar1,2
    1UMRAM, Bilkent University, Ankara, Turkey, 2Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey
    A novel hybrid gradient coil consisting of two conventional actively-shielded x/y gradient coils and a programmable active-shield z-gradient array coil is introduced and simulated. The proposed gradient coil can dynamically provide different magnetic profiles for diverse applications.
    Figure 1: A quarter geometry of the proposed gradient system. Z-gradient array consists of 12 pairs of wire bundles, each with 10 copper wires of 2 mm diameter. The diameter/height of the main and shield array coils are 24/30 and 30/35 cm, respectively. Copper sheets of 2 mm thickness and 40 cm height for the main coil and square copper wires (2x2 mm2) are used for shield coils. The diameter of the main/shield coil for the x- and y-gradient coils are 20/26 and 22/28 cm, respectively. The aluminum shell of 90 cm diameter (the warm cryostat) is not shown here.
    Figure 4: The same gradient system with the second configuration for the z-gradient array profile. In this case, the FOV size is reduced to 60 mm and it is shifted up 30 mm along the z-axis. Z-gradient strength is doubled to be 250 mT/m within the shifted FOV region with less than 8.9% deviation from linearity. The contour lines indicate 1 mT separations. The residual eddy current is 0.007%.
  • Mechanical Tilt-Induced Gradient Fields for Low-Field Spokes-and-Hub Permanent Magnet MR Imagers
    Irene Kuang1, Jason Stockmann2,3, Elfar Adalsteinsson1,4, and Jacob White1
    1Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 2A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, United States
    We present a mechanically-tilted-magnet approach for generating gradient fields in a spokes-and-hub magnet topology for point-of-care and educational low-field permanent magnet MR systems without use of traditional MR gradient coils.
    Figure 4. (a) Diagram of spokes-and-hub magnet with no tilt (blue outline) and a 0.25° tilt (orange outline) about the x-axis, creating a y-axis gradient field; (b) Simulated receive signal spectrum with and without tilt; (c) Measured spin echo obtained from 8mm diameter test tube of water using dithered ultrasound RF pulse with 100 kHz bandwidth with 0° (blue) and 0.25° (orange) tilted magnet; (d) Measured received signal spectrum with (orange) and without (blue) tilt.
    Figure 5. Repeat of experiment in Figure 4, but with an 8mm diameter phantom containing two distinct tubes of water. In (a), tubes are side-by-side (along the y axis and in line with the tilt-induced gradient field); and in (b), tubes are stacked (along the z axis, orthogonal to the gradient field). Note the clear double hump in (a) showing spatial separation.
  • Systematic, Linear Algebra-based Dimensional Analysis of Gradient Inductance Scaling with Coil Radius
    Matt A Bernstein1 and Seung-Kyun Lee2
    1Radiology, Mayo Clinic, Rochester, MN, United States, 2Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of

    Applying systematic dimensional analysis, based linear algebraic methods, we derive the important scaling relation for MR gradient design: gradient coil inductance L scales as the square gradient efficiency (measured in T/m/A) and the fifth power of radius a.

  • Acceleration of magnetic field calculation of permanent magnet arrays for their optimizations
    Ting-Ou Liang1, Yi-Dan Chen2, Shao Ying Huang2, and Erping Li1
    1Zhejiang University, Hangzhou, China, 2Singapore University of Technology and Design, Singapore, Singapore
    Two current-model-based approaches are presented to accelerate the calculation of the magnetic field of permanent-magnet-array (PMA) for the popular body-part-dedicated portable MRI. It enables the acceleration of the optimization of a PMA.
    The illustration of (a) Nested for-loop, and (b) Approach-1 and (c) Approach-2 for acceleration of calculation of the magnetic fields of PMAs.
    The configuration of (a) Halbach (60 blocks) [6] and (b) IO ring-pair PMA (1200 blocks) [5]
  • Design and Development of a Hybrid Helmholtz Coil System for Production of Low Magnetic Field System (up to 7 mT)
    Yenal Gokpek1 and Ozkan Doganay1
    1Institute of Health Sciences, Ege University, IZMIR, Turkey
    The design, construction and numerical modelling of a 4-coil system were investigated for production of homogeneous magnetic field over a large volume of interest and the magnetic field created by the 4‑coil system was found to be B=2.5182±0.0035 mT.
    Theoretical calculations in 2D consisting of four coils (a) and comparison of magnetic field values with experimental results from A to B (b).
    3D representation (a) and measurement points (b) of constructed system
  • D-T2 Distribution Obtained Using CPMG-only Sequence Compared with Traditional SE-CPMG Sequence on the Single-sided NMR Device
    Ziyi Pan1, Jieying Zhang1, Hai Luo2, Weiqian Wang2, Xiao Chen2, Ziyue Wu2, and Hua Guo1
    1Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China, 2Marvel Stone Healthcare, Wuxi, Jiangsu, China
    The newly proposed CPMG-only sequence without diffusion editing can achieve comparable D-T2 distributions for water/oil separation in single-sided NMR systems, compared with the traditional diffusion editing SE-CPMG sequence.
    Fig. 5 First column: the unimodal distribution of the 0.5mmol/L MnCl2 solution. Diffusion and relaxation coefficients measured by CPMG-only are identical to the results of SE-CPMG. Second column: D-T2 distribution of the peanut oil. Two different fat components are demonstrated. However the distributions, especially for the T2 values are different. Third column: D-T2 distribution for the combination of 0.5mmol/L MnCl2 solution and peanut oil. Despite the differences in the D-T2 distributions, both (c) and (f) show clear separation of fat and water components (three peaks).
    Fig. 2 Details of the single-sided NMR experimental device. (a) The arched magnet with the RF coil in the magnet center. The geometry is designed to position the device at the waist abdomen for liver detection. (b) Two different plastic bottles are placed inside the magnet, filled with peanut oil and 0.5 mmol/L MnCl2 solution, respectively. The experiment is designed to explore the sequences for separate characterization of oil and water simultaneously.
Back to Top
Digital Poster Session - New Shim Systems & Other Advanced MR Engineering
Engineering/Interventional/Safety
Wednesday, 19 May 2021 17:00 - 18:00
  • In-vivo B0 shimming of the liver using a local array of shim coils in combination with 2nd order spherical harmonics at 7T
    Lieke van den Wildenberg1, Quincy van Houtum1, Wybe van der Kemp1, Catalina Arteaga de Castro1, Alex Bhogal1, Paul Chang2, Sahar Nassirpour2, and Dennis Klomp1
    1Radiology Department, UMC Utrecht, Utrecht, Netherlands, 2MR Shim GmbH, Reutlingen, Germany
    The B0 field inhomogeneity in the liver at 7T can be reduced using a local shim coil array combined with 2nd order spherical harmonics B0 shimming correction methods which can potentially lead to an inhomogeneity reduction of 44%.
    Figure 5: In-vivo B0 field mapping of the liver in a healthy volunteer during free breathing with two different B0 shimming methods: 1) scanner shim (2nd order) and 2) array of shim coils in combination with the conventional second scanner shim. A: anterior, P: posterior, R: right, L: left, H: head, F: feet, and std: standard deviation.
    Figure 1: A) The array of shim coils B) used around the body array in the 7T MR system
  • “RF transparent” local B0 shim coil
    Xinqiang Yan1,2
    1Vanderbilt University Institute of Imaging Science, Nashville, TN, United States, 2Department of Radiology and Radilogical Science, Vanderbilt University Medical Center, Nashville, TN, United States
    “RF transparent” local B0 shim coil have minimal crosstalk with RF coils without using massive bulky RF chokes.
    Figure 1 Circuit diagrams of RF-only coil, RF coil + DC coil, and RF coil + “RF transparent” DC coil. Note that in practice, feed wires of DC coils are typically twisted to cancel the Lorentz force. For simplicity, straight feed wires were shown here. For Bazooka-feedline [5] few-turn “RF transparent” DC coil, the shield of twisted wires (could be one shield for each wire or one shield for both wires) was shorted at the position that is 1/4 wavelength away from the DC coil, forming open circuits and block RF signal going through.
    Figure 2 Circuit diagrams of many-turn “RF transparent” DC coil using float trap design. For many-turn DC coil, it may resonate close to the RF frequency even the feed port is broken using the bazooka feedline method (A). In this case, it still can be seen as a coupled lossy resonator and leads to SNR loss. The float trap [6] is a concentric resonator that inductively couples and traps the RF signal going through it. It is expected to break every turn of the DC coil that goes through the trap and thus avoid unwanted resonate modes.
  • Shim Coils Tailored for Correction of B0 Inhomogeneity in the Human Brain (SCOTCH) at Ultra High Field
    Bruno Pinho-Meneses1, Jason Stockmann2,3, Edouard Chazel1, Paul-François Gapais1, Eric Giacomini1, Franck Mauconduit1, Alexandre Vignaud1, Michel Luong4, and Alexis Amadon1
    1Université Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France, 2Athinoula A. Martinos Center for Biomedical Imaging, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Université Paris-Saclay, CEA, Institut de Recherche sur les Lois Fondamentales de l'Univers, Gif-sur-Yvette, France
    A 2-layer 36-channel optimized Multi-Coil Array for Bshimming of the human brain was designed. A prototype was built and in-vivo GRE and EPI acquisitions were performed, confirming high inhomogeneity reduction and artifact mitigation after shimming with the optimized MCA.
    Figure 1: a) From left to right: Ideal channel disposition and geometry for first and second layers, both layers resulting from the SVD-based MCA optimization; and ideal 2-layer 36-channel system. b) The two prototype layers and the assembled shimming system with RF housing in the interior of the shim layers. c) Simulation models of generic 24-ch (left) and 48-ch (right) M-MCA used for comparison. Circular coil radii are 4.5-cm and 3.5-cm, respectively.
    Figure 5: a) Brain mask overlay (green) considered for optimal current calculations for global shimming and selected slices (blue) presented in b) rs-EPI acquisitions. Baseline, global and slicewise (dynamic) 2-layer 36-channel SCOTCH shimming are shown. Gradient Echo rS-EPI sequence parameters were 2-mm in-plane resolution, fifteen 3-mm-thick slices, TE/TR=44/84000ms. Yellow arrows highlighting zones with significant improvement.
  • Designing a high-density combined RF/B0 shim coil for imaging the brain at 7T
    Paul Chang1, Sahar Nassirpour1, Kaizad Rustomji2, Elodie Georget2, Ingmar Voogt3, Aidin Haghnejad3, Evita Wiegers4, Jannie Wijnen4, and Dennis Klomp4
    1MR Shim GmbH, Reutlingen, Germany, 2Multiwave Imaging, Marseille, France, 3WaveTronica, Utrecht, Netherlands, 4UMC Utrecht, Utrecht, Netherlands
    To minimize RF interference, the Rx loops should be larger than shim coils with an optimal gap of 4cm. An optimized distribution of the shim loops tailored to the brain anatomy achieved significant B0 shimming improvement (43% compared to 2nd order) with at max 1.2dB loss in MR sensitivity.
    Figure 1 Three RF loops (84mm diameter each) were arranged in the configuration shown on the left. The four different settings for the shim coildiameter/positioning used in the numerical simulation are also shown. The reference S-parameter curves (in the absence of the shim coils) are shown on theright.
    Figure 4 Optimal arrangement of 24 shim coils on a cylindrical holder (units in [m]).
  • Shielded coaxial cable coils for transmit, receive and B0 shimming in a 7T neck array
    Vincent O. Boer1, Jan Ole Pedersen2, Hørður Andreasen3, Sadri Güler1,4, Vitaliy Zhurbenko3, Jason Stockmann5, Irena Zivkovic6, and Esben Thade Petersen1,4
    1Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital, Hvidovre, Denmark, 2Philips Healthcare, Copenhagen, Denmark, 3Department of Electrical Engineering, Technical University of Denmark, Kongens Lyngby, Denmark, 4Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kongens Lyngby, Denmark, 5Athinoula A Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 6Electrical Engineering department, Technical University of Eindhoven, Eindhoven, Netherlands
    Shielded coaxial RF coils show very low coupling between elements, and are ideal for building transmit-receive arrays. In this work we extended a 6 element transceive array with B0 shim capabilities. 
    Figure 3 – The 6 channel transceive array facilitates good coverage over the cervical spine (a), although there is a large B0 background field present (b). The transceive array with B0 shim capabilities can provide additional B0 fields (c, sagittal). The six channels show non-overlapping RF fields (d) as well as complementary B0 fields (e) even for 1A current.
    Figure 1 – Extension of RF coils with DC circuits. In a traditional loop coil (a) the conductor is broken up with capacitors whereas (b) shielded coaxial coils (SSC) do not require this. Adding a DC path (red) therefore typically requires the addition of several RF chokes in a traditional coil (c) compared to only two on an SSC coil (d). Addition of RF chokes is only required on the feeding points (e) and not limited additional space is needed on the coil array (f). Note the coil array shows virtually no coupling between the elements despite lack of overlap between elements (data not shown).
  • Improving MRI Near Metal with Local B0 Shimming using a Unified Shim-RF Coil (UNIC): First Case Study, Hip Prosthesis in Phantom.
    Fardad Michael Serry1, Junzhou Chen1,2, Anthony G Christodoulou1,2, Yuheng Huang1,2, Fei Han3, Won Bae4,5, Christine Chung4,5, Richard Handlin1, John Stager1, Matthew Dausch1, Yubin Cai1, Yujie Shan1, Yucen Liu1, Yibin Xie1, Xiaoming Bi3, Rohan Dharmakumar1,2, Zhaoyang Fan1,6, Debiao Li1,2, Hsin-Jung Yang1, and Hui Han1
    1Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, CA, United States, 2Department of Bioengineering, UCLA, Los Angeles, CA, United States, 3Siemens Medical Solutions USA, Inc., Los Angeles, CA, United States, 4University of California, San Diego, San Diego, CA, United States, 5VA Medical Center, San Diego, CA, United States, 6University of Southern California Department of Radiology, Los Angeles, CA, United States
    Novel unified coil (UNIC) array 3D B0 shim reduced metal-induced signal void artifact, increasing the visible area around a hip prosthesis in phantom by up to 50% in some slices. The sequence-agnostic technology was tested with a 3D GRE, generally more susceptible to metal artifact than SE.
    Figure 3. Phantom axial GRE images of cross sections at the femoral head (a,b) and near the femoral insertion end (c,d) of the metal hip prosthesis at TE=4.80ms before (left) and after (right) adding UNIC shim to scanner shim. Signal void is stronger at the longer echo time due to additional dephasing between the two echos. UNIC shimming reduced signal void artifact area by 9% (b versus a) and 9% (d versus c), and extended the visible area closer to the metal, increasing it by 50% (b versus a); and 8 % (d versus c); see text for ROI definition. Fiducial markers are visible in a and b.
    Figure 5. B0 field homogeneity improvement with UNIC shimming. Histograms of Larmor frequency deviation from the scanner center frequency (~123MHz), of the masked volume of pixels in the 3D slab scanned before and after adding UNIC shim to scanner shim (a), and B0 field magnitude maps (b) for the slice in Figures 3c,d before (b. left) and after (b. right) adding UNIC shim to scanner shim.
  • Shimming-Toolbox: An open-source software package for performing realtime B0 shimming experiments
    Alexandre D'Astous1, Ryan Topfer1, Gaspard Cereza1, Eva Alonso-Ortiz1, Lincoln Craven-Brightman2, Jason Stockmann2,3, and Julien Cohen-Adad1,4
    1NeuroPoly Lab, Institute of Biomedical Engineering, Ecole Polytechnique, Montreal, QC, Canada, 2Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Functional Neuroimaging Unit, Centre de recherche de l'Institut universitaire de gériatrie de Montréal, Montreal, QC, Canada
    Shimming Toolbox is an open-source software package that aims to provide an easy-to-use, streamlined platform to perform static, dynamic and realtime $$$B_0$$$ shimming experiments.
    Figure 2: Realtime z-shimming overview. Training session: Step 1: The pressure and fieldmaps are acquired. Step 2: Shimming Toolbox’s optimizer computes a time series of Gz maps. A linear regression is computed between the field gradient and the pressure. Imaging session: Step 3: A dummy GRE scan is run in order to obtain the target image. Step 4: The $$$G_z$$$ maps are resampled onto the target GRE images and $$$G_{z, static}$$$ and c for each slice are sent to the scanner. Step 5: Run the custom realtime z-shimming GRE sequence.
    Figure 1: Pneumatic phantom with a long stick, which terminates by a slightly ferromagnetic object that oscillates back and forth creating a time-varying magnetic field.
  • Magnetic fields produced by simple coils inside finite-length cylindrical passive shields with end-caps
    Richard W. Bowtell1, Michael Packer 2, Peter Hobson 2, James Leggett1, Niall Holmes1, Paul Glover 1, Matthew Brookes1, and Mark Fromhold2
    1Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom, 2School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom
    We show how to analytically calculate the fields that are produced by simple coils inside a finite-length, cylindrical mu-metal shield with end-caps.
    Figure 1 (a) Schematic diagram cylindrical shield with end-caps of radius, b, and length, 2Ls, and coil of radius, a. Illustrative coil wirepath and direction of current flow indicated. (b) Multiple reflections of the coil due to the end-caps (dotted lines indicate reflected versions of the wirepath).
    Figure 2 (a) Experimental set-up: fields from loop and saddle coils measured using a fluxgate magnetometer inside a 30 cm diameter 1 m long cylinder formed from 1.5 mm thick mu-metal. (b) Helmholtz-type coil formed from two loops of radius a, separated by a distance 2d, carrying current in the same sense.
  • Hardware developed for phase and frequency locking of interleaved MRI and DMI studies
    Terence W. Nixon1, Yanning Liu1, Henk M. DeFeyter1, Scott McIntyre1, and Robin A. de Graaf1
    1Yale University, New Haven, CT, United States
    Hardware was developed to allow 2H DMI data to be interleaved into a 1H MRI experiment thereby reducing scan time. 2H signal was upconverted to and acquired as 1H data using an RF mixer with a frequency and phased locked local oscillator.

    Figure 3 – Phase variation during 2H MRS interleaved with a 7-slice spin-echo MRI. The phantom consisted of three bottles containing ~0.1% D2O and various amounts of DMSO-D6 (~0.02 – 0.05%). A small (0.5 mL) phantom containing pure D2O served as an external off-resonance reference signal. (A) Phase variation in the absence of a phase lock. (B) In addition to the strong linear phase roll there is also a smaller sequence-dependent phase modulation. (C) In the presence of a phase lock, each spectrum has an identical phase. The 1D spectra shown in (A) and (C) are extracted from repetition 200.

    Figure 1 – (A) Four channel interleaved mixer. Inputs for the mixer are from their respective preamplifiers. The 1H signal is applied to one arm of the RF switch. The 2H signal is sent to an RF mixer and upconverted to 170MHz by mixing with the 144MHz LO. The upconverted signal is filtered to remove the unwanted sideband, amplified and sent to the RF switch. The PPG controls when each nucleus is sent to the scanner’s 1H receiver. (B) Each channel is isolated from outside interference and each other by using a dedicated RF shielded enclosure built onto the board.

  • Simultaneous transmit/receive for Bloch-Siegert encoding: a feasibility study
    Baosong Wu1, Sajad Hosseinnezhadian2, Yonghyun Ha2, Kartiga Selvaganesan2, Charles Rogers III2, Kasey Hancock2, Gigi Galiana2, and R. Todd Constable2
    1Department of Radiology and Biomedical Imaging, Yale School of Medicine, NEW HAVEN, CT, United States, 2Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, CT, United States
    The feasibility of simultaneous transmit/receive was studied at 1MHz. This study demonstrates survival of the NMR signal  in nanovolt level in strong radio frequency interference.
    Figure 1. (a) The open MRI system. (b) The diagram of simultaneous transmit and receive. The phantom is above the coils. The coil on the right side is for the use of picking up NMR signal at 1MHz, and the coil on the right side is irradiating offset field.
    Figure 5. Experiments for simultaneous transmit and receive. (a) The diagram of a multiple spin echo sequence with continuous offset irradiation in the front five acquisition window. (b) Experimental data acquired with the sequence above with echo spacing (TE) 1ms. It shows three echo trains under different offset-field irradiations.
  • RF sinc pulse distortion compensation using multiple square pulses at 1 MHz.
    Yonghyun Ha1, Kartiga Selvaganesan1, Baosong Wu1, Kasey Hancock1, Charles Rogers III1, Sajad Hosseinnezhadian1, Gigi Galiana1, and R. Todd Constable1
    1Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, CT, United States
    Compensation pulses are shown that were optimized using a series of square pulses. The SNRs of the echo signal acquired using compensated pulse was compared with those of signal obtained with uncompensated pulses and showed significant improvements of 51.5%.
    Figure 4. Measured actual pulses (vertical dashed lines indicate the input pulse ends.) of the sinc pulses (top row) and compensated sinc pulses calculated with NS values of 13 (second row), 25 (third row), 37 (bottom row) and bandwidths of 30 kHz (left column), 90 kHz (middle column), and 150 kHz (right column).
    Figure 2. (a) Schematic diagram of an open, table-top, MRI system with photos of (b) the RF solenoid coil and (c) the magnet. (d) A spin-echo sequence with pre-polarization.
  • Automatic 3D B1 field mapping using 3D printer and digital oscilloscope for gradient-free MRI system
    Yonghyun Ha1, Kartiga Selvaganesan1, Sajad Hosseinnezhadian1, Baosong Wu1, Kasey Hancock1, Gigi Galiana1, and R. Todd Constable1
    1Department of Radiology and Biomedical Imaging, Yale School of Medicine, New Haven, CT, United States
    A method to automatically measure the B1 field is introduced.3-axis probe were used to measure the B1 field in three orthogonal axes simultaneously. The position of the probe was controlled by 3D printer, and the peak-to-peak voltage of the signal was automatically saved using oscilloscope.
    Figure 1. (a) Scheme and (b) photo of the 3-axis probe. The probe consists with three perpendicular square loops.
    Figure 2. Experimental set up.
  • Inhomogeneity and ramping effects in field-cycled quantitative molecular MRI
    Matthew A. McCready1, William B. Handler1, Francisco Martinez2,3, Timothy J. Scholl2,3, and Blaine A. Chronik1,2
    1Physics and Astronomy, Western University, London, ON, Canada, 2Medical Biophysics, Western University, London, ON, Canada, 3Robarts Research Institute, Western University, London, ON, Canada
    The dreMR method assumes field strength is homogenous and coils reach field instantly. We have shown that these assumptions result in errors within dreMR images, which can be solved by use of improved homogeneity coils and higher slew rates.
    Figure 2. (LEFT) Percent difference in dreMR image of VivoTrax, in uniform concentration of 160 μM, using field of previously constructed coil. Percent difference is taken from center point where field is at desired value. (RIGHT) Same simulation using field from new design method coil. Colour axis chosen to match the left.
    Figure 1. (LEFT) Simulated dreMR image of 4 cylinders of VivoTrax, in concentrations of 20, 80, 120, and 160 μM, using field of previously constructed coil. Note change in signal across cylinders. Sequence uses 300mT field shift in a 0.5T system. (RIGHT) Same simulation using field from new design method coil. Signal does not noticeably change across cylinders. Data points are given in both figures at locations with no contrast agent.
  • Very low field MRI for brain imaging
    Samson Lecurieux Lafayette1 and Claude Fermon1
    1SPEC - CEA Saclay - Université Paris Saclay, Saclay, France
    We have developed a MRI device working at 10mT for brain imaging. Parallel acquisition and rapid sequences have been implemented in order to speed up the acquisition time. Strengths and limitations compared to high field are given.
    Figure 3: The very low field MRI in the magnetically shielded room.
    Figure 1: The eight channels sensor. Each channel is matched and tuned, and decoupled from emission radiofrequency.
  • Comparison of SNR between a low-field (0.26T) Tabletop-MRI and a clinical high-field (3T) scanner
    Robert Kowal1, Enrico Pannicke2, Marcus Prier1, Ralf Vick2, Georg Rose3, and Oliver Speck1
    1Department of Biomedical Magnetic Resonance, Otto von Guericke University, Magdeburg, Germany, 2Chair of Electromagnetic Compatibility, Otto von Guericke University, Magdeburg, Germany, 3Chair in Healthcare Telematics and Medical Engineering, Otto von Guericke University, Magdeburg, Germany
    A low-field (0.26T) Tabletop-MR-system was compared to a high-field (3T) clinical scanner by evaluating the SNR of simple FID time domain signals. Direct comparison of the SNR led to a measured relative performance of 11.3% for the Tabletop before compensation for experimental conditions.
    Measurement setup on the Tabletop-system. Left side: OCRA-console for controlling the system, supplying the components and connecting to the computer. Right side: Opened system with the sample in the carrier between the neodymium magnets with connected signal and shimming lines.
    Measurement setup on the Skyra-system. The test tube sample was positioned with the coil perpendicular to the static magnetic field fixed by wooden planks on the patient table. The connection to the MR system was established via the T/R-switch and TIM-connector. Shown on the bottom is an image taken with the patient camera.
  • Initial Experience of Body Imaging at 5T
    Zhenhua Shen1, Xuchen Zhu1, Shihong Han1, Fuyi Fang1, Wei Luo1, Shao Che1, Zidong Wei1, Jinguang Zong1, Yongquan Ye2, Bo Li1, Shuheng Zhang1, Anthony Vu2, Weiguo Zhang2, and Guobin Li1
    1United Imaging Healthcare, Shanghai, China, 2UIH America, Inc., Houston, TX, United States
    With multi-channel RF parallel transmission hardware architecture and static RF shimming techniques, the uniformity of the RF transmission field is shown to be well controlled for imaging quality guarantee. Preliminary results show great promise for body imaging at 5T.
    Figure 2 Abdominal imaging at 5T. a) and c) B1+ sensitivity map and T2W FSE image with CP mode excitation. b) and d) B1+sensitivity map and T2W FSE image after static RF shimming optimization. The CV values were measured for bothcases.The red arrows show the most significant improvement area.
    Figure 3 Pelvic imaging at 5T. a) and c) B1+ sensitivity map and T2W FSE image with CP mode excitation. b) and d) B1+sensitivity map and T2W FSE image after static RF shimming optimization. The CV values were measured for bothcases.The red arrows show the most significant improvement area.
  • A sample temperature control system for post mortem MRI
    Sebastian Walter Rieger1, Karla Miller1, Peter Jezzard1, and Wenchuan Wu1
    1Wellcome Centre for Integrative Neuroimaging, University of Oxford, Oxford, United Kingdom
    In post mortem MRI, RF absorption can cause significant heating in the sample. This can interfere with scanning and accelerate decomposition of unfixed samples. Here, a temperature control system is presented which enables prolonged scanning at a stable temperature while preserving tissue.
    Cooling system schematic. A: Pre-cooling / setup configuration - The chiller is operating and pre-cooling the bulk of the fluid to the desired temperature, while the cooling pad and temperature sensor(s) are being used for sample preparation. B: Normal operation configuration - The temperature probe(s) are affixed to the sample, and the cooling pad is wrapped around it. A layer of cloth around the outside provides thermal insulation to prevent excessive condensation, and the pad is connected to the chiller.
    Example T2- and diffusion weighted images. Top: T2 weighted images were acquired at 0.4mm isotropic resolution using a 3D-SPACE sequence with FOV=150x150x96mm3, TR/TE=2300/586ms, 6 repetitions. Bottom: averaged diffusion weighting images acquired at 0.8mm isotropic resolution using a spin-echo diffusion weighted readout-segmented EPI sequence with FOV=150x150x104mm3, TR/TE=10.8s/113ms, multiband factor 2, GRAPPA factor 2, b=9000 s/mm2.
  • Compact MRI bioreactor for real-time monitoring 3D printed tissue-engineered constructs.
    Jean-Lynce GNANAGO1,2,3,4,5,6, Tony GERGES1,2,3,4,5,6, Laura Chastagnier1,2,4,6,7,8,9,10, Emma Petiot2,4,6,7,8,9,10, Vincent SEMET1,2,4,5,6, Philippe Lombard1,2,3,4,5,6, Christophe Marquette1,2,4,6,7,8,9,10, Michel Cabrera1,2,3,4,5,6, and Simon Auguste Lambert1,2,3,4,5,6
    1Université Claude Bernard Lyon 1, VILLEURBANNE, France, 2INSA LYON, VILLEURBANNE, France, 3Ecole Centrale Lyon, Ecully, France, 4CNRS, VILLEURBANNE, France, 5AMPERE UMR 5005, VILLEURBANNE, France, 6Université de Lyon, VILLEURBANNE, France, 73d.FAB, VILLEURBANNE, France, 8CPE Lyon, VILLEURBANNE, France, 9ICBMS, VILLEURBANNE, France, 10UMR 5246, VILLEURBANNE, France
    An MRI bioreactor is built using novel 3D printing techniques and plastronics. The integration of an MRI probe function within the bioreactor opens up possibilities of real time monitoring in the tissue engineering field.
    Figure 1: Design and conception of the bioreactor. (A) 3D model of the bioreactor and its support. (B) Detailed view of the bottom of the 3D coil circuit and its integration. (C) Detailed view of the top of the 3D coil circuit and its integration
    Figure 3: (A) Picture of the bioprinted tissue after construction. (B) 2D TURBO RARE image of the tissue using the bioreactor acquired on a 7T MRI scanner with a 75 µm in plane resolution.
  • Characterization of 3AM diffusion MRI phantoms via microscopy, and phase-contrast micro-CT
    Farah Mushtaha1,2, Tristan K. Kuehn1,3, Omar El-Deeb4, Seyed A. Rohani3, Luke W. Helpard3, Hanif Ladak2,3,5, Amanda Moehring6, Corey A. Baron1,2,3,7, and Ali R. Khan1,2,3,7
    1Robarts Research Institute, London, ON, Canada, 2Medical Biophysics, Schulich School of Medicine and Dentistry, Western University, London, ON, Canada, 3School of Biomedical Engineering, Western University, London, ON, Canada, 4Neuroscience, Western University, London, ON, Canada, 5Department of Electrical and Computer Engineering, Western University, London, ON, Canada, 6Biology, Western University, London, ON, Canada, 7The Brain and Mind Institute, Western University, London, ON, Canada
    Pore diameters of the 3AM phantoms measured using  microscopy and micro-CT were in agreement with a mean pore diameter of ~ 8 μm, and fits of both kurtosis and ball and stick models were in agreement between simulation and acquired dMRI data.
    Figure 3. Synchrotron micro-CT scan data at two zoom levels, transformed to align lines of material with the viewing planes. Upper row: View of the entire scan ROI. Lower row: Detailed view of a short length of five stacked lines of material. The direction of print-head motion was left-right in the left- and right-most columns, and perpendicular to the image in the center column.
    Figure 1. a) Confocal microscopy z-stack image of a stained phantom slide. Elastomeric matrix (red) and pores (black) are visible. Each white outline indicates an individual line of material. b) 2D projection of a 3D microscopy volume acquired with confocal microscopy. Outlined in yellow are larger pores caused by the printing pattern of the phantom. In both a and b, the image plane is perpendicular to the long axis of the pores.
  • A phantom system for evaluating the effect of lipid and iron composition on qMRI parameters
    Rona Shaharabani1, Shir Filo1, Oshrat Shtangel1, and Aviv Mezer1
    1The Edmond and Lily Safra Center for Brain Science, The Hebrew University of Jerusalem, Jerusalem, Israel, Jerusalem, Israel
    We describe a phantom system composed of lipids and iron for the assessment of their unique contribution to quantitative MRI parameters. We found that both changes in lipid concentrations (lipid-water fraction) and iron concentrations affected R2*, for all lipid types.
    Fig. 1: Schematic representation of the MLV preparation method. (a) Soy PC, alone or as a mixture, was dissolved in chloroform (b) put in a vacuum rotary evaporator to create a uniform lipid film. (c) Ammonium bicarbonate was added to dissolve the lipid film. (d) The solution was then lyophilized and (e) rehydrated with a desired iron ion buffer to create the iron ion encapsulated MLV. (f) The samples were imaged in a clinical scanner.
    Fig. 3: qMRI relaxation rate R2* with increasing concentration of iron ion Fe2+. R2* is affected by the lipid-water fraction and iron concentration. R2* values increase with increasing concentration of Fe2+ for all lipid types. Each color represents a different lipid-water fraction MLV phantom. (a) Lipid MLV phantom PC-Cholesterol with Fe2+. (b) Lipid MLV phantom PC with Fe2+. (c) Lipid MLV phantom PC-SM with Fe2+.