Novel Acquisitions
Acq/Recon/Analysis Monday, 17 May 2021
Oral

Oral Session - Novel Acquisitions
Acq/Recon/Analysis
Monday, 17 May 2021 12:00 - 14:00
  • Magnetic Resonance Coherence Pathway Unraveling
    Nikolai Mickevicius1 and Eric Paulson2
    1Medical College of Wisconsin, Milwaukee, WI, United States, 2Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI, United States
    This is a framework to simultaneously acquire and reconstruct spin echo images arising from different coherence pathways.
    (a) Acquisition timing diagram between (top) applied RF pulses (middle) crusher gradients $$$G_C$$$, and the crusher gradient-induced phase for the five pathways relevant to this 4-pulse experiment. (b) The RF phase cycling scheme used in this work modulates the phase of $$$E_3$$$ and $$$E_5$$$ by $$$\pi$$$ between k-space segments, but does not modulate the phase of $$$E_2$$$ and $$$E_4$$$. Here, $$$F^H$$$ represents the adjoint Fourier transform operation, and $$$\Phi^H_{E_e}$$$ adds the complex conjugate of the RF-induced phase to the $$$e^{th}$$$ pathway.
    In vivo MR-CPU results from the brain of a healthy volunteer. Multi-coherence pathway images were reconstructed successfully with both Cartesian and radial k-space trajectories. Each image is windowed separately for display. The arrow indicates the presence of barely visible streaking artifacts in the radial image that are not present in the Cartesian image.
  • VUDU: motion-robust, distortion-free multi-shot EPI
    Jaejin Cho1,2, Avery JL Berman1,2, Borjan Gagoski2,3, Congyu Liao4, Jason Stockmann1,2, Jonathan R Polimeni1,2, and Berkin Bilgic1,2
    1Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children's Hospital, Boston, MA, United States, 4Radiological Sciences Laboratory, Stanford University, Palo Alto, CA, United States
    We propose VUDU for motion-robust, distortion-free multi-shot EPI acquisition for GRE and DWI. VUDU acquires two EPI shots for each slice in quick succession and alternates the phase encoding polarity to encode B0 information.
    Fig1. Proposed VUDU sequence and reconstruction for GRE contrast. Using FLEET ordering, two shots of EPI are acquired successively for each slice, before proceeding to the next slice. Phase encoding polarity is alternated between the shots. Variable flip angle (vfa) excitation maximizes the signal. A fieldmap estimated from interim blip-up and -down reconstructions is incorporated into parallel imaging model to jointly reconstruct two shots with low-rank regularization to eliminate distortion. In SE/diffusion variant, 180° refocusing pulses succeed each excitation pulse.

    Fig2. shows individual SENSE reconstructions for each shot of vfa-FLEET acquisition with GRE contrast. At TRslice=80 msec, it took vfa-FLEET 160 msec to encode each slice, and 5.1 sec for the entire volume.

    Horizontal lines help compare the geometric fidelity of distortion-free VUDU reconstruction relative to each shot. Yellow and blue arrows point to distortion artifacts in the sagittal views of SENSE reconstructions, which were eliminated in the VUDU result.

  • Maxwell Compensation for Spiral Turbo-Spin-Echo Imaging
    John P. Mugler1, Adrienne E. Campbell-Washburn2, Rajiv Ramasawmy2, Josef Pfeuffer3, and Craig H. Meyer1
    1University of Virginia, Charlottesville, VA, United States, 2Cardiovascular Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, MD, United States, 3Siemens Healthcare GmbH, Erlangen, Germany
    An interleaved-spiral T2-weighted 2D-TSE sequence was developed that incorporates gradient waveform modifications to achieve compensation of self-squared Maxwell terms at both the echoes and over echo spacings.  This approach provided substantial improvement in image quality at 0.55T.
    Figure 4. Axial spiral-in-out images of a healthy volunteer without (top row) and with (bottom row) Maxwell compensation, acquired at 0.55T. Substantial artifacts seen without compensation are suppressed with compensation. Parameters included: TR/TEeff: 3200/~130 ms, FOV: 250 mm, slice thickness: 5 mm, slices: 11, matrix: 320, interleaves: 90, trajectory: self-retraced 7.2-ms spiral in-out, averages: 4 with T2-decay compensation (8 total), Maxwell compensation: 4.4 ms per echo spacing, acquisition time: 2.6 min.
    Figure 2. Self-retraced spiral in-out waveform and associated Maxwell integrals illustrating compensation achieved by adapting trapezoidal waveforms (red arrow) and adding bipolar waveforms (green arrows). Without compensation, Maxwell integrals at times A and C are not equal and that at B is not zero, whereas values at A and C are equal and that at B is zero with compensation. A small bipolar is sometimes needed because gradient duration is a discrete variable, and so the exact change in Maxwell integral required is not achieved by only modifying the trapezoidal waveform.
  • Low-Angle Combined-Echo (LACE) Imaging in Highly Inhomogeneous B0 Magnetic Fields
    Sebastian Theilenberg1, Chathura Kumaragamage2, Scott McIntyre2, Terry W. Nixon2, Christoph Juchem1,3, and Robin A. de Graaf2
    1Biomedical Engineering, Columbia University, New York, NY, United States, 2Department of Radiology and Biomedical Imaging, Magnetic Resonance Research Center, Yale University School of Medicine, New Haven, CT, United States, 3Radiology, Columbia University Medical Center, New York, NY, United States
    A proposed low-angle combined-echo sequence is capable of producing high quality images in the presence of strong B0 magnetic field inhomogeneity with good contrast for human gray and white brain matter.
    Figure 5: LACE brain images of 5 human subjects in the presence of B0 inhomogeneity induced by a paper clip (FoV 256x192x128 mm3 at 2x2x4 mm resolution). 3D gradient echo images show large areas of signal dropout near the paper clip (top row). LACE images recover signal in this area, but exhibit image deformations due to the extreme inhomogeneity (middle row). Multi-Coil generated LACE images show largely the same image quality than the ones acquired with standard gradient coils (lower row). No image distortion correction was applied.
    Figure 1: LACE sequence. A) Sequence diagram. B) Phase graph. The sequence consists of two broadband excitation pulses separated by TE/2. Signal is acquired at the time of the spin echo of the two RF pulses (red crosses in B), which in the steady-state comprises a combination of signal from spin and stimulated echoes. Crusher gradients symmetrically around the 2nd RF pulse and at the end of TR ensure a consistent signal independent of the background B0 field.
  • Interleaved MRI and DMI on human brain in vivo
    Yanning Liu1, Henk M. De Feyter1, Scott McIntyre1, Terence W. Nixon1, and Robin A. de Graaf1
    1MRRC Yale University, New Haven, CT, United States
    We propose an interleaved MRI+DMI routine, including the hardware and sequence modifications to integrate DMI with clinical MRI. Using interleaved FLAIR+DMI, we demonstrate that MR image quality, DMI sensitivity, information content are preserved in phantoms and in the human brain in vivo.
    Interleaved FLAIR and DMI of human brain 75 min following the oral administration of [6,6’-2H2]-glucose. DMI was acquired as a 13 x 9 x 11 matrix (YXZ) over a 260 x 180 x 220 mm3 FOV, whereas the FLAIR images were acquired as 14 slices and a 256 x 192 matrix over a 256 x 192 mm2 FOV (TR/TI/TE = 8800/2200/90 ms). During the ~ 7 min FLAIR acquisition, the DMI data could be acquired twice (i.e. NA = 2). (A-C) FLAIR images spanning circa 40 mm, covering two axial planes from the 3D DMI dataset as shown in (A) and (C).
    Interleaved FLAIR MRI+DMI sequence. (A) 1H pulse acquisition scheme. 7 slices are inverted consecutively followed by a fast spin-echo (FSE) acquisition (TR/TI/TE = 8800/2200/90 ms). The (B) 2H pulse acquisitions are placed in the dead time during TI and right after FSE with equal TR (= 314 ms). (E) Shows the gradients used during acquisition. Note the plot only shows the first half of FLAIR TR, where half of the slices (odd number slices) were excited. The pulse acquisition scheme is repeated on the even number slices in the second half of TR. 28 DMI acquisitions are achieved in one FLAIR TR.
  • Accelerated diffusion and relaxation-diffusion magnetic resonance imaging using time-division multiplexing echo-planar imaging (TDM-EPI)
    Yang Ji1,2, Borjan Gagoski 2,3, W. Scott Hoge 1,2, Yogesh Rathi 1,2, and Lipeng Ning 1,2
    1Brigham and Women’s Hospital, Boston, MA, United States, 2Harvard Medical School, Boston, MA, United States, 3Boston Children’s Hospital, Boston, MA, United States
     we propose a time-division multiplexing based echo-planar imaging (TDM-EPI) sequence, which can accelerate relaxation-diffusion MRI and standard dMRI by 2 or 3 folds.
    Figure1: Diagrams of the proposed TDM-2s (B), TDM-2e (B) and TDM-3e EPI (C) sequences. The red gradient pulses alternately dephase and rephase the echoes to separate the k-space of two TDM slices, and gray gradients compensate for the phase dispersion induced by the slice-selection gradients. Numbers listed on or below the gradients signify the relative value of gradient area (i.e. zeroth moment) and corresponding subscripts denote that gradients those have same subscripts share the same relative unit standard.
    Figure 2: Evaluation of signal leakage for two representative shifting factors of 2.0 rad/mm (A) and 4.8 rad/mm (B) in TDM-3e EPI sequence. To purely acquire the leaked signal of high spatial frequencies from one slice to the others, only one pair of excitation and refocusing RF pulses was enabled with the other two disabled in a single sequence cycle. The numbers in the bottom right of the images represent the corresponding enlargement factor.
  • Fast 2D J-resolved MRSI combining echo planar imaging acquisition and turbo spin echo train evolution
    Ke Dai1, Qingjia Bao2,3, Hao Chen1, Yiling Liu1, and Zhiyong Zhang1
    1School of Biomedical Engineering, Shanghai, China, 2Wuhan United Imaging Life Science Instruments Co., Ltd, Wuhan, China, Wuhan, China, 3Department of Chemical and Biological Physics, Weizmann Institute of Science, Rehovot, Israel
    We propose a novel fast fully sampled 2D J-resolved MRSI, termed as J-resolved xSPEN spectroscopy, combining echo planar imaging acquisition and turbo spin echo train evolution.
    Figure 2 (a)The representative 2D J-resolved spectrum from the spatial location marked by the red dot. Please noting the tilting feature of the spectrum due to the mixing chemical shift and J splittings along the spectral dimension. (b) 2D J-resolved spectrum with J-decoupled chemical shift dimension.
    Figure 1 (a) 2D J-resolved xSPEN MRSI sequence performing multiple scans with increasing time (t1) evolution to achieve the chemical shift dimension, while the J couplings are obtained by repeating the distortion-free echo-planar imaging-based acquisition of xSPEN in a turbo spin echo train evolution way. (b) 2D J-resolved xSPEN MRSI sequence with pure chemical shift dimension.
  • Dual Spin-Echo Proton Density-Weighted and T2-Weighted Knee Imaging with Asymmetric Spiral In-out Trajectories
    Dinghui Wang1, Francis I. Baffour1, Daniel D. Borup2, Tzu-Cheng Chao1, and James G. Pipe1
    1Radiology, Mayo Clinic, Rochester, MN, United States, 2Royal Philips, MR R&D, Rochester, MN, United States
    A spiral dual echo spin-echo sequence with asymmetric in-out trajectories has been implemented and tested for simultaneous PDW and T2W knee imaging. Roughly a 20-36% gain in SNR can be achieved for the T2W images.
    Figure 2 Comparison of spiral T2W water images of a right knee. Data were collected with difference spiral-in and spiral-out combinations. The ratio of SNR from left to right is 1: 1.22: 1.36 estimated inside the region shown by the yellow box.
    Figure 3 Sagittal PDW and T2W images of a right knee. Cartesian reference images, spiral water-fat combined images and spiral water only images are shown from left to right. In T2W spiral imaging, the fat only image (not shown) was partially (18%) combined with the water only image to form the water-fat combined image in the middle column. Spiral in/out lengths are 8.5/17ms. The yellow arrows point to flow artifacts seen in the Cartesian images, which are mitigated in the spiral images.
  • Whole-heart CMRA non-rigid motion compensation with autofocus virtual 3D iNAV
    Alina Psenicny1, Gastao Cruz1, Camila Munoz1, Reza Hajhosseiny1, Thomas Kuestner1, Karl P Kunze2, Radhouene Neji1,2, René M Botnar1, and Claudia Prieto1
    1School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom, 2MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom
    A novel autofocused-based virtual 3D iNAV approach to enable improved 3D beat-to-beat and intra-bin non-rigid respiratory motion corrected reconstruction for free-breathing 3D cardiac MRI is proposed.
    Reformatted images of 3 representative patients using the following reconstruction frameworks: a) no motion correction (NMC), b) 2D iNAV translational correction (2D iNAV TC), c) v3D iNAV translational correction (v3D iNAV TC), d) 2D iNAV non-rigid motion compensation (2D iNAV NRMC) and e) the proposed v3D iNAV non-rigid motion compensation (v3D iNAV NRMC). Improvements when using v3D iNAV can be observed in all the cases, although in some cases (patient 1) these are minor.
    In the first step of proposed autofocus-based virtual 3D iNAV non-rigid motion correction reconstruction framework, the FH and RL motion is estimated from the 2D iNAVs. RL and AP motion is estimated with autofocus approach, assuming linear relationship between the FH iNAV signal and AP movement of the heart. Respiratory binning using the FH motion estimated from the iNAVs is performed and beat-to-beat 3D translational respiratory motion is applied to each bin. 3D non-rigid motion is then estimated from respiratory-resolved bin images and used to produce motion-corrected datasets.
  • Multi-Band Multi-Slab 3D Multi-Echo Acquisition for Simultaneous Time-of-Flight MR Angiography and Susceptibility-Weighted Imaging at 3T
    Misung Han1, Brian L Burns2, Suchandrima Banerjee2, and Janine M Lupo1,3
    1Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA, United States, 2Applications and Workflow, GE Healthcare, Menlo Park, CA, United States, 3UCSF-UC Berkeley Graduate Program in Bioengineering, University of California, San Francisco and University of California, Berkeley, San Francisco, CA, United States
    We demonstrate the feasibility of combining multi-band acceleration with 3D multi-slab, multi-echo acquisition for simultaneous time-of-flight MR angiography and susceptibility-wieghted imaging  acquisition in assessing vascular injury due to radiation therapy. 
    Figure 2. (a, e) TOF-MRA images, (b, f) TOF-MRA MIP-8 mm images, (c, g) SWI images, and (d, h) SWI mIP-8 mm images, from original and MB-accelerated acquisitions. No visible aliasing artifacts were observed with MB acceleration and overall image contrast was similar between original and MB-accelerated images. CBMs were well visualized in SWI mIP-8 mm images, as depicted by arrows.
    Figure 4. SWI mIP-8 mm images focusing on four CMBs from two patient from original and MB accelerated acquisitions were shown. The profiles over a line (with 5.9 mm length) across the center of CMBs have similar patterns between the two acquisitions, while the minimum signal levels were slightly higher with MB acceleration, possibility due to an increased noise level.
Back to Top
Digital Poster Session - Data Acquisition Sampling Trajectories
Acq/Recon/Analysis
Monday, 17 May 2021 13:00 - 14:00
  • Fast 3D whole Heart Imaging using low-discrepancy Sampling Strategies
    Tobias Speidel1, Dagmar Bertsche1, Patrick Metze1, Kilian Stumpf1, Wolfgang Rottbauer1, and Volker Rasche1
    1Internal Medicine II, Ulm University Hospital, Ulm, Germany
    Qualitative 3D isotropic heart images were acquired within 250 heartbeats (ECG gating), using specifically generated k-space interleaves with low-coherent and low-discrepancy sampling properties, based on a generalized form of the previously introduced Seiffert-Spirals. 
    Figure 2: Axial, coronal and sagittal anatomical planes of the 3D $$$\zeta$$$-based acquisition. A,B,C) was reconstructed using the entire dataset of 1,000 heart beats while D,E,F) consists of the first 250 heart beats.
    Figure 5: 3D MRI cross-sections in addition to segmented LCA, RCA and RCX as well as a segmented left ventricle, right ventricle and right atrium. A,C) was reconstructed using the entire dataset of 1,000 heart beats while B,D) is based on the first 250 heart beats.
  • Reducing clustering of readout trajectories in non-Cartesian CINE MRI
    Datta Singh Goolaub1,2 and Christopher Macgowan1,2
    1Medical Biophysics, University of Toronto, Toronto, ON, Canada, 2Translational Medicine, The Hospital for Sick Children, Toronto, ON, Canada
    A method to reduce clustering of readout trajectories for CINE reconstructions is developed and demonstrated through simulations. Improved image quality is presented.
    Representative reconstructions at HR = 137 bpm and 126 bpm for trajectories without inter-block additional angles Ɛ (i.e. Ɛ = 0), random Ɛ (between 0 and 2π), and optimized Ɛ. The trajectories in k-space are depicted on the top right corner of each reconstruction. The arrows denote worst edge (blue) and vessel (red) conspicuities with Ɛ = 0. Improvements are noticeable with random Ɛ. Further edge improvement is observed with optimized Ɛ. bpm: beats per minute.
    Framework to reduce effects of clustering. After a block of N spokes spanning TRT, additional angles (εi) are played to reduce coherence during cardiac gating. The k-space from the first block corresponding to N = 64 spokes is shown with the blue lines denoting the trajectories. Within each block the angular increment for the trajectories is Δϑ allowing adequate k-space coverage for reconstruction of real-time series.
  • Fast $$$T_{2}^{*}$$$ Mapping Using Complementary Poisson Disk Sampling and ADMM Reconstruction
    Charles Iglehart1, Ali Bilgin1, Evan Levine2, and Manojkumar Saranathan3
    1Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States, 2Microsoft Research, Redmond, WA, United States, 3Department of Medical Imaging, University of Arizona, Tucson, AZ, United States
    We have successfully shown proof of concept of a novel means of accelerating $$$T_{2}^{*}$$$ mapping via CPD-based undersampling and ADMM reconstruction. Parameter estimates from reconstructed images exhibit excellent performance at $$$R=6$$$ and acceptable performance at $$$R=8$$$
    Figure 1. Sampling via CPD. Each $$$(k_{y},k_{z})$$$ point is acquired exactly once along the echo dimension, save for a fully sampled 24 by 8 central region. This allows full sampling of k-space with incoherent sampling at each TE while promoting temporal sparsity. For a single echo train (white arrows), successive echoes are acquired within a constrained k-space neighborhood to minimize eddy currents. This results in approximate $$$R=ETL$$$ acceleration. To obtain different acceleration rates, new sampling patterns are generated and set unioned with the existing pattern.
    Figure 2. Sampling and reconstruction. (a) ME GRE images ($$$\mathbf{x}$$$) are multiplied by coil sensitivities ($$$\mathbf{S}$$$) and Fourier Transformed ($$$\mathbf{F}$$$). CPD masks ($$$\mathbf{M}$$$) are applied to produce the undersampled data array. (b) Singular Value Decomposition on 5000 signal evolutions with $$$T_{2}^{*}$$$ selected between $$$0.1$$$ and $$$2000$$$ ms generates a temporal basis. The $$$K$$$ principal components ($$$\mathbf{\phi}$$$) with the largest singular values are selected as a truncated basis. (c) Locally low rank regularization.
  • Optimizing the fixed angular increment between k-space spokes can lead to improved SNR in radial imaging
    S Sophie Schauman1, Thomas W Okell1, and Mark Chiew1
    1Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom
    Golden ratio radial sampling is not always the best way to acquire dynamic MRI data for reconstruction at multiple temporal resolutions. Higher uniformity and SNR can be achieved if the angular increment between spokes is optimised for the reconstructed temporal resolutions.
    Fig 1 - SILVER optimizes a single parameter, the angular increment, α, in the window sizes that are included in the optimisation set, S (A). In this project a TURBINE acquisition scheme, (B), was used (fig modified from [4]), and the data were reconstructed using 2D non-Cartesian SENSE (C).
    Fig 2 - Predicted sampling efficiency for different numbers of spokes per frame. SILVER optimised regions highlighted in gray.
  • Boosting the SNR-efficiency of Free Gradient Waveform Diffusion MRI using Spiral Readouts and Ultra-Strong Gradients
    Lars Mueller1, Maryam Afzali1, Malwina molendowska1, Chantal M.W. Tax1,2, Fabrizio Fasano3,4, Suryanarayana Umesh Rudrapatna1, and Derek K. Jones1,5
    1CUBRIC, Cardiff University, Cardiff, United Kingdom, 2Image Sciences Institute, University Medical Center Utrecht, Utrecht, Netherlands, 3Siemens Healthcare Ltd, Camberly, United Kingdom, 4Siemens Healthcare GmbH, Erlangen, Germany, 5Mary McKillop Institute for Health Research, Faculty of Health Sciences, Australian Catholic University, Melbourne, Melbourne, Australia
    This work demonstrates the feasibility of estimating microstructure parameters from a free gradient waveform diffusion sequence with spiral readouts in a biomimetic phantom.
    Parameters estimated from LTE and STE spiral images. The mean diffusivity (MD) is higher in the free water than in the vials with microstructure. The microscopic isotropic diffusion coefficent (Diso) is lower than MD in the microstructure. The fractional anisotropy (FA) is higher than the microscopic FA ($$$\mu$$$FA). The estimates of $$$\mu$$$FA in the free water is relatively high which is due to the fact there was only noise left for the higher b-value.
    Sequence diagram for the free gradient waveform diffusion sequence with spiral readout.
  • Variable Density Phase Encoding for High Resolution Single-Shot EPI
    Mark Chiew1
    1Wellcome Centre for Integrative Neuroimaging, FMRIB, Nuffield Department of Clinical Neurosciences, University of Oxford, Oxford, United Kingdom
    A variable density phase encoding scheme for high-resolution EPI is shown to reduce T2* blurring and residual artefacts compared to a ¾ partial Fourier conventional EPI sampling scheme.
    Figure 1 - (a) Conventional EPI phase encoding scheme showing uniform density coverage of k-space. (b) Proposed variable density EPI phase encoding, with larger PE blips at the beginning of the readout (yellow lines), and smaller blips towards the end (blue lines). (c) Conventional PE showing an exponential signal decay envelope. (d) Variable density PE with linear signal decay envelope ($$$\alpha$$$=T2*).
    Figure 4 - Example reconstructions and zoomed insets. (a,d,g): Zero-filled reconstruction of the 3/4PF sampling, (b,e,h): VC reconstruction of the 3/4PF sampling, (c,f,i): VC reconstruction of the vdPE sampling. Green arrows denote a feature that is sharp in the VC reconstructions, and blurry in the zero-filled. Red arrows denote a ringing artefact visible in both 3/4PF data, but is not present in the vdPE reconstruction.
  • Mono-planar T-Hex EPI
    Maria Engel1, Lars Kasper1, and Klaas Prüssmann1
    1Institute for Biomedical Engineering, ETH Zurich, Zurich, Switzerland
    In this work, we show high-resolution stacks of EPIs on a tilted hexagonal grid. The scheme provides flexibility in balancing readout and scan time, thereby allowing for high-quality images in a temporal resolution regime suitable for fMRI. 0.7 mm whole-brain coverage is achieved in below 5s.
    Example of mono-planar T-Hex scheme. Red and yellow dots mark those points of the hexagonal grid which are covered by two subsequent shots. Green lines in the background show that all points of the hexagonal grid lie on nodes of a finer, rectilinear lattice, from which the shift in the 1st phase-encoding (PE) direction of each EPI shot can be derived. Subsequent shots exhibit multiples of the initial shift c, modulo the PE step size d. This shift also determines the necessary echo-time shift ΔTEN = shiftN/d · TAQline, where TAQline denotes the acquisition time of one k-space line.
    Images acquired with T-Hex EPI as depicted in Fig.1. The whole brain is covered with 0.7x0.7x2mm3 resolution in 4.3s, TE = 20ms.
  • Whole-brain fMRI at 5 frames per second using T-Hex spiral acquisition
    Maria Engel1, Lars Kasper1, Franz Patzig1, Samuel Bianchi1, and Klaas Prüssmann1
    1Institute for Biomedical Engineering, ETH Zurich, Zurich, Switzerland
    In this work we show high-temporal resolution fMRI using T-Hex spiral-in trajectories. 3mm-resolved whole-brain volumes are acquired at a frame rate of 5Hz.
    Results from run B. 1) Time course of raw data, mean over ten exemplarily chosen, activated motor cortex voxel. Gray lines mark right hand tapping, dotted line marks pause (no tapping). 2) Spectrum of (1). 3) Spectrum of 12 exemplarily chosen voxel in CSF. 4) and 5) Output from cardiac and respiratory monitoring.
    T-Hex spiral used in run A. Upper panel: k-space time course of one shot. Lower panel: Cross-section through T-Hex stack, colour encoding different shots (left) and acquisition time (center). Parametric view of one shot, colour encoding again acquisition time (right).
  • High Resolution EPI with Multi-spoke Parallel Transmission and Virtual Coil Reconstruction
    Zhipeng Cao1
    1Vanderbilt University, Nashville, TN, United States
    By designing multi-spoke parallel transmit excitation pulses targeting optimal phase determined based on multichannel receive coil, the acceleration ratio of parallel imaging can be drastically increased that enables distortion-minimized single and multi- shot high resolution EPI.
    Figure 1. Demonstration of VCC-ssEPI compared to conventional ssEPI. With an 8Tx array and an 8.7ms 3-spoke pTx pulse, the VCC-ssEPI jointly achieved a relatively homogeneous flip angle distribution and the target VCC phase optimized by the 32Rx array, showed no obvious SNR degradation due to flip angle homogeneity, and achieved same SENSE g-factor with x6 acceleration as the conventional ssEPI with x3 acceleration.
    Figure 2. Demonstration of VCC-msEPI compared to shuttered msEPI, showing the ideal excitation patterns, achievable excitation patterns from pTx pulse design, and the 1/g factor map from different excitation strategies. Off-target excitation between the bands cannot be minimized with limited pulse duration and RF power limit, and the shuttered msEPI showed degraded 1/g map and thus reduced SNR compared to VCC msEPI.
  • Elliptical shells: a fast single-shot 3D readout for Arterial Spin Labeling perfusion imaging
    Joseph G. Woods1, Divya Bolar1, and Eric C. Wong1
    1Department of Radiology, University of California San Diego, San Diego, CA, United States
    We introduce an elliptical shells trajectory with spherically symmetrical sampling density for accelerated single shot FSE based whole brain ASL imaging. This trajectory is faster than a stack of spirals, and is demonstrated in a human subject.
    Generation of elliptical shells trajectory. Each spiral is confined to the surface of an ellipsoid of revolution about the Z axis. Ellipsoid i is defined by axes Kzmaxi and Kxymaxi. The dimensions of the voronoi patch are dkr*dkθ, as shown. For a given set of Kzmax and Kxymax, dkr is determined by the local geometry of the elliptical shells, and the spiral pitch dkθ is adjusted to produce a voronoi patch of the desired area for the local sampling density (see text).
    PCASL difference images from the two trajectories shown in Figure 2, as well as a fully sampled segmented stack of spiral acquisition.
  • 3D choline metabolite imaging in the liver by MRSI with selective excitation using spectral-spatial RF pulses at 7T
    Lieke van den Wildenberg1, Arjan Hendriks1, Wybe van der Kemp1, Dennis Klomp1, and Jeanine Prompers1
    1Radiology Department, UMC Utrecht, Utrecht, Netherlands
    With the application of a spectral-spatial RF pulse, we were able to map total choline using a 3D 1H-MRSI sequence with full liver coverage at maximum intrinsic sensitivity (short TE and TR).
    Figure 5: In-vivo 3D 1H-MRSI SPSP liver data of one volunteer. A) Transverse, coronal and sagittal MR image of the liver overlaid with the 1H-MRSI grid. Spectra of three different voxels indicated in panel A are shown in the green (B­-SPSP water, E-SPSP choline), orange (-SPSP water, F-SPSP choline) and blue (D-SPSP water, G-SPSP choline) frames below.
    Figure 3: Coronal image of the spherical phantom (d=10cm) (A) and the liver (E) with a normal excitation pulse, but without slice selection, and with the spectral-spatial pulse (B and F), showing the transversal slice excited by the spectral-spatial pulse. Panel D shows a normal coronal image of the body with coronal slice selection as a reference. Line profiles through the images acquired with the spectral-spatial pulse (blue lines in B and F) in the phantom (C) and in the liver (G).
  • Towards Accelerating 3D 1H-MRSI Using Randomly Undersampled Spatial and Spectral Spirals with Low-rank Subspaces
    Yamin Arefeen1, Borjan Gagoski2,3, and Elfar Adalsteinsson1,4,5
    1Massachusetts Institute of Technology, Cambridge, MA, United States, 2Fetal-Neonatal Neuroimaging and Developmental Science Center, Boston Children’s Hospital, Boston, MA, United States, 3Department of Radiology, Harvard Medical School, Boston, MA, United States, 4Harvard-MIT Health Sciences and Technology, Cambridge, MA, United States, 5Institute for Medical Engineering and Science, Cambridge, MA, United States
    A randomly undersampled, spiral-based trajectory yields 2-3x acceleration  over fully sampled 3D-spiral acquisitions in 3D Magnetic Resonance Spectroscopy by inducing incoherent aliasing that can be resolved with SPICE-based low rank priors.
    Figure 4: Metabolites maps, RMSE on image support, and example spectra from subspace reconstructions of fully sampled , uniformly undersampled (R = 2), and undersampled with the proposed trajectory (R = 2) on a spectroscopy phantom containing physiological concentrations of the major brain metabolites. The proposed trajectory yields similar spectra and metabolite maps to the fully sampled acquisition and achieves ~2x reduction in RMSE across the metabolites in comparison to uniform undersampling.
    Figure 1: Schematic view of (a) the fully sampled 3D-spiral MRSI trajectory and (b) our proposed randomly undersampled trajectory. For demonstration purposes, we assume 1 temporal interleaf. In (a), the same angular interleaf and $$$k_z$$$ point is sampled at each timepoint in a TR. The proposed (b) trajectory samples random angular interleaves and $$$k_z$$$ points during each TR and can be accelerated by reducing the number of TR’s. Our proposed trajectory takes advantage of the 4D-encoding space by spreading incoherent aliasing across spatial and spectral dimensions.
  • Lipid and water separation through an SMS-like approach in 7T to reduce SAR in brain EPI
    Amir Seginer1, Edna Furman-Haran2,3, Ilan Goldberg4, and Rita Schmidt3,5
    1Siemens Healthcare, Rosh Ha'ayin, Israel, 2Life Sciences Core Facilities, Weizmann Institute of Science, Rehovot, Israel, 3The Azrieli National Institute for Human Brain Imaging and Research, Weizmann Institute of Science, Rehovot, Israel, 4Deparment of Neurology, Wolfson medical center, Holon, Israel, 5Neurobiology Department, Weizmann Institute of Science, Rehovot, Israel
    • SAR may limit volume coverage and min. TR for fMRI at 7T.
    • Removing the fat suppression pulse from GRE-EPI greatly reduces SAR.
    • To compensate, an SMS-like reconstruction separates lipid and water images.
    • Simulations, phantom experiments, and fMRI experiments support the method.
    Figure 1: Extended formulation unifying three parallel imaging aspects: (i) in-plane PE acceleration, (ii) SMS acceleration and (iii) lipid-water separation. The steps show how an EPI image (far right) is equivalent to an SMS of water and lipid “slices” (Iw and Ilip) and to an in-plane PE acceleration (Ilip&w #2 vs. Iw+Ilip shifted). The formulation can be extended to any number of slices and fat/water images.
    Figure 5: Resting-state fMRI. a),b) Examples of simultaneously acquired slices. Left to right: standard reconstruction, water image after lipid separation, and a fat suppressed image. Arrows mark artifacts from the lipids. (c) tSNR on three orthogonal planes. SNR and tSNR in main text where estimated from marked regions. SAR was 33% without fat-suppression and 97% with fat-suppression (reference amplitude of 240 V for a 1 ms 180° hard pulse). Scan parameters: in-plane accelerations ×3, SMS ×2, FOV=220×220 mm2, resolution = 1.7×1.7 mm2, slice thickness = 1.7 mm, TR/TE = 1500/22 ms.
  • On the impact of B0 map resolution and undersampling on reconstructions using expanded encoding models
    Paul Dubovan1,2, Lars Kasper3, Kamil Uludag3,4, and Corey Baron1,2
    1Medical Biophysics, Western University, London, ON, Canada, 2Center for Functional and Metabolic Mapping, Robarts Research Institute, London, ON, Canada, 3Techna Institute, University Health Network, Toronto, ON, Canada, 4Medical Biophysics, University of Toronto, Toronto, ON, Canada
    We investigate the effects of signal undersampling and static field map resolution for single-shot spiral diffusion MRI acquisitions, which showed optimized image quality for a moderate acceleration factor and the highest field map resolution.   
    FIG. 3. Mean DWI maps of a sample slice created from 6 different diffusion-encoding directions (b = 1000 s/mm2), acquired with acceleration factors R = 2-5, and varying field map resolutions.
    FIG. 4. Noise SD maps for acceleration factors of R = 2-5, and isotropic field map resolutions of 3 mm, 2 mm, and 1.5 mm. Maps created by calculating the SD of 100 b0 image replicas and scaled inversely by the square root of R. Numbers below maps indicate the mean/maximum noise SD of brain and skull components. All values shown are truncated and are to be multiplied by a factor of 10-6 to show the full values.
  • Pushing the acceleration performance of WAVE-CAIPI using a single-axis gradient insert – a simulation study
    Alejandro Monreal Madrigal1, Edwin Versteeg1, and Jeroen Siero1,2
    1Radiology, University Medical Center Utrecht, Utrecht, Netherlands, 2Spinoza Centre for Neuroimaging, Amsterdam, Netherlands
    A simulation study was performed by retrospectively under-sampling data with different Wave-CAIPI parameters. A 16-fold acceleration was simulated without noticeable decrease in image quality. Furthermore, the high gradient performance allowed for a 5-fold shorter readout time.
    Figure 2: Image reconstructions. a) Low WC parameters R=4x4, b) Reference image, c) Very High WC parameters R=4x4
    Table 2: Conventional vs Insert gradient reconstruction results
  • The Design of an Acentric Cartesian Spiral Sampling with Partial Fourier for Highly Accelerated Single-breath-hold Isotropic 3-D Cardiac Cine
    Yu Ding1, Qi Liu1, Jingyuan Lyu1, Yuan Zheng1, and Jian Xu1
    1UIH America, Inc, Houston, TX, United States
    A combination of novel acentric spiral pattern and partial Fourier acquisition method was designed and tested in 3-D cardiac cine imaging with isotropic spatial resolution and whole heart coverage. 
    Figure 1. The acentric Cartesian spiral pattern in PE-SPE plane, with matrix size 128 x 64, 75% partial Fourier in both directions.
    Figure 2. The upper row are eight short axis slices at the end of systole; the lower row are the corresponding eight slices at the end of diastole. The distance between two adjacent slices shown is 8.4 mm (four slice thickness).
  • Optimization of scan parameters to reduce acquisition time for resolve-based DKI in NPC
    Yaoyao He1, Hao Chen1, Huiting Zhang2, Robert Grimm2, Cecheng Zhao3, Xiaofang Guo1, Yulin Liu1, and Zilong Yuan1
    1Hubei Cancer Hospital, Wuhan, China, 2Siemens Healthcare, Erlangen, Germany, 3Huazhong Agricultural University, Wuhan, China
    With the advantages of stable image quality and time saving, we recommended that the DKI imaging of (200, 400, 800, 2000) with 5/8 RPF can be applied to the clinical DKI research in NPC.
    Figure 1. Representative images from a patient, including T2w with fat suppression (FS), T1w with contrast, MK and MD based on DKI with 5b-values (200, 400, 800, 1500, 2000).
    Table 1. The b-value combinations and simulated acquisition time used in this study.
  • Strategically Acquired Gradient Echo imaging with Compressed Sensing: Comparison of quantitative images with different acceleration factors
    bingbing gao1, Yuhan Jiang1, Nan Zhang1, Yanwei Miao1, Qingwei Song1, Ailian Liu1, and Peng Sun2
    1the First Affiliated hospital of Dalian Medical university, Dalian, China, 2Philips Healthcare, BeiJing, China
    Combing Strategically Acquired Gradient Echo (STAGE) imaging with compressed sensing (CS) technique to find an optimal acceleration factor.
  • Optimization of Compressed Sensing Acceleration Factors for Lumbosacral Plexus 3D MRI
    Renwang PU1, Qingwei SONG2, Ailian LIU1, Hao nan ZHANG1, Nan ZHANG1, and Jiazheng WANG3
    1the First Affiliated Hospital of Dalian Medical University, Dalian, China, 2the First Affiliated Hospital of Dalian Medical University, DALIAN, China, 3Philips Healthcare, BEIJING, China
    Three-fold acceleration rate with compressed sensing reached an acceptable balance between imaging time and image quality for lumbosacral plexus 3D-T2*-FFE imaging, taking the traditional 2-fold SENSE acceleration as a gold standard, with the scan time cut by 43% from the baseline. 
    one-slice images of the 3D-T2*-FFE.A:Coronal image of lumbosacral plexus.The areaof interest in the nerve and muscle. B: Conventional 3D-T2*-FFE with SENSE factor of 2.C-G: 3D-T2*-FFE with CS factors of 2, 3, 4,5 and 6, respectively. Data were collected from a healthy volunteer (male, 24 years old).
    The scan parameters of 3D-T2*-FFE with SENSE factor of 2, 3D-T2*-FFE with CS factors of 2, 3, 4,5 and 6, respectively.
  • Rapid and robust DTI using a turboPROP technique with blade sharing and whole-blade acquisition
    Zhiqiang Li1 and John P Karis1
    1Neuroradiology, Barrow Neurological Institute, Phoenix, AZ, United States
    ssEPI is widely used in clinical DTI but suffers geometric distortion artifacts. In this work we develop a whole-blade acquisition mode and a blade sharing strategy for turboPROP to achieve comparable scan time as clinical ssEPI-based DTI scan. In vivo results demonstrate good image quality.
    Fig. 5. Comparison of quantitative measurement (FA maps) between ssEPI and turboPROP.
    Fig. 4. Demonstration of of geometric distortion artifacts, which are observed in the trace images from ssEPI but minimized in turboPROP.
Back to Top
Digital Poster Session - Novel Encoding Strategies
Acq/Recon/Analysis
Monday, 17 May 2021 13:00 - 14:00
  • Nonlinear projection imaging with the Bloch-Siegert shift in an inhomogeneous B0 at low-field
    Kartiga Selvaganesan1, Yonghyun Ha2, Basong Wu2, Kasey Hancock3, Charles Rogers III2, Sajad Hosseinnezhadian2, Gigi Galiana2, and Todd Constable2
    1Biomedical Engineering, Yale University, New Haven, CT, United States, 2Radiology and Biomedical Imaging, Yale University, New Haven, CT, United States, 3Electrical Engineering, Yale University, New Haven, CT, United States
    This work shows that the Bloch-Siegert shift, which provides an RF alternative to spatial encoding with gradient coils, can be used to image in a low-field, open MRI system. We present a new approach to evaluate and optimize nonlinear RF encoding schemes generated by transmit phased arrays.
    Figure 3. The correlation between pairs of phasors was calculated for the initial set of N phasors (1). The phasors with the smallest correlation was added to the "input optimal set" (2). The sum of correlations squared was calculated between the remaining $$$(N-i)$$$ phasors, and the phasors in the "input optimal set", for a total of $$$(N-i)$$$ β values. The phasor with the smallest $$$β$$$ value was then added to the "output optimal set" (3). The ”output optimal set” became the new input, and steps 2 and 3 were repeated until the final optimal set contained the desired number of phasors (4).
    Figure 5. Brain images reconstructed from the simulated phasor patterns. Image reconstruction was performed using 128 encoding patterns in a 128x128 matrix. The overall image resolution is improved by applying a higher Bloch-Siegert pulse amplitude (C vs. B).
  • B1-gradient based MRI using Frequency-modulated Rabi Encoded Echoes (FREE)
    Efraín Torres1,2, Taylor Froelich2, Lance DeLaBarre2, Michael Mullen2, Gregory Adriany2, Alberto Tannús3, Daniel Cosmo Pizetta3, Mateus Jose Martins3, and Michael Garwood2
    1Biomedical Engineering, University of Minnesota, Minneapolis, MN, United States, 2Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States, 3Centro de Imagens e Espectroscopia por Ressonância Magnética - CIERMag - Sao Carlos Physics, São Carlos, Brazil
    Here, a B1-encoding approach is presented that allows for phase-encoding without the use of Bgradient coils. This presents the possibility of designing systems without B0 gradient coils, greatly reducing cost and increasing portability. 
    Fig. 1a) Multi-shot approach of FREE 1b) is a single-shot approach. First double spin-echo pair has the greatest R-difference. Subsequent π pulse pairs have a difference of exactly ∆R incrementally adding phase at each echo acquired. 1c) Multi-shot version of FREE, in which all (odd and even) echoes are acquired. All pulses, except for the first, have Rbase + ∆R, where Rbase is any adiabatic R value. Further shots reduce the R value of the first pulse (Ro) by integer values of ∆R. The k-space matrix is created by concatenating the data columns (shot number) with rows as echo number.
    Phantom is a NMR tube with ~15 cm water doped with copper sulfate. Structure was created in the phantom by purposefully creating a flip angle range. Comparison of the GRE projection and FREE reconstruction. The x-axis represents nutation frequency which has positional dependence due to the surface coil’s B1 gradient. The plane of surface coil is located on the right where is the largest. Linearity of the B1 gradient decreases with distance from the coil. Good agreement between GRE and FREE reconstructions is achieved except at the deeper location.
  • Rapid simultaneous T1 and T2 quantification (RAS-Q T1T2) of the myocardium using transient bSSFP with variable flip angles
    Céline Marquet1,2,3, Jihye Jang1,2,4, Andrew J. Powell1,2, and Mehdi H. Moghari1,2
    1Department of Pediatrics, Harvard Medical School, Boston, MA, United States, 2Department of Cardiology, Boston Children's Hospital, Boston, MA, United States, 3Department of Informatics, Technical University Munich, Munich, Germany, 4Philips Healthcare, Boston, MA, United States
    RAS-Q T1T2 combines transient bSSFP with variable flip angles for simultaneous myocardial T1 and T2 quantification at a rapid scan time (≤ 4s). It shows accurate T2 quantification, lower T2 precision, and lower T1 accuracy and precision.
    Figure 1: RAS-Q T1T2 pulse sequence acquisition scheme. Four images are acquired in the same diastolic phases within 4 consecutive heartbeats (t). Each image is acquired with a different flip angle (α = [10°, 50°, 90°, 130°]) after ten linear startup pulses (ST). All further sequence parameters are kept constant (TR/TE: 2.96/1.48ms, CS-SENSE: 3). Image acquisition is performed during breath-hold in ≤4s.
    Figure 2: Magnetization curves for different T1/T2 ratios. Graphs show normalized magnetization evolution over flip angles from 10°-130° for 9 different tubes of the T1MES phantom.15 Measured mean pixel intensity (black, dashed curves) is compared to theoretical values of RAS-Q T1T2 (red curve) and the transient bSSFP function (blue, dotted curves). Results are depicted for heartrate 60bpm. Measured RAS-Q T1T2 pixel intensities demonstrate a close alignment to expected theoretical values over different T1/T2 ratios and heartrates.
  • Susceptibility artifact-insensitive ultrafast 3D gradient-echo imaging by combination of head-tilting and ERASE acquisition
    Jaeyong Yu1,2, Seulki Yoo1,2, Jae-Kyun Ryu3, Seung-Kyun Lee1,2,4, and Jang-Yeon Park1,2,3
    1Department of Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of, 2Department of Intelligent Precision Healthcare Convergence, Sungkyunkwan University, Suwon, Korea, Republic of, 3Biomedical Institute for Convergence at SKKU, Sungkyunkwan University, Suwon, Korea, Republic of, 4Center for Neuroscience Imaging Research, Institute for Basic Science (IBS), Suwon, Korea, Republic of
    We proposed a new method for ERASE sequence with head-tilted scan to improve B0 inhomogeneity in comparison with gradient-echo EPI.
    Figure 3. Axial views of the reference GRE magnitude, EPI and ERASE images for a single subject. The solid red lines indicate the manually drawn masks as a guide for comparison of image quality. Signal loss and image degradation in prefrontal cortex of the normal-orientation scans (white arrows) are reduced in tilted scans (blue arrows and yellow arrows).
    Figure 2. Measured B0 maps at normal and tilted head orientations at 3T and 7T. Black arrows indicate B0 inhomogeneity near the prefrontal region. In head-tilted scan (c-d, g-h), B0 field distribution is spread more indicating that local gradient is smaller than the normal orientation scan (a-b, e-f).
  • Optimization of Flip Angle and RF Pulse Phase in MP-SSFP for MRI in Inhomogeneous Magnetic Fields
    Naoharu Kobayashi1 and Michael Garwood1
    1CMRR, Radiology, University of Minnesota, Minneapolis, MN, United States
    Flip angle and RF phase in MP-SSFP were numerically optimized. The optimal RF setting improved SNR under the fixed SAR conditions and provided strong image contrasts in brain tissues with linear gradient inhomogeneous fields at 3T.
    Figure 5. Brain MP-SSFP images with 1.5-mm isotropic resolution. Images acquired with αstd = 6º, optimal flip angle and even phase (A) provided stronger image contrasts between gray and white matter as compared to those with αstd=20º, optimal flip angle and odd phase (B), but as predicted in simulation and phantom experiments, the noise level is noticeably higher in the images acquired with αstd=6º.
    Figure 3. Brain MP-SSFP images acquired with αstd=6º (A) and 20º (B). With αstd=6º, strong image contrasts between gray and white matters were detected, which is consistent in even and odd RF phase images. With αstd=20º, odd RF phase provided stronger image contrasts in brain tissues than even RF phase. Although SNR increased along with an increase of NMP, image contrasts were consistent in each RF pulse setting.
  • Spectro-Dynamic MRI: Characterizing Bio-Mechanical Systems on a Millisecond Scale
    Max H. C. van Riel1,2, Niek R. F. Huttinga1, and Alessandro Sbrizzi1
    1Department of Radiotherapy, Computational Imaging Group for MR diagnostics and therapy, UMC Utrecht, Utrecht, Netherlands, 2Department of Biomedical Engineering, Medical Image Analysis, Eindhoven University of Technology, Eindhoven, Netherlands
    We propose spectro-dynamic MRI as a method for the characterization of dynamical systems directly from k-space data. A measurement model and dynamical model are used to estimate the motion fields and dynamical parameters of the system at a millisecond time scale from heavily undersampled data.
    Figure 3: The estimated displacements. The black lines indicate the estimated displacements for the experiments with (a) a simple pendulum, (b) coupled pendula, (c) a spherical pendulum, and (d) coupled spherical pendula. For the 2D acquisitions in (c) and (d), the $$$x$$$- and $$$y$$$-displacements are plotted separately. The red line is the reference line as determined in Fig. 2.
    Figure 1: (a) In conventional MRI, a different line in k-space is sampled every repetition. (b) In spectro-dynamic MRI, one k-space line is repeated $$$N_\textrm{rep}$$$ times, reducing the time interval between the acquisition of the same data point in k-space. (c) Schematic representations of the four dynamical systems used for our experiments. (d) The experimental setup with two pendula, each containing a gel-filled vial (red arrow), which are connected with a spring (light blue arrow).
  • 2.5D MRI of the Vocal Fold Oscillation using Single Point Imaging with Rapid Encoding
    Johannes Fischer1, Ali Caglar Özen1,2, Matthias Echternach3, Louisa Traser4, Bernhard Richter4, and Michael Bock1
    1Dept.of Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Freiburg Site, German Cancer Research Center (DKFZ), Heidelberg, Germany, 3Division of Phoniatrics and Pediatric Audiology, Department of Otorhinolaryngology, Head and Neck Surgery, Ludwig-Maximilians-University, Munich, Germany, 4Institute of Musicians' Medicine, Freiburg University Medical Center, Germany Faculty of Medicine, University of Freiburg, Freiburg, Germany
    Single point imaging with rapid encoding (SPIRE) is used image the vocal fold oscillations in the coronal plane and readout in AP direction allows for the reconstruction in multiple slices. Electroglottography and MR navigators are used to gate and motion-correct data prior to reconstruction.
    Figure 4: Vocal fold oscillation in the coronal plane reconstructed into 4 slices and 12 Frames resulting in a temporal resolution Δt = 890μs (f0 = 92Hz). Although motion of the vocal folds is visible, no full contact can be observed. The vocal fold oscillations cause the EGG electrodes to vibrate on the skin, which can be seen on the edges of each frame.
    Figure 1: Sequence diagram of the 2.5D SPIRE method. Rapidly switched phase encoding gradients immediately follow $$$G_\mathrm{SS}$$$ and slice rephasing is performed simultaneously with the frequency encoding in slice direction (dark grey areas indicate equal gradient moments). PE gradients are rewound and a spoiler is applied with opposite sign to the readout gradient to reduce switching when ramping to $$$G_\mathrm{SS}$$$ at the beginning of the next acquisition block. Dotted lines indicate the gradient shapes for the largest phase encoding moments.
  • Zero Echo Time Imaging Using Low Resolution k-Space Interleaves
    Hanna Frantz1, Thomas Huefken1, Patrick Metze1, Kilian Stumpf1, Tobias Speidel1, and Volker Rasche1
    1Department of Internal Medicine II, Ulm University Medical Center, Ulm, Germany
    This abstract presents an approach that relies on the interleaved combination of ZTE read-outs with different resolutions combined with a Compressed Sensing reconstruction, to acquire more information around the k-space center, for high-quality ZTE without prolonged acquisition times.
    Figure 1: Coronal slice of a dedicated rubber phantom acquired at 15 mT/m gradient strength (A, B) and 24 mT/m (C, D). Images A and C were acquired with the modified approach including the low-strength gradients.
    Figure 3: Coronal slice of a wrist acquired at 15 mT/m gradient strength (A, B) and 24 mT/m (C, D). Images A and C were acquired with the modified approach including the low-strength gradients.
  • Arterial calcification imaging using ZTE on PET/MR
    Edwin Eduard Gert Willem ter Voert1, Florian Wiesinger2, Graeme McKinnon3, Mathias Engström4, Jose Fernando de Arcos5, Marlena Hofbauer1, Ronny R Buechel1, and Philipp A Kaufmann1
    1Department of Nuclear Medicine, University Hospital Zurich, Zurich, Switzerland, 2GE Healthcare, Munich, Germany, 3GE Healthcare, Milwaukee, WI, United States, 4GE Healthcare, Stockholm, Sweden, 5GE Healthcare, Little Chalfont, Amersham, United Kingdom
    In the current study we demonstrate that ZTE imaging on PET/MR, combined with Deep Learning reconstructed ZTE, can likely be applied to detect calcifications in peripheral vasculature.
    Figure 2: Example coronal slices and a 3D volume rendering in the abdominal region. a) prior obtained CT, b) Deep Learning reconstructed ZTE, c) ZTE calcifications (red) overlaid on the contrast enhanced water map MRI (LAVA-Flex based), d) 3D volume rendering showing the ZTE calcifications (cyan).
    Figure 4: Example coronal slices and a 3D volume rendering in the head/neck region. a) prior obtained CT with indicated (red arrow) calcification, b) Deep Learning reconstructed ZTE, c) ZTE calcifications (red) overlaid on the contrast enhanced water map MRI (LAVA-Flex based), d) 3D volume rendering showing the ZTE calcifications (cyan).
  • Compressed Sensing PETRA MRI
    Serhat Ilbey1, Johannes Fischer1, Michael Bock1, and Ali Caglar Ozen1,2
    1Dept. of Radiology, Medical Physics, Medical Center University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany, 2German Consortium for Translational Cancer Research Partner Site Freiburg, German Cancer Research Center (DKFZ), Heidelberg, Germany
    csPETRA enables 3D imaging with isotropic sub-millimeter resolution within only a few minutes, e.g., for (0.5 mm)3 voxel size with (20 cm)3 field of view, which is demonstrated with different acceleration factors for high-resolution imaging of the knee.
    Fig. 3: Coronal (first row) and sagittal view (third row) of PETRA (Acc=1) and csPETRA (Acc>1) images. Yellow numbers show the Acc factor used for the SPI acquisition. Second and fourth rows show difference images subtracted with Acc=1 images (far left images) of coronal and sagittal views, respectively.
    Table 1: Examples of SPI acquisition times w.r.t. the parameters $$$\text{BR}$$$, $$$\text{TE}$$$, and $$$T_\text{enc}$$$ with the corresponding maximum gradient amplitude (|G|). Default parameters were: $$$\text{BR}$$$=512, $$$\text{TE}$$$=50 µs, $$$T_\text{enc}$$$=500 us, TR=2 ms, when they were not used as the sweep parameter.
  • Single echo reconstruction for rapid and silent MRI
    Sairam Geethanath1
    1Columbia MR Center, Columbia University, New York, NY, United States
    This work demonstrates the the first, rapid 128 x 128 imaging without phase encoding using a 64-channel coil. This was applied to T2 weighted imaging of 11 slices with (i) Acquisition time for TE (80ms) = 1.8s (ii) Total RF deposition for both TEs = 10.8W (iii) PNS % 12.09 (iv) no blurring artifacts
    Fig.4: Comparison of SER with other methods for eleven slices - the top row shows vendor-provided gold standard spin-echo (SE), a turbo SE with an ETL of 7 and a GRAPPA factor of 6, a half-Fourier acquisition single-shot TSE (HASTE) at echo times (TE) shown in red font. The corresponding acquisition times (Tacq) are shown in yellow font and were recorded from the vendor’s user interface. The bottom row shows the corresponding images at TEs close to 80ms allowed by the vendor. SER images acquired using pypulseq at similar slice locations, do not suffer from saturation or blurring artifacts.
    Fig. 5: Single echo reconstruction (SER) imaging performance a) SER provides the fastest acquisition time for the four methods, depends on echo time (TE) rather than repetition time (TR) and phase encoding steps, and is faster than TSE + GRAPPA by an order of magnitude; b) SER delivers the lowest RF power to the phantom among the methods, due to the one-time use of the 900 and 1800 pulse; c) SER is the most silent scan with the least peripheral nerve stimulation (PNS) percentage due to the one-time use of the readout gradient
  • Investigations on accelerated imaging at 9.4T with electronically modulated time-varying receive sensitivities
    Felix Glang1, Kai Buckenmaier1, Jonas Bause1, Alexander Loktyushin1, Nikolai Avdievich1, and Klaus Scheffler1,2
    1High-field Magnetic Resonance Center, Max-Planck Institute for Biological Cybernetics, Tübingen, Germany, 2Department of Biomedical Magnetic Resonance, Eberhard Karls University Tübingen, Tübingen, Germany
    Time-varying receive sensitivities can be achieved by introducing variable capacitance diodes to the receive loops, which allow to dynamically tune B1-. This offers an additional degree of freedom for MR image encoding, with the potential for improved parallel imaging performance.
    Figure 1. Schematic of the proposed receive loop geometry and circuitry. (a) Simulation setup for one of the four modulatable receive loops placed circumferentially around a cylindrical phantom. (b) Circuit diagram of the receive loop with tuning and matching capacitors as well as six adjustable varactor diodes (each in a range between 2 and 20pF) placed on the long sides of the loop, which are used to modulate receive sensitivity profiles B1-.
    Figure 4. Sensitivity modulation during readout (RO) of k-space lines. k-space configuration weightings for (a) linear ramp, (b) 8 and (c) 64 sinusoidal oscillations between C1 and C2 along each RO line. The latter is equivalent to switching between C1 and C2 for each ADC sample. (d), (e), (f) G-factors for the respective sensitivity switching patterns. (g) Maximum g-factor for sinusoidal sensitivity switching versus number of oscillations, compared to static sensitivity sets of C1, C2 and combined C1 & C2.
  • Feasibility of a Novel Sampling/Reconstruction Method Ensuring a SNR Benefit Over the Traditional Sampling Approach
    Samuel Perron1, Matthew S. Fox1,2, Hacene Serrai1, and Alexei Ouriadov1,2,3
    1Physics and Astronomy, The University of Western Ontario, London, ON, Canada, 2Lawson Health Research Institute, London, ON, Canada, 3School of Biomedical Engineering, The University of Western Ontario, London, ON, Canada
    The proposed software-based under-sampling and reconstruction method has been shown to yield improved SNR for FGRE, x-Centric, and FE-Sectoral imaging schemes, with possible implementation in existing MRI systems without any hardware modifications.
    Figure 4. Representative resolution phantom images reconstructed from retroactively under-sampled k-spaces corresponding to FGRE, x-Centric and FE Sectoral. AF were 7, 10 and 14. FGRE SNR was 20, x-Centric SNR was 28 and Sectoral SNR was 25. The FE Sectoral images show much less image distortions at all three AF compared to FGRE and x-Centric.
    Figure 1. Depicts three different k-space under-sampling patterns corresponding to AF=10, used for each wash-out image retroactively in wash-out or SNR attenuation direction.
  • A Deselecting Alias Approach to Volumetric Zoomed Imaging
    Nicolas Arango1, Molin Zhang1, Jason Stockmann2,3, Jacob White1, and Elfar Adalsteinsson1,4
    1Department of Electrical Engineering and Computer Science, Massachusetts Institute of Technology, Cambridge, MA, United States, 2A. A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Harvard Medical School, Boston, MA, United States, 4Institute for Medical Engineering and Science, Massachusetts Institute of Technology, Cambridge, MA, United States
    We demonstrate simulations of a new method for ∆B0 selective excitation for rFOV imaging by only deselecting signal that would otherwise alias into the target FOV.
    Figure 1. Excited volume from one-sided zoomed imaging excitation and optimized b) ΔB0 field. The Hyperboloid excitation volume does not take advantage of the don't care regions introduced by the alias deselection mask.
    Figure 3. Sagital slice of excitation (a,b,c) and rFOV aliasing pattern (d,e,f) from Linear gradient optimized and linear gradients plus multicoil shim optimized rFOV excitation. (a,d) Linear gradients can select an off-axis slab between aliases with a sufficiency large rFOV. (b,e) Tighter rFOVs cannot be achieved with linear gradients only.(c,f) Multicoil shim array in conjunction with linear gradients achieves tighter rFOV.
  • Efficient NUFFT Backpropagation for Stochastic Sampling Optimization in MRI
    Guanhua Wang1, Douglas C. Noll1, and Jeffrey A. Fessler2
    1Biomedical Engineering, University of Michigan, Ann Arbor, MI, United States, 2Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI, United States
    Fast approximation method for calculating the Jacobian matrix of the NUFFT operations.
    Figure 3. The learned trajectories by different methods. The initialization is an under-sampled radial sampling pattern.
    Figure 1. The gradient w.r.t. $$$\boldsymbol{x}$$$. We denote auto-differentiation of NUFFT as auto-diff, exact NDFT as exact, and our approximation method as approx. Three calculation methods achieve similar results.
  • A digital MRI RF-receiver using an ordinary GPU
    Annalena Erbrecht1, Enrico Pannicke1, and Christoph Dinh1
    1Otto-von-Guericke University, Magdeburg, Germany
    Our studies showed that a CIC based decimation filter is suitable to realize a digital MRI console which requires low calculation effort and high noise attenuation. This filter can be used to directly convert the acquired signal to the k-space.
    Figure 1: Block diagram of an MRI RF-receiver
    Figure 2: Block diagram of a five stages CIC filter combined with a compensation filter
  • Feasibility of a single low dose dual temporal resolution DCE-MRI method for whole-brain, high-spatial resolution parametric mapping
    Ka-Loh Li1, Daniel Lewis2, David J Coope3, Federico Roncaroli3, Omar N Pathmanaban2, Andrew T King2, Sha Zhao1, Erjon Agushi1, Alan Jackson1, Timothy Cootes1, and Xiaoping Zhu1
    1Division of Informatics, Imaging and Data Sciences, The University of Manchester, Manchester, United Kingdom, 2Department of Neurosurgery, Salford Royal NHS Foundation Trust, Manchester, United Kingdom, 3Division of Neuroscience and Experimental Psychology, The University of Manchester, Manchester, United Kingdom
    A new dual-temporal resolution DCE-MRI analysis technique (LEGATOS) combined with a single-injection low-dose interleaved protocol permits the derivation of accurate, tissue-validated high-spatial resolution kinetic parameter estimates following a single low contrast agent dose. 
    Figure 3: Representative images from a patient with a large left sided sporadic VS imaged using a single low-dose (LD) interleaved high-temporal (HT) (frame duration of Dt = 1.46 s) and high-spatial resolution (Dt = 6.04 s) DCE-MRI acquisition on a 1.5 T Philips scanner. Kinetic maps derived using the HT segments alone (LDHT, left column) or the LEGATOS method are shown. Note the increased vascularity around the tumor capsule, visible on both the high-spatial T2W DRIVE acquisition (voxel size = 0.5 x 0.5 x 0.5mm3) and LEGATOS derived vp parameter maps (*).

    Figure 4: Comparison of LEGATOS derived vp and Ktrans estimates from the in vivo single-injection low-dose DTR Study against tissue derived parameters

    (A) Inter-tumor scatterplot comparison of LEGATOS derived mean tumor vp estimates against mean CD31 % microvessel surface area (SA).

    (B) Inter-tumor scatterplot analysis of LEGATOS derived mean tumor Ktrans (min-1) against mean CD31 % microvessel surface area (SA).

    (C) Representative imaging and histology from a patient with a growing highly vascular VS (top row) and a comparatively less vascular static VS (bottom row) are shown.


  • B1+ inhomogeneity mitigation using adiabatic refocusing RF pulses for diffusion weighted imaging at 7T
    Shahrokh Abbasi-Rad1, Martijn Cloos1,2, Jin Jin3, Kieran O'Brien3, and Markus Barth1,2,4
    1Centre for Advanced Imaging, University of Queensland, Brisbane, Australia, 2ARC Training Centre for Innovation in Biomedical Imaging Technology, The University of Queensland, Brisbane, Australia, 3Siemens Healthcare Pty Ltd, Brisbane, Australia, 4School of Electrical Engineering and Information Technology, The University of Queensland, Brisbane, Australia
    We showed that the use of two adiabatic refocusing pulses (TR-FOCI) could mitigate the B1+ inhomogeneity problem commonly observed at 7T in a twice-refocussed spin echo diffusion weighted images.
    Figure 4. Shows the comparison of b0 images acquired by SLR versus adiabatic refocusing pulses. The B1+ insensitivity of the adiabatic TR-FOCI pulse mitigates the B1+ inhomogeneity leading to retrieval of spin-echo signal at the temporal lobe and cerebellum. The first column shows the relative FA map indicating the low B1+ areas and the last column shows the relative SNR map. The SNR map indicates that in the B1+ starved areas of the brain SNR is improved up to 10 times. The TR value (30000 ms) was kept identical between the two measurements so that signal intensities could be comparable.
    Figure 3. (a) shows the localizer view of the phantom with the location of the acquired slice. (b) shows the slice profile images of the SLR refocusing pulse versus TR-FOCI refocusing pulse with different scale factors for the power of the pulse. With the scale factor of one, the pulse is applied with the power of 9 T, which is derived from the simulations as the minimum value required for the adiabaticity of the pulse. (c), (d), and (e) show the slice profile across the superior, center, and the inferior regions of the slice.
  • RF-encoding for improved multi-voxel separation in MR spectroscopy
    Adam Berrington1, Penny Gowland1, and Richard Bowtell1
    1Sir Peter Mansfield Imaging Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, United Kingdom
    Phase and amplitude signal modulation was shown to improve the g-factor of reconstruction for 5 voxels in simulation. Phantom spectra from two-voxels were acquired over M=1…5 encoding steps with 30° phase increments and were of higher SNR than reconstructing using coil-encoding alone.
    Fig.1: RF-encoding simulated with unique spectral peaks assigned to 5 voxel ‘locations’ (coloured regions). A) Encoding scheme applied to each voxel at each encoding step, $$$m=1…5$$$. Amplitude was weighted randomly between 0.8 and 1 and phase modulated between 0 and 180$$$^{\circ}$$$. B) Receive sensitivities (shown for two coils), which are combined with (A) to generate weighted signals at each channel for each encoding step, $$$m$$$. Signals are concatenated over M=5 steps. C) Inversion of the system matrix, $$$\bf S$$$, reveals reconstructed signals from each voxel.
    Fig. 4: Data acquired from two voxels (voxel 1=blue; voxel 2=orange) in a phantom (A) using proposed RF-encoded technique with an increasing number of encoding steps, M. (B) Combined signal (not unfolded) from all receive coils shows shot-to-shot variation in phase from (0°,30°,60°,90°,120°) modulation on voxel 2 using the dual-band pulses. (C) Reconstructed data using RF-encoding (steps M=2-5) are shown below coil-encoding only (M=1). The spectral SNR and effective g-factor values are stated next to each encoding.
  • Extended Accelerated Systematic Tracking using Experimental Radiology: an Encephalo-Graphic Generation
    S. T. Claus1, Y. Eti2, and E. S. Terbuny3
    1Dept. of Presents, North Pole Research Agency, Rovaniemi, Finland, 2Atopof Amountain, Himalaya, Bhutan, 3Dept. for Sweets, Greenfield Institute, Ostereistedt, Germany
    Don't let your chocolate lie around.
    Figure 1: Sequence diagram of the UTE se-quence with the rabbit feigning (RF) pulse shape. Note that the gradients in the lower part require fast switching capabilities and a small amount of time travel. Allergy advice: sequence may contain traces of nuts.
    Figure 3: POTS reconstructed image of the EASTER-EGG sequence being presented to the CNN (or $$$CNN$$$ or CNN). Rabbit detection is completely impossible for a human being, but the CNN (or $$$CNN$$$ or CNN) was able to ignore it in less than 4 h.